K081814 · Sterilmed, Inc. · ERL · Dec 17, 2008 · Ear, Nose, Throat
Device Facts
Record ID
K081814
Device Name
REPROCESSED ENT SHAVERS
Applicant
Sterilmed, Inc.
Product Code
ERL · Ear, Nose, Throat
Decision Date
Dec 17, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The reprocessed ENT shavers are intended to be used with a reusable hand piece and are designed for use in various ENT, head and neck surgeries, and otoneurologic procedures.
Device Story
Reprocessed ENT shavers are powered dissectors used for bone and tissue removal during ENT, head and neck, and otoneurologic surgeries. The device is inserted into a reusable hand piece (not included). Operation involves mechanical cutting/shaving of tissue. Used in clinical settings by surgeons. Reprocessing restores the device to original functional specifications. Benefits include providing a cost-effective, sterile, and functional surgical instrument for tissue debridement and bone removal.
Clinical Evidence
Bench testing only. Functional testing and process validation were performed to verify cleaning, sterilization, and packaging integrity. Visual and functional testing were conducted on all products produced to ensure performance equivalence.
Technological Characteristics
Powered surgical dissector (shaver) for bone and tissue removal. Materials and construction are identical to the original manufacturer's specifications. Device is intended for use with a reusable hand piece. Sterilization is performed as part of the reprocessing cycle.
Indications for Use
Indicated for patients undergoing ENT, head and neck, and otoneurologic surgeries, including sinus applications (polypectomy, ethmoidectomy/sphenoethmoidectomy, septoplasty, antrostomy, endoscopic DCR, frontal sinus trephination/irrigation/drill out, septal spur removal, trans-sphenoidal procedures), head and neck procedures (rhinoplasty, lipodebridement, soft tissue shaving, acoustic neuroma removal), nasopharyngeal and laryngeal procedures (tonsillectomy, tracheal procedures, adenoidectomy, laryngeal lesion de-bulking, laryngeal polypectomy), and otology procedures (mastoidotomy, mastoidectomy).
Regulatory Classification
Identification
An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.
Predicate Devices
Gyrus Diego
Medtronic Xomed XPS
Submission Summary (Full Text)
{0}------------------------------------------------
K081814
Image /page/0/Picture/1 description: The image shows the logo for SterilMed, INC. The logo is in black and white and features the company name in a bold, sans-serif font. Below the company name, there is smaller text that reads "Medical Device Reprocessing" and "Small Equipment & Instrument Repair."
510(K) PREMARKET NOTIFICATION SUBMISSION JUNE 25, 2008 FOR REPROCESSED ENT SHAVERS
#### SUMMARY AND CERTIFICATION II.
DEC 1 7 2008
#### 510(k) Summary A.
| Submitter: | SterilMed, Inc. |
|------------------------|------------------------------------------------------------------------------------------------------------|
| Contact Person: | Joshua Clarin<br>11400 73rd Avenue North<br>Maple Grove, MN 55369<br>Ph: 612-644-8402<br>Fax: 763-488-3350 |
| Date Prepared: | June 25, 2008 |
| Trade Name: | Reprocessed ENT Shavers |
| Classification Name: | Ear, Nose and Throat Electrical or Pneumatic Surgical Drill |
| Classification Number: | Class II, 21 CFR 874.4250 |
| Product Code: | ERL |
| Predicate<br>Devices: | The reprocessed ENT shavers are substantially equivalent to Gyrus Diego® and Medtronic<br>Xomed XPS® shavers. |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Description: | SterilMed's reprocessed ENT shavers are powered dissectors inserted into a reusable hand<br>piece and designed to be used in the removal of bone and tissue in various ENT, head and neck<br>surgeries, and otoneurologic procedures. These devices were originally manufactured by Gyrus<br>and Medtronic. |
| | Note: Only the shaver is the subject of this submission, the reusable hand piece and any other<br>related equipment are not included in the scope of this submission. |
| Intended Use: | The reprocessed ENT shavers are intended to be used with a reusable hand piece and are<br>designed for use in various ENT, head and neck surgeries, and otoneurologic procedures. |
| Functional and<br>Safety Testing: | Representative samples of reprocessed ENT shavers were tested to demonstrate appropriate<br>functional characteristics. Process validation testing was performed to validate the cleaning and<br>sterilization procedures as well as device packaging. In addition, the manufacturing process<br>includes visual and validated functional testing of all products produced. |
| Conclusion: | The reprocessed ENT shavers are substantially equivalent to Gyrus Diego® and Medtronic<br>Xomed XPS® blades. |
| | This conclusion is based upon the devices' similarities in functional design (principle of<br>operation), materials, indications for use and methods of construction. |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for SterilMed INC. The logo is in bold, black font. Underneath the company name, it says "Medical Device Reprocessing" and "Small Equipment & Instrument Repair".
