K082050 · DiaSorin, Inc. · LOL · Nov 3, 2008 · Microbiology
Device Facts
Record ID
K082050
Device Name
LIAISON HAV IGM ASSAY, LIAISON CONTROL HAV IGM
Applicant
DiaSorin, Inc.
Product Code
LOL · Microbiology
Decision Date
Nov 3, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3310
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K082050 · Nov 3, 2008
LIAISON HAV IGM ASSAY, LIAISON CONTROL HAV IGM
DiaSorin, Inc.
Clinical laboratory samples from patients undergoing routine HAV testing; Retrospective clinical samples from patients with documented acute Hepatitis A infection
The clinical performance of the LIAISON® HAV IgM assay was evaluated using both prospective and retrospective clinical samples to demonstrate diagnostic agreement with a predicate ELISA method.
Individuals with documented acute Hepatitis A infection; Sample Size: 123
ETI-HA-IGMK Plus (ELISA)
Percent agreement with comparator
Prospective clinical study
Individuals sent to the laboratory for Hepatitis A testing; Sample Size: 500
ETI-HA-IGMK Plus (ELISA)
Negative percent agreement
Indications for Use
The LIAISON® HAV IgM assay is an in vitro chemiluminescent immunoassay intended for the qualitative detection of IgM antibodies to hepatitis A virus (IgM anti-HAV) in human serum and sodium heparin plasma using the LIAISON® Analyzer. Assay results, in conjunction with other serological and clinical information, may be used for testing specimens from individuals who have signs and symptoms consistent with acute hepatitis as an aid in the laboratory diagnosis of acute or recent HAV infection. This assay is not intended for screening blood or solid or soft tissue donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients. The user is responsible for establishing their own assay performance characteristics in these populations. The LIAISON® Control HAV IgM (negative and Reactive) is intended for use as assayed quality control samples to monitor the performance of the LIAISON® HAV IgM assay.
Device Story
LIAISON® HAV IgM is an in vitro diagnostic chemiluminescent immunoassay (CLIA) for detecting IgM anti-HAV in serum or heparinized plasma. Performed on LIAISON® Analyzer; uses magnetic particles coated with mouse monoclonal IgG to human IgM. Process: 1) capture of patient IgM on magnetic beads; 2) addition of HAV antigen and isoluminol-labeled mouse monoclonal anti-HAV conjugate; 3) wash cycles to remove unbound material; 4) induction of flash chemiluminescence via starter reagents. Photomultiplier measures relative light units (RLU) proportional to IgM anti-HAV concentration. Used in clinical laboratories; operated by trained personnel. Output provides qualitative results (Index Value) to assist physicians in diagnosing acute or recent Hepatitis A infection. Benefits include automated processing and faster turnaround compared to manual ELISA methods.
Clinical Evidence
Clinical performance evaluated via prospective and retrospective studies. Prospective study (n=739) of HAV testing and at-risk populations showed 99.9% negative percent agreement. Pediatric study (n=108) showed 100% negative percent agreement. Retrospective study (n=123) of acute HAV infection cases showed 96.7% positive percent agreement. Analytical studies included precision/reproducibility (5-day and 20-day studies), cross-reactivity (195 samples, no cross-reactivity observed), and interference testing (hemolysis, lipemia, icterus, etc.).
Technological Characteristics
CLIA assay; magnetic particles coated with mouse monoclonal IgG to human IgM; isoluminol-antibody conjugate; flash chemiluminescence detection via photomultiplier. Automated processing on LIAISON® Analyzer. Sample volume 20 μL; total incubation 40 minutes. Storage 2-8°C.
Indications for Use
Indicated for qualitative detection of IgM anti-HAV in human serum and sodium heparin plasma for individuals with signs/symptoms of acute hepatitis to aid in diagnosis of acute or recent HAV infection. Not for donor screening. Performance not established for immunocompromised/immunosuppressed patients.
