ANEROID SPHYGMOMANOMETER, MODEL KT-A01

K081951 · Wenzhou Kindcare Import & Export Co., Ltd. · DXQ · Aug 15, 2008 · Cardiovascular

Device Facts

Record IDK081951
Device NameANEROID SPHYGMOMANOMETER, MODEL KT-A01
ApplicantWenzhou Kindcare Import & Export Co., Ltd.
Product CodeDXQ · Cardiovascular
Decision DateAug 15, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1120
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

Aneroid Sphygmomanometer Model KT-A01 with Stethoscope is a non-automated, mechanical blood pressure monitor that is used for the indirect measurement (noninvasive) and display of arterial blood pressure. It can be used by professionals as well as trained individual users at hospitals or at home to monitor both systolic and diastolic pressure. This device is sold with an adult D size cuff and suitable for use on adults.

Device Story

Aneroid Sphygmomanometer Model KT-A01 with Stethoscope is a manual, non-automated blood pressure measurement system. Components include an adjustable adult D-ring cuff, an aneroid manometer (300 mmHg gauge), and an attached stethoscope. Operation involves wrapping the cuff around the patient's arm, inflating the cuff, and using the stethoscope to detect Korotkoff sounds over the brachial artery. The user manually observes the gauge to identify systolic and diastolic pressure values. Used by professionals or trained individuals in hospitals or home environments. Provides immediate visual and auditory feedback for clinical monitoring of arterial blood pressure.

Clinical Evidence

Bench testing only. Performance verified through compliance with ANSI/AAMI SP10:2002 and ANSI/AAMI SP10:2002/A1:2003 standards for electrical, mechanical, and environmental safety. Biological evaluation conducted per ISO 10993:2003.

Technological Characteristics

Manual, non-automated mechanical sphygmomanometer. Components: adult D-ring cuff, aneroid manometer (300 mmHg), stethoscope. Materials evaluated per ISO 10993:2003. Performance standards: ANSI/AAMI SP10:2002 and ANSI/AAMI SP10:2002/A1:2003. No electronic or software components.

Indications for Use

Indicated for adult patients for non-invasive, indirect measurement and display of arterial blood pressure (systolic and diastolic). Suitable for use by healthcare professionals or trained individuals in clinical or home settings.

Regulatory Classification

Identification

A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Kodi951 # AUG 1 5 2008 #### 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date: June 24, 2008 1. Company and Correspondent making the submission: | Name | Wenzhou Kindcare Import & Export Co., Ltd. | |-----------|----------------------------------------------------------------------| | Address | F407, Zhannan Commercial & Trading Bldg, Wenzhou, Zhejiang,<br>China | | Telephone | +86-577-8676 7197 | | Fax | +86 577-8678 0748 | | Contact | Mr. Robin Miao / Manager | | Email | Robin@Kindcare.Cn or Robin@Kindmed.Com | ## 2. Device: | Trade/proprietary name: | Aneroid Sphygmomanometer Model KT-A01 with Stethoscope | |-------------------------|--------------------------------------------------------| | Common Name: | Blood Pressure Cuff | | Classification Name: | Blood Pressure Cuff | Substantial Equivalence was demonstrated by compliance to ANSI/AAMI SP10: 2002 & ANSI/AAMI SP10: 2002/A1: 2003, ISO 10993: 2003 Biological Evaluation of Medical Devices - 3. Classifications Names & Citations: 21CFR 870.1120, DXQ, Blood Pressure Cuff, Class II {2}------------------------------------------------ ### 4. Description: - 4.1 General The Aneroid Sphygmomanomefer Model KT-A01 with Stethoscope is a non-invasive blood pressure measurement system for monitoring blood pressure levels. This Non-automated Sphygmomanometer uses an occluding cuff, an aneroid manometer to measure pressure, and a stethoscope for detecting Korotkoff sounds. - 4.2 Direction As discussed in the General description, the Wenzhou Kindcare Import & Export Co., Ltd., Aneroid Sphygmomanometer Model KT-A01 with Stethoscope is relatively simple to use. The attached stethoscope is placed on the inner arm above the bend in the elbow of the patient by detecting the pulse of the brachial artery. After inflation of the cuff, the user does auditory monitoring with the stethoscope to evaluate systolic and diastolic pressure. The two values are usually recorded as a ratio of the two measurements: systolic over diastolic. - The Aneroid Sphygmomanometer Model KT-A01 with Stethoscope contains: 4.3 - 1. Adjustable D-ring Cuff (Adult Size) - 2. Stethoscope (Attaches to the cuff) - 3. Non-stop rotary pin, 300 mmHg gauge - 4. Instruction booklet and record - 5. Carrying case - 5. Indication for use: Aneroid Sphygmomanometer Model KT-A01 with Stethoscope is a non-automated, mechanical blood pressure monitor that is used for the indirect measurement (noninvasive) and display of arterial blood pressure. It can be used by professionals as well as trained individual users at hospitals or at home to monitor both systolic and diastolic pressure. This device is sold with an adult D size cuff and suitable for use on adults. - 5. Comparison with predicate device: Wenzhou Kindcare import & Export Co., Ltd., believes that the Aneroid Sphygmomanometer Model KT-A01 with Stethoscope, is substantially equivalent to Wenzhou Longwan, Model AS101 BP Cuff. The devices have the same components, usage, patient population and functions. The packaging has the same claims, but varies only in the submitted device has more advertising verbiage regarding BP monitors. - 14 - {3}------------------------------------------------ - 7. Safety and Performance Data: Electrical, mechanical, environmental safety and performance testing according to standard AAMI SP10:2002. Specific testing relating to Aneroid Sphygmomanometer equipment was use to verify performance to recognized standards. All test results were satisfactory. #### 8. Conclusions: in accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Wenzhou Kindcare Import & Export Co., Ltd. concludes that the Aneroid Sphygmomanometer Model KT-A01 with Stethoscope is safe and effective and complies with the testing validations defined in AAMI /ANSI SP10 standard so that it is substantially equivalent to predicate devices as described herein. - 9. Wenzhou Kindcare Import & Export Co., Ltd. will update and include in a summary any other information deemed reasonably necessary by the FDA. END \$\frac{2}{3}\$ {4}------------------------------------------------ Public Health Service Image /page/3/Picture/2 description: The image shows a black and white logo for the Department of Health & Human Services. The logo features a stylized image of an eagle with three swooping lines representing its wings. The eagle is positioned to the right of a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 1 5 2008 Wenzhou Kindcare Import & Export Co., Ltd. c/o Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313 Re: K081951 Trade/Device Name: Aneroid Sphygmomanometer Model KT-A01 with Stethoscope Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II (Two) Product Code: DXQ Dated: July 28, 2008 Received: July 29, 2008 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {5}------------------------------------------------ comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ #### Indications for Use 510(k) Number (if known): `K081951 Device Name: Aneroid Sphygmomanometer Model KT-A01 with Stethoscope Indications for Use: Aneroid Sphygmomanometer Model KT-A01 with Stethoscope is a non-automated, mechanical blood pressure monitor that is used for the indirect measurement (noninvasive) and display of arterial blood pressure. It can be used by professionals as well as trained individual users at hospitals or at home to monitor both systolic and diastolic pressure. This device is sold with an adult D size cuff and suitable for use on adults. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use ブ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ed.mall (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number ~ 12~
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...