K081815 · Cytophil, Inc. · MIX · Jan 5, 2009 · Ear, Nose, Throat
Device Facts
Record ID
K081815
Device Name
VOCALIS GEL
Applicant
Cytophil, Inc.
Product Code
MIX · Ear, Nose, Throat
Decision Date
Jan 5, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3620
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Vocalis Gel is indicated for vocal fold medialization and vocal fold insufficiency that may be improved by injection of a soft tissue bulking agent. Vocalis Gel injection augments the size of the displaced or deformed vocal fold so that it may meet the opposing fold at the midline for improved phonation. Vocal fold insufficiency associated with serious aspiration difficulties may be an urgent indication. The product is intended to be durable for a minimum of one month.
Device Story
Vocalis Gel is a sterile, latex-free, non-pyrogenic, isotonic, clear gel injectable implant; composed of sterile water, glycerin, mannitol, and carbomer (USP). Device functions as a soft tissue bulking agent for vocal fold medialization; injected into displaced or deformed vocal folds to augment size and facilitate midline closure against the opposing fold. Used in clinical settings by physicians during injection laryngoplasty procedures; guided by nasopharyngoscopy. Provides durable augmentation for at least one month; allows for tissue infiltration over time. Benefits patients by improving phonation and addressing aspiration difficulties associated with vocal fold insufficiency.
Clinical Evidence
Bench testing only. Preclinical safety studies and animal implant studies conducted to evaluate biocompatibility, including irritation, sensitization, cytotoxicity, acute/sub-chronic toxicity, genotoxicity, and hemolysis. Results indicate the device is non-irritant and non-toxic.
Technological Characteristics
Injectable gel implant; components: sterile water (USP), glycerin (USP), mannitol (USP), carbomer (USP). High yield strength, cohesive, shear-thinning for fine-needle injection. Sterilized via steam (computer-controlled autoclave, overkill methodology to 100 SAL).
Indications for Use
Indicated for vocal fold medialization and insufficiency in patients requiring soft tissue bulking for improved phonation or management of aspiration difficulties.
Regulatory Classification
Identification
Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.
{1}------------------------------------------------
K081815
# JAN - 5 2009
# SECTION 4
# 510(k) Summary of Safety and Effectiveness
Vocalis Gel Trade Name: Common Name: Vocal Cord Medialization System, Vocal Cord Medialization Classification Name: Official Contact Name: Greg Johnson President & CEO Cytophil, Inc. Address: 5546 N Santa Monica Blvd Whitefish Bay, WI 53217 Phone: 414-961-7372 414-961-6167 Fax:
6/28/2008
#### 4.1 Intended Use
Date Prepared:
E-mail:
Vocalis Gel is indicated for vocal fold medialization and vocal fold insufficiency that may be improved by injection of a soft tissue bulking agent. Vocalis Gel injection augments the size of the displaced or deformed vocal fold so that it may meet the opposing fold at the midline for improved Vocal fold insufficiency associated with serious aspiration phonation. difficulties may be an urgent indication. The product is intended to be durable for a minimum of one month.
gjohnson@cytophil.com
#### 4.2 Product Description
Sterile, latex free, non-pyrogenic, high yield strength, isotonic, clear gel injectable implant. The gel consists primarily of sterile water for injection (USP), glycerin (USP) and mannitol (USP). The high yield strength is created by small amounts of carbomer (USP). The gel carrier allows tissue infiltration over time. All components are listed as GRAS (Generally Recognized as Safe 21 CFR 182). The character of the gel allows it to be very thick and cohesive but sheer to be easily injected through very fine needles with minimal force.
Cytophil, Inc.
Section 4
4-1
{2}------------------------------------------------
# 4.3 Substantial Equivalence
The following is the predicate device that is substantially equivalent to Vocalis Gel:
#### K080956
Modification to VF Gel Coapt Systems, Inc. 1820 Embarcadero Rd. Palo Alto, CA 94303
## K013243
Coaptite Laryngeal Augmentation System BioForm Medical, Inc. 4133 Courtney Road, Suite 10 Franksville, WI 53126
#### K070090
Radiesse Laryngeal Implant BioForm Medical, Inc. 1875 South Grant St., Suite 110 San Mateo, CA 94402
K071663 VF Long Term Coapt Systems, Inc. 1820 Embarcadero Rd. Palo Alto, CA 94303
### 4.4 Biocompatibility Evaluations
The battery of preclinical safety studies and animal implant studies show that Vocalis Gel is biocompatible when injected into soft tissues.
#### 4.5 Sterilization
Vocalis Gel is sterilized using steam. Processing is preformed by a contract sterilization company, Haemonetics, using a computer controlled autoclave system. Cycle parameters were validated using an overkill methodology to 100 SAL. Sterilization by the user is not required.
Cytophil, Inc.
Section 4
4-2
{3}------------------------------------------------
# 4.6 Pre-Clinical Tests Performed
In vivo and in vitro tests were performed to address irritation, sensitization, cytotoxicity, acute and sub-chronic toxicity, genotoxicity and hemolysis. Results identified Vocalis Gel as a nonirritant, and nontoxic with no concerns for long-term safety.
## 4.7 Risk Assessment
਼ ਸ
The primary risks with Vocalis Gel have been identified through a risk assessment procedure in accordance with EN 1441. The risks identified are primarily associated with nasopharyngoscopy and injection laryngoplasty.
#### 4.8 Summary
Vocalis Gel is a safe and effective implant used as a space filling material for soft tissue augmentation in laryngeal procedures for vocal fold medialization and augmentation.
Section 4
4-3
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the department's name around the perimeter. Inside the circle is an emblem featuring an abstract image of an eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Cytophil, Inc. c/o Mr. Greg Johnson President & CEO 5546 N Santa Monica Blvd Whitefish Bay, WI 53217
AN - 5 2009
Re: K081815
Trade/Device Name: Vocalis Gel Regulation Number: 21 CFR 874.3620 Regulation Name: Ear, Nose, and Throat Synthetic Polymer Material Regulatory Class: Class II Product Code: MIX Dated: December 12, 2008 Received: December 15, 2008
Dear Mr. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 -- Mr. Greg Johnson
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Malina B. Egleston, and
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K0818/5
# SECTION 2
# Indications for Use
# 510(k) Number: K081815
Device Name: Vocalis Gel
Indications for Use:
Vocalis Gel is indicated for vocal fold medialization and vocal fold insufficiency that may be improved by injection of a soft tissue bulking agent. Vocalis Gel injection augments the size of the displaced or deformed vocal fold so that it may meet the opposing fold at the midline for improved phonation. Vocal fold insufficiency associated with serious aspiration difficulties may be an urgent indication. The product is intended to be durable for a minimum of one month.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use or Over-the-Counter Use X (Per 21 CFR 801.109
Daniel Cearron
(Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices
510(k) Number K081815
Section 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.