Regulatory Classification
Identification
HAV serological assays are devices that consist of antigens and antisera for the detection of hepatitis A virus-specific IgM, IgG, or total antibodies (IgM and IgG), in human serum or plasma. These devices are used for testing specimens from individuals who have signs and symptoms consistent with acute hepatitis to determine if an individual has been previously infected with HAV, or as an aid to identify HAV-susceptible individuals. The detection of these antibodies aids in the clinical laboratory diagnosis of an acute or past infection by HAV in conjunction with other clinical laboratory findings. These devices are not intended for screening blood or solid or soft tissue donors.
Special Controls
*Classification.* Class II (special controls). The special control is “Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Hepatitis A Virus Serological Assays.” See § 866.1(e) for the availability of this guidance document.
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number: K082050
B. Purpose for Submission: Initial Premarket Notification
C. Measurand: IgM Antibody to Hepatitis A Virus in serum and heparinized plasma
D. Type of Test: Qualitative, Chemiluminescent Immunoassay (CLIA)
E. Applicant: Diasorin, Inc.
F. Proprietary and Established Names: LIAISON® HAV IgM
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LOL : Hepatitis A Test (IgM Antibody) | Class II | 866.3310 | Microbiology (83) |
## H. Intended Use:
The LIAISON® HAV IgM assay is an in vitro chemiluminescent immunoassay intended for the qualitative detection of IgM antibodies to hepatitis A virus (IgM anti-HAV) in human serum and heparinized plasma using the LIAISON® Analyzer. Assay results, in conjunction with other serological and clinical information, may be used for testing specimens from individuals who have signs and symptoms consistent with acute hepatitis as an aid in the laboratory diagnosis of acute or recent HAV infection.
This assay is not intended for screening blood or solid or soft tissue donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients. The user is responsible for establishing assay performance characteristics in these populations.
Caution: U.S. Federal Law restricts this device to sale by or on the order of a physician.
3) Special conditions for use statement(s):
For Prescription use only
4) Special instrument requirements: LIAISON® Analyzer with Chemiluminescence reader
{2}
I. Device Description: The LIAISON® HAV IgM assay is an in vitro diagnostic device that consists of magnetic particles, calibrators, specimen diluent, conjugate and HAV antigen. The kit contains 100 tests. Additionally, controls are required to run the assay. The LIAISON® HAV IgM immunoassay is performed on the LIAISON® Analyzer (Model 15970) that was previously cleared (K062473, K071480). Magnetic particles are beads coated with mouse monoclonal antibodies (IgG) to human IgM. During first incubation, the magnetic particles are incubated with serum or heparinized plasma to capture IgM antibodies to Hepatitis A Virus. During the second incubation, the antibody conjugate reacts with HAV antigen just added and the immune complex thus formed reacts with IgM already bound to the solid phase. The unbound material is removed with a wash cycle after each incubation. Subsequently, a flash chemiluminescence reaction is induced by adding the starter reagents. The light signal, and hence the amount of isoluminol-antibody conjugate, is measured by a photomultiplier as relative light units (RLU) that is indicative of IgM anti-HAV present in the samples.
