← Product Code [DQA](/productcode/DQA) · K081659

# RAINBOW ADHESIVE CO-OXIMETRY SENSOR, MODELS RI-25-L, RI-25, RI-20 AND RI-20-L (K081659)

_Masimo Corporation · DQA · Sep 9, 2008 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K081659

## Device Facts

- **Applicant:** Masimo Corporation
- **Product Code:** [DQA](/productcode/DQA.md)
- **Decision Date:** Sep 9, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The Rainbow Adhesive Sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), and/or total hemoglobin (SpHb). The Rainbow Adhesive Sensors are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.

## Device Story

Disposable adhesive sensors; attach to finger, hand, foot, or toe; interface with Masimo Rainbow SET pulse CO-Oximeter monitors. Utilize 12-wavelength optical emitters and detectors to measure tissue light absorption; noninvasive assessment of hemoglobin species and pulse rate. Used by clinicians or patients in hospitals, mobile, or home environments. Output displayed on compatible monitor; provides real-time physiological data for clinical decision-making regarding oxygenation and hemoglobin status. Benefits include continuous monitoring without invasive blood sampling.

## Clinical Evidence

Clinical studies on healthy adult volunteers (induced hypoxia/motion) and hospitalized neonates. Validated against laboratory CO-Oximeter. Accuracy: SpO2 <2% (70-100% range), <3% (60-80% range); SpCO <3% (1-40% range); SpMet <1% (1-15% range); SpHb <1 g/dL (8-17 g/dL range). Neonatal SpO2 accuracy <3% (70-100% range), SpMet <1% (0-2.5% range).

## Technological Characteristics

12-wavelength optical sensor; disposable adhesive form factor for finger/hand/foot/toe attachment. Compatible with Masimo Rainbow SET monitors. Materials identical to predicate K071024. Non-sterile, single-patient use. Operates via noninvasive optical tissue assessment.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- Rainbow DCI and Direct Connect Pulse CO-Oximetry Sensors ([K080238](/device/K080238.md))
- Rainbow Adhesive CO-Oximetry Sensors ([K071024](/device/K071024.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K081659

# 510(k) SUMMARY

## SEP 0 9 2008

| Submitted by:                         | Masimo Corporation<br>40 Parker<br>Irvine, CA 92618<br>949-297-7000<br>FAX 949-297-7001                                                                                             |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact:                      | Marguerite Thomlinson, Manager of Regulatory Affairs                                                                                                                                |
| Date Summary Prepared:                | June 9, 2008                                                                                                                                                                        |
| Trade Name                            | Rainbow Adhesive Pulse CO-Oximeter Sensors                                                                                                                                          |
| Common Name                           | Oximeter Sensor                                                                                                                                                                     |
| Classification Name and Product Code: | Oximeter (74DQA) (870.2700)<br>Cable, Transducer and Electrode (74DSA) (870.2900)<br>Carbon Monoxide Test System (JKS) (862.3220)                                                   |
| Substantially Equivalent Devices:     | Masimo Rainbow SET Radical 7/ Rad 87/ Rad 57t Pulse<br>CO-Oximeters and Accessories, 510(k) Number K080238<br>Masimo Rainbow Adhesive CO-Oximetry Sensors,<br>510(k) Number K071024 |

#### Device Description

The Rainbow Adhesive Sensors are fully compatible disposable sensor for use with Masimo Rainbow SET and Masimo Rainbow SET compatible pulse CO-Oximeter monitors. They represent a design change to the Masimo Rainbow DCI and Direct Connect (Reusable) Sensors in the K080238 filing and the Rainbow Adhesive CO-Oximetry Sensors in the K071024 filing.

The Rainbow Adhesive Sensors in this filing have the same intended use/indications for use and performance specifications as the Rainbow Reusable Sensors in the K080238 filing. However, The Rainbow Adhesive Sensors in this filing are similar in construction to the Rainbow Adhesive CO-Oximetry Sensors in the K071024 filing. The main difference is that the sensors in the K071024 filing have 8wavelengths and the sensors in this filing have 12-wavelengths.

Similar to the 8-wavelength sensors in this filing include the emitter and detector assemblies connecting to the flex circuit, and an adhesive bandage to allow the sensor to be attached to the patient's finger, hand, foot or toe. Sensor sizing for use with adult, pediatric, infant, and neonatal patients are the

{1}------------------------------------------------

## 510(k) SUMMARY

same for the sensors in the K071024 and for the sensors in this filing. The patient-contact materials in the Rainbow Adhesive Sensors in this filing are the same that is used in the K071024 filing. Also similar to the sensors in the K071024 filing, the Rainbow Adhesive Sensors in this filing are supplied non-sterile for single patient use.

