K081508 · Shenyang Canta Medical Tech. Co., Ltd. · CAW · Aug 27, 2008 · Anesthesiology
Device Facts
Record ID
K081508
Device Name
HG OXYGEN CONCENTRATOR, MODELS HG5-W & HG5-WN
Applicant
Shenyang Canta Medical Tech. Co., Ltd.
Product Code
CAW · Anesthesiology
Decision Date
Aug 27, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
HG Oxygen concentrator is intended to provide supplemental oxygen to patient continuously in home, hospital or health care facility environments. It is not for supporting or sustaining life.
Device Story
HG Oxygen Concentrator (models HG5-W, HG5-WN) provides supplemental oxygen; uses Pressure Swing Adsorption (PSA) technology. Device draws room air through two chambers containing molecular sieve material to separate nitrogen; cycles compressed air between chambers to produce concentrated oxygen (≥90%); stores output in cylinder for delivery. Used in homes, hospitals, or care facilities; operated by patients or healthcare staff. Oxygen delivered via nasal cannula or mask (user-supplied). HG5-WN includes additional interface for nebulizer air supply. Device provides supplemental oxygen to improve patient comfort/status; not intended for life support.
Indicated for patients requiring supplemental low flow oxygen (≥90%) in home, hospital, or healthcare facility settings. Not for life support.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
K0Y1508
Image /page/0/Picture/1 description: The image shows a dark, abstract shape that resembles a triangle or a mountain range. The shape is composed of dark, overlapping lines and areas, creating a sense of depth and texture. The background is lighter and more textured, providing contrast to the dark shape. The overall effect is mysterious and intriguing.
AUG 2 7 2008
### Shenyang Canta Medical TECH. Co., Ltd
# Chapter III 510(k) Summary
As Required by CFR 807.92
The assigned 510(k) Number is:
- 1. Date Prepared: May 27, 2008
- 2. Sponsor Information
Shenyang Canta Medical Tech. Co., Ltd No.127 Nujiang Street, Huanggu District Shenyang 110036, P.R.China
Contact Person: Mr. Qiu Xiao, Quality Manager Tel: +86-24-86728299 Fax: +86-24-86728298 E-Mail: qiuxiao1971@163.com
- 3. Submission Correspondent
Ms. Diana Hong Mr. Lee Fu Shanghai Mid-Link Business Consulting Co., Ltd Suite 8D, Zhongshan Zhongxin Mansion No.19, Lane 999, Zhongshan No.2 Road(S) Shanghai, 200030, China
- 4. Device Name and Classification:
Device Trade Name: HG Oxygen Concentrator Modes: HG5-W & HG5-WN Device Common Name: Oxygen Concentrator Device Classification Name: generator, oxygen, portable
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a dark, blurry figure against a lighter background. The figure appears to be somewhat triangular in shape, with a broader base and a narrower top. The background is speckled with light and dark spots, creating a textured effect. Due to the image quality, specific details of the figure are difficult to discern.
### Shenyang Canta Medical TECH. Co., Ltd
Review Panel: Anesthesiology Product Code: CAW Regulation Number:868.5440 Device Class: II
- 5. Predicate Device Identification: Delphi Portable Oxygen Concentrator K-number: K073242
- 6. Intended Use:
HG Oxygen concentrator is intended to provide supplemental oxygen to patient continuously in home, hospital or health care facility environments. It is not for supporting or sustaining life.
- 7. Device Description:
The subject device, HG Oxygen Concentrator, contains two models, HG5-W and HG5-WN. These two models follow the same design principle, same raw material, same main function and same specification. The only difference between the two models is that HG5-WN oxygen concentrator provide an addition interface on the front panel, which can be connected to the nebulizer and supply air to the nebulizer.
The subject device, HG Oxygen Concentrator, is intend to provide ≥90% supplemental low flow oxygen, which is separated from the room air, to the patient in the home, nursing homes, patient care facilities, etc. Oxygen is delivered to the patient via a nasal cannula or oxygen mask, but these accessories are not supplied with the concentrator, the user shall select appropriate and legally marketed accessories themselves. The subject device is not intended to support or sustain life.
The subject device, HG Oxygen Concentrator, uses Pressure Swing Adsorption technology to deliver concentrated oxygen. Two chambers and a valve allow compressed air to enter the embedded sieve, which will separate the nitrogen from the air. When one chamber is receiving compressed air, the other is expelling nitrogen back to the air. The cycle is repeated continuously. The concentrated oxygen created at each cycle is stored in an oxygen store cylinder to be delivered to a patient. This working principle is widely used in the oxygen concentrator, it will not arise new question of safety and effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a black and white symbol that appears to be a stylized design. The symbol consists of a triangle shape at the top, with a curved line or arc below it. The overall design has a somewhat tribal or mystical appearance, with the black ink contrasting against the white background. The image quality is grainy, which gives the symbol a textured look.
Shenyang Canta Medical TECH. Co., Ltd
Test Conclusion 8.
> Laboratory testing was conducted to validate and verify that HG Oxygen Concentrator met all design specifications and was substantially equivalent to the predicate device.
- 9. Substantially Equivalent Conclusion:
The subject device, HG Oxygen Concentrator, is substantially equivalent to the predicate device.
3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and three wavy lines extending from its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Shenyang Canta Medical TECH Company, Limited C/O Ms. Diana Hong General Manager Shanghai Mid-Link Business Consulting Company, Limited Suite 8D, Zhongshan Zhongxin Masion No.19, Lane 999, Zhongshan No.2 Road(S) Shanghai, 200030 CHINA
AUG 2 7 2008
Re: K081508
Trade/Device Name: HG Oxygen Concentrator Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: May 27, 2008 Received: May 29, 2008
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Ms. Hong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
JK. Ammels-Find my for //
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{5}------------------------------------------------
# Indication for Use
510(k) Number:
Device Name: HG Oxygen Concentrator
Indications for Use:
HG Oxygen concentrator is intended to provide supplemental oxygen to facility hospital or continuously in health care home, patient environments. It is not for supporting or sustaining life.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mohr
Page _1_ of __1
(Division Sign-Off) Division of Anesthesiology, General Hospital infection Control, Dental Devices
10(k) Number: K041508
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.