K081744 · Optimedica Corporation · GEX · Sep 9, 2008 · General, Plastic Surgery
Device Facts
Record ID
K081744
Device Name
PASCAL SYNTHESIS DELIVERY SYSTEM
Applicant
Optimedica Corporation
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 9, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PASCAL Synthesis Delivery System is a stand alone scanning delivery system that, when connected to compatible commercially available laser photocoagulators, allows fully functional scanning Photocoagulation. When connected to a compatible laser system, the PASCAL Synthesis Delivery System is indicated for use in the treatment of ocular pathology in both the posterior and anterior segments, retinal photocoagulation, pan retinal photocoagulation, focal photocoagulation and grid photocoagulation for vascular and structural abnormalities of the retina and choroid including: Proliferative and nonproliferative diabetic retinopathy . Macular edema . Choroidal neovascularization . Branch and central retinal vein occlusion . Age-related macular degeneration . Lattice degeneration . Retinal tears and detachments . Iridotomy, iridectomy and trabeculoplasty in angle closure and open angle . glaucoma.
Device Story
Scanning laser delivery system; connects to external laser photocoagulators to enable multi-spot pattern scanning. Components: slit lamp with LCD/touchscreen GUI; scanner control module with power supply, electronics, and optical/electrical connections. Operated by ophthalmologists in clinical settings. Physician selects patterns via GUI; device controls laser emission and sequential scanning of spots. Enables faster retinal photocoagulation compared to single-spot methods. Treatment aborted via footswitch. Benefits: precise, rapid delivery of laser patterns for retinal/choroidal vascular and structural abnormalities.
Clinical Evidence
Bench testing only. Performance verification conducted by connecting device to Zeiss Visulas 532s and Alcon Ophthalas 532 Eyelite. Evaluated power delivery, pulse widths, scan patterns, spot size, and beam uniformity. Results demonstrated equivalence to PASCAL Photocoagulator predicate. System-level testing included software V&V, life testing, environmental/shipping, and EMC/EMI/electrical safety compliance.
Technological Characteristics
Scanning laser delivery system; integrates with slit lamp. Features LCD/touchscreen GUI. Operates in Single Spot or Pattern Mode. Delivers laser energy from external 532nm laser sources. Connectivity: optical/electrical interface to compatible laser systems. Sterilization: N/A (non-patient contacting). Software: pattern scanning control algorithm.
Indications for Use
Indicated for patients requiring treatment of ocular pathology in posterior/anterior segments, including proliferative/nonproliferative diabetic retinopathy, macular edema, choroidal neovascularization, retinal vein occlusions, age-related macular degeneration, lattice degeneration, retinal tears/detachments, and glaucoma (iridotomy, iridectomy, trabeculoplasty).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
1744
#### 2.0 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT | OptiMedica Corporation |
|----------------|--------------------------------------------------------|
| TRADE NAME: | PASCAL Synthesis™ Delivery System |
| COMMON NAME: | Scanning Photocoagulator Delivery System |
| DEVICE | GEX |
| CLASSIFICATION | 21 CFR 878.4810<br>Laser instrument, surgical, powered |
#### 2.1 SUBSTANTIALLY EQUIVALENT
The PASCAL Synthesis Delivery System is substantially equivalent in intended use and has the same technological characteristics as the delivery system incorporated into PASCAL Photocoagulator (K043486). The PASCAL Synthesis Delivery System is also substantially equivalent to the Infinitech Multi-Spot Slit Lamp Laser Adapter (K971950) by Infinitech, Inc. in functionality and technology in that when this device is connected to another laser, it enables the user to perform multi-spot laser photocoagulation.
#### 2.2 DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION
The PASCAL Synthesis Delivery System is a scanning laser delivery system that enables the use of proprietary pattern scanning technology of the PASCAL multi-spot photocoagulator (a predicate device) with other commercially available laser platforms. This offers existing commercially available laser platforms the ability to deliver a full spectrum of pattern scanning options. The PASCAL Synthesis Delivery System is intended for use by trained ophthalmologists for diagnosis and treatment of ocular pathology.
The PASCAL Synthesis Delivery System consists of the following system components:
- 1) Scanning laser delivery system integrated into an Optimedica slit lamp with LCD/Touchscreen GUI that can be positioned by the user for optimal access and viewing.
- 2) Scanner control module with scanner controls, power supply, electronics, optical fiber & electrical connections integrated into a slit lamp table.
{2}------------------------------------------------
The PASCAL Synthesis Delivery System is compatible with the following commercially available laser systems:
- Alcon Ophthalas 532 Eyelite Laser Photocoagulator ●
- Zeiss Visulas 532s .
# INDICATION FOR USE 2.3
The PASCAL Synthesis Delivery System is a stand alone scanning delivery system that, when connected to compatible commercially available laser photocoagulators, allows fully functional scanning Photocoagulation.
When connected to a compatible laser system, the PASCAL Synthesis Delivery System is indicated for use in the treatment of ocular pathology in both the posterior and anterior segments, retinal photocoagulation, pan retinal photocoagulation, focal photocoagulation and grid photocoagulation for vascular and structural abnormalities of the retina and choroid including:
- Proliferative and nonproliferative diabetic retinopathy .
- Macular edema .
- Choroidal neovascularization ●
- Branch and central retinal vein occlusion .
- Age-related macular degeneration .
- Lattice degeneration .
- Retinal tears and detachments .
