The T2™ Spinal System is a vertebral body replacement system intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). The T2TM components consist of a series of end caps, baskets, and covers which must be attached to a T2 XVBR™ expanding centerpiece to form a complete construct. The final construct is to be used with supplemental fixation. The T2 SCEPTOR™ components also serve as a vertebral body replacement device for the same intended use in the thoracolumbar spine. The T2 SCEPTOR™ end caps and endcleats must be attached to a PYRAMESH-C® device to form a complete construct. Both constructs (T2 XVBR™ and T2 SCEPTOR™) must be used with supplemental fixation to form a final construct. Specifically, the construct is to be used with the Medtronic ZPLATE II™ Anterior Fixation System, the DYNALOK CLASSIC® Spinal System, the VANTAGE™ Anterior Fixation System, the TSRH® Spinal System, the CD HORIZON® Spinal System, or their successors. Additionally, the T2TM Spinal System construct is intended to be used with bone graft.
Device Story
Distractible vertebral body replacement system for thoracolumbar (T1-L5) corpectomy procedures; construct inserted between vertebral bodies and expanded to stabilize spine. Components include expanding centerpiece (T2 XVBR or PYRAMESH-C), end caps, baskets, and covers. Not for standalone use; requires supplemental anterior/posterior fixation (e.g., ZPLATE-II, DYNALOK CLASSIC, VANTAGE, TSRH, CD HORIZON) and bone graft. Used by surgeons in clinical settings to restore spinal column integrity following tumor or trauma.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Construct materials: titanium alloy, cobalt chrome, nitinol. Distractible/expanding cage design with modular end caps, baskets, and covers. Mechanical stabilization device; non-electronic; non-software.
Indications for Use
Indicated for patients requiring vertebral body replacement in the thoracolumbar spine (T1-L5) due to tumor or trauma-related collapse, damage, or instability. Must be used with supplemental fixation and bone graft.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Reference Devices
ZPLATE-II™ Anterior Fixation System
DYNALOK CLASSIC® Spinal System
VANTAGE® Anterior Fixation System
TSRH® Spinal System
CD HORIZON® Spinal System
PYRAMESH-C®
Submission Summary (Full Text)
{0}------------------------------------------------
K082.112 (pg. 1 of 2)
## AUG 2 7 2008
## T2TM Spinal System 510(k) Summary July 2008
| I. | Company: | Medtronic Sofamor Danek, Inc.<br>1800 Pyramid Place<br>Memphis, Tennessee 38132<br>Telephone: (901) 396-3133<br>Fax: (901) 346-9738 |
|-----|-----------------|-------------------------------------------------------------------------------------------------------------------------------------|
| | Contact: | Melisa Lansky, M.B.A.<br>Sr. Regulatory Affairs Specialist |
| II. | Product Name: | T2T™ Spinal System |
| | Classification: | MQP |
- III. Description: The T2TM Spinal System is a distractible system used in corpectomy procedures. This construct is inserted between two vertebral bodies in the thoracic and/or lumbar spine and is expanded to aid in the surgical correction and stabilization of the spine. The construct is not intended to be used as a stand alone device. The construct is intended to be used with either anterior and/or posterior supplemental spinal fixation systems already cleared for thoracic and lumbar spine stabilization.
The T2 XVBR™ expanding cage is made of titanium alloy, cobalt chrome, and nitinol. The T2 XVBR™ end caps, baskets, and covers are attached to the T2 XVBR™ expanding cage to accommodate the individual anatomical requirements of the vertebral space created by the corpectomy.
T2™ Spinal System constructs may not be used with stainless steel supplemental fixation devices. Titanium constructs comprised from one of the following Medtronic spinal systems or their successors must be used with the T2TM Spinal System.
| | Anterior | Posterior |
|-------------------------------------|----------|-----------|
| ZPLATE-II™ Anterior Fixation System | X | |
| DYNALOK CLASSIC® Spinal System | X | X |
| VANTAGE® Anterior Fixation System | X | |
| TSRH® Spinal System | X | X |
| CD HORIZON® Spinal System | X | X |
{1}------------------------------------------------
$$\begin{bmatrix} \mathbf{z} \ \mathbf{f} \ \mathbf{z} \end{bmatrix} \text{ and } \mathbf{z} \text{ are } \mathbf{g} \times \mathbf{b}$$
- Indications for Use: The T2™ Spinal System is a vertebral body replacement system IV. intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). The T2TM components consist of a series of end caps, baskets, and covers which must be attached to a T2 XVBR™ expanding centerpiece to form a complete construct. The final construct is to be used with supplemental fixation. The T2 SCEPTOR™ components also serve as a vertebral body replacement device for the same intended use in the thoracolumbar spine. The T2 SCEPTOR™ end caps and endcleats must be attached to a PYRAMESH-C® device to form a complete construct. Both constructs (T2 XVBR™ and T2 SCEPTOR™) must be used with supplemental fixation to form a final construct. Specifically, the construct is to be used with the Medtronic ZPLATE IITM Anterior Fixation System, the DYNALOK CLASSIC® Spinal System, the VANTAGE™ Anterior Fixation System, the TSRH® Spinal System, the CD HORIZON® Spinal System, or their successors. Additionally, the T2™ Spinal System construct is intended to be used with bone graft.
- Substantial Equivalence: Documentation was provided which demonstrated that the V. T2TM Spinal System components are substantially equivalent to previously cleared vertebral body replacement devices, including itself.
C2008 Medtronic Sofamor Danek
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its head turned to the left and its wings outstretched. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle. The text is arranged so that it follows the curve of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 7 2008
Medtronic Sofamor Danek % Ms. Melisa Lansky, Senior Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132
Re: K082112
> Trade Name: T2 Spinal System Regulation Number(s): 21 CFR 888.3060 Regulation Names: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: MQP Dated: July 23, 2008 Received: July 28, 2008
Dear Ms. Lansky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Melisa Lansky
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):
4082112
Device Name: J T2TM Spinal System
## Indications for Use:
The T2™ Spinal System is a vertebral body replacement system intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). The T2TM components consist of a series of end caps, baskets, and covers which must be attached to a T2 XVBR™ expanding centerpiece to form a complete construct. The final construct is to be used with supplemental fixation. The T2 SCEPTOR™ components also serve as a vertebral body replacement device for the same intended use in the thoracolumbar spine. The T2 SCEPTOR™ end caps and endcleats must be attached to a PYRAMESH-C® device to form a complete construct. Both constructs (T2 XVBR™ and T2 SCEPTOR™) must be used with supplemental fixation to form a final construct. Specifically, the construct is to be used with the Medtronic ZPLATE II™ Anterior Fixation System, the DYNALOK CLASSIC® Spinal System, the VANTAGE™ Anterior Fixation System, the TSRH® Spinal System, the CD HORIZON® Spinal System, or their successors. Additionally, the T2TM Spinal System construct is intended to be used with bone graft.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Darlene Muelm
estorative, Division of Geners and Neurological Devices
510(k) Number.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.