← Product Code [GGP](/productcode/GGP) · K082205

# FACTOR VIII ANTIBODY SCREEN (K082205)

_Genetic Testing Institute · GGP · Nov 20, 2008 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K082205

## Device Facts

- **Applicant:** Genetic Testing Institute
- **Product Code:** [GGP](/productcode/GGP.md)
- **Decision Date:** Nov 20, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.7290
- **Device Class:** Class 2
- **Review Panel:** Hematology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K082205 · Nov 20, 2008 | FACTOR VIII ANTIBODY SCREEN | Genetic Testing Institute | Retrospective clinical serum and plasma samples from hemophiliac and non-hemophiliac donors | Retrospective clinical samples were used to determine the assay cutoff and to demonstrate accuracy through method comparison against the predicate Factor VIII Inhibitor Assay and the Bethesda assay. | Method comparison; Retrospective clinical samples; Assay validation |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Method Comparison Study; Retrospective method comparison | Hemophiliac donors with and without Factor VIII antibodies; Sample Size: >200 samples; Number of Sites: 2 (Internal GTI study and external study at Mayo Clinic) | Factor VIII Inhibitor Assay and Bethesda assay | Co-positivity, co-negativity, and percent agreement |

## Indications for Use

The GTI Diagnostics Factor VIII Antibody Screen assay is a qualitative solid phase enzyme linked immunosorbent assay (ELISA) designed to detect IgG antibodies reactive with recombinant human factor VIII (FVIII) in human serum and plasma.

## Device Story

The GTI Diagnostics Factor VIII Antibody Screen is an ELISA-based diagnostic kit used in clinical laboratories to detect IgG antibodies against recombinant human Factor VIII. The device utilizes a 96-well microwell plate coated with recombinant FVIII. Patient serum or plasma is added to the wells; if anti-FVIII IgG antibodies are present, they bind to the immobilized FVIII. After washing, an alkaline phosphatase-conjugated anti-human IgG reagent is added, followed by a p-nitrophenyl phosphate (PNPP) substrate. The resulting colorimetric reaction is measured via optical density at 405nm. The assay output is a qualitative determination of antibody presence based on a comparison of the sample's optical density against a kit control. This information assists clinicians in identifying patients with FVIII inhibitors, which can impact the management of hemophilia A.

## Clinical Evidence

No clinical trials were performed. Evidence consists of analytical bench testing and method comparison studies. Precision testing (n=8 samples) showed within-run CV ≤20% and between-run CV ≤24%. Interference testing confirmed no impact from hemoglobin, lipemia, bilirubin, IVIG, or Rituxan. Method comparison against the predicate (n=137) showed 89 positive/40 negative agreement, with 7 false negatives and 1 false positive. Comparison against the Bethesda assay (n=81) showed 35 positive/40 negative agreement, with 2 false negatives and 4 false positives. Matrix equivalence was established for serum, ACD plasma, and 3.2% Na citrate plasma.

## Technological Characteristics

Solid-phase ELISA with colorimetric endpoint. Components: recombinant human Factor VIII (Kogenate FS) coated on low-volume flat-bottom microwells, alkaline phosphatase-labeled anti-human IgG conjugate, PNPP substrate, 3M NaOH stop solution. Connectivity: None (standalone). Energy source: Spectrophotometer (external). Materials: Tris-buffered saline, bovine serum albumin, sodium azide. Sterilization: Not specified.

## Predicate Devices

- GTI Factor VIII Inhibitor Assay ([K993553](/device/K993553.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE

A. 510(k) Number:
K0822905

B. Purpose for Submission:
Clearance of a new device

C. Measurand:
Factor VIII

D. Type of Test:
ELISA

E. Applicant:
Genetic Testing Institute, Inc. (GTI)

F. Proprietary and Established Names:
Factor VIII Antibody Screen

G. Regulatory Information:
1. Regulation section:
21 CFR 864.7290
2. Classification:
Class II
3. Product code:
GGP
4. Panel:
81 Hematology

{1}

H. Intended Use:

1. Intended use(s):

The GTI Diagnostics Factor VIII Antibody Screen assay is a qualitative solid phase enzyme linked immunosorbent assay (ELISA) designed to detect IgG antibodies reactive with recombinant human factor VIII (FVIII) in human serum and plasma.

2. Indication(s) for use:

3. Special conditions for use statement(s):

4. Special instrument requirements:

I. Device Description:

The GTI Diagnostics Factor VIII Antibody Screen assay is an ELISA with a colorimetric endpoint. The assay kit is made of a 96 welled microwell plate (in 1 X 8 strips), alkaline phosphatase conjugated goat antibody to human immunoglobulin G, p-nitrophenyl phosphate substrate, positive and negative control, Stop Solution, diluents, buffers, wash solutions, and plate sealers.

