Proclear Multifocal Toric XR (omafilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic or hyperopic, possess astigmatism of 3.00 diopters or less, and are presbyopic. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only). Proclear Multifocal XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic or hyperopic and are presbyopic. The lenses may be worn by persons who exhibit astigmatism of 0.75 diopters or less that does not interfere with visual acuity. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only). Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic or hyperopic, possess astigmatism of 5.00 diopters or less. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only). Daily wear replacement schedules may vary from patient and should be decided by the eye care practitioner in consultation with their patients. The lenses are to be cleaned, rinsed and disinfected each time they are removed from the patients eye and discarded after the recommended wearing period prescribed by the eye care practitioner. The lenses may be disinfected using a chemical disinfection system.
Device Story
Soft (hydrophilic) contact lenses made of omafilcon A polymer; designed for daily wear to correct myopia, hyperopia, astigmatism, and presbyopia. Lenses feature specific geometries: back surface toric for astigmatism correction; multifocal designs with aspherical front surfaces for presbyopia. Lenses are tinted with Vat Blue 6 for visibility. Prescribed by eye care practitioners; patients perform daily cleaning, rinsing, and chemical disinfection. Lenses provide vision correction and may improve comfort for patients with mild dry eye symptoms. No electronic or software components.
Clinical Evidence
No clinical data required. Substantial equivalence established through physiochemical, optical, and chemical property studies, alongside in-vivo and in-vitro toxicological studies confirming biocompatibility.
Technological Characteristics
Material: Omafilcon A (copolymer of 2-hydroxy-ethylmethacrylate and 2-methacryloyloxyethyl phosphorylcholine crosslinked with ethyleneglycol dimethacrylate). Water content: 59%. Refractive index: 1.40. Oxygen permeability: 21.05 x 10^-11 (cm^2/sec)(ml O2/ml x mmHg) at 35°C. Design: Back surface toric, multifocal (aspherical front/spherical back). Tint: Vat Blue 6. Sterilization: Steam validated autoclave. Packaging: Blister pack.
Indications for Use
Indicated for daily wear correction of visual acuity in aphakic/non-aphakic patients with non-diseased eyes (myopic/hyperopic). Proclear Multifocal Toric XR: presbyopic patients with astigmatism <= 3.00D. Proclear Multifocal XR: presbyopic patients with astigmatism <= 0.75D. Proclear Toric XR: patients with astigmatism <= 5.00D. All indicated for patients experiencing mild discomfort/dryness associated with Evaporative or Aqueous Tear Deficiency (non-Sjogren's).
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for CooperVision. The logo is black and white and features the company name in a stylized font. There is a registered trademark symbol to the right of the name.
SEE BEYOND THE OR
## 510(k) Summary
| 510(k) SUMMARY | | |
|----------------|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. | Applicant Name & Address: | CooperVision, Inc.<br>711 North Road<br>Scottsville, NY 14546 |
| 2. | Contact: | Bonnie Tsymbal<br>Telephone (585) 264-3210<br>Facsimile (585) 889-5688 |
| 3. | Date Prepared: | June 17, 2008 |
| 4. | Device Identification: | |
| | Trade Name: | Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lenses<br>Proclear Multifocal XR (omafilcon A) Soft (Hydrophilic) Contact Lenses<br>Proclear Multifocal Toric (omafilcon A) Soft (Hydrophilic) Contact Lenses |
| | Common Name: | Soft Contact Lenses |
| | Classification Name: | Lenses, Soft Contact, Daily Wear |
| | Device Classification: | Class II (21 CFR 886.5925) |
| | Product Code: | LPL |
#### Intended Use: 5.
Proclear Multifocal Toric (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for the correction of visual acuity in aphakic persons with non-diseased eyes that are myopic or hyperopic, possess astigmatism of 10.00 diopters or less, and are presbyopic. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only).
