K081915 · Fujirebio Diagnostics Texas, Inc. · JJX · Aug 8, 2008 · Clinical Chemistry
Device Facts
Record ID
K081915
Device Name
FDTX GLUCOSE CONTROL SOLUTION
Applicant
Fujirebio Diagnostics Texas, Inc.
Product Code
JJX · Clinical Chemistry
Decision Date
Aug 8, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Intended Use
The FDTX Glucose Control Solution is intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of Bayer Ascensia Contour Blood Glucose Monitor.
Device Story
FDTX Glucose Control Solution is an aqueous, viscosity-adjusted liquid containing a known quantity of D-Glucose; used to verify performance of Bayer Ascensia Contour Blood Glucose Monitor. Packaged in plastic dropper bottles; includes red dye for visual confirmation of application to test strips. Used by healthcare professionals or patients at home. User applies solution to test strip; monitor reads glucose level; user compares result to acceptable range provided in labeling. Benefits patient by ensuring accuracy of blood glucose monitoring system.
Clinical Evidence
No clinical data. Bench testing only. Stability determined via real-time studies; shelf-life 24 months unopened, 90 days open-vial. Traceability established via in-house glucose preparation; acceptable ranges assigned via repeat analysis using three lots of test strips.
Technological Characteristics
Buffered aqueous solution containing D-Glucose, viscosity modifier, preservatives, and non-reactive ingredients. Packaged in plastic dropper-tipped bottles. Red colorant added for visual confirmation. Standalone quality control material; no software or electronic components.
Indications for Use
Indicated for healthcare professionals and people with diabetes mellitus to assess the performance of the Contour Blood Glucose Monitor. For in vitro diagnostic, external use only.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
Ascensia Microfill Control Solution (k023657)
Related Devices
K080273 — AB-TROL AND PRIVATE LABEL GLUCOSE CONTROL SOLUTION · Fujirebio Diagnostics Texas, Inc. · Aug 8, 2008
K082395 — GLUCOSE METER-CHECK CONTROL SOLUTION FOR BAYER ASCENSIA BLOOD GLUCOSE METER, MODEL: MC002 · Bionostics, Inc. · Sep 18, 2008
{0}------------------------------------------------
ﮐﺎ | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
## 5 510(k) Summary
AUG - 8 2008
| Introduction: | According to the requirements of 21 CFR 807.92, the<br>following information provides sufficient detail to<br>understand the basis for a determination of<br>substantial equivalence. |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Fujirebio Diagnostics Texas, Inc.<br>940 Crossroads Blvd<br>Seguin, TX 78155<br>(830) 372-1391 ex. 210<br>Establishment Registration Number: 1643621 |
| Contact Person: | John C. Gormley |
| Device Name: | FDTX Glucose Control Solution |
| Common Name: | Single Analyte Control Solution, All Types (Assayed<br>and Unassayed) |
| Classification Name: | Quality Control Material (assayed and unassayed). |
| Classification: | Class I per 21 CFR 862.1660 |
| Product Code: | 75 JJX |
| Panel: | Chemistry |
| Predicate Devices: | Name: Ascensia Microfill Control<br>Solution<br>Manufacturer: Bayer Healthcare<br>510(k) No.: K023657 |
| | Name: Liberty Normal Glucose Control<br>Solution<br>Manufacturer: Liberty Healthcare Group<br>510(k) No.: K060706 |
| Device Description: | The FDTX Glucose Control Solution consists of a<br>viscosity-adjusted, aqueous liquid control solution<br>containing a known quantity of glucose. The product<br>is packaged in plastic dropper tipped bottles for easy<br>application of the control solutions to the test strips<br>and a red coloration to aid the user to visually confirm |
{1}------------------------------------------------
> application of the control. The product is nonhazardous and contains no human or animal derived materials.
Intended Use: The FDTX Glucose Control Solution is intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of Bayer Ascensia Contour Blood Glucose Monitor.
## Comparison to Predicate Devices:
| Characteristic/<br>Aspect | Predicate Device No.<br>1 | Predicate Device No.<br>2 | New Product |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Name | Ascensia<br>Contour/Microfill<br>Control Solution | Liberty Normal Glucose<br>Control | FDTX Glucose Control<br>Solution |
| 510(k), Date | K023657<br>05/12/2003 | K060706, 04/28/2006 | |
| Number of<br>Levels | 1 | 1 | 1 |
| Analyte | Glucose | Glucose | Glucose |
| Target Range<br>(mg/dL) | $100 - 143^{(1)}$ | $90 - 130^{(2)}$ | 100 - 145 |
| Container | Plastic bottle with<br>dropper-tip | Plastic bottle with<br>dropper-tip | Plastic bottle with<br>dropper-tip |
| Fill Volume | 2.5 mL | 3.6 mL | 3.6 mL |
| Color | Red | Red | Red |
| Matrix | Aqueous Glucose<br>solution. | Buffered aqueous<br>solution of D-Glucose,<br>a viscosity modifier,<br>preservatives, and<br>other non-reactive<br>ingredients | Buffered aqueous solution<br>of D-Glucose, a viscosity<br>modifier, preservatives,<br>and other non-reactive<br>ingredients |
| Indications for<br>Use | For use with the<br>Ascensia Contour<br>Blood Glucose Meter<br>and the Ascensia<br>MICROFILL Test Strips<br>as a quality control<br>check. | To check the<br>performance of<br>OneTouch Ultra,<br>OneTouch FastTake,<br>Accu-Chek Active, and<br>Ascensia Contour<br>Blood Glucose<br>Systems. | To check the performance<br>of the Ascensia Contour<br>Blood Glucose System. |
| Target<br>Population | Professional and home<br>use | Professional and home<br>use | Professional and home<br>use |
(1) Estimated from published control ranges assigned by the manufacturer for several lots of Contour test strips.
(2)Estimated from published control ranges assigned by the manufacturer for several lots of Liberty Normal Glucose Control.
{2}------------------------------------------------
- Performance Studies: Tests were performed to verify specific performance characteristics:
- 1. Accelerated Stability
- 2. Open Vial
- 3. Test precision
- Conclusion: Comparison of the performance characteristics, formulation and intended use support the claim of substantial equivalence.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (USA). The seal features a stylized eagle with three tail feathers, representing the department's commitment to health, services, and people. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Fujirebio Diagnostics Texas, Inc c/o Mr. John Gormley, Director of Quality & Regulatory Affairs 940 Crossroads Boulevard Seguin. TX 78155
AUG - 8 2008
Re: K081915 Trade/Device Name: FDTX Glucose Control Solution Regulation Number: 21 CFR 862.1660 Regulation Name: Quality Control Material (assayed and unassayed) Regulatory Class: Class I , reserved Product Code: JJX Dated: June 27, 2008 Received: July 3, 2008
Dear Mr. Gormley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jean M. Cooper, M.S., D.V.M.
Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 4 Indications for Use Statement
510(k) Number (if known): 081915
Device Name: FDTX Glucose Control Solution
Indications for Use:
For in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the Contour Blood Glucose Monitor.
Prescription Use (21 CFR Part 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR Part 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Signature
Div sign /Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K081915
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.