← Product Code [KWQ](/productcode/KWQ) · K081568

# ANTEGRA-T (K081568)

_Synthes Spine Co.Lp · KWQ · Sep 2, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K081568

## Device Facts

- **Applicant:** Synthes Spine Co.Lp
- **Product Code:** [KWQ](/productcode/KWQ.md)
- **Decision Date:** Sep 2, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Synthes Antegra-T System is indicated for use via the lateral or anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels as an adjunct to fusion. This system is indicated in the treatment of lumbar and lumbosacral (LI - SI) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), pseudoarthrosis, spondylolysis, spondylolisthesis, scoliosis, lordotic deformities of the spine, spinal stenosis, or a failed previous spine surgery. Warning: This device is not intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.

## Device Story

Synthes Antegra-T System comprises one-level titanium lumbar/sacral plates and cancellous screws with locking heads; used for anterior/anterolateral spinal stabilization as adjunct to fusion. Plates range 37-53mm; sacral plates feature underside step for precise placement. System utilizes bone screws from predicate Antegra System. Intended for surgical implantation by physicians in clinical settings to treat spinal instability. Device provides mechanical fixation to stabilize vertebral bodies; benefits patient by facilitating fusion and correcting spinal alignment. Not for posterior pedicle fixation.

## Clinical Evidence

No clinical data was required or provided for this device; substantial equivalence was demonstrated through non-clinical bench testing.

## Technological Characteristics

Titanium alloy spinal intervertebral body fixation system. Includes one-level plates (37-53mm) and locking cancellous screws. Mechanical fixation device; no energy source or software components. Sterilization method not specified.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Synthes Anterior Tension Band (ATB) System ([K022791](/device/K022791.md))
- Synthes Antegra System ([K063158](/device/K063158.md))
- Synthes VentroFix MIS System ([K031100](/device/K031100.md))
- Synthes Small Stature USS System ([K994121](/device/K994121.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Section 5 - 510(k) Summary

| 510(k) Summary       |                                                                                |
|----------------------|--------------------------------------------------------------------------------|
| Name of Firm:        | Synthes Spine                                                                  |
|                      | 1302 Wrights Lane East                                                         |
|                      | West Chester, PA 19380                                                         |
| 510(k) Contact:      | Susan Lewandowski                                                              |
|                      | Manager, Spine Regulatory Affairs                                              |
|                      | Telephone: 610-719-5712<br>Facsimile: 610-719-5102                             |
|                      | Email: lewandowski.susan@synthes.com                                           |
| Date Prepared:       | June 3, 2008                                                                   |
| Trade Name:          | Synthes Antegra-T System                                                       |
| Classification:      | 21 CFR 888.3060 - Spinal Intervertebral Body Fixation Orthosis                 |
|                      | Class II                                                                       |
|                      |                                                                                |
|                      | Orthopaedic and Rehabilitation Devices Panel<br>Product Code KWQ               |
| Predicates:          | Synthes Anterior Tension Band (ATB) System (K022791)                           |
|                      | Synthes Antegra System (K063158)                                               |
|                      | Synthes VentroFix MIS System (K031100)                                         |
|                      | Synthes Small Stature USS System (K994121)                                     |
| Device               |                                                                                |
| Description:         | The Synthes Antegra-T System consists of one-level titanium lumbar and         |
|                      | sacral plates and cancellous screws with locking heads. The plates attach to   |
|                      | the anterior portion of the lumbar and lumbosacral spine (L1-S1). These        |
|                      | one-level plates are available in lengths ranging from 37 mm to 53 mm in       |
|                      | 2mm increments. The sacral plates have a step on the underside that allows     |
|                      | for precise sacral placement. The Antegra-T System uses the bone-screws        |
|                      | from the Synthes Antegra System (K063158).                                     |
| Intended Use/        | The Synthes Antegra-T System is indicated for use via the lateral or           |
| Indications for      | anterolateral surgical approach above the bifurcation of the great vessels or  |
| Use:                 | via the anterior surgical approach, below the bifurcation of the great vessels |
|                      | as an adjunct to fusion. This system is indicated in the treatment of lumbar   |
|                      | and lumbosacral (LI - SI) spine instability as a result of fracture (including |
|                      | dislocation and subluxation), tumor, degenerative disc disease (defined as     |
|                      | back pain of discogenic origin with degeneration of the disc confirmed by      |
|                      | patient history and radiographic studies), pseudoarthrosis, spondylolysis,     |
|                      | spondylolisthesis, scoliosis, lordotic deformities of the spine, spinal        |
|                      | stenosis, or a failed previous spine surgery.  Warning: This device is not     |
|                      | intended for screw attachment or fixation to the posterior elements            |
|                      | (pedicles) of the cervical, thoracic, or lumbar spine.                         |
| Comparison of the    | The Synthes Antegra-T System is a result of design modifications to the        |
| device to predicate  | predicate devices. It is substantially equivalent to the predicates in design, |
| device(s):           | function, material, and intended use.                                          |
| Performance Date     | Non-Clinical Performance and Conclusions:                                      |
| (Non-Clinical and/or | Bench testing results demonstrate that the Synthes Antegra-T System is         |
| Clinical):           | substantially equivalent to the predicate devices.                             |
|                      | Clinical Performance and Conclusions:                                          |
|                      | Clinical data and conclusions were not needed for this device.                 |

Traditional 510(k)
Synthes Antegra-T System

{1}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like bird with outstretched wings, positioned to the right of a circular arrangement of text. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular fashion around the bird. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Synthes Spine Company, LP % Ms. Susan Lewandowski Manager, Spine Regulatory Affairs 1302 Wrights Lane East West Chester, Pennsylvania 19380

SEP - 2 2008

Re: K081568

Trade/Device Name: Synthes Antegra-T System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWO Dated: June 3, 2008 Received: June 4, 2008

Dear Ms. Lewandowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Ms. Susan Lewandowski

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Presmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Mark N. Millerson

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Section 4 - Indications for Use Statement

510(k) Number: (if known)

KO81568

Device Name: Synthes Antegra-T System

Indications for Use:

The Synthes Antegra-T System is indicated for use via the lateral or anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels as an adjunct to fusion. This system is indicated in the treatment of lumbar and lumbosacral (LI - Sl) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), pseudoarthrosis, spondylolysis, spondylolisthesis, scoliosis, lordotic deformities of the spine, spinal stenosis, or a failed previous spine surgery.

Warning: This device is not intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.

X Prescription Use (21 CFR 801 Subpart D) AND / OR

Over-the-Counter Use (21 CFR 801 Subpart C)

f

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Ev

Vaubare Bneam
(Division Sign-Off)

Division of General, Restorative, and Neurological Devices

Traditional 510(k) Synthes Antegra-T System

510(k) Number K081568

---

**Source:** [https://fda.innolitics.com/device/K081568](https://fda.innolitics.com/device/K081568)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
