SYSMEX XT-SERIES IG PARAMETER,SYSMEX XT-SERIES RET-HE PARAMETER

K081890 · Sysmex America, Inc. · GKZ · Oct 10, 2008 · Hematology

Device Facts

Record IDK081890
Device NameSYSMEX XT-SERIES IG PARAMETER,SYSMEX XT-SERIES RET-HE PARAMETER
ApplicantSysmex America, Inc.
Product CodeGKZ · Hematology
Decision DateOct 10, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5220
Device ClassClass 2

Indications for Use

The IG parameter on the Sysmex® XT-Series, Automated Hematology Analyzer, is intended for in vitro diagnostic use to classify and count immature granulocytes cells in EDTA anti-coagulated blood. The RET-He parameter on the Sysmex® XT-Series, Automated Hematology Analyzer, determines the hemoglobin of reticulocytes for in vitro diagnostic use in EDTA anti-coagulated blood.

Device Story

The Sysmex XT-Series is an automated hematology analyzer used in clinical laboratories. It processes random whole blood samples to provide Immature Granulocyte (IG) counts and Reticulocyte Hemoglobin (RET-He) values. The device utilizes flow cytometry in the DIFF channel, measuring side scatter, forward scatter, and fluorescent intensity of nucleated cells to identify and count leukocyte populations, including immature granulocytes. Reticulocyte parameters are derived from forward scattered light signals using a proprietary calculation equation. The system requires XT-Pro software and specific master programs (IG Master, RET Master) to function. Healthcare providers use these outputs to assess hematological status. The device benefits patients by providing automated, standardized hematological data, reducing manual labor and improving diagnostic consistency compared to manual methods.

Clinical Evidence

Bench testing only. Precision/reproducibility evaluated over 20 days using 3-level QC materials and residual samples; results met manufacturer specifications. Method comparison study (n=272) across two sites included diverse disease states (anemia, thalassemia, abnormal lymphocytes, etc.), showing high correlation (r ≥ 0.90) for IG# (0.9846), IG% (0.9833), and RET-He (0.9277). Reference intervals verified against XE-2100.

Technological Characteristics

Automated hematology analyzer; flow cytometry; semiconductor laser; hydrodynamic focusing; SLS-hemoglobin method; polymethine dye staining. Software: XT-Pro, IG Master Software, RET Master Software. Specimen: EDTA anti-coagulated blood.

