K082149 · Mego Afek AC , Ltd. · JOW · Oct 6, 2008 · Cardiovascular
Device Facts
Record ID
K082149
Device Name
LYMPHA PRESS OPTIMAL, MODEL# 1201AP
Applicant
Mego Afek AC , Ltd.
Product Code
JOW · Cardiovascular
Decision Date
Oct 6, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Lympha Press Optimal is intended for the following Indications for Use: Primary lymphedema (for example, congenital lymphedema/ milroy's disease) Secondary lymphedema (for example, post-mastectomy, chronic edema, post-traumatic edema) Venous disorders (for example, venous insufficiency, varicose veins, venous stasis ulcers) Dysfunction of the muscle pump (for example, promotion of wound recovery, reduction of edema and lower limb pain following trauma and sports injuries) The device is intended to be used by the patient at home, as well as by physicians at clinics or hospitals.
Device Story
Lympha Press Optimal Model 1201AP is a compressible limb sleeve system for lymphatic and venous therapy. Device consists of a main console with an air compressor and compression garments (sleeves) containing up to 12 overlapping pressure cells. Operation involves sequential inflation and deflation of sleeve cells to move interstitial fluid, improve circulation, and support muscle pump function. System is operated by patients at home or by clinicians in hospitals/clinics. Console uses electro-mechanical regulation, including pressure sensors and a rotating disc to control air outflow through hoses to the garment. Therapy helps treat lymphedema, venous insufficiency, and edema. Output is mechanical compression of the limb; healthcare providers use the device to manage fluid-related disorders and promote wound healing.
Clinical Evidence
Bench testing only. The device underwent extensive safety, performance, and software validation testing to ensure compliance with functional specifications and industry safety standards. No clinical trial data was provided.
Technological Characteristics
System comprises an air compressor console and multi-cell (up to 12) compression garments. Employs electro-mechanical regulation with pressure sensors and a rotating disc for air distribution. Sequential pneumatic compression principle. Software-controlled inflation/deflation cycles. Standalone device.
Indications for Use
Indicated for patients with primary lymphedema (e.g., congenital/Milroy's disease), secondary lymphedema (e.g., post-mastectomy, chronic/post-traumatic edema), venous disorders (e.g., venous insufficiency, varicose veins, venous stasis ulcers), and muscle pump dysfunction (e.g., wound recovery, edema reduction, pain relief post-trauma/sports injury). Intended for home, clinic, or hospital use.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
### SPECIAL 510(K) SUMMARY
## LYMPHA PRESS OPTIMAL MODEL 1201AP
# COMPRESSIBLE LIMB SLEEVE DEVICE
### 510(k) Number K082149
| 510(k) Number K082149 | |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant's Name: | Mego Afek AC Ltd.<br>Kibbutz Afek<br>Post Afek<br>Phone: (972)-4-8784277<br>Fax: (972)-4-8784148<br>E-mail: moshe@megoafek.co.il |
| Contact Person: | Ahava Stein/ Ofer Hornick<br>A. Stein - Regulatory Affairs Consulting<br>20 Hata'as St.<br>Kfar Saba 44425<br>Israel<br>Tel. + 972-9-7670002<br>Fax. +972-9-7668534<br>e-mail: ahava@asteinrac.com or ofer@asteinrac.com |
| Date Prepared: | July 2008 |
| Trade Name: | Lympha Press Optimal Model 1201AP Compression Therapy<br>Device |
| Device Common or Usual Names: | Compressible Limb Sleeve |
| Classification Name: | CFR Classification section 870.5800 (Product code JOW) |
| Classification: | Class II medical Device |
| Predicate Device: | The Lympha Press Optimal Model 1201AP Compression Therapy<br>Device is substantially equivalent to the following predicate device |
・
.
{1}------------------------------------------------
- Lympha Press Plus Model 1033AL (K013331) manufactured also by Mego Afek AC, Ltd (Israel). Lympha Press Plus is a compressible limb sleeve, similar to the Lympha Press Optimal Compression Therapy Device.
