← Product Code [IYN](/productcode/IYN) · K082374

# VIVID I AND VIVID Q (K082374)

_Ge Medical Systems Israel, Ultrasound, Ltd. · IYN · Sep 2, 2008 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K082374

## Device Facts

- **Applicant:** Ge Medical Systems Israel, Ultrasound, Ltd.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Sep 2, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Indications for Use

Ultrasound imaging, measurement and analysis of the human body as follows: Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transesophageal; Transvaginal; Intraoperative (obdominal, thoracic, and vascular), Intra-cardiac and Inra-luminal.

## Device Story

Compact, portable diagnostic ultrasound system; features integrated keyboard, fold-up LCD display, interchangeable electronic-array transducers. Inputs: ultrasonic signals from transducers. Processing: digital acquisition, signal processing, image reconstruction. Outputs: B-mode, M-mode, PW/CW Doppler, Color/Power Doppler, Harmonic imaging, Coded pulse imaging. Used in clinical settings by physicians for diagnostic imaging/measurement. Output displayed on integrated GUI; assists in clinical diagnosis/patient management. Benefits: portable, real-time diagnostic imaging for diverse clinical applications.

## Clinical Evidence

Bench testing only. Evaluated for acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, and mechanical safety. Conforms to applicable medical device safety standards. No clinical data required.

## Technological Characteristics

Compact portable ultrasound system; electronic-array transducers. Modes: B, M, PW/CW Doppler, Color/Power Doppler, Harmonic, Coded Pulse. Connectivity: digital acquisition/display. Quality systems: 21 CFR 820, ISO 9001, ISO 13485. Software: includes standard GE ultrasound imaging algorithms.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- GE Vivid-i Ultrasound System ([K061525](/device/K061525.md))

## Reference Devices

- Vivid 7 system ([K051449](/device/K051449.md))
- LOGIQ-ile system ([K072797](/device/K072797.md))
- Vivid-e system ([K072797](/device/K072797.md))
- AcuNav 10F Transducer ([K033650](/device/K033650.md))
- Various transducers ([K033139](/device/K033139.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Y082374

SEP - 2 2008

### Section 2 Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92(c).

Image /page/0/Picture/4 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circular border. The letters and the border are black, contrasting with the white background.

2

GE Healthcare

General Electric Company P.O. Box 414, Milwaukee, WI 53201

### Section a):

| Submitter:      | GE Medical Systems Israel, Ultrasound LTD<br>Haetgar Str. 4<br>Tirat Carmel, ISRAEL 39120                            |
|-----------------|----------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Israel Citron,<br>Quality Assurance and Regulatory Affairs manager<br>Telephone: 972-4-8519-555; Fax: 972-4-8519-500 |
| Date Prepared:  | July 31, 2008                                                                                                        |

2.

| Device Name: | GE Vivid-i and Vivid q Diagnostic Ultrasound System<br>Ultrasonic Pulsed Doppler Imaging System, 21 CFR 892.1550, 90-IYN<br>Ultrasonic Pulsed Echo Imaging System, 21 CFR 892.1560, 90-IYO<br>Diagnostic Ultrasonic Transducer, 21 CFR 892.1570, 90-ITX |
|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

3. Marketed Device: GE Vivid-i Ultrasound System, K061525 currently in commercial distribution.

4. Device Description: The GE Vivid-i and Vivid-q is compact and portable diagnostic ultrasound system with integrated keyboard, fold-up LCD type display and interchangeable electronic-array transancess. It has an overall size approximately 36 cm wide, 31.5 cm deep and 6 cm high in transport configuration and provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, an intuitive layout of specialized controls, color GUI display and Doppler audio.

5. Indications for Use: The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transesophageal; Transvaginal; Intraoperative (obdominal, thoracic, and vascular), Intra-cardiac and Inra-luminal.

6. Comparison with Predicate Device: The modified GE Vivid-i is of a comparable type and substantially equivalent to the currently marketed GE Vivid-i. It is a compact and readily portable unit having the same design, construction, and materials; is comparable in key safety and effectiveness features. It has the same intended uses as the predicate device and additional software features are identical to that of other cleared GE Ultrasound systems.

### Section b):

1. Non-clinical Tests: The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and has been found to conform with applicable medical device safety standards.