- Iridotomy, iridectomy and trabeculoplasty in angle closure and open angle . glaucoma.
## 2.4 CONTRAINDICATIONS
Laser surgery is contraindicated when an appropriate procedure cannot be performed safely. This occurs when tissue targets cannot be visualized properly. Under such circumstances, an important tissue structure adjacent to a target tissue might be photocoagulated inadvertently. Corneal opacities, cataract formation, and vitreous hemorrhage can all interfere with the laser surgeon's view of appropriate target structures. Treatment should be delayed until visualization has adequately improved.
## 2.5 TECHNICAL CHARACTERISTICS
All user parameters and technological characteristics of the PASCAL Synthesis Delivery System are consistent with those of the predicate device, the OptiMedica PASCAL Photocoagulator. Specifically, the PASCAL Synthesis Delivery System is the same delivery system as the delivery system in the cleared PASCAL Photocoagulator (K043486). Both delivery systems utilize a scanner to enable faster retinal photocoagulation. Both devices may operate in an equivalent Single Spot Mode as well as in a Pattern Mode, where the aim beam is rapidly scanned allowing the physician to visualize a user-selected pattern of spots at the appropriate location for treatment. Similar to the second predicate device, the Infinitech Multi-Spot laser adapter, the PASCAL
{3}------------------------------------------------
Synthesis Delivery System is a stand alone unit that can perform multiple spot photocoagulation when connected to a compatible external laser system. Both the PASCAL Synthesis Delivery System and the Infinitech Multi-Spot laser adapter deliver laser power from the separate 532mm laser system they are connected to. Once activated by the user, the PASCAL Synthesis Delivery System delivers a predetermined pattern by sequentially scanning the placement of the laser spots and controlling the emission of the individual pulses of laser light. Likewise, treatment may be aborted by releasing the footswitch.
The PASCAL Synthesis Delivery System has been tested and found to be compatible with the Zeiss Visulas 532s (K013402) and the Alcon Ophthalas 532 Eyelite Laser Photocoagulator (K962592) laser systems. At this time, no other systems have been tested for use with the PASCAL Synthesis Delivery System.
## 2.6 PERFORMANCE DATA
System performance verification testing was completed to verify that measured system performance complies with specifications and requirements. This was accomplished by connecting the PASCAL Synthesis Delivery System to the Zeiss Visulas 532s and the Alcon Ophthalas 532 Eyelite Laser Photocoagulator laser systems and performing functionality testing to verify that measured system performance complies with defined system specifications and performance requirements. This testing and evaluation was performed at OptiMedica's facility in Santa Clara CA and at outside test facilities as required.
Extensive system level testing was also completed, including system and software verification & validation testing, system life testing, system environmental & shipping testing, EMC/EMI and Electrical safety compliance testing along with system performance testing. System performance testing included evaluation of power delivered, pulse widths, scan patterns, spot size & beam uniformity. All test results demonstrated that performance of the PASCAL Synthesis Delivery System, when connected to these two commercially-available laser systems, was equivalent to the performance of the PASCAL Photocoagulator that incorporates the same delivery system. Verification testing conducted on the PASCAL Synthesis Delivery System also demonstrated compliance with system performance requirements and specifications: this testing further confirmed that this device consistently and accurately produces laser outputs that are equivalent to those produced by the PASCAL Photocoagulator predicate device.
## 2.7 BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE
The indications for use for the PASCAL Synthesis Delivery System are the same as those cleared for the predicate device, the PASCAL Photocoagulator (K043486). The materials used for the manufacture of the PASCAL Synthesis Delivery System and the delivery system of the predicate device, the PASCAL Photocoagulator, are also the same. Testing of the PASCAL Synthesis Delivery System has demonstrated that it is functionally equivalent to the predicate device.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" arranged in a circular pattern around the symbol. The caduceus is black, and the text is in a smaller, sans-serif font.
SEP 0 9 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OptiMedica Corporation % ClinReg Consulting Services, Inc. Judy F. Gordon, D.V.M. 733 Bolsana Drive Laguna Beach, California 92651
Re: K081744
Trade/Device Name: PASCAL Synthesis™ Delivery System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: June 18, 2008 Received: June 19, 2008
Dear Dr. Gordon: .
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Judy F. Gordon, D.V.M.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark M Mulkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): _ |< 08 | 7 44
Device Name: PASCAL Synthesis" Delivery System
Indications for Use:
The PASCAL Synthesis Delivery System is a stand alone scanning delivery system that, when connected to compatible commercially available laser photocoagulators, allows fully functional scanning photocoagulation.
When connected to a compatible laser system, the PASCAL Synthesis Delivery System is indicated for use in the treatment of ocular pathology in both the posterior and anterior segments, retinal photocoagulation, pan retinal photocoagulation, focal photocoagulation and grid photocoagulation for vascular and structural abnormalities of the retina and choroid including:
- Proliferative and nonproliferative diabetic retinopathy �
- . Macular edema
- Choroidal neovascularization .
- Branch and central retinal vein occlusion .
- Age-related macular degeneration .
- Lattice degeneration .
- Retinal tears and detachments .
- Iridotomy, iridectomy and trabeculoplasty in angle closure and open . angle glaucoma.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X | AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |
|-------------------------------------------------|---|--------|------------------------------------------------|--|
|-------------------------------------------------|---|--------|------------------------------------------------|--|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE
OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative, Page 1 of 1
and Neurological Devices
| 510(k) Number | K081744 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.