Indications for Use

Indicated for in vitro diagnostic classification and counting of immature granulocytes and determination of reticulocyte hemoglobin in EDTA anti-coagulated blood samples.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k081890 B. Purpose for Submission: Addition of IG and RET-He parameters to the Sysmex® XT-Series, Automated Hematology Analyzer. C. Measurand: IG%/#: Immature Granulocyte Percent and Count RET-He: Reticulocyte Hemoglobin D. Type of Test: IG: Metamyelocytes, Myelocytes, and Promyelocytes RET-He: hemoglobin of reticulocytes E. Applicant: Sysmex America, Inc. F. Proprietary and Established Names: Sysmex® XT-Series IG Parameter Sysmex® XT-Series RET-He Parameter G. Regulatory Information: 1. Regulation section: 21 CFR 864.5220, Automated Differential Cell Counter 2. Classification: II 3. Product code: GKZ, Counter, differential cell 4. Panel: Hematology, 81 H. Intended Use: 1. Intended use(s): The IG parameter on the Sysmex® XT-Series, Automated Hematology Analyzer, is intended for in vitro diagnostic use to classify and count immature granulocytes cells in EDTA anti-coagulated blood. The RET-He parameter on the Sysmex® XT-Series, Automated Hematology Analyzer, determines the hemoglobin of reticulocytes for in vitro diagnostic use in EDTA anti-coagulated blood. - IG%/#: Immature Granulocyte Percent and Count - RET-He: Reticulocyte Hemoglobin 2. Indication(s) for use: Same as the Intended Use 3. Special conditions for use statement(s): For use with EDTA anti-coagulated blood 4. Special instrument requirements: For use only on Sysmex® XT analyzers with: - IG%/#: Sysmex® XT-Pro and IG Master Software - RET-He: Sysmex® XT-Pro and RET Master Software {1} I. Device Description: The XT-Series is an automated hematology analyzer using Hydrodynamic Focusing, flow cytometry method, and SLS-hemoglobin method for analysis of IG and RET-He in EDTA anti-coagulated blood samples. The IG parameter is measured in the DIFF channel, cells are detected by a combination of side scattered, forward scatter, and fluorescent intensity. Abnormal and immature cells are clustered and indistinguishable in the DIFF scattergram. The RET-He parameter is measured in the reticulocyte channel; cells are stained with polymethine dye which stains RNA differentiating reticulocytes from mature RBCs. The reticulocyte parameters are derived from the reticulocyte histogram, which is generated by the forward light signals. J. Substantial Equivalence Information: 1. Predicate device name(s): Sysmex® XE-2100, IG parameter Sysmex® XE-2100, RET-He parameter 2. Predicate K number(s): k032039 k050589 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device: Sysmex® XT-Series | Predicate: Sysmex® XE-2100 | | Intended Use | The IG parameter is intended for in vitro diagnostic use to classify and count immature granulocytes cells, the RET-He parameter determines the hemoglobin of reticulocytes. | Same | | Methodology | IG: measured in the DIFF channel using the combination of side scatter, forward scatter, and fluorescent intensity of nucleated cells. RET-He: reticulocyte parameters are derived using the reticulocyte measurement channel and a propriety Sysmex calculation equation. | Same | | Specimen Type | EDTA anti-coagulated blood | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Software | XT-Pro & IG Master Software XT-Pro & RET Master Software | XE-Pro & IG Master Software XE-Pro & RET Master Software | K. Standard/Guidance Document Referenced (if applicable): None provided {2} 3 L. Test Principle: IG parameter: When blood is diluted and stained with Stomatolyzer-4DL, mature leukocytes and red blood cells are selectively lysed, leaving immature granulocytes intact due to its lower phospholipid content. In the optical detector block, the sample is analyzed via flow cytometry method utilizing a semiconductor laser. RET-He parameter: When blood is stained with polymethine dye in the reticulocyte channel, the sample is analyzed by automated fluorescent flow cytometry method. The analyzer measures the mean value of the forward light scatter histogram of mature red blood cells and reticulocytes. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Quality control and residual samples were used. E-Check Quality Control materials, 3 levels, were assayed for 20 days. Five residual samples were assayed 10 consecutive times. Results met manufacturer specification and testing criteria as follows: | Parameter | Specification | Testing Criteria | | --- | --- | --- | | IG% | CV% ≤25%, or within ±1.5 IG% | IG% ≥ 2%, WBC ≥ 4.00 x 10³ /μL | | IG# | CV% ≤25%, or within ±0.12 x 10³/μL | IG#: 0.1 x 10³ /μL | | RET-He | CV% ≤5% | RET#: ≥ 0.020 x 10⁶ /μL | b. Linearity/assay reportable range Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Stability: Samples should be stored at room temperature (18-26°C) and refrigerated temperature (2-8°C). The variation from 0 hour after the blood collection to 24 hours was calculated and met manufacturer specification. - 30 normal samples were used in the IG stability study. The variation IG from was within ± 2 IG% - 30 samples with RET-He ≥ 0.010 x 10⁶ /μL were used in the RET-He stability study. The variation for RET-He was within ± 8%. d. Detection limit: Not applicable. e. Analytical specificity: Not applicable. f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: 272 samples were collected and analyzed at two sites. The study included abnormal samples with a variety of disease states such as anemia, iron deficiency, thalassemia, immature granulocytes, blast forms, left shift, atypical/abnormal lymphocytes, high WBC counts, lipemia, NRBCs, etc. Results met manufacturer specification, r ≥ 0.90, as follows: {3} | Parameter | Correlation Coefficient | Slope | Intercept | | --- | --- | --- | --- | | IG # | 0.9846 | 0.947 | -0.0103 | | IG % | 0.9833 | 1.024 | -0.13822 | | RET-He | 0.9277 | 1.049 | 1.43 | b. Matrix comparison: Not applicable. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: 40 specimens were tested and compared to verify the existing reference intervals established for the Sysmex XE-2100. Sysmex recommends that each laboratory establish its own expected reference intervals based upon the laboratory's patient population and use NCLSI C28-A2 for guidelines. | | IG% | IG# | RET-He | | --- | --- | --- | --- | | Mean | 0.2 | 0.016 | 34.265 | | SD | 0.159 | 0.0139 | 2.2 | | Range | 0-0.518 | 0-0.0438 | 29.9-38.7 | N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...