2. Clinical Tests: None required.

3. Oncelusion: Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines, and established methods of patient examination. The design and development process of the manufacturer conforms with 21 CFR 820, ISO 9001 and ISO13485 quality systems. The device conforms to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore, it is the opinion of GE Medical Systems that the GE Vivid-i and Vivid -d Diagnostic Ultrasound is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three legs and a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" is written around the border of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

GE Medical Systems, Ultrasound % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

SEP - 2 2008

Re: K082374

Trade/Device Name: Vivid - i and Vivid - q Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: TYN, IYO, and ITX Dated: August 15, 2008 Received: August 18, 2008

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the Vivid - i and Vivid - q, as described in your premarket notification:

#### Transducer Model Number

| Vivid i/q with M4S-RS                | Vivid i/q with 3C-RS   | Vivid i/q with 3S-RS     | Vivid i/q with 9T-RS |
|--------------------------------------|------------------------|--------------------------|----------------------|
| Vivid i with 4C-RS                   | Vivid i/q with 8C-RS   | Vivid i/q with 7S-RS     | Vivid i/q with P2D   |
| Vivid i with 12L-RS                  | Vivid i/q with 8L-RS   | Vivid i/q with 10S-RS    | Vivid i/q with P6D   |
| Vivid i/q with 5S-RS                 | Vivid i/q with i12L-RS | Vivid i/q with 6T/6Tc-RS | Vivid i/q with 6S-RS |
| Vivid i/q with AcuNav         TM 10F |                        |                          |                      |

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any

{2}------------------------------------------------

Page 2 - Mr. Mark Job

Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special renor is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-201 or (240) 27-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

If you have any questions regarding the content of this letter, please contact Ms. Lauren Hefner at (240) 276-3666.

Sincerely yours,

Vorin M. Whay

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{3}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form GE Vivid-i and Vivid-g Diagnostic Ultrasound System

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                 |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|-----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse* | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Fetal                                              | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               |       |
| Abdominal[1]                                       | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               |       |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               |       |
| Small Organ (specify)[2]                           | P                 | P | P             | P             | P                | E                  | P                | P                 | P                   | P               |       |
| Neonatal Cephalic                                  | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               |       |
| Adult Cephalic                                     | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               |       |
| Cardiac[3]                                         | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P               |       |
| Peripheral Vascular                                | P                 | P | P             | P             | P                | E                  | P                | P                 | P                   | P               |       |
| Musculo-skeletal Conventional                      | P                 | P | P             | P             | P                | E                  | P                | P                 | P                   | P               |       |
| Musculo-skeletal Superficial                       | P                 | P | P             | P             | P                | E                  | P                | P                 | P                   | P               |       |
| Other[4]                                           | P                 | P | P             | P             | P                | E                  | P                | P                 | P                   | P               |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Transesophageal                                    | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                 |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Intraoperative (specify)[5]                        | P                 | P | P             | P             | N                | P                  | P                | P                 | P                   |                 |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Intracardiac and Intraluminal                      | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   |                 |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes Renal.

[2] Small organ includes breast, testes, thyroid.

[3] Cardiac is Adult and Pediatric.

[4] Other use includes Urology.

[5] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV).

- ["] Combined modes are B/M, B/PWD, B/CWD, B/Color/PWD, B/Amplitude/PWD.
{"} Coded Pulse includes Coded Octave Imaging (COI), and Coded Phase Inversion (CPI).

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division of Reproductive, Abdominal, and Radiological Dev 510(k) Number

{4}------------------------------------------------

# Diagnostic Ultrasound Indications for Use Form GE Vivid i/q with M4S-RS Transducer

| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |  |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|--|
| Ophthalmic                                         |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Fetal                                              | N | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |       |  |
| Abdominal [1]                                      | N | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |       |  |
| Pediatric                                          | N | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |       |  |
| Small Organ (specify)                              |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Neonatal Cephalic                                  |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Adult Cephalic                                     | N | N | N             | N             | N                | N                  | N                | N                 | N                   | N              | N     |  |
| Cardiac [2]                                        | N | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |       |  |
| Peripheral Vascular                                |   |   |               |               |                  |                    |                  |                   | N                   |                |       |  |
| Musculo-skeletal Conventional                      |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Musculo-skeletal Superficial                       |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Other (specify)                                    |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Exam Type, Means of Access                         |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transesophageal                                    |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transrectal                                        |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transvaginal                                       |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transuretheral                                     |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative (specify)                           |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative Neurological                        |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intracardiac and Intraluminal                      |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Laparoscopic                                       |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |

Intended Use: Ultrasound imaging, measurement and analysis of the human body as foll

N = new indication (transducer previously cleared with Vivid 7 system K051449); P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes Renal

[2] Cardiac is Adult and Pediatric.

{"] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

["] Coded Pulse includes Coded Octave Imaging (COI), and Coded Phase Inversion (CP).