J. Substantial Equivalence Information:
a) Predicate device name (s):
PMA: Diasorin Inc. Enzyme Immunoassay for detection of IgM Antibody to Hepatitis A Virus (IgM anti-HAV) in human serum or plasma (ETI-HA-IGMK Plus)
b) Predicate Numbers (s):
P890014/S02
Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | LIAISON® HAV IgM | ETI-HA-IGMK PLUS PMA No. P890014/S02 |
| Intended Use | The LIAISON® HAV IgM assay is an in vitro chemiluminescent immunoassay intended for the qualitative detection of IgM antibodies to hepatitis A virus (IgM anti-HAV) in human serum and heparinized plasma using the LIAISON® Analyzer. Assay results, in conjunction with other serological and clinical information, may be used for | ETI-HA-IGMK PLUS is an in vitro enzyme immunoassay intended for use in the qualitative determination of IgM antibody to hepatitis A virus (IgM antiHAV) in human serum or plasma. |
{3}
3
| | testing specimens from individuals who have signs and symptoms consistent with acute hepatitis as an aid in the laboratory diagnosis of acute or recent HAV infection. | |
| --- | --- | --- |
| Controls | 2 (Negative and Positive) | 2 (Negative and Positive) |
| Sample Matrix | Serum or heparinized plasma | Serum or plasma |
| Reagent Storage | 2-8°C, On-board or in Refrigerator | 2-8°C Refrigerator only |
| Differences | | |
| Item | LIAISON® HAV IgM | ETI-HA-IGMK PLUS PMA No. P890014/S02 |
| Type of Assay | Chemiluminescence Immunoassay | Enzyme Immunoassay |
| Sample Handling/processing | Automated | Manual |
| Cutoff | Index Value 1.0 | Cutoff = Calibrator x absorbance + 0.250 |
| Calibrators | Two | One |
| Detector | Mouse monoclonal antibody to HAV conjugated to an isoluminol derivative | Horseradish peroxidase-labeled mouse monoclonal antibody to HAV |
| Capture Reagent | Magnetic particles coated with IgG to human IgM (mouse monoclonal) | Microwells coated with mouse monoclonal antibody to human IgM |
| Equivocal zone | Index Value of >0.90 and <1.1 | Absorbance within 80 – 100% of assay cutoff |
| Sample Volume | 20 μL | 10 μL |
| Measurement System | Photomultiplier (flash chemiluminescence reader) | Spectrophotometer (EIA Microtiter plate reader) |
| Total incubation | 40 minutes | 3.5 hours (210 minutes) |
K. Standard/Guidance Document Referenced (if applicable): NA
1. Class II Special Controls Guidance Document: Hepatitis A Virus Serological Assays, Feb. 9, 2006. http://www.fda.gov/cdrh/ode/guidance/1536.pdf
2. Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005. http://www.fda.gov/cdrh/ode/guidance/337.pdf
L. Test Principle: The method for qualitative determination of IgM anti-HAV is an antibody capture chemiluminescence immunoassay (CLIA). IgG to human IgM (mouse monoclonal) is used for coating magnetic particles (solid phase). A mouse monoclonal antibody to HAV is linked to an isoluminol derivative (isoluminol-
{4}
antibody conjugate). During the first incubation, IgM antibodies present in calibrators, samples or controls bind to the solid phase. During the second incubation, the antibody conjugate reacts with HAV antigen and the immune complex thus formed reacts with the IgM already bound to the solid phase. After each incubation, the unbound material is removed with a wash cycle. Subsequently, the starter reagents are added and a flash chemiluminescence reaction is thus induced. The light signal, and hence the amount of isoluminol-antibody conjugate, is measured by a photomultiplier as relative light units (RLU) and is indicative of IgM anti-HAV present in the calibrators, samples or controls.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
a. Precision/Reproducibility: A five-day reproducibility study was conducted at three external laboratories. A coded panel comprised of 12 frozen contrived serum samples was prepared by DiaSorin S.p.A. and provided to the sites. The coded panel samples were prepared by spiking reactive samples into negative samples to achieve high negative, low reactive and high reactive results. The two negative panel samples were not spiked. The coded panel was tested at all three sites, one run per day using four replicates per run for five operating days. The mean Index value and coefficient of variation (%CV) of the results were computed for each of the tested specimens for each of the sites.
