← Product Code [LLZ](/productcode/LLZ) · K082336

# XR-EXPRESS (K082336)

_New Mexico Software, Inc. · LLZ · Oct 15, 2008 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K082336

## Device Facts

- **Applicant:** New Mexico Software, Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Oct 15, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

XR-EXpress is a PACS and teleradiology system used to receive DICOM images, scheduling information and textual reports, organize and store them in an internal format, and to make that information available across a network via web and customized user interfaces. XR-EXpress is for hospitals, imaging centers, nursing homes, radiologist reading practices, and any user who requires and is granted access to patient image, demographic, and report information. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Moreover, it is the user's responsibility to ensure monitor quality, ambient light conditions, and image compression ratios are consistent with the clinical application.

## Device Story

XR-EXpress is a PACS/teleradiology system for image management and distribution. Inputs include DICOM images, scheduling info, and textual reports received via network or direct DICOM connection. System components include Capture (data ingestion), Server (registration, archiving, association with prior cases), and Viewer (web-based or dedicated workstation for diagnosis). Users include radiologists, hospital staff, and authorized clinicians. System organizes data into worklists; supports dictation and transcription of reports. Output is electronic or faxed reports and rendered images for clinical review. Physician intervention is required for all interpretations. Benefits include centralized access to patient imaging and reports across clinical settings, facilitating remote reading and efficient workflow.

## Clinical Evidence

No clinical data. Validation testing consisted of bench-level stress testing of system components to ensure safe and effective operation as a PACS for teleradiology services.

## Technological Characteristics

System architecture includes Capture, Server, and Viewer components. Operates on general-purpose computing hardware. Connectivity via network/web browser. Supports DICOM standard. Software-based image rendering and management. No specific materials or energy sources defined beyond standard IT infrastructure.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Advanced 3D Web PACS ([K063221](/device/K063221.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for New Mexico Economic Development. The logo consists of a stylized evergreen tree on the left, followed by the words "New Mexico" in a bold, sans-serif font. The words "Economic Development" are to the right of "New Mexico" in a smaller font. The text is somewhat blurry.

OCT 1 5 2008

# 510(k) Summary

(in accordance with 21 CFR 807.92)

510(k) Number K_O&236

#### I. Applicant Information

| Applicant:                 | NEW MEXICO SOFTWARE INC.<br>5021 Indian School Road NE, Suite 100<br>Albuquerque, NM 87110<br>U.S.A.                             |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:            | Rafael Rubio<br>Technology & Development Vice President<br>Tel: (505) 255-1999<br>Fax: (505) 255-7201<br>e-mail: rafael@nmxs.com |
| Application Correspondent: | EMERGO GROUP INC.<br>1705 S. Capital of Texas Hwy., Suite 500<br>Austin, TX 78746<br>U.S.A.                                      |
| Contact Person:            | Neal Kolber<br>Project Manager<br>Tel: (512) 327-9997<br>Fax: (512) 327-9998<br>e-mail: neal@emergogroup.com                     |
| Date Prepared:             | August 5, 2008                                                                                                                   |

Device Name and Classification

| Proprietary Name:     | XR-EXpress                                     |
|-----------------------|------------------------------------------------|
| Common/Usual Name:    | Soft-copy reading and acquisition system       |
| Classification Name:  | Picture Archiving Communications System (PACS) |
| Regulation Number:    | 892.2050                                       |
| Product Codes:        | LLZ                                            |
| Classification:       | Class II                                       |
| Classification Panel: | Radiology Devices                              |

Traditional 510(k) -- XR-EXpress

II.

CONFIDENTIAL

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for New Mexico. The logo consists of a stylized Zia symbol on the left, followed by the word "New Mexico" in a sans-serif font. The Zia symbol is a sun symbol that represents the four cardinal directions, the four seasons, and the four stages of life. The logo is simple and recognizable, and it is often used on official documents and websites.