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription User (Per 21 CFR 801.109)

Aorai In Mitz

(Division Sign Division of Reproductive, Abdomina and Radiological Devic 510(k) Number

{5}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form GE Vivid-i with 4C-RS Transducer

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal                                              | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P              |       |
| Abdominal[1]                                       | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P              |       |
| Pediatric                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Small Organ (specify)                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other[2]                                           | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P              |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intracardiac and Intraluminal                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

N = new indication; P = previously cleared by FDA (K061525); E = added under Appendix E

Notes: [1] Abdominal includes Renal;

[2] Other use includes Urology.

["] Combined modes are B/M, B/PWD, B/CWD, B/Color/PWD, B/Amplitude/PWD.

["] Coded Pulse includes Coded Octave Imaging (COI), and Coded Phase Inversion (CPI).

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription User (Per 21 CFR 801.109)

Domi In Wh

(Division Sign-Off) Division of Reproductive and Radiological Dev

510(k) Number

{6}------------------------------------------------

# Diagnostic Ultrasound Indications for Use Form

## GE Vivid-i with 12L-RS Transducer

Mode of Operation Clinical Application PW CW Color Color M Power Combined Harmonic B M Coded Other Anatomy/Region of Interest Doppler Doppler Doppler Doppler Doppler Modes Imaging Pulse Ophthalmic Fetal Abdominal Pediatric Small Organ(1) P P p P E P P P P Neonatal Cephalic Adult Cephalic Cardiac Peripheral Vascular P P P P P m P P P Musculo-skeletal Conventional P P P P ח P P P P Musculo-skeletal Superficial P P P P E P P P P Other (specify) Exam Type, Means of Access Transesophageal Transrectal Transvaginal Transuretheral Intraoperative (specify) Intraoperative Neurological Intracardiac and Intraluminal Laparoscopic

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

N = new indication; P = previously cleared by FDA (K061525); E = added under Appendix E

Notes: [1] Small organ includes breast, testes, thyroid.

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

["] Coded Pulse includes Coded Octave Imaging (COI), and Coded Phase Inversion (CPI),

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription User (Per 21 CFR 801.109)

*(Division Sign-Off)*

Division of Reproductive, Abdominal, and Radiological Device

510(k) Number

{7}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form GE Vivid-i/q with 5S-RS Transducer

| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |  |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|--|
| Ophthalmic                                         |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Fetal                                              | P | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |       |  |
| Abdominal                                          |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Pediatric                                          | P | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |       |  |
| Small Organ (specify)                              |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Neonatal Cephalic                                  |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Adult Cephalic                                     |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Cardiac[1]                                         | P | P | P             | P             | P                | P                  | P                | P                 | P                   | P              | P     |  |
| Peripheral Vascular                                |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Musculo-skeletal Conventional                      |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Musculo-skeletal Superficial                       |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Other                                              |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Exam Type, Means of Access                         |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transesophageal                                    |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transrectal                                        |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transvaginal                                       |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transuretheral                                     |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative (specify)                           |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative Neurological                        |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intracardiac and Intraluminal                      |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Laparoscopic                                       |   |   |               |               |                  |                    |                  |                   |                     |                |       |  |

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

N = new indication; P = previously cleared by FDA (K061525); E = added under Appendix E

Notes: [1] Cardiac is Adult and Pediatric.

["] Combined modes are B/M, B/PWD, B/CWD, B/Color/PWD, B/Amplitude/PWD.

["] Coded Pulse includes Coded Octave Imaging (COI), and Coded Phase Inversion (CPI).

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Reproductive, Abdomina and Radiological Device 510(k) Number

{8}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form GE Vivid-i/q with 3C-RS Transducer

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal                                              | P                 | P | P             |               | P                | E                  |                  | P                 | P                   | P              |       |
| Abdominal[1]                                       | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P              |       |
| Pediatric                                          | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P              |       |
| Small Organ (specify)                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other[2]                                           | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P              |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intracardiac and Intraluminal                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

N = new indication; P = previously cleared by FDA (K033139); E = added under Appendix E

Notes: [1] Abdominal includes Renal.

[2] Other use includes Urology.

["] Combined modes are B/M, B/PWD, B/CWD, B/Color/PWD, B/Amplitude/PWD.

["] Coded Pulse includes Coded Octave Imaging (COI), and Coded Phase Inversion (CPI).

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription User (Per 21 CFR 801.109)

forgath why

(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Dev 510(k) Number

{9}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form GE Vivid-i/q with 8C-RS Transducer

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal                                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Abdominal                                          | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P              |       |
| Pediatric                                          | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P              |       |
| Small Organ (specify)[1]                           | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P              |       |
| Neonatal Cephalic                                  | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P              |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Peripheral Vascular                                | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P              |       |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other                                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intracardiac and Intraluminal                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA (K033139); E = added under Appendix E

Notes: [1] Small organ includes breast, testes, thyroid.

[*] Combined modes are B/M, B/PWD, B/CWD, B/Color/PWD, B/Amplitude/PWD.