| Sample | | N | Mean | Within run | Between runs | Total (by site) | Between sites | Overall | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| ID# | Matrix | | Index | %CV | %CV | %CV | %CV | SD | %CV |
| NC | Buffer | 60 | 0.15 | 14.2 | 15.6 | 19.6 | 63.6 | 0.08 | 54.2 |
| PC | Buffer | 60 | 2.17 | 5.2 | 4.7 | 6.5 | 10.3 | 0.23 | 10.7 |
| HAMu-el | Serum | 60 | 0.73 | 4.4 | 9.1 | 9.7 | 24.9 | 0.17 | 23.5 |
| HAMu-e2 | Serum | 60 | 0.81 | 6.9 | 9.9 | 11.6 | 23.2 | 0.18 | 22.6 |
| HAMu-n1 | Serum | 60 | 0.49 | 4.9 | 7.8 | 9.1 | 23.8 | 0.11 | 22.7 |
| HAMu-n2 | Serum | 60 | 0.42 | 4.4 | 7.0 | 8.0 | 23.3 | 0.09 | 21.2 |
| HAMu-P1 | Serum | 60 | 6.87 | 5.2 | 7.5 | 9.2 | 10.3 | 0.87 | 12.7 |
| HAMu-P2 | Serum | 60 | 4.48 | 6.7 | 9.7 | 11.3 | 15.8 | 0.78 | 17.4 |
| HAMu-P3 | Serum | 60 | 2.45 | 5.3 | 6.7 | 9.5 | 19.4 | 0.46 | 18.8 |
| HAMu-P4 | Serum | 60 | 2.17 | 4.1 | 7.9 | 8.4 | 14.6 | 0.34 | 15.5 |
| HAMu-P5 | Serum | 60 | 1.95 | 4.0 | 8.7 | 8.9 | 10.4 | 0.25 | 12.6 |
| HAMu-P6 | Serum | 60 | 1.53 | 4.0 | 7.1 | 7.6 | 18.9 | 0.27 | 17.4 |
| HAMu-P7 | Serum | 60 | 1.31 | 4.8 | 9.5 | 10.0 | 16.2 | 0.22 | 16.8 |
| HAMu-P8 | Serum | 60 | 1.24 | 5.6 | 12.6 | 12.8 | 22.4 | 0.28 | 22.5 |
{5}
A twenty day precision study was performed at DiaSorin Inc. The same coded panel tested in the 5 day study, was also used in the 20 day study. The mean, standard deviation, and coefficient of variation (%CV) of the results were computed for each of the tested specimens.
| Panel ID | N | Mean Index | Within run | | Between day | | Between runs | | Total Index | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| NC | 80 | <<0.10 | NA | NA | NA | NA | NA | NA | NA | NA |
| PC | 80 | 2.20 | 0.148 | 6.9 | 0.173 | 7.9 | 0.034 | 1.6 | 0.214 | 9.8 |
| HAMu-el | 80 | 0.77 | 0.028 | 3.6 | 0.093 | 12.0 | 0.005 | 0.7 | 0.096 | 12.4 |
| HAMu-e2 | 80 | 0.87 | 0.039 | 4.6 | 0.078 | 8.9 | 0.006 | 0.7 | 0.086 | 9.9 |
| HAMu-n1 | 80 | 0.46 | 0.014 | 3.1 | 0.041 | 9.1 | 0.002 | 0.4 | 0.043 | 9.5 |
| HAMu-n2 | 80 | 0.41 | 0.015 | 3.6 | 0.091 | 21.5 | 0.079 | 19.3 | 0.108 | 26.4 |
| HAMu-P1 | 80 | 7.08 | 0.242 | 3.5 | 0.764 | 10.8 | 0.019 | 0.3 | 0.795 | 11.2 |
| HAMu-P2 | 80 | 4.65 | 0.194 | 4.5 | 0.385 | 8.3 | 0.058 | 1.2 | 0.435 | 9.4 |
| HAMu-P3 | 80 | 2.73 | 0.074 | 2.7 | 0.221 | 8.1 | 0.015 | 0.5 | 0.228 | 8.4 |
| HAMu-P4 | 80 | 2.39 | 0.080 | 3.3 | 0.172 | 7.2 | 0.008 | 0.4 | 0.185 | 7.7 |
| HAMu-P5 | 80 | 2.01 | 0.062 | 3.1 | 0.13 | 6.5 | 0.019 | 0.9 | 0.144 | 7.1 |
| HAMu-P6 | 80 | 1.68 | 0.063 | 3.8 | 0.111 | 6.6 | 0.017 | 1.0 | 0.123 | 7.3 |
| HAMu-P7 | 80 | 1.40 | 0.053 | 3.8 | 0.117 | 8.4 | 0.013 | 1.0 | 0.126 | 9.0 |
| HAMu-P8 | 80 | 1.30 | 0.065 | 5.3 | 0.122 | 9.4 | 0.014 | 1.1 | 0.14 | 10.8 |
b. Linearity/assay reportable range: NA
c. Traceability, Stability, Expected values (controls, calibrators, or methods): NA
d. Detection limit: This device is for the qualitative detection of HAV- IgM antibody and the limit of detection was not determined.