Device Description: Mego Afek's Lympha Press Optimal Model 1201AP Compression Therapy Device is a modification of the original Lympha Press Plus device. It utilizes a software controlled air compression pump, which sequentially inflates and deflates cells within a compression garment (sleeve) that is put around a limb. This helps to push excessive interstitial fluid in the treated limb, back into the venous and lymphatic systems; improve limb circulation; and thus treat the symptoms of a variety of lymphatic disorders, venous disorders and dysfunction of the "muscle pump". The device consists of a main console and compression garments. The main console contains an air compressor that is regulated by an electro-mechanical mechanism, including pressure sensors and a rotating disc controlling air outflow. The regulated compressed air is transferred via an air distributor through a series of hoses to the sleeve garments. In the Model 1201 device, each garment contains up to 12 overlapping pressure cells. The sleeve fits on the affected limb and can be easily adjusted to any limb size within the sleeve tolerance.
Intended Use / Indication for Use: The Lympha Press Optimal is intended for the following Indications for Use: Primary lymphedema (for example, congenital lymphedema/ milroy's disease)
Secondary lymphedema (for example, post-mastectomy, chronic edema, post-traumatic edema)
Venous disorders (for example, venous insufficiency, varicose veins, venous stasis ulcers)
{2}------------------------------------------------
Dysfunction of the muscle pump (for example, promotion of wound recovery, reduction of edema and lower limb pain following trauma and sports injuries)
The device is intended to be used by the patient at home, as well as by physicians at clinics or hospitals.
#### Performance Standards:
There are no performance standards under the Federal Food, Drug and Cosmetic Act, for a compressible limb sleeve device.
The Lympha Press Optimal Model 1201AP Compression Therapy Test Data: device has been subjected to extensive safety, performance testing, and validation before release, as required by the risk analysis performed for device modifications. Final testing of the Model 1201AP Compression Therapy device included various performance tests and software validation tests, designed to ensure that the device met all its functional specifications. Tests have been performed to ensure the device complies with industry and safety standards.
Substantial Equivalence: The Lympha Press Optimal Model 1201AP Compression Therapy device is similar to currently distributed Compression Therapy devices intended for treatment of venous and lymphatic disorders and dysfunction of the "muscle pump". The device uses sequential inflation and deflation of cells within compression sleeves put around a limb. Inflation/ Deflation pressures and sequences are similar to those of predicate devices. Operating modes are similar to those of predicate devices. All of the above features are similar to these features in the predicate devices.
{3}------------------------------------------------
### Conclusions:
:
:
The conclusions drawn from the above Performance Testing and comparison to predicate devices is that the Lympha Press Optimal Model 1201AP compression therapy device is substantially equivalent in safety and efficacy to the predicate devices listed above.
:
.
.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the eagle. The eagle is rendered in black, and the text is also in black.
OCT 0 6 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mego Afek AC LTD c/o Dr. Ofer Hornick Regulatory Affairs Consultant A. Stein Regulatory Affairs Consulting Beit Hapaamon Suite (Box 124) 20 HATAAS Str. Kfar Saba 44425 Israel
K082149 Re:
Lympha Press Optimal™, Model # 1201AP Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II (two) Product Code: JOW Dated: July 27, 2008 Received: July 30, 2008
Dear Dr. Hornick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Dr. Ofer Hornick
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely yours,
summer D. Willing
/ \ Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known): KO8 Z 149
Device Name:
Lympha Press Optimal (Model 1201AP) Compression Therapy device
Indications for Use:
- Primary lymphedema (for example, congenital/ milroy's disease) 트
- Secondary lymphedema (for example, post-mastectomy, chronic 트 edema, post-traumatic edema)
- Venous disorders (for example, venous insufficiency, varicose veins, l venous stasis ulcers)
- 트 Dysfunction of the muscle pump (for example, promotion of wound recovery, reduction of edema and lower limb pain following trauma and sports injuries)
The device is intended to be used by the patient at home, as well as by physicians at clinics or hospitals.
Prescription Use V OR Over-The-Counter Use (Per 21 C.F.R. 801 Subpart D) (Optional Format Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
9
— — — — — — — —
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nuna R. Evans
(Division Sign-Off) (Division Sign-Silvascular Devices
510(k) Number_ko 8214
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.