Proclear Multifocal XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic or hyperopic and are presbyopic. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency of from Aqueous Tear Deficiency (non-Sjogren's only).
Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic or hyperopic; possess astigmatism of 10.00 diopters or less. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only).
Daily wear replacement schedules may vary from patient and should be decided by the eye care practitioner in consultation with their patients. The lenses are to be cleaned, rinsed and time they
1
{1}------------------------------------------------
## 510(k) Summary
are removed from the patients' eve and discarded after the recommended wearing period prescribed by the eve care practitioner. The lenses may be disinfected using a chemical disinfection system.
## 6. Device Description
## Proclear Toric XR
Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lens is a back surface toric. The lenses are made of polymer of 2-hydroxy-ethylmethacrylate and 2-methacryloyloxyethyl phosphoryicholine crosslinked with ethylenedycol dimethacrylate. The lenses are tinted edge for visibility purposes with the color additive Vat Blue 6. Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are available as astigmatic (toric) lenses with the following dimensions:
- I Chord Diameter:
- 트 Center Thickness
- Base Curve: 트
- 트 Spherical Powers:
- I Cylinder Powers:
- 지 Axis
#### Proclear Multifocal XR
Proclear Multifocal XR (omafilic) Contact Lenses are available as a multifecal lens with an aspherical front surface and scherical back surface for the correction of visual acuity in presbyopic persons who are myopic or hyperopic. The Proclear Multifocal XR is designed with two multifocal zones, as well as the edge shape being optimized to provide comfort without sacrificing tensile strength. The Proclear Multifocal XR / D (dominant) has a spherical central zone for the correction of distance vision and an aspherical annular zone for the correction of intermediate and near vision. The Proclear Multifccal XR / N (non-dominant) has a spherical central zone for the correction of near vision and an aspherical annular zone for the mediate and distance vision. The lenses are tinted edge for visibility purposes with the color additive Vat Blue 6.
The lens material, omafilcon A is a copolymer of 2-hydroxy-ethylmethacryloyloxyethyl phosphorycholine crosslinked with ethylenedycol dimethacrylate. The Proclear UltraVue/D and Proclear UltraVue/N (omatilcon A) Soft (Hydrophilic) Contact Lenses are flexible transparent hemispherical shells of the following dimensions:
- Chord Diameter: 트
- Center Thickness (minus):
- エ Base Curve:
- 트 Spherical Powers:
- Add Powers:
- Central Zone Diameter: ■
13.6 to 15.2 mm 0.09 mm to 0.65 mm 8.3 mm to 8.9mm -20.00 D to +20.00 D +1.00 to +4.00 D 2.3 mm to 2.6 mm (Proclear Multifocal XR / D) 1.7 mm to 2.0 mm (Proclear Multifocal XR / N)
Proclear Multifocal Toric
13.6 to 15.2 mm 0.09 mm to 0. 65 mm 8.0 mm to 9.3 mm -20.00 D to +20.00 D -0.75 to -5.00 D 1° to 180°
{2}------------------------------------------------
GoperVisiona®
## 510(k) Summary
Proclear Multifocal Toric (omafilcon A) Soft (Hydrophilic) Contact Lenses are made of polymer of 2-hydroxyethylmethacrylate and 2-methacryloyloxycholine crosslinked with ethyleneglycol dimethacrylate. The lenses are tinted edge to edge for visibility purposes with the color additive Vat Blue 6.
The front surface of the Proclear Multifocal Toric (omafilic) Contact Lenses is aspherical, with the anterior surface having a toric generated surface for the purpose of correcting vision in an eye that is astigmatic. The Proclear Multifocal Toric contact lenses are designed with two multifocal zones, as well as the edge shape being optimized to provide comfort without sacrificing tensile strength.
The Proclear Multifocal Toric (omafilic) Contact Lenses are available in two versions. The Proclear Multifocal Toric / D with a spherical central zone for the correction of distance vision and an aspherical annular zone for the correction of internediate and near vision. The Proclear Multifocal Toric / N with a spherical central zone for the correction of near vision and an aspherical annular zone for the correction of intermediate and distance vision.
