A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable vinyl synthetic examination gloves; powdered yellow color. Worn on examiner's hands or fingers during medical examinations. Acts as physical barrier to prevent contamination between patient and examiner. Used in clinical settings. No electronic or mechanical components.
Clinical Evidence
Bench testing only. Testing included physical and dimension testing (ASTM D5250-06, AQL 4.0), watertight testing (ASTM D5151-06, AQL 2.5), and biocompatibility (primary skin irritation and skin sensitization). No clinical data provided.
Technological Characteristics
Material: Vinyl synthetic. Standard: ASTM D5250-06. Form factor: Disposable examination glove, powdered, yellow. Sterilization: Not specified. No software or electronic components.
Indications for Use
Indicated for use by medical examiners to prevent cross-contamination between the examiner and the patient.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Assured Medical Supply Company, Ltd. Vinyl Examination Glove (K932054)
Related Devices
K071123 — DISPOSABLE POWDER FREE VINYL EXAM GLOVE, YELLOW COLOR · Xuzhou Full Sun Medical Products , Ltd. · Jul 30, 2007
{0}------------------------------------------------
### 510(k) SUMMARY
This summary of 510(K) safety and effectiveness information is being submitted in Accordance with requirements of SMDA 1990 and 21 CFR 807.92.
K082448 The assigned 510(K) number is:
DEC 1 0 2008
# 1. Submitter's Identification:
Ms. Michelle Shih Rich Mountain Medical Products Inc .. No. 8, Jiang Yin Road, YiTang Town Pizhou County, jiangsu Province -China
Date Summary Revised: Nov. 21, 2008
# 2. Name of the Device:
Rich Mountain Medical Products Inc. Disposable Vinyl Synthetic Examination Gloves, Powdered Yellow Color
# 3. Predicate Device Information:
Assured Medical Supply Company, Ltd. Vinyl Examination Glove (K932054)
# 4. Device description:
Classified by FDA's General and Plastic Surgery Device panel as Class 1, 21 CFR 880.6250, Powdered Vinyl Patient Examination Glove, 80LYZ, and meets all requirements of ASTM Standard D5250-06
# 5. Intended Use:
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
7 - 1
3/4
{1}------------------------------------------------
### 6.Comparison to Predicate Devices:
Rich Mountain Medical Products Inc.'s Disposable Vinyl Synthetic Examination Gloves. Powdered Yellow Color is substantially equivalent to the device manufactured by Assured Medical Supply Company, Ltd. (K932054).
# 7. Discussion of Non-Clinical tests Performed for Determination of Substantial Equivalence are as follows:
The standards used for Rich Mountain Medical Products Inc.' s glove production are Based on ASTM D5250-06. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection level S-2, AQL 4.0.
The FDA 1000 ml Watertight Test based on ASTM D-5151-06 was also conducted with samplings of AQL 2.5, Inspection level I, meeting these requirements. Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.
There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.
# 8. Discussion of Clinic Tests Performed:
Not applicable – There is no Hypoallergenic Claim
#### 9. Conclusions:
Rich mountain Medical Products Inc.' s Disposable Vinyl Synthetic Examination Gloves, Powdered Yellow Color conform fully to ASTM D-5250-06 standard as well as applicable 21CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# DEC 1 0 2008
Ms. Michelle Shih Coordinator Rich Mountain Medical Products, Incorporated No. 8, Jiang Yin Road, Yitang Town Pizhou County, Jiangsu CHINA
Re: K082448
Trade/Device Name: Disposable Vinyl Synthetic Examination Gloves, Powdered Yellow Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LYZ Dated: November 21, 2008 Received: November 25, 2008
Dear Ms. Shih:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2 - Ms. Shih
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice. requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
QRS
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### RICH MOUNTAIN MEDICAL PRODUCTS INC. Applicant:
510(k) Number (if known): K082448
Device Name: Disposable Vinyl Synthetic Examination Gloves, Powdered
Yellow Color
Indications for Use:
A PATIENT EXAMINATION GLOVES IS A DISPOSABLE DEVICE INTENDED FOR MEDICAL PURPOSE THAT IS WORN ON THE EXAMINER'S HAND OR FINGERS TO PREVENT CONTAMINATION BETWEEN PATIENT AND EXAMINER.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER
PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
)ivision of Anesthesiology, General Hospital fection Control, Dental Devices
10(k) Number: K 0
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.