K082248 · Diagnostica Stago, Inc. · JPA · Dec 12, 2008 · Hematology
Device Facts
Record ID
K082248
Device Name
STA SATELLITE AUTOMATED MULTI-PARAMETER ANALYZER
Applicant
Diagnostica Stago, Inc.
Product Code
JPA · Hematology
Decision Date
Dec 12, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5425
Device Class
Class 2
Intended Use
The STA Satellite™ Automated Multi-Parametric AnalyzerSatellite® Automated Multi-Parametric Analyzer is a fully automatic clinical instrument indicated and intended for the performance of tests on human plasmas, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.
Device Story
STA Satellite™ is a fully automated bench-top clinical analyzer for coagulation testing. Input: human plasma samples loaded into cuvettes via automated needle aspiration. Operation: system uses photometric and chronometric detection; photometric method measures absorbance of monochromatic light through cuvette; chronometric method measures oscillation amplitude of a metal ball within the cuvette, where amplitude decrease indicates coagulation. Output: automated analysis and reporting of coagulation test results. Used in clinical laboratory settings; operated by laboratory personnel. Healthcare providers use results to diagnose coagulation abnormalities or monitor anticoagulant therapy (e.g., Coumadin, UFH, LMWH). Benefits: provides standardized, automated, and efficient coagulation assessment compared to manual methods.
Clinical Evidence
Bench testing only. Validation study conducted at one internal and two external sites comparing STA Satellite to STA-R predicate. Samples analyzed in duplicate. Statistical analysis included mean, standard deviation, 95% confidence intervals, and correlation. Results demonstrated correlation within acceptance criteria for all assays; no significant difference between instruments or sites.
Technological Characteristics
Fully automatic bench-top system. Sensing principles: chronometric (viscosity-based oscillation amplitude) and photometric (optical density). Energy source: electrical. Connectivity: central processing unit for data management and reporting. Reagent management via barcode or manual entry. Software-controlled automated analysis.
Indications for Use
Indicated for use on human plasma samples to aid in the diagnosis of coagulation abnormalities or monitoring of anticoagulant therapy. Intended for clinical laboratory use.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
K983460 — STAR AUTOMATED MULTI-PARAMETRIC ANALYZER · American Bioproducts Co. · Nov 12, 1998
K093001 — STA-R EVOLUTION EXPERT SERIES HEMOSTASIS SYSTEM AUTOMATED MULTI-PARAMETRIC ANALYZER · Diagnostica Stago, Incorporated · Dec 18, 2009
K151867 — STA R MAX · Diagnostica Stago S.A.S · Aug 7, 2015
K961579 — STA COMPACT AUTOMATED MULTI-PARAMETRIC ANALYZER · American Bioproducts Co. · Jul 10, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "Stago" in a serif font, with the first letter capitalized and the rest in lowercase. Above the word, there is a stylized logo consisting of three curved shapes arranged in a circular fashion. The logo is black, and the word "Stago" is also black.
## Attachment - 6
## 510(k) SUMMARY
DEC 1 2 2008
# Diagnost
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:
Umberto V. Parrotta Diagnostica Stago, Inc. Five Century Drive Parsippany, New Jersey 07054
Phone: (973) -631-1200, x-2044 Facsimile: (973) -695-0095
Contact Person: Umberto V. Parrotta
Date Prepared: August 06, 2008
#### Name of Modified Device and Name/Address of Sponsor:
STA Satellite®
Diagnostica Stago, Inc. Five Century Drive Parsippany, New Jersey 07054
#### Common or Usual Name:
- IVD Coagulation Device/Instrument. ●
- . Automated and Semi-Automated Hematology Device.
- Multi-Parametric Analyzer ●
#### Classification Name:
System, Multipurpose for In Vitro Coagulation Studies
#### Predicate Devices:
- Diagnostica Stago SAS' STA-R® Automated Multi-Parametric Analyzer . (K983460).
- Sigma Diagnostic Inc.'s AMAX Destiny Coagulation Analyzer (K021162). .
File: DSI-Satellite-Attachment-06v4-510(k)Summary.doc
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "Stago" in a serif font, with the first letter capitalized. Above the word is a logo consisting of three curved shapes arranged in a circular pattern. The logo is black, and the word "Stago" is also black.
CONFIDENTIAL
Attachment - 6
## 510(k) SUMMARY
## Predicate Device Manufacturer:
Diagnostica Stago SAS
#### Purpose of the Special 510(k) notice:
The STA Satellite® is a modification to the company's own legally marketed device, STA-R®.
#### Indication/Intended Use:
The STA Satellite™ Automated Multi-Parametric AnalyzerSatellite® Automated Multi-Parametric Analyzer is a fully automatic clinical instrument indicated and intended for the performance of tests on human plasmas, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.