e. Analytical sensitivity:
{6}
6
# 1. Analytical Sensitivity as Seroconversion Panel Performance
Five commercially available HAV seroconversion panels were tested using LIAISON® HAV IgM and the FDA approved comparator assay to determine when is to detect IgM antibodies after infection and how long IgM can be detected after infection. The results are summarized in the following table:
| Panel ID | DiaSorin LIAISON HAV IgM* | | Comparator Assay* | | Difference in Days from Last Reactive Result |
| --- | --- | --- | --- | --- | --- |
| | Post Bleed Day of Earliest Reactive Result | Post Bleed Day of Last Reactive Result | Post Bleed Day of Earliest Reactive Result | Post Bleed Day of Last Reactive Result | |
| PHT901 seroconversion | 12 | 17 | 12 | 17 | 0 |
| PHT902 seroconversion | 16 | 21 | 16 | 21 | 0 |
| RP004 seroconversion | 6 | 62 | 6 | 62 | 0 |
| RP013 seroconversion | 8 | 189 | 8 | 189 | 0 |
| HAV01 seroconversion | 0 | 77 | 0 | 91 | 14 |
* Only reactive results were used; equivocal results were not used to determine a reactive result.
The time point of the serial bleed when HAV IgM was first detected and the bleed when it was no longer detected by the LIAISON® HAV IgM was equivalent to the comparator assay in four out of five seroconversion panels. In one panel out of five the last day of detection between the two assays differed by 14 days.
2. Cross Reactivity: The cross-reactivity study for the LIAISON® HAV IgM assay was designed to evaluate potential interference from other viruses that may cause symptoms similar to HAV infection (EBV, CMV, Rubella, Measles, Mumps, HBV, HCV), other organisms that may cause infectious disease (VZV, HSV, HIV, Toxoplasma gondii) and from other conditions that may result from atypical immune system activity (i.e. rheumatoid factor, RF, antinuclear autoantibodies, ANA, human anti-mouse antibodies). A total of 195 samples were tested and no cross reactivity was observed. Results are presented in the following table.
