SPARC SYSTEM / MONARC, MONARC C, MONARC + SYSTEMS

K081613 · American Medical Systems, Inc. · OTN · Sep 17, 2008 · General, Plastic Surgery

Device Facts

Record IDK081613
Device NameSPARC SYSTEM / MONARC, MONARC C, MONARC + SYSTEMS
ApplicantAmerican Medical Systems, Inc.
Product CodeOTN · General, Plastic Surgery
Decision DateSep 17, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Intended for the placement of pubourethral sling for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and /or intrinsic sphincter deficiency. Intended for the placement of suburethral mesh for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and /or intrinsic sphincter deficiency.

Device Story

The Sparc and Monarc systems are sterile, single-use surgical kits designed for the treatment of female stress urinary incontinence. Each kit consists of two stainless steel, curved needle passers and a mesh sling assembly. During the surgical procedure, the physician uses the needle passers to position the mesh sling suburethrally or pubourethrally to provide support to the urethra. This mechanical support addresses urethral hypermobility or intrinsic sphincter deficiency, thereby reducing incontinence. The device is intended for use in a clinical or surgical setting by a qualified physician.

Clinical Evidence

Bench testing only. Biocompatibility and performance requirements were evaluated and found to be substantially equivalent to the predicate devices.

Technological Characteristics

Sterile, single-use surgical kit. Components include stainless steel curved needle passers and a polymeric surgical mesh sling. Class II device (21 CFR 878.3300).

Indications for Use

Indicated for female patients with stress urinary incontinence (SUI) caused by urethral hypermobility and/or intrinsic sphincter deficiency.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # K081613 ## SEP 1 7 2008 | Submitter: | American Medical Systems (AMS)<br>10700 Bren Road West<br>Minnetonka, MN 55343 | |------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Sarah Peterson<br>Phone: 952.930.6431<br>Fax: 952.930.5785 | | Device Common Name: | Surgical Mesh | | Device Trade Name: | SPARCTM System, Monarc® System,<br>Monarc® + System, and Monarc® C System | | Device Classification/<br>Classification Name: | Class II, 21 CFR Part 878.3300<br>Surgical Mesh, polymeric (OTN) | | Predicate Device: | SPARCTM System (K041948), Monarc® System,<br>Monarc® + System, and Monarc® C System<br>(K051530) | | Indications for Use | | | Sparc System: | Intended for the placement of pubourethral sling for the<br>treatment of female stress urinary incontinence (SUI) resulting<br>from urethral hypermobility and /or intrinsic sphincter<br>deficiency | #### 510(k) Summary Statement a maa | Monarc®,<br>Monarc® +,<br>Monarc® C<br>Systems: | Intended for the placement of suburethral mesh for the<br>treatment of female stress urinary incontinence (SUI) resulting<br>from urethral hypermobility and /or intrinsic sphincter<br>deficiency. | |-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| #### Device Description The Sparc, Monarc®, Monarc® +, and Monarc® C Systems are sterile, single use procedure kits that consist of two stainless steel, curved needle passers and a mesh sling assembly. #### Summary of Testing The components of the Sparc, Monarc, Monarc +, and Monarc C Systems have been tested for biocompatibility and performance requirements and found to be substantially equivalent to the predicate devices. {2}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular emblem with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a bird in flight, composed of three curved lines. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 SEP 2 8 20:2 American Medical Systems, Inc. % Ms. Sarah J. P. Meyer Regulatory Affairs Specialist 10700 Bren Road West MINNETONKA MN 55343 Re: K081613 > Trade/Device Name: Sparc System, Monarc® System, Monarc® + System, and Monarc® C System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: OTN Dated: September 4, 2008 Received: September 5, 2008 Dear Ms. Meyer: This letter corrects our substantially equivalent letter of September 17, 2008. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {3}------------------------------------------------ comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K081613 pg 1 of 2 ### Indications for Use Statement | 510(k) Number:<br>(if known) | | |------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | Monarc® System, Monarc® + System, and Monarc® C<br>System | | Indications For Use: | Intended for the placement of suburethral mesh for<br>the treatment of female stress urinary incontinence<br>(SUI) resulting from urethral hypermobility and /or<br>intrinsic sphincter deficiency. | Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) (Divisibr Sign Off) Division of General, Restorative, and Neurological Devices **510(k) Number** `li 081 617 {5}------------------------------------------------ ي الا ## Indications for Use Statement | 510(k) Number: (if known) | | |---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | Sparc System | | Indications For Use: | Intended for the placement of pubourethral sling for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and /or intrinsic sphincter deficiency. | Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED ### Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Pecorative, and Neur 510(k) Number L0811613 Page 3 of 3
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...