Both lenses are a flexible transparent hemispherical shell of the following dimensions:
| Chord Diameter: | 13.6 to 15.2 mm |
|---------------------------|-----------------------------------------------------------------------------------------------------|
| Center Thickness (minus): | 0.035 mm to 0.96 mm |
| Base Curve: | 8.3 mm to 8.9mm |
| Spherical Powers: | -20.00 D to +20.00 D |
| Cylinder Powers: | -0.75 to -3.00 D |
| Add Powers: | +1.00 to +4.00 |
| Central Zone Diameter: | 2.3 mm to 2.6 mm (Proclear Multifocal Toric / D)<br>1.7 mm to 2.0 mm (Proclear Multifocal Toric/ N) |
The physical properties of the lenses are:
| Refractive Index at 25º C | 1.40 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Light Transmittance | >90% |
| Water Content | 59 % |
| Oxygen Permeability* | 21.05 x 10-11<br>*(cm²/sec) (ml O2/ml x mm Hg) at 35°C. as measured by 201T<br>Permeometer connected to a curved Rehder guard ring polarographic<br>cell |
{3}------------------------------------------------
# 510(k) Summary
# 7. Substantial Equivalence Table:
| | SUBJECT DEVICE | PREDICATE DEVICE<br>K050717 | PREDICATE DEVICE<br>K061948 |
|-------------------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| | Proclear Toric XR<br>Proclear Multifocal XR<br>Proclear Multifocal Toric | Proclear UltraVue Toric<br>XR<br>Proclear UltraVue<br>Multifocal XR<br>Proclear UltraVue<br>Multifocal Toric | Proclear Sphere/Asphere 1-Day<br>Proclear Toric 1-Day<br>Proclear Multifocal 1-Day<br>Proclear Multifocal Toric 1-Day |
| Material<br>USAN Name | Omafilcon A | Omafilcon A | Omafilcon A |
| FDA Category<br>(Group) | Group II Non-Ionic<br>High Water | Group II Non-Ionic<br>High Water | Group II Non-Ionic<br>High Water |
| Water Content | 59% | 59% | 60% |
| Light<br>Transmittance | >90% | >90% | >90% |
| Index of<br>Refraction | 1.40 | 1.40 | 1.40 |
| Oxygen<br>Permeability | 21.05 | 21.05 | 20.45 |
| Manufacturing<br>Method | Finished Inside<br>Polymerization System | Finished Inside<br>Polymerization System | Cast Molded |
| Sterilization | Steam Validated Autoclave | Steam Validated Autoclave | Steam Validated Autoclave |
| Packaging | Blister Pack | Blister Pack | Blister Pack |
| Shelf-Life | 5 Yrs. | 5 Yrs. | 5 Yrs. |
| Visibility Tint | VAT Blue #6 | C.I. Reactive Blue #4 | VAT Blue #6 |
| Tinting Process | Entrapment | Reactive | Entrapment |
| | SUBJECT DEVICE | PREDICATE DEVICE<br>K050717 | PREDICATE DEVICE<br>K061948 |
| | Proclear Toric XR | Proclear UltraVue Toric | Proclear Toric |
| Lens<br>Design | Back Surface Toric | Back Surface Toric | Back Surface Toric |
| Intended<br>Use | Correction of visual acuity in<br>patients with myopia, hyperopia<br>and are astigmatic | Correction of visual acuity in<br>patients with myopia, hyperopia<br>and are astigmatic | Correction of visual acuity in<br>patients with myopia, hyperopia<br>and are astigmatic |
| | SUBJECT DEVICE | PREDICATE DEVICE<br>K05717 | PREDICATE DEVICE<br>K061948 |
| | Proclear Multifocal XR | Proclear UltraVue Multifocal | Proclear Multifocal |
| Lens<br>Design | Multifocal | Multifocal | Multifocal |
| Intended<br>Use | Correction of visual acuity in<br>patients with myopia or<br>hyperopia, and are presbyopic | Correction of visual acuity in<br>patients with myopia or<br>hyperopia, and are presbyopic | Correction of visual acuity in<br>patients with myopia or<br>hyperopia, and are presbyopic |
| | SUBJECT DEVICE | PREDICATE DEVICE<br>K05717 | PREDICATE DEVICE<br>K061948 |
| Lens<br>Design | Multifocal Toric | Multifocal Toric | Multifocal Toric |
| Intended<br>Use | Correction of visual acuity in<br>patients with myopia or<br>hyperopia, and are presbyopic | Correction of visual acuity in<br>patients with myopia or<br>hyperopia, and are presbyopic | Correction of visual acuity in<br>patients with myopia or<br>hyperopia, and are presbyopic |