#### Technological Characteristics:
The STA Satellite® Automated Multi-Parametric Analyzer is designed as a fully automatic bench-top system. Samples and test reagents are loaded into the instrument where sample handling, reagent delivery, analysis, and reporting of results are performed automatically. A central processing unit controls instrument functions such as, management of patient results, quality control, system supervision, support for instrument maintenance, and work load optimization. The STA Satellite® is the bench-top version of the company's STA-R®, legally marketed predicate device.
The instrument utilizes Diagnostica Stago reagents in addition to open adaptation of other currently available reagents. Barcoding of test reagents, calibrators, and controls facilitate their use on the system and permits reagent management simple. Manual entry of reagent information enables the use of non-barcoded reagents.
The instrument performs multiple test methodologies in random access as selected by the user. These include clotting time or clot-based tests (i.e. Chronometricchronometric) measurements and photometric assays (at specific wave lengths) on plasma samples. The principle of the chronometric method consists in measuring the variation of the oscillation amplitude of the ball (in the cuvette). A decrease in oscillation amplitude corresponds to an increase in the viscosity of the media (i.e. ., coagulation). The principle of the photometric measurements on the instrument is based on measured absorbance (also referred
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "Stago" in a serif font, with the "S" capitalized and the rest of the letters in lowercase. Above the word "Stago" is a stylized graphic element that appears to be composed of three curved shapes arranged in a circular or triangular pattern. The graphic is solid black, contrasting with the white background.
CONFIDENTIAL
## Attachment - 6
## 510(k) SUMMARY
to as Optical Density, or OD) of monochromatic light of predetermined wavelength passing through the cuvette as a (clotting) reaction takes place.
### Performance Data:
Comparison of the device and predicate device performance characteristics was conducted during the performance validation study. Validation of the STA -Satellite® with the STA-R® predicate device was performed at an internal site and two external sites.
Test samples were analyzed in duplicate on the STA -Satellite and the STA-R. Results were analyzed by standard statistical methods (mean, standard deviation, 95% confidence intervals, correlation between instruments and comparison to acceptance criteria).
Correlation data between the STA -Satellite and the STA-R were within the acceptance criteria for all assays and all sites (see Table below for assays performed). There was no significant difference or clinically significant difference between the results from the instruments between the sites and instruments.
In addition, the STA Satellite® is CE Marked and has been in commerce in France since September 2004. There have been no adverse performance-related events or analogous complaints resulting in recalls with regards to its design during this time that would raise questions of safety and effectiveness.
In conclusion, performance validation study comparison results between the STA Satellite® and the STA-R device instrument and predicate device indicate statistically similar results. , Analysis of which demonstrates substantial equivalence.
#### Substantial Equivalence:
The STA Satellite® has the same intended use and similar indications, principles of operation, and technological characteristics as the STA-R® and the AMAX predicate devices. The differences in the external architecture, software, and chronometric viscosity detection do not raise any new questions of safety or effectiveness. Performance data and risk assessment demonstrate that the STA Satellite® is as safe and effective as STA-R@the predicate devices. Thus, the STA Satellite® is substantially equivalent to its predicate devices.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The text is in all caps and is relatively small compared to the caduceus symbol.
#### Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Diagnostica Stago, Inc. C/o Umberto V. Parrotta Five Century Drive Parsippany, New Jersey 07054
## DEC 1 2 2008
Re: k082248
Trade/Device Name: STA Satellite® Automated Multi-Parametric Analyzer Regulation Number: 21 CFR 864.5425 Regulation Name: Multipurpose System for In Vitro Coagulation Studies Regulatory Class: Class II Product Code: JPA Dated: October 03, 2008 Received: October 06, 2008
#### Dear Mr. Parrotta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR). Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed
{4}------------------------------------------------
Page 2 - Mr. Parrotta
predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Marian Chan, PhD
Acting Division Director
Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Stago" in a serif font, with the first letter capitalized and the rest in lowercase. Above the word "Stago" is a stylized graphic element. The graphic element is composed of three curved shapes that are arranged to suggest a circular or spherical form with a missing section at the top.
CONFIDENTIAL Attachment - 2A
## Indications for Use Form
510(k) Number (if known): KO822248
Device Name:
Satellite™ Automated Multi-Parametric Analyzer
Indications for Use:
The Satellite™ is a fully automatic clinical instrument designed for the performance of tests on human plasmas, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.
### (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off
Office of In Vitro Diagnostic Device
Office of In Vitro Diamostic Device Evaluation and Safely
Prescription Use (Per 21 C.F.R. 801.109)
510(k) OR K082246 Over-The-Counter Use
(Optional Format 1-2-96)
File: DSI-Satellite-Attachment-2A-IndicationsForUseStatementForm.doc
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.