{7}
| Organism/Condition | N | Comparator HAV IgM Assay | LIAISON® HAV IgM Reactive | LIAISON® HAV IgM Negative | LIAISON® HAV IgM Equivocal |
| --- | --- | --- | --- | --- | --- |
| IgG anti-Measles | 3 | Negative | 0 | 3 | 0 |
| IgG anti-Mumps | 8 | Negative | 0 | 8 | 0 |
| IgG anti-VCA | 3 | Negative | 0 | 3 | 0 |
| IgG anti-EA | 3 | Negative | 0 | 3 | 0 |
| IgG anti-CMV | 3 | Negative | 0 | 3 | 0 |
| IgG anti-Rubella | 2 | Negative | 0 | 2 | 0 |
| IgG anti-Toxoplasma | 3 | Negative | 0 | 3 | 0 |
| IgG anti-HSV-1/2 | 1 | Negative | 0 | 1 | 0 |
| IgG anti-HSV-2 | 6 | Negative | 0 | 6 | 0 |
| IgG anti-syphilis | 4 | Negative | 0 | 4 | 0 |
| Anti-VZV | 3 | Negative | 0 | 3 | 0 |
| Anti-HTLV I/II | 3 | Negative | 0 | 3 | 0 |
| Anti-HCV | 4 | Negative | 0 | 4 | 0 |
| Anti-Borrelia | 4 | Negative | 0 | 4 | 0 |
| Anti-HBs | 3 | Negative | 0 | 3 | 0 |
| Anti-HIV | 10 | Negative | 0 | 10 | 0 |
| Anti-Parvovirus B19 | 4 | Negative | 0 | 4 | 0 |
| IgM anti-HBc | 4 | Negative | 0 | 4 | 0 |
| IgM anti-Borrelia | 5 | Negative | 0 | 5 | 0 |
| IgM anti-CMV | 5 | Negative | 0 | 5 | 0 |
| IgM anti-EBV | 5 | Negative | 0 | 5 | 0 |
| IgM anti-HSV | 7 | Negative | 0 | 7 | 0 |
| IgM anti-Rubella | 6 | Negative | 0 | 6 | 0 |
| IgM anti-Toxoplasma | 5 | Negative | 0 | 5 | 0 |
| IgM anti-VZV | 6 | Negative | 0 | 6 | 0 |
| Anti-Influenza virus | 3 | Negative | 0 | 3 | 0 |
| HBsAg | 3 | Negative | 0 | 3 | 0 |
| HBeAg | 6 | Negative | 0 | 6 | 0 |
| Nucleotides | 4 | Negative | 0 | 4 | 0 |
| ENA | 4 | Negative | 0 | 4 | 0 |
| Rheumatoid Factor | 17 | Negative | 0 | 17 | 0 |
| γ-globulin | 36 | Negative | 0 | 36 | 0 |
| HAMA | 12 | Negative | 0 | 12 | 0 |
| Total | 195 | | 0 | 195 | 0 |
{8}
3. Interference: Controlled studies were performed to determine whether the presence of hemoglobin, lipemia, bilirubin, serum albumin and gamma globulin affect assay performance. The highest concentrations which were considered not to impact results are as follows: hemolysis (at 1000 mg/dL hemoglobin), lipemia (at 3000 mg/dL triglycerides), icterus (at 20 mg/dL bilirubin), serum albumin (at 5 g/dL), γ-Globulin (at 4 g/dL).
f. Assay cut-off: NA
2. Comparison studies:
a. Method comparison with reference method:
Diasorin Inc. Enzyme Immunoassay for detection of IgM Antibody to Hepatitis A Virus (IgM anti-HAV) in human serum or plasma (ETI-HA-IGMK Plus)
b. Matrix comparison: Fifty sample sets of matched serum/plasma were used in the study to evaluate the risk of a potential nonspecific reaction related to the use of heparinized plasma. A graph of the serum and corresponding heparinized plasma result for each borderline sample is presented below. The samples are grouped by "Negative", "Equivocal", and "Low Reactive" category, as established by the serum result. The graph confirms that the heparinized plasma results are similar to the serum results.

{9}
9
3. Clinical studies:
a. Clinical Sensitivity: NA
b. Clinical specificity: NA
Performance Characteristics
Summary of Clinical Studies
Prospective and Retrospective studies were performed to evaluate the performance of the LIAISON® HAV IgM assay using samples sent to the testing laboratory for Hepatitis A or other testing, from patients which were at high risk for viral hepatitis, showed signs and symptoms of hepatitis infection, or were receiving an HAV vaccination.
The prospective studies consisted of 500 samples sent to the laboratory of Hepatitis A testing, 239 samples from individuals at risk for viral hepatitis, and 150 samples from pediatric patients.
The retrospective study consisted of 123 samples from individuals with documented acute Hepatitis A infection.
