K081738 · Health & Life Co., Ltd. · CAF · Sep 17, 2008 · Anesthesiology
Device Facts
Record ID
K081738
Device Name
ULTRASONIC NEBULIZER SYSTEM, MODEL HL100
Applicant
Health & Life Co., Ltd.
Product Code
CAF · Anesthesiology
Decision Date
Sep 17, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Intended Use
The HL100 Ultrasonic Nebulizer System utillizes the state-of-the-art elecctrospray technology that sprays liquid medication in aerosol form and deliveries it directly to the adult and pediatric patients who suffer from asthma, Chronic Obstructive Pulmonary Disease (COPD) such as emphysema and chronic bronchitis, or other respiratory diseases that are characterized by obstruction to air flow for breathing.
Device Story
Mesh screen ultrasonic nebulizer; converts liquid medication into aerosol droplets for inhalation. Operates via ultrasonic piezoelectric member vibrating an aperture plate (mesh) when energized by high-frequency alternating voltage. Components include main unit (control circuitry) and single-patient reusable medication chamber. Powered by 115 VAC adapter or two AA alkaline batteries. Patient interface includes mouthpiece or optional mask. Used in clinical or home settings; operated by patient or caregiver. Provides aerosolized medication directly to respiratory tract to treat airflow obstruction; benefits patients by facilitating delivery of inhaled therapies.
Clinical Evidence
No clinical data. Substantial equivalence based on nonclinical bench testing comparing performance to the predicate device.
Technological Characteristics
Mesh screen ultrasonic nebulizer; piezoelectric transducer; aperture plate mesh. Powered by 115 VAC or 2x AA alkaline batteries. Single-patient reusable medication chamber. Form factor includes main unit with control circuitry and patient interface (mouthpiece/mask).
Indications for Use
Indicated for adult and pediatric patients with asthma, COPD (emphysema, chronic bronchitis), or other respiratory diseases characterized by airflow obstruction.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
# SEP 1 7 2008
510(k) Summary - Health and Life Model HL100 Ultrasonic Nebulizer System
# MANUFACTURER IDENTIFICATION 0.0
| Manufacturer: | Health and Life Corporation |
|---------------|-------------------------------------|
| Address: | 9F, No. 186, Jian-Yi Rd |
| City, State: | Chung Ho City, Taipei Hsien, Taiwan |
#### 1.0 PRODUCT NAME
| Common or usual name | |
|-------------------------|--------------------------------------|
| or classification name: | Nebulizer (Direct Patient Interface) |
| Trade or proprietary | |
| or model name: | HL100 Ultrasonic Nebulizer System |
#### 2.0 PRODUCT CLASSIFICATION
| Device Class: | II |
|-----------------------|----------------|
| Product Code: | 73 CAF |
| C.F.R. Section: | |
| Classification Panel: | Anesthesiology |
#### 3.0 PREDICATE DEVICE
Omron Model NE-U22V MircroAir Nebulizer (K062263)
### 4.0 DEVICE DESCRIPTION
The Health and Life Model HL100 is a mesh screen ultrasonic nebulizer that operates in an identical fashion as other mesh screen ultrasonic nebulizers, including the referenced predicate device. The device creates aerosols of liquid medication of droplets from a mesh (aperture plate) vibrated at ultrasonic frequencies. As with other mesh screen ultrasonic nebulizers, the mesh is caused to vibrate by fluidic contact with a ultrasonic piezoelectric member that is energized by application of a high frequency alternating voltage.
The HL100 Ultrasonic Nebulizer System consists primarily of a Main Unit and a Medication Chamber. The Main Unit and the Medication Chamber are single patient re-usable. The Main Unit contains all control circuitry and is powered by a cable connecting to a 115 VAC power source with the AC adapter or two AA alkaline batteries. No medication comes into contact with the Main Unit, only the Medication Chamber. The patient interface consists of a mouthpiece or an optional mask. When the device is turned on, the ultrasonic piezoelectric member vibrates, causing the mesh to vibrate. Liquid medication in the reservoir is caused to be aerosolized by
{1}------------------------------------------------
Image /page/1/Figure/1 description: This image shows a diagram of a nebulizer machine and its various components. The diagram labels the mouthpiece, mask adapter, medication chamber, inhalation mask, connecting tube, AC adapter, and main unit. The main unit has a low-power indicator, power indicator, and start/stop button. The diagram also shows the bottom and rear views of the main unit, highlighting the battery cover, push button, and AC adapter inlet.
the vibrating mesh into the patient interface where the aerosol particles can be inhaled by the patient.
## 5.0 INTENDED USE
The HL100 Ultrasonic Nebulizer System utillizes the state-of-the-art elecctrospray technology that sprays liquid medication in aerosol form and deliveries it directly to the adult and pediatric patients who suffer from asthma, Chronic Obstructive Pulmonary Disease (COPD) such as emphysema and chronic bronchitis, or other respiratory diseases that are characterized by obstruction to air flow for breathing.
#### 6.0 COMPARISON TO PREDICATE DEVICE
The Health and Life Model HL100 Ultrasonic Nebulizer System is substantially equivalent to the Omron Model NE-U22V MircroAir Nebulizer (K062263). Nonclinical tests have been performed comparing the HL100 Ultrasonic Nebulizer System to the predicate device and have shown that the performance of the HL100 Ultrasonic Nebulizer System is substantially equivalent to the performance of the predicate device. No clinical tests were conducted
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its wings spread, symbolizing protection and care. The eagle is positioned in the center of a circular emblem. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 7 2008
Health & Life Company, Limited C/O Mr. Samuel Piper President/Chief Executive Officer Piper Medical, Incorporated P.O. Box 993 Carmichael, California 95609
Rc: K081738
Trade/Device Name: HL100 Ultrasonic Nebulizer System Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: June 17, 2008 Received: June 19, 2008
Dear Mr. Piper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Piper
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Parl 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Thomaels-Lindens foxll
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: HL100 Ultrasonic Nebulizer System
Indications For Use: The HL100 Ultrasonic Nebulizer System utillizes the state-of-the-art elecctrospray technology that sprays liquid medication in aerosol form and deliveries it directly to the adult and pediatric patients who suffer from asthma, Chronic Obstructive Pulmonary Disease (COPD) such as emphysema and chronic bronchitis, or other respiratory diseases that are characterized by obstruction to air flow for breathing.
Prescription Use X
AND/OR Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mr. 76
Page 1 of 1
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.