.
.
.
## 510(K) PREMARKET NOTIFICATION SUBMISSION JUNE 25, 2008 FOR REPROCESSED ENT SHAVERS
K081814
:
-
#### B. Premarket Notification Statement
11 - 11 - 11
In lieu of a Premarket Notification Statement, a 510(k) Summary has been provided. Refer to Section II.A for 510(k) summary and certification.
:
: .
:
.
: 上一篇:
1999 - 1999
:
:
CONFIDENTIAL & PROPRIETARY INFORMATION OF STERILMED, INC.
:
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
## Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SterilMed. Inc. c/o Dennis Toussaint Director Regulatory Affairs 11400 73rd Avenue, N. Suite 100 Maple Grove, MN 55369
# DEC 1 7 2008
K081814 Re: Trade/Device Name: Reprocessed ENT Shavers Regulation Number: 21 CFR 874.4250 Regulation Name: Ear, Nose, and Throat electric or pneumatic surgical drill Regulatory Class: Class II Product Code: ERL Dated: December 12, 2008 Received: December 15, 2008
Dear Mr. Toussaint:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M. B. Egolints, mD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K081814
# Indications for Use
510(k) Number (if known):_K081814
Device Name: Reprocessed ENT Shavers
### Indications for Use:
Indications tor Use:
The reprocessed ENT shavers are intended to be used with a reusable hand piece and are designed for use in various ENT, head and neck surgeries and otoneurologic procedures.
Sinus Applications:
- Polypectomy .
- Ethmoidectory/Sphenoethmoidectomy
- Septoplasty
- Antrostomy
- Endoscopic DCR .
- Frontal Sinus Trephination and Irrigation
- Frontal Sinus Drill Out
- Septal Spur Removal
- Trans-Sphenoidal Procedures
## Head and Neck Procedures
- Rhinoplasty .
- Lipodebridement in the Maxillary and Mandibular Region .
- Soft Tissue Shaving
- Acoustic Neuroma Removal
Nasopharyngeal and Laryngeal Procedures
- Tonsillectomy .
- Tracheal Procedures .
- Adenoidectomy
- Laryngeal Lesion De-Bulking
- Laryngeal Polypectomy .
# Otology Procedures
- Mastoidotomy .
- Mastoidectomy
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sir Nadler
(Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices
510(k) Number k081814
{5}------------------------------------------------
# K081814
:
:
. And the same of
-
. . . . . . .
Carlos Concession
:
.
| Manufacturer | Model # |
|-----------------|-------------|
| Gyrus | 7013-8100 |
| | 7013-8000 |
| | 7013-8035 |
| | 7013-8001 |
| | 7013-8002 |
| | 7013-8003 |
| Medtronic Xomed | 18-84004HR |
| | 18-84004 |
| | 18-83504HR |
| | 18-83504 |
| | 18-82904HRE |
| | 18-82904 |
| | 18-84002HRE |
| | 18-84002 |
| | 18-83502HRE |
| | 18-83502 |
| | 18-82902HRE |
| | 18-82902 |
| | 18-84005HRE |
| | 18-84005 |
| | 18-82905HRE |
| | 18-82905 |
| | 18-82040HR |
| | 18-82040 |
| 18-82940HR | |
| 18-82940 | |
:
.
.
、
List of Devices included in this Premarket Notification Submission - 510(k) K081814
:
and the state of the states of the states
:
:
:
and the comments of the comments of the country of
:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.