{4}------------------------------------------------
@operVisiona®
510(k) Summary
## 8. Lens Design and Indications Table:
#### ഗ് Physiochemical Studies:
Resuits from physical, optical and chemical properties show substantial equivalency with the predicate devices, and are within established specifications for the lenses.
{5}------------------------------------------------
## 510(k) Summary
## 10. Toxicology:
Results from in-vivo and in-vitro studies were conducted and verify that the lenses remain non-toxic and are biocompatible with the ocular environment.
## 11. Clinical Studies:
The technical characteristics, formulation, manufacturing, and sterilization processes of this lens are equivalent to omafilcon A soft contact lenses currently marketed by CooperVision, therefore no clinical data is required.
### 12. Conclusion:
The device will be manufactured according to specified process controls and an established quality assurance program. The device will undergo the same manufacturing, packaging and sterilization procedures to devices currently marketed by CooperVision, Inc. Scottsville, NY manufacturing facility. Being similar with respect to indications for use, the risks of the subject device are the same as those normally attributed to the wearing of soft (hydrophilic) contact lenses on a daily wear basis.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is an abstract emblem resembling an eagle or bird in flight, rendered in black.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 9 2008
CooperVision Bonnie Tsymbal Director, Regulatory Affairs and Quality Assurance. 711 North Road Scottsville NY 14546
Re: K081865
Trade/Device Name: Proclear Multifocal Toric XR, Proclear Multifocal XR, Proclear Toric XR (omafilcon A) Daily Wear Soft (hydrophilic) Contact Lenses
Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL Dated: June 30, 2008 Received: September 15, 2008
Dear Ms. Tsymbal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M. B. Egelston, m.D.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Ooper Visione
## Indication for Use Statement
| 510(k) Number: | K081865 |
|----------------|------------------------------------------------------------------------------|
| Device Name: | Proclear Multifocal Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lenses |
| | Proclear Multifocal XR (omafilcon A) Soft (Hydrophilic) Contact Lenses |
| | Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lenses |
### Indication for Use:
Proclear Multifocal Toric XR (omafilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic or hyperopic, possess astigmatism of 3.00 diopters or less, and are presbyopic. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only).
Proclear Multifocal XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic or hyperopic and are presbyopic. The lenses may be worn by persons who exhibit astigmatism of 0.75 diopters or less that does not interfere with visual acuity. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only).
Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic or hyperopic, possess astigmatism of 5.00 diopters or less. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only).
Daily wear replacement schedules may vary from patient and should be decided by the eye care practitioner in consultation with their patients. The lenses are to be cleaned, rinsed and disinfected each time they are removed from the patients eye and discarded after the recommended wearing period prescribed by the eye care practitioner. The lenses may be disinfected using a chemical disinfection system.
Prescription Use X (Per 21 CFR Subpart D) AND/OR
Over-The-Counter (Per 21 CFR Subpart C)
PLEASE DO NO WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Myra Smith
ivision of Ophthalmic Ear. Nose and Throat Devises
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.