AMS SPECTRA CONCEALABLE PENILE PROSTHESIS
K082006 · American Medical Systems, Inc. · FAE · Oct 20, 2008 · Gastroenterology, Urology
Device Facts
| Record ID | K082006 |
| Device Name | AMS SPECTRA CONCEALABLE PENILE PROSTHESIS |
| Applicant | American Medical Systems, Inc. |
| Product Code | FAE · Gastroenterology, Urology |
| Decision Date | Oct 20, 2008 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.3630 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The AMS Spectra™ Concealable Penile Prosthesis is a sterile, nonpyrogenic, single-use implant that is intended for use in the treatment of chronic, organic, erectile dysfunction (impotence) in men who are determined to be suitable candidates for implantation surgery.
Device Story
AMS Spectra™ is a sterile, single-use, concealable penile prosthesis; surgically implanted into corpora cavernosa to provide penile rigidity. Device consists of two cylinders featuring articulating polymer segments held by internal cable and spring assemblies; exterior surface made of silicone. Patient manually positions device for concealment or intercourse; friction/rigidity provided by internal mechanical assembly. Rear tip extenders (RTEs) allow length adjustment. Used in clinical setting by physicians. Benefits include restoration of penile rigidity for patients with organic erectile dysfunction.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including cycle life, springback angle, column strength, bend force, bond strength, packaging, sterilization, biocompatibility, and cadaveric evaluation.
Technological Characteristics
Implantable penile prosthesis; silicone exterior; articulating polymer segments; internal cable/spring assembly. Available in 9.5-, 12-, and 14-mm diameters; 12-, 16-, and 20-cm lengths. Adjustable via rear tip extenders (RTEs). Mechanical operation; no energy source. Sterile, single-use.
Indications for Use
Indicated for men with chronic, organic erectile dysfunction (impotence) who are candidates for implantation surgery.
Regulatory Classification
Identification
A penile rigidity implant is a device that consists of a pair of semi-rigid rods implanted in the corpora cavernosa of the penis to provide rigidity. It is intended to be used in men diagnosed as having erectile dysfunction.
Special Controls
The special control for this device is the FDA guidance entitled “Guidance for the Content of Premarket Notifications for Penile Rigidity Implants.”
Predicate Devices
- DURA II® Penile Prosthesis (K953640)
- AMS 600MTM Malleable Penile Prosthesis (K912935)
- AMS 650TM Malleable Penile Prosthesis (K951716)
Reference Devices
- AMS 700™ CXR (D970012)
- AMS 700™ CX/LGX (D970012)
Related Devices
- K090663 — AMS SPECTRA CONCEALABLE PENILE PROSTHESIS · American Medical Systems, Inc. · Apr 9, 2009
- K181673 — Rigi10 Malleable Penile Prosthesis · Rigicon, Inc. · Apr 15, 2019
- K040959 — MENTOR GENESIS PENILE PROSTHESIS · Mentor Corp. · Oct 29, 2004
- K183619 — Tactra Penile Prosthesis, 9.5 mm Device, Tactra Penile Prosthesis, 11 mm Device, Tactra Penile Prosthesis, 13 mm Device · Boston Scientific Corporation · Apr 16, 2019
- K963328 — AMS AMBICOR PENILE PROSTHESIS · American Medical Systems, Inc. · Dec 4, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
| Submitter | American Medical Systems Inc. (AMS)<br>10700 Bren Road West<br>Minnetonka, MN 55343 | OCT 20 2008 |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Contact Person | Stephanie George<br>Phone (952)-930-6389<br>Fax (952)-930-5785 | |
| Preparation Date | July 14, 2008 | |
| Device Common Name | Penile Prosthesis | |
| Device Trade Name | AMS Spectra™ Concealable Penile Prosthesis | |
| CFR Number | 21 CFR Part 876.3630 | |
| Regulatory Class | Class II (special controls) | |
| Product Codes | 78 FAE (penile prosthesis) | |
| Predicate Devices | DURA II® Penile Prosthesis (K953640)<br>AMS 600MTM Malleable Penile Prosthesis (K912935)<br>AMS 650TM Malleable Penile Prosthesis (K951716) | |
#### 7.3 510(k) Summary Statement (21CFR 807.92)
### Indications for Use
The AMS Spectra™ Concealable Penile Prosthesis is a sterile, nonpyrogenic, single-use implant that is intended for use in the treatment of chronic, organic, erectile dysfunction (impotence) in men who are determined to be suitable candidates for implantation surgery.
