K081884 · Major Prodotti Dentari S.P.A. · EBI · Oct 8, 2008 · Dental
Device Facts
Record ID
K081884
Device Name
MAJOR.BASE 20
Applicant
Major Prodotti Dentari S.P.A.
Product Code
EBI · Dental
Decision Date
Oct 8, 2008
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.3760
Device Class
Class 2
Indications for Use
Major.Base 20 is a denture base polymer for dental prosthesis. Poly-methylmethacrylate based. Heat-processed polymer. Powder and liquid. It is used for: - dental prosthesis - partial prosthesis - clasp prosthesis -
Device Story
Major.base 20 is a heat-cure denture base material consisting of poly(methyl methacrylate) (PMMA) powder and a liquid monomer (methyl methacrylate). Used by dental practitioners in a clinical or laboratory setting to fabricate total, partial, or clasp dental prostheses. The material is processed via heat-curing techniques. The final product provides a structural base for dental appliances, aiding in patient oral rehabilitation. No automated processing or software is involved.
Clinical Evidence
No clinical data provided. Safety is supported by 50 years of clinical history, literature, and established use of similar PMMA formulations in dentistry. Bench testing performed by NIOM Laboratory confirmed physical properties: flexural strength 78 MPa, flexural modulus 2390 MPa, water absorption 22.0 µg/mm3, water solubility 1.5 µg/mm3, and residual monomer 1.8%.
Technological Characteristics
Heat-cure denture base material; poly(methyl methacrylate) (PMMA) polymer powder and methyl methacrylate liquid. Complies with ISO 1567:1999 and ADA/ANSI 12:2002. Cadmium-free pigments. Physical properties: 78 MPa flexural strength, 2390 MPa flexural modulus.
Indications for Use
Indicated for use as a denture base material for the fabrication of total, partial, and clasp dental prostheses. For use by dental practitioners only; not for OTC use.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
Predicate Devices
Paladon® 65 by Heraeus Kulzer GmbH & Co. (K901789)
{0}------------------------------------------------
Kof1884
# OCT 0 8 2008
# 510(K) SUMMARY E.
Image /page/0/Picture/3 description: The image is a logo for Major Dental. The logo is a large, stylized letter "M" in white against a black background. Below the "M" are the words "MAJOR" and "DENTAL" stacked on top of each other, also in white.
MAJOR Prodotti Dentari S.p.A. Via Luigi Einaudi, 23, 10024 Moncalieri (TO) - ITALY Tel +39.011.6400.211 - Fax +39.011.6400.222 E-mail: info@majordental.com - Internet: www.majordental.com
#### SUBMITTER
| Name | MAJOR Prodotti Dentari S.p.A. |
|---------------------------------|-----------------------------------------------------------|
| Address | Via Luigi Einaudi, 23<br>I-10024 Moncalieri (TO)<br>ITALY |
| Facility registration<br>number | 4565 |
| Contact person | Monica Funai |
| Phone number | +39.011.6400.211 |
| Fax number | +39.011.6400.222 |
| e-mail | m.funai@majordental.com |
| Document issued on | June 26, 2008 |
### NAME OF DEVICE
| Proprietary name | Major.base 20 |
|---------------------|------------------------------------------------------------------------------------------------|
| Common name | Denture base material |
| Classification name | Denture relining, reparing, or rebasing resin<br>(21 CFR 872.3760, Product Code EBI) |
| Predicate Devices | Paladon® 65 by Heraeus Kulzer GmbH & Co. K901789<br>Probase Hot by Ivoclar Vivadent AG K913655 |
## DESCRIPTION OF THE DEVICE
Major.base 20 is a denture base, poly(methyl methacrylate)-based polymers for dental prosthesis. It can be used for:
- Total prosthesis .
- Partial prosthesis .
- Clasp dentures e
{1}------------------------------------------------
Major base 20 is a heat cure denture base material composed of a poly-methylmethacrylate polymer powder and a liquid consisting of methylmethacrylate and other ingredients solution.
