← Product Code [MBI](/productcode/MBI) · K081511

# 2.0 PK SUTURE ANCHOR T, 2.0 PK SUTURE ANCHOR S (K081511)

_Smith & Nephew, Inc. · MBI · Aug 26, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K081511

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Aug 26, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The 2.0 PK Suture Anchor T and 2.0 PK Suture Anchor S is intended for use for the reattachment of soft tissue to bone for the following indications: Shoulder: Capsular stabilization- Bankart repair, Anterior shoulder instability, Slap lesion repair, Capsular shift or capsulolabral reconstruction, Acromilavicular separation repairs, Deltoid repairs, Rotator cuff tear repairs and Biceps tenodesis. Foot/Ankle: Hallux valgus repairs, Medial or lateral instability repairs/reconstruction. Achilles tendon repairs/reconstruction, Mid-foot reconstruction, Metatarsal ligament/tendon repairs/recontructions and Bunionectomy. Elbow, Wrist, and Hand: Biceps tendon reattachment, Ulnar or radial collateral ligament reconstructions, Lateral epicondylitis repair, and Scapholunate ligament reconstruction, Knee: Extra-capular repairs -Medial collateral ligament, Lateral ligament, Posterior oblique ligament, Patellar realignment and tendon repairs - Vastus medialis obliquous advancement and Illoiotibial band tenodesis.

## Device Story

Non-absorbable suture anchor with attached non-absorbable sutures; preassembled to stainless steel insertion device. Used by surgeons in clinical/OR settings for soft tissue-to-bone reattachment. Ribbed/threaded design provides mechanical fixation. Single-use, sterile device. Benefits patient by securing soft tissue during orthopedic repairs, facilitating healing.

## Clinical Evidence

Bench testing only. Performance testing demonstrated that insertion and fixation properties are substantially equivalent to predicate devices.

## Technological Characteristics

Non-absorbable suture anchor; stainless steel insertion device; ribbed/threaded design for bone fixation; sterile; single-use.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Smith & Nephew BIORAPTOR 2.3 PK Suture Anchor ([K071586](/device/K071586.md))
- Arthrex Mini BioSuture Tak ([K050749](/device/K050749.md)/[K061863](/device/K061863.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K0815" pyf 'f2

## AUG 2 6 2008

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION

as required by the Safe Medical Devices Act of 1990 and codified in 21 CFR 807.92 upon which the substantial equivalence is based.

2.0 PK Suture Anchor T and 2.0 PK Suture Anchor S

Date Prepared: May 23, 2008

### A. Submitter's Name:

SECTION IV

Smith & Nephew, Inc., Endoscopy Division 150 Minuteman Road Andover, MA 01810

#### B. Company Contact

Janice Haselton Sr. Regulatory Affairs Specialist Phone: 978-749-1494 Fax: 978-749-1443

#### C. Device Name

| Trade Name:          | 2.0 PK Suture Anchor T<br>2.0 PK Suture Anchor S   |
|----------------------|----------------------------------------------------|
| Common Name:         | Fastener, fixation, non-degradable, soft tissue    |
| Classification Name: | Smooth or threaded metallic bone fixation fastener |
| Product Code:        | MBI                                                |
| Regulation Number:   | 21 CFR §888.3040                                   |

#### D. Predicate Devices

The 2.0 PK Suture Anchor T and the 2.0 PK Suture Anchor S are substantially equivalent to the following legally marketed device in commercial distribution: Smith & Nephew BIORAPTOR 2.3 PK Suture Anchor cleared in K071586 and Arthrex Mini BioSuture Tak™ cleared in K050749/K061863.

> Smith & Nephew, INC 2.0 PK Anchors Pg. 25 of 54

{1}------------------------------------------------

#### E. Description of Device

The 2.0 PK Suture Anchor T and the 2.0 PK Suture Anchor S consist of a non-absorbable suture anchor with attached non-absorbable suture(s) preassembled to a stainless steel insertion device, and is provided sterile, for single use only. Both anchor models achieve fixation via ribbed/threaded design characteristics that are consistent with the repair of soft tissue to bone.