Prospective
Individuals sent to the Lab for Hepatitis A testing (HAV testing population)
A total of 500 samples collected from the Northeastern U.S. were included in this study. Of the samples from individuals sent to the lab for HAV testing, 59.8% were female (n=299) ranging in age from 20 - 101 yrs. and 40.2% were male (n=201) ranging in age from 17 to 89.
Individuals At Risk for Viral Hepatitis
A total of 239 individuals at risk for viral hepatitis due to lifestyle, behavior or occupation were included in this study. The 239 individuals were from the following at risk groups: homosexual males (n=38), healthcare workers (n=10), commercial sex workers (n=34), drug users (n=77), prison inmates (n=49), dialysis patients (n=25) and hemophiliacs (n=6). Of the at risk individuals, 29.7% were females (n=71), ranging in age from 17 to 79, and 43.1% were males (n=103) ranging in age from 16 to 79. The age and gender were unknown for the remaining 27.2% (n=65).
The data for the combined prospective populations are shown in the following table.
{10}
10
# Comparison of the HAV Testing Population and the At Risk Population with the LIAISON® HAV IgM and the Comparator ELISA
| LIAISON®
HAV IgM | Comparator ELISA | | | Total |
| --- | --- | --- | --- | --- |
| | Reactive | Borderline | Negative | |
| Reactive | 0 | 0 | 1 | 1 |
| Equivocal | 0 | 0 | 0 | 0 |
| Negative | 0 | 0 | 738 | 738 |
| Total | 0 | 0 | 739 | 739 |
| | # of samples | Negative Percent Agreement | Exact 95% Confidence Intervals |
| --- | --- | --- | --- |
| Negative | 738/739 | 99.9% | 99.4 – 100% |
# Pediatric Population
One hundred and eight (108) prospectively collected pediatric samples were tested. The 108 pediatric samples were collected from children in the United States. Of these 108 samples 57.4% were females (n=62) and 42.6% were males (n=46), ranging in age from 2 to 17.
The table below shows the comparison of the LIAISON® HAV IgM and the Comparator ELISA for the 108 samples from the “pediatric” prospective population.
# Comparison of the Pediatric Population with the LIAISON® HAV IgM and the Comparator ELISA
| LIAISON®
HAV IgM | Comparator ELISA | | | Total |
| --- | --- | --- | --- | --- |
| | Reactive | Borderline | Negative | |
| Reactive | 0 | 0 | 0 | 0 |
| Equivocal | 0 | 0 | 0 | 0 |
| Negative | 0 | 0 | 108 | 108 |
| Total | 0 | 0 | 108 | 108 |
| | # of samples | Negative Percent Agreement | Exact 95% Confidence Intervals |
| --- | --- | --- | --- |
| Negative | 108/108 | 100% | 97.3 – 100% |
{11}
11
# Acute HAV Infection (Retrospective)
The retrospective population consisted of 123 samples from individuals who had an (Acute) HAV infection. Of these 123 samples, 32.5% females (n=40) and 51.2% males (n=63) ranging in age from 4 to 51. For 15.5% of the samples (n=19) gender and age were unknown. One sample (0.8%) was age 18, but gender was unknown. Of the 123 samples 42 were acutely infected pediatric patients
The 42 pediatric samples that had an acute infection were collected from children in Egypt. Of the pediatric samples, 30.9% were females (n=13) and 69.1% were males (n=29) all ranging in age from 4 to 17.
## Comparison of the Acute HAV Infection Population with the LIAISON® HAV IgM and the Comparator ELISA
| LIAISON® HAV IgM | Comparator ELISA | | | Total |
| --- | --- | --- | --- | --- |
| | Reactive | Borderline | Negative | |
| Reactive | 119 | 4 | 0 | 123 |
| Equivocal | 0 | 0 | 0 | 0 |
| Negative | 0 | 0 | 0 | 0 |
| Total | 119 | 4 | 0 | 123 |
| | # of samples | Percent Agreement | Exact 95% Confidence Interval |
| --- | --- | --- | --- |
| Reactive | 119/123 | 96.7% | 92.7 – 98.9% |
## 4. Clinical cut-off: NA
## 5. Expected values:
The expected prevalence results of the LIAISON® HAV IgM assay were determined in 802 apparently healthy adults from the Western (historically high prevalence) and the Eastern (historically lower prevalence) regions of the U.S. Three hundred and one (301) samples were from the Western U.S. and 501 were samples from the Eastern U.S.