Major.base 20 is substantially equivalent to predicate denture base system presently on the USA market and safety and effectiveness are well documented in the dental literature. All pigments used are approved for alimentary or similar use and are Cadmium free.
Polymerized material technical data (test performed by NIOM Laboratory - Norway) are the following:
- . Flexural strength: 78 MPa
- Flexural modulus: 2390 MPa .
- Water absorption: 22,0 µg/mm3 .
- Water solubility: 1,5 µg/mm3 .
- Residual monomer: 1,8 % .
Major.base 20 is inherently safe when used according to the instructions for use. It is for use only by dental practitioners; it is not intended for OTC use.
#### TECHNOLOGICAL CHARACTERISTICS (compared to the predicate device)
Major base 20 has the same technological characteristics as the predicate devices since all the devices are complying with ISO 1567:1999 or ADA/ANSI 12:2002 (Adoption of the ISO 1567:1999) and have:
- Same intended use .
- Same polymers composition .
- Same working technique .
Finally our laboratory experience in using the listed devices is that you cannot find substantial differences: they are equivalent to major.base 20.
#### NONCLINICAL TESTING AND CONCLUSIONS
Taking in account that:
- the products and their formulation and use are widely described and totally accepted in . dentistry (McKabe, J., Applied Dental Materials, 7th, Blackwell Scientif. Publ.; Phillips, J., Skinner's Science of Dental Materials, Saunders Co .; Graig and Oth., Dental Materials, Properties and Manipulation, 6th, Mosby Publ .; O' Brien, W. J., Dental Materials, Properties and Selection, Quintessence Books);
- a large amount of literature has assessed the clinical liability of these product types and their formulation, dedicating the attention to the denture base (see for instance articles registered on MedLine);
{2}------------------------------------------------
- these product types and their formulation are universal in the last fifty years manufacturing and clinical experience, counting for almost the totality of the denture bases in the world;
- the composition and philosophy of the evaluated products are exactly within what is described in most general literature;
- our company manufactures and sells these products, under similar formulations and processes, since 1966, without any record of human incompatibility or clinical evidence of adverse effects, having totaled, directly or through connected companies or transferred technologies, more than 150 tons of powders in the last ten years;
we can assess these statements:
- the materials with which the products are manufactured, own a 50 years history of 피 clinical research, testing and literature, which is largely sufficient not to request any further clinical research or testing;
- 트 the products can then be considered clinically tested to be safe;
- . the clinical monitoring is effected independently by the general literature and it is largely sufficient not to request any further research or testing.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged in a circular fashion around the symbol. The logo is in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 0 8 2008
Ms. Monica Funai Ouality and Regulatory Affairs MAJOR Prodotti Dentari S.p.A. Via Luigi Einaudi 23, Moncalieri (Torino) ITALY I-10024
Re: K081884
Trade/Device Name: Major.Base 20 Regulation Number: 21 CFR 872.3760 Regulation Name: Denture, Relining, Repairing, or Rebasing Resin Regulatory Class: II Product Code: EBI Dated: September 10, 2008 Received: September 15, 2008
Dear Ms. Funai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{4}------------------------------------------------
Page 2 - Ms. Funai
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Klemule Lendner forll
Chiu S. Lin, Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo with a stylized letter "M" at the top, which appears to be a graphical representation. Below the "M", the word "MAJOR" is printed in a bold, sans-serif font. The word "DENTAL" is printed below "MAJOR", also in a bold, sans-serif font. The logo is set against a black background, with the letter and text appearing in white.
# Indications for Use
510(k) Number (if known):
Device Name: major.base 20
Indications for Use: Major.Base 20 is a denture base polymer for dental prosthesis. Poly-methylmethacrylate based. Heat-processed polymer. Powder and liquid.
It is used for:
- dental prosthesis -
- partial prosthesis
- clasp prosthesis -
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Susan Reese
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K051884
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.