### Device Description
The AMS Spectra™ Concealable Penile Prosthesis consists of a pair of cylinders which are surgically implanted, one into each corpus cavernosum, to provide penile rigidity. Each device consists of two cylinders and may include rear tip extenders (RTEs) for additional length. All components consist of implantable, biocompatible materials.
The cylinder consists of a malleable section of articulating polymer segments. A cable extends through the center of the articulating segments. The proximal and distal ends of the cable are each connected to a spring that is encased in a metal housing. The entire outside surface of the cylinder is made of silicone.
{1}------------------------------------------------
AMS Spectra™ Concealable Penile Prosthesis Special 510(k)
> The articulating segments, held together by the cable and spring assemblies, provide sufficient friction and rigidity. This allows the patient to position the device for concealment or for intercourse.
Spectra cylinders are available in 9.5-, 12-, and 14-mm diameters. Each cylinder diameter is available in three lengths: 12-, 16-, and 20-cm.
The total cylinder length can be adjusted by adding rear tip extenders (RTEs) to the proximal cylinder end. A range of RTE lengths is included to accommodate the patient's total intracorporal length.
Rear tip extenders may be attached to the Spectra™ cylinders in 0.5-cm increments, and may extend the cylinder lengths from 0.5- to 7.5-cm, with the exception of 7-cm. The RTEs are the same as those used with the inflatable AMS 700™ CXR and CX/LGX models (D970012.) These RTE components' functional performance has been demonstrated in use with the 700™ IPP products.
### Substantial Equivalence
The AMS Spectra™ Concealable Penile Prosthesis was subjected to performance tests to evaluate its function. These tests included evaluations of cycle life, springback angle, column strength (rigidity), bend force, bond strength, packaging, sterilization, biocompatibility, and cadaveric evaluation with physicians. For each test, acceptance criteria were established prior to testing. The outcome of each test was that the acceptance criteria were met by the Spectra™ product, demonstrating substantial equivalence to the predicate products.
Because the intended use and the fundamental scientific technology of the predicate DURA II® and AMS 600M™ and 650TM Malleables were maintained with the AMS Spectra™, a clinical study was not conducted for SpectraTM.
The nonclinical performance tests demonstrated that the Spectra™ device is as safe, as effective + and performs as well as or better than the legally marketed DURA II® and AMS 600MTM and 650TM Malleable penile prostheses.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, representing health and human services.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# OCT 20 2008
Ms. Stephanie George Regulatory Affairs Specialist American Medical Systems 10700 Bren Road West MINNETONKA MN 55343
K082006 Re:
Trade/Device Name: AMS Spectra" Concealable Penile Prosthesis Regulation Number: 21 CFR §876.3630 Regulation Name: Penile rigidity implant Regulatory Class: II Product Code: FAE Dated: October 14, 2008 Received: October 15, 2008
Dear Ms. George:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jozue M. Whang
Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
AMS Spectra™ Concealable Penile Prosthesis Special 510(k)
#### 7.2 Statement of Indications for Use
# Indications for Use
510(k) Number (if known):
K082006 Unknown
Device Names:
Indications For Use:
AMS Spectra™ Concealable Penile Prosthesis
The AMS Spectra™ Concealable Penile Prosthesis is a sterile, non-pyrogenic, single-use implant that is intended for use in the treatment of chronic, organic, erectile dysfunction (impotence) in men who are determined to be suitable candidates for implantation surgery.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED)
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number J
CONFIDENTIAL