XTAG RESPIRATORY VIRAL PANEL

K081483 · Luminex Molecular Diagnostics, Inc. · OCC · Jun 25, 2008 · Microbiology

Device Facts

Record IDK081483
Device NameXTAG RESPIRATORY VIRAL PANEL
ApplicantLuminex Molecular Diagnostics, Inc.
Product CodeOCC · Microbiology
Decision DateJun 25, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 866.3980
Device ClassClass 2

Intended Use

The xTAG™ Respiratory Viral Panel (RVP) is a qualitative nucleic acid multiplex test intended for the simultaneous detection and identification of multiple respiratory virus nucleic acids in nasopharyngeal swabs from individuals suspected of respiratory tract infections. The following virus types and subtypes are identified using RVP: Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza B, Respiratory Syncytial Virus subtype A, Respiratory Syncytial Virus subtype B, Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus. Human Metapneumovirus, Rhinovirus, and Adenovirus. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection if used in conjunction with other clinical and laboratory findings. It is recommended that specimens found to be negative for Influenza B, Respiratory Syncytial Virus subtype A and B, Parainfluenza 2, Parainfluenza 3 and Adenovirus, after examination using RVP be confirmed by cell culture. Negative results do not preclude respiratory virus infection and should not be used as the sole basis for diagnosis, treatment or other management decisions. Positive results do not rule out bacterial infection, or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing (e.g. bacterial culture, immunofluorescence, radiography) and clinical presentation must be taken into consideration in order to obtain the final diagnosis of respiratory viral infection. Due to seasonal prevalence characteristics for Influenza A/H1 were established primarily with retrospective specimens. The RVP assay cannot adequately detect Adenovirus species C, or scrotypes 7a and 41. The RVP primers for detection of thinovirus cross-react with enterovirus. A rhinovirus reactive result should be confirmed by an alternate method (e.g. cell culture). Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary. If infections with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to a state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Device Story

xTAG Respiratory Viral Panel (RVP) is a multiplex nucleic acid assay; detects viral nucleic acids in nasopharyngeal swabs. Components include multiplex PCR primer mix, target-specific primer extension (TSPE) primers, coupled bead mix, buffer, and TDAS RVP-I data analysis software. Device identifies Influenza A (H1, H3), Influenza B, RSV (A, B), Parainfluenza (1, 2, 3), Human Metapneumovirus, Rhinovirus, and Adenovirus. Used in clinical laboratory settings to aid diagnosis; results must be interpreted alongside clinical presentation and other laboratory findings (e.g., bacterial culture, radiography). Negative results for specific viruses require confirmation by cell culture. Rhinovirus results require confirmation due to enterovirus cross-reactivity. Device performance is dependent on circulating viral strains; novel Influenza A strains require public health authority screening. Benefits include rapid identification of multiple respiratory pathogens to inform clinical management.

Clinical Evidence

No clinical data provided in the submission; performance characteristics were established in the predicate device (K063765).

Technological Characteristics

Multiplex nucleic acid assay using PCR and target-specific primer extension (TSPE). Includes coupled bead mix for detection. Software-based data analysis (TDAS RVP-I).

Indications for Use

Indicated for individuals suspected of respiratory tract infections exhibiting signs and symptoms of infection. Used for qualitative detection and identification of respiratory viral nucleic acids in nasopharyngeal swabs.

Regulatory Classification

Identification

A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.