Of the Western U.S. individuals 53.8% were females (n=162) ranging in age from 9 to 87 and 46.2% were males (n=139) ranging in age from 16 to 76. The majority of the individuals were Caucasian (60.8%), with other ethnic groups were represented as follows: Hispanic (17.6%), African Americans (15.3%), Asian (6.0%) and Middle Eastern (0.3%). In the study group from the Western region, none of the individuals were found to be reactive for HAV IgM antibodies.
Of the Eastern U.S. individuals 46.5% were females (n=233) ranging in age from 17 to 83, and 53.5% were males (n=268) ranging in age from 17 to 82. The majority of the individuals were Caucasian (69.9%), with other ethnic groups represented as follows: Hispanic (14.0%), African American (12.1%) and Asian (4.0%). In the study group from the Eastern region none of the individuals were found to be reactive for HAV IgM antibodies. The expected results for the Western
{12}
and Eastern regions of the U.S. are presented in the tables below.
Expected Results for the LIAISON® HAV IgM Assay from the Western U.S. Population (n=301)
| | N | Negative | Equivocal | Reactive | Reactive Prevalence |
| --- | --- | --- | --- | --- | --- |
| Total | 301 | 301 | 0 | 0 | NA |
| Gender | | | | | |
| Female | 162 | 162 | 0 | 0 | NA |
| Male | 139 | 139 | 0 | 0 | NA |
| Age range (years) | N | (-) | (E) | (+) | |
| ≤18 | 12 | 12 | 0 | 0 | NA |
| <10 | 1 | 1 | 0 | 0 | NA |
| 10 - 19 | 15 | 15 | 0 | 0 | NA |
| 20 - 29 | 81 | 81 | 0 | 0 | NA |
| 30 - 39 | 68 | 68 | 0 | 0 | NA |
| 40 - 49 | 52 | 52 | 0 | 0 | NA |
| 50 - 59 | 48 | 48 | 0 | 0 | NA |
| 60 - 69 | 31 | 31 | 0 | 0 | NA |
| ≥70 | 5 | 5 | 0 | 0 | NA |
Expected Results for the LIAISON® HAV IgM Assay from the East U.S. Population (n=501)
| | N | Negative | Equivocal | Reactive | Reactive Prevalence |
| --- | --- | --- | --- | --- | --- |
| Total | 501 | 501 | 0 | 0 | NA |
| Gender | | | | | |
| Female | 233 | 233 | 1 | 44 | NA |
| Male | 268 | 268 | 0 | 56 | NA |
| Age range (years) | N | (-) | (E) | (+) | |
| ≤18 | 46 | | | | |
| <10 | 0 | | | | |
| 10 - 19 | 49 | 49 | 0 | 0 | NA |
| 20 - 29 | 39 | 39 | 0 | 0 | NA |
| 30 - 39 | 78 | 78 | 0 | 0 | NA |
| 40 - 49 | 107 | 107 | 0 | 0 | NA |
| 50 - 59 | 142 | 142 | 0 | 0 | NA |
| 60 - 69 | 52 | 52 | 0 | 0 | NA |
| ≥70 | 34 | 34 | 0 | 0 | NA |
{13}
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. The LIAISON® HAV IgM Assay including Calibrators and LIAISON® Control HAV IgM (Part numbers: 310210, 310211) on the LIAISON® Analyzer are substantially equivalent to the ETI-HA-IGMK Plus) for the qualitative determination of IgM antibodies to Hepatitis A Virus in serum and heparinized plasma.
13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.