#### Predicate Devices

Rainbow DCI and Direct Connect Pulse CO-Oximetry Sensors (K080238) and Rainbow Adhesive CO-Oximetry Sensors (K071024).

#### Intended Use

The Rainbow Adhesive Sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO₂), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), and/or total hemoglobin (SpHb). The Rainbow Adhesive Sensors are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.

### Technology Comparison

The Rainbow Adhesive Sensors are substantially equivalent in intended use, design, principles of operation, materials, and performance to predicate sensors and operate on identical principles of noninvasive optical assessment of tissue oxygenation using emitters and detectors.

The Rainbow Adhesive Sensors are designed, configured, and manufactured for full compatibility with Masimo Rainbow SET and Masimo Rainbow SET compatible pulse CO-Oximeters. The Rainbow Adhesive Sensors are constructed of similar materials and components of equivalent specifications as used in the predicate devices.

The accuracy of the Rainbow Adhesive Sensors is equivalent to those of the predicate devices.

#### Performance Testing

#### Biocompatibility

Test results of all patient-contact materials used in the Rainbow Adhesive Sensors demonstrated that the materials were non-toxic, non-irritating, and non-sensitizing.

#### Environmental

Applicable environmental testing per the Reviewers Guidance for Premarket Submissions -November 1993, i.e. electrical, mechanical and environmental were performed and all tests passed.

#### Clinical

Clinical studies were performed using Masimo Rainbow SET technology with Rainbow Adhesive Sensors on healthy adult volunteer subjects during motion conditions who were

{2}------------------------------------------------

# 510(k) SUMMARY

subjected to a progressive induced hypoxia and measuring the arterial hemoglobin saturation value with the instruments against the arterial hemoglobin oxygen determined from arterial blood samples with a laboratory CO-Oximeter.

Clinical testing of the Rainbow Adhesive sensors resulted in an accuracy of less than 2% SpO2 Авмя in the range of 70%-100% SaO2 and less than 3% SpO2 AcMs in the range of 60%-80% SaO2 for adults, pediatrics, and infants.

The Masimo Rainbow SET technology with Rainbow Adhesive Sensors have been validated in human blood studies on adult volunteers against a laboratory CO-Oximeter from 1-40% for carboxyhemoglobin, 1-15% for methemoglobin, and 8-17 g/dL for total hemoglobin. Clinical testing of the Rainbow Adhesive sensors resulted in an accuracy of less than 3% SpCO Aggs in the range of 1%-40% SaCO, an accuracy of less than 1% SpMet Aggs in the range of 1%-15% SaMet, and an accuracy of less than 1 g/dL SpHb ARMs in the range of 8-17 g/dL tHb.

Additional clinical studies were performed using Masimo Rainbow SET technology with Rainbow Adhesive Sensors on hospitalized neonatal patients resulted in an accuracy of less than 3% SpO2 ARMs in the range of 70%-100% SaO2 and less than 1% SpMet ARMs in the range of 0% -2.5% SaMet.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, clutching a staff entwined with a serpent. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 0 9 2008

Ms. Marguerite Thomlinson Manager of Regulatory Affairs Masimo Corporation 40 Parker Irvine, California 92618

Re: K081659

Trade/Device Name: Rainbow Adhesive Pulse CO-Oximeter Sensors Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA, JKS, DPZ Dated: August 28, 2008 Received: September 2, 2008

Dear Ms. Thomlinson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Thomlison

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Thimothy Penley

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

### Indications for Use

510(k) Number (if known):

Device Name: Rainbow Adhesive Pulse CO-Oximeter Sensors

Indications For Use:

The Rainbow Adhesive Sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), and/or total hemoglobin (SpHb). The Rainbow Adhesive Sensors are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals,

hospital-type facilities, mobile, and home environments.

shl

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K081659

| Prescription Use _X_           | AND/OR | Over-The-Counter Use _         |
|--------------------------------|--------|--------------------------------|
| (Per 21 CFR 801.109 Subpart D) |        | (Per 21 CFR 801.109 Subpart C) |

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

### Concurrence of CDRH, Office of Device Evaluation (ODE)

---

**Source:** [https://fda.innolitics.com/device/K081659](https://fda.innolitics.com/device/K081659)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