["] Coded Pulse includes Coded Octave Imaging (COI), and Coded Phase Inversion (CPI).

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Arzu M. Whay

(Division Sign-Off) Division of Reproductive, Al and Radiological Devices 510(k) Number .

{10}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form GE Vivid i/q with 8L-RS Transducer

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                 |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|-----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse* | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Fetal                                              |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Abdominal                                          | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P               |       |
| Pediatric                                          | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P               |       |
| Small Organ (specify)[1]                           | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P               |       |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Peripheral Vascular                                | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P               |       |
| Musculo-skeletal Conventional                      | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P               |       |
| Musculo-skeletal Superficial                       | P                 | P | P             |               | P                | E                  | P                | P                 | P                   | P               |       |
| Other                                              |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Intracardiac and Intraluminal                      |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                 |       |

N = new indication; P = previously cleared by FDA (K033139); E = added under Appendix E

Notes: [1] Small organ includes breast, testes, thyroid.

["] Combined modes are B/M, B/PWD, B/CWD, B/Color/PWD, B/Amplitude/PWD.

["] Coded Pulse includes Coded Octave Imaging (COI), and Coded Phase Inversion (CPI).

(PLEASE DO NOT WRITE BELOW THIS LINE . CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

---

(Division Sign-Off)

Prescription User (Per 21 CFR 801.109)

Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

{11}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form GE Vivid i/q with i12L-RS Transducer

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

| Clinical Application<br>Anatomy/Region of Interest | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
|                                                    | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal                                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Abdominal                                          | P                 | P | P             |               | P                | E                  | P                | P                 | P                   |                |       |
| Pediatric                                          | P                 | P | P             |               | P                | E                  | P                | P                 | P                   |                |       |
| Small Organ (specify)[1]                           | P                 | P | P             |               | P                | E                  | P                | P                 | P                   |                |       |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac [2]                                        | P                 | P | P             |               | P                | E                  | P                | P                 | P                   |                |       |
| Peripheral Vascular                                | P                 | P | P             |               | P                | E                  | P                | P                 | P                   |                |       |
| Musculo-skeletal Conventional                      | P                 | P | P             |               | P                | E                  | P                | P                 | P                   |                |       |
| Musculo-skeletal Superficial                       | P                 | P | P             |               | P                | E                  | P                | P                 | P                   |                |       |
| Other                                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)[3]                        | P                 | P | P             |               | P                | E                  | P                | P                 | P                   |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intracardiac and Intraluminal                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA (K033139); E = added under Appendix E

Notes: [1] Small organ includes breast, testes, thyroid.

[2] Cardiac is Adult and Pediatric.

[3] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV).

[*] Combined modes are B/M, B/PWD, B/CWD, B/Color/PWD, B/Amplitude/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Reproductive. Abdominal. and Radiological Devices 510(k) Number .

{12}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form GE Vivid i/q with 3S-RS Transducer

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal                                              | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |       |
| Abdominal(1)                                       | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |       |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |       |
| Small Organ (specify)                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                     | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |       |
| Cardiac[2]                                         | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |       |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other                                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intracardiac and Intraluminal                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

N = new indication; P = previously cleared by FDA (K033139); E = added under Appendix E

Notes: [1] Abdominal includes Renal.

[2] Cardiac is Adult and Pediatric.

[*] Combined modes are B/M, B/PWD, B/CWD, B/Color/PWD, B/Amplitude/PWD.

["] Coded Pulse includes Coded Octave Imaging (COI), and Coded Phase Inversion (CPI),

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Angel McWhay

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

{13}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form GE Vivid i/q with 7S-RS Transducer

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal                                              | E                 | E | E             | E             | E                | E                  |                  | E                 | E                   |                |       |
| Abdominal                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Small Organ (specify)                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Neonatal Cephalic                                  | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac[1]                                         | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Peripheral Vascular                                | E                 | E | E             | E             | E                | E                  | E                | E                 | E                   |                |       |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other                                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intracardiac and Intraluminal                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA (K033139); E = added under Appendix E

Notes: [1] Cardiac is Adult and Pediatric.

["] Combined modes are B/M, B/PWD, B/CWD, B/Color/PWD, B/Amplitude/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Arqui M. Why

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number .

{14}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form GE Vivid i/q with 10S-RS Transducer

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal                                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Abdominal                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Small Organ (specify)                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Neonatal Cephalic                                  | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac[1]                                         | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |       |
| Peripheral Vascular                                | E                 | E | E             | E             | E                | E                  | E                | E                 | E                   |                |       |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other                                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |…

---

**Source:** [https://fda.innolitics.com/device/K082374](https://fda.innolitics.com/device/K082374)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