#### III. Predicate Device

The XR-EXpress device is substantially equivalent to the following FDA cleared predicate device with regard to indications for use, performance and technological characteristics:

| 510(k) Number:       | K063221                                        |
|----------------------|------------------------------------------------|
| Trade Name:          | Advanced 3D Web PACS TM                        |
| Manufacturer:        | Nexsys Electronics, Inc. DBA Medweb LLC        |
| Common/Usual Name:   | Soft-copy reading and acquisition system       |
| Classification Name: | Picture Archiving Communications System (PACS) |
| Regulation Number:   | 892.2050                                       |
| Product Codes:       | LLZ                                            |
| Classification:      | Class II                                       |

#### IV. Description and Intended Use of the Device

XR-EXpress is a PACS and teleradiology system used to receive DICOM images, scheduling information and textual reports, organize and store them in an internal format, and to make that information available across a network via web and customized user interfaces. XR-EXpress is for hospitals, imaging centers, nursing homes, radiologist reading practices, and any user who requires and is granted access to patient image, demographic, and report information.

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Moreover, it is the user's responsibility to ensure monitor quality, ambient light conditions, and image compression ratios are consistent with the clinical application.

The software does not contact the patient, nor does it control any lifesustaining devices. A physician has ample opportunity for competent human intervention while interpreting images and clinical information.

#### V. Summary of the Technical Characteristics

The XR-EXpress system is comprised of a capture component (XR-EXpress Capture), a central server component (XR-EXpress Server), and a viewer component (XR-EXpress Viewer) that is used on general purpose computing hardware. Patient information, procedure information, and image acquisition data is captured by XR-EXpress Capture and sent to the XR-EXpress server or

CONFIDENTIAL Traditional 510(k) - XR-EXpress

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the New Mexico state logo. The logo consists of a stylized Zia symbol on the left, followed by the words "New Mexico" in a sans-serif font. To the right of the state name are some blurred words.

alternatively, it is accepted directly from a DICOM compliant device. After receiving image objects and associated data, the XR-EXpress server registers the incoming data and images with the archive including any associations with prior cases. The cases are presented to users with the correct permissions from a work list as controlled by a managing user. The image data is transmitted and rendered on the users workstation using a standard web browser for management views or the XR-EXpress Viewer for diagnosis. After using the workstation to view the images, the user optionally dictates a report into the system, after which, a user can play back the dictation and transcribe it to text. Once the XR-EXpress Server registers a report, the report is available to the referring physician or client service where it available electronically or via fax.

#### VI. Testing

New Mexico Software has conducted extensive validation testing of the XR-EX System, as a PACS system that is capable of providing reliable teleradiology services over the Internet. All of the different components of the XR-EX system have been stress tested to ensure that the system as a whole provides all the capabilities necessary to operate safely and effectively.

#### VII. Safety & Effectiveness Conclusions

Based on the intended use and technological characteristics, the XR-EXpress system is substantially equivalent to the Advanced 3D Web PACSTM device manufactured by Nexsys Electronics, Inc. DBA Medweb LLC (K063221). The XR-EXpress raises no new safety or effectiveness issues.

{3}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, topped by a pair of wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

### OCT 1 5 2008

New Mexico Software, Inc. % Mr. Neal Kolber Project Manager Emergo Group, Inc. 1705 S. Capital of Texas Hwy., Suite 500 AUSTIN TX 78746

Re: K082336

Trade/Device Name: XR-EXpress Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 5, 2008 Received: August 15, 2008

Dear Mr. Kolber:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

### Page 2

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

hope M. Whang

Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the words "New Mexico" in a stylized font. To the left of the words is a symbol that looks like a stylized pine tree. The words "Department of Health" are written in a smaller font to the right of "New Mexico."

## 4. Indication for Use Statement

510(k) Number (if known):

K082326

Device Name:

XR-EXpress

Indications for Use:

XR-EXpress is a PACS and teleradiology system used to receive DICOM images. scheduling information and textual reports, organize and store them in an internal format, and to make that information available across a network via web and customized user interfaces. XR-EXpress is for hospitals, imaging centers, nursing homes, radiologist reading practices, and any user who requires and is granted access to patient image, demographic, and report information.

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Helena

Page 1 of

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K082336](https://fda.innolitics.com/device/K082336)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