J. Substantial Equivalence Information:

1. Predicate device name(s):

GTI Factor VIII Inhibitor Assay

2. Predicate 510(k) number(s):

K993553

3. Comparison with predicate:

{2}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended Use | To detect IgG antibodies reactive with recombinant human factor VIII (FVIII) in human serum and plasma. | same  |
|  Technology | ELISA with a colorimetric endpoint | same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Reagents | Tris buffer containing NaCl, Na azide, bovine serum albumin for specimen diluent | Tris buffer containing NaCl and Na azide as the specimen diluent  |
|  Formulation | Low-volume, flat bottom microwells with Kogenate FS as the source of Factor VIII. Microwells are not blocked | Starwells with Recombinant as the Factor VIII source. Microwells are blocked with bovine serum albumin  |
|  Control | Kit control used to determine positive samples | Negative Control used to determine cutoff for positive samples  |
|  Sample Matrix | Plasma collected in ACD or 3.2% Na citrate, and serum | Plasma collected in ACD or 3.2% Na citrate  |

# K. Standard/Guidance Document Referenced (if applicable):

CLSI EP-5A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline.

CLSI - EP7A2, Interference testing in Clinical Chemistry; Approved Guideline

CLSI - EP12A2, User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline

CLSI - H21A5, Collection, Transport, Processing of Blood Specimens for Testing Plasma Based Coagulation Assays and Molecular Hemostasis Assays; Approved Guideline

{3}

CLSI EP9-A2; Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline

BS EN 13612:2002; Performance Evaluation of in vitro Diagnostic Medical Devices

BS EN 13640:2002; Stability Testing of in vitro Diagnostic Reagents

## L. Test Principle:

Diluted patient sample is added to microwells coated with recombinant FVIII, and antibody, if present will bind. Unbound material is washed away, and an alkaline phosphatase labeled anti-human immunoglobulin reagent (anti IgG) is added to the wells and incubated. Unbound anti-IgG is washed away and then PNPP (p-nitrophenyl phosphate) substrate is added. After a 30 minute incubation period, the reaction is stopped, and the optical density of the color that develops is measured at 405nm.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

a. Precision/Reproducibility:

8 samples were tested in duplicate in 10 separate assays, and analyzed by ANOVA for agreement within and between runs. Results demonstrated a within run imprecision of ≤20% CV and a between run CV of ≤24%.

Lot-to Lot Reproducibility was demonstrated by testing 12 samples (7 neg, 5 pos) in duplicate in 3 separate assays, on 3 different kit lots. Results demonstrated 100% agreement between reportable results for all 3 kit lots.

b. Linearity/assay reportable range:

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

d. Detection limit:

e. Analytical specificity:

3 samples (negative, medium, and high FVIII antibody reactivity) were used

{4}

to demonstrate interference of hemoglobin, bilirubin, lipid, Gammagaurd (IVIG) and Rituxan (rituximab). Samples were spiked with interferents and tested in replicates of 10. Results showed no interference in hemoglobin and lipemia at 500mg/dL,bilirubin at 10 mg/dL, , IVIG 200mg.dL, Rituxan 10μg/mL.

f. Assay cut-off:

The cutoff value for the Factor VIII Antibody Screen is determined by the kit control. Any sample with an average OD value &gt; than the average OD value of the kit control is positive. Any sample with an average OD value ≤ the average OD value of the kit control is negative.

2. Comparison studies:

a. Method comparison with predicate device:

Accuracy was demonstrated by a 2 site study in which the Factor VIII Antibody Screen was compared to the Factor VIII Inhibitor Assay and the Bethesda Assay.

Factor VIII Inhibitor Assay

|  Factor VIII | Factor VIII Inhibitor Assay  |   |   |
| --- | --- | --- | --- |
|   |  Pos | Neg | Total  |
|  Antibody Screen | Pos | 89 | 1  |
|   | Neg | 7 | 40  |
|   | Total | 96 | 41  |

b. Matrix comparison:

|  Factor VIII | Bethesda Assay  |   |   |
| --- | --- | --- | --- |
|   |  Pos | Neg | Total  |
|  Antibody Screen | Pos | 35 | 4  |
|   | Neg | 2 | 40  |
|   | Total | 37 | 44  |

Normal and spiked sample studies were conducted. Paired 3.2% Na citrate and serum samples was collected from 59 normal donors. Serum, 3.2% Na citrate, and ACD plasma samples were collected form 14 normal donors.

{5}

Antibody positive plasma was spiked in 12 serum and 3.2% Na citrate samples. 24 ACD plasma and 3.2% sodium citrate pairs were spiked with FVIII antibody positive plasma. All studies demonstrated no difference was observed between the reportable results obtained from samples collected as 3.2% sodium citrate plasma, ACD Plasma or as serum.

3. Clinical studies:
a. Clinical Sensitivity:
b. Clinical specificity:
c. Other clinical supportive data (when a. and b. are not applicable):

4. Clinical cut-off:

5. Expected values/Reference range:

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

6

---

**Source:** [https://fda.innolitics.com/device/K082205](https://fda.innolitics.com/device/K082205)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