#### F. Intended Use

The 2.0 PK Suture Anchor T and 2.0 PK Suture Anchor S is intended for use for the reattachment of soft tissue to bone for the following indications:

Shoulder: Capsular stabilization- Bankart repair, Anterior shoulder instability, Slap lesion repair, Capsular shift or capsulolabral reconstruction, Acromilavicular separation repairs, Deltoid repairs, Rotator cuff tear repairs and Biceps tenodesis.

Foot/Ankle: Hallux valgus repairs, Medial or lateral instability repairs/reconstruction. Achilles tendon repairs/reconstruction, Mid-foot reconstruction, Metatarsal ligament/tendon repairs/recontructions and Bunionectomy.

Elbow, Wrist, and Hand: Biceps tendon reattachment, Ulnar or radial collateral ligament reconstructions, Lateral epicondylitis repair, and Scapholunate ligament reconstruction,

Knee: Extra-capular repairs -Medial collateral ligament, Lateral ligament, Posterior oblique ligament, Patellar realignment and tendon repairs - Vastus medialis obliquous advancement and Illoiotibial band tenodesis.

#### G. Comparison of Technological Characteristics

The performance testing conducted demonstrates substantial equivalence to the Arthrex Mini BioSuture Tak™suture anchor, cleared in K050749/K061863. The testing also demonstrates that the differences in the new device and the predicate device do not raise any new issues of safety and efficacy

#### H. Summary Performance Data

The performance testing conducted demonstrates that the insertion and fixation properties of the 2,0 PK Suture Anchor T and the 2.0 PK Suture Anchor S are substantially equivalent to the Smith & Nephew 2.3 BIORAPTOR and the Arthrex Mini BioSuture TakTM anchors.

> Smith & Nephew INC 2.0 PK Anchors Pg. 26 of 54

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# AUG 2 6 2008

Smith & Nephew, Inc. Endoscopy Division % Ms. Janice Haselton Senior Regulatory Affairs Specialist 150 Minuteman Road Andover, Massachusetts 01810

Re: K081511

Trade/Device Name: 2.0 PK Suture Anchor "S" and 2.0 PK Suture Anchor "T" Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI, HWC, JDR Dated: May 28, 2008 Received: May 29, 2008

Dear Ms. Haselton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Janice Haselton.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark M. Millican

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Indications for Use

510(k) Number (if known):

K0815-11

Device Name: ____2.0 PK Suture Anchor S and 2.0 PK Suture Anchor T

Indications For Use:

The 2.0 PK Suture Anchor S and 2.0 PK Suture Anchor T is intended for use for the reattachment of soft tissue to bone for the following indications:

Shoulder: Capsular stabilization- Bankart repair, Anterior shoulder instability, Slap lesion repair, Capsular shift or capsulolabral reconstruction, Acromilavicular separation repairs, Deltoid repairs, Rotator cuff tear repairs and Biceps tenodesis.

Foot/Ankle: Hallux valgus repairs, Medial or lateral instability repairs/reconstruction. Achilles tendon repairs/reconstruction, Mid-foot reconstruction, Metatarsal ligament/tendon repairs/recontructions and Bunionectomy.

Elbow, Wrist, and Hand: Biceps tendon reattachment, Ulnar or radial collateral ligament reconstructions, Lateral epicondylitis repair, and Scapholunate ligament reconstruction,

Knee: Extra-capular repairs -Medial collateral ligament, Lateral ligament, Posterior oblique ligament, Patellar realignment and tendon repairs - Vastus medialis obliquous advancement and Illoiotibial band tenodesis.

Prescription Use

AND/OR

(Per 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE F NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Barbara Brown

Division of Ger and Neurological Devices

510 k) Number_

Smith & Nephew, INC 2.0 PK Anchors Pg. 16 of 54

---

**Source:** [https://fda.innolitics.com/device/K081511](https://fda.innolitics.com/device/K081511)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