Special Controls

*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;” (2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and (3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K081483 ### 510(k) Summary ## JUN 2 5 2008 #### 1.0 Submitted By: Gloria Lee, Ph.D. Manager, Regulatory Affairs (Global Submissions) Luminex Molecular Diagnostics Inc. 439 University Ave. Toronto, Ontario M5G 1Y8 Canada Tel: 416.593.4323 x374 Fax: 416.593.1001 Email: glee@luminexcorp.com #### 2.0 Date Submitted June 24, 2008-06-23 #### 3.0 Device Name(s): Proprietary Name: xTAG™ Respiratory Viral Panel Classification Name: Respiratory Viral Panel Multiplex Nucleic Acid Assay [866.3980] #### Legally Marketed Device 4.0 xTAG™ Respiratory Viral Panel claims substantial equivalence to the xTAGTM Respiratory Viral Panel originally cleared under FDA 510(k) k063765. #### 5.0 Device Description The xTAGTM Respiratory Viral Panel includes the following components: - Multiplex PCR primer mix (without dNTPs) . - Multiplex target specific primer extension (TSPE) primers (includes dNTPs) . - Coupled bead mix ● - 10x buffer . - xTAG™ Data Analysis Software (TDAS RVP-I) . #### Intended Use 6.0 The xTAG™ Respiratory Viral Panel (RVP) is a qualitative nucleic acid multiplex test intended for the simultaneous detection and identification of multiple respiratory virus nucleic acids in nasopharyngeal swabs from individuals suspected of respiratory tract infections. The following virus types and subtypes are identified using RVP: Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza B, Respiratory Syncytial Virus subtype A, Respiratory Syncytial Virus subtype B, Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus. Human Metapneumovirus, Rhinovirus, and Adenovirus. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection if used in conjunction with other clinical and laboratory findings. It is recommended that specimens found to be negative for Influenza B, Respiratory Syncytial Virus subtype A and B, Parainfluenza 2, Parainfluenza 3 and {1}------------------------------------------------ Adenovirus, after examination using RVP be confirmed by cell culture. Negative results do not preclude respiratory virus infection and should not be used as the sole basis for diagnosis, treatment or other management decisions. Positive results do not rule out bacterial infection, or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing (e.g. bacterial culture, immunofluorescence, radiography) and clinical presentation must be taken into consideration in order to obtain the final diagnosis of respiratory viral infection. ## Due to seasonal prevalence characteristics for Influenza A/H1 were established primarily with retrospective specimens. The RVP assay cannot adequately detect Adenovirus species C, or scrotypes 7a and 41. The RVP primers for detection of thinovirus cross-react with enterovirus. A rhinovirus reactive result should be confirmed by an alternate method (e.g. cell culture). Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary. If infections with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to a state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. ### 7.0 Comparison to the Predicate (Description of the Modification to the Legally Marketed Device) The xTAG™ Respiratory Viral Panel performance parameters remain unchanged. #### 8.0 Summary of Performance Data Not applicable. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its body and wings. The eagle is positioned to the right of the department's name, which is arranged in a circular pattern around the logo. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUN 2 5 2008 Gloria Lee, Ph.D. Senior Regulatory Associate Luminex Molecular Diagnostics 439 University Ave. Toronto, Ontario M5G 1Y8 Re: k081483 Trade/Device Name: xTAGTM RVP (Respiratory Viral Panel) Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory viral panel multiplex nucleic acid assay Regulatory Class: Class II Product Code: OCC, OEM, OEP Dated: May 27, 2008 Received: May 28, 2008 Dear Dr. Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Sally attym Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indication for Use Statement 510(k) Number (if known): k081483 Device Name: xTAGTM Respiratory Viral Panel Indication For Use: The xTAG™ Respiratory Viral Panel (RVP) is a qualitative nucleic acid multiplex test intended for the simultaneous detection and identification of multiple respiratory virus nucleic acids in nasopharyngeal swabs from individuals suspected of respiratory tract infections. The following virus types and subtypes are identified using RVP: Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza B, Respiratory Syncytial Virus subtype A, Respiratory Syncytial Virus subtype B, Parainfluenza I, Parainfluenza 2, and Parainfluenza 3 virus, Human Metapneumovirus, Rhinovirus, and Adenovirus. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection if used in conjunction with other clinical and laboratory findings. It is recommended that specimens found to be negative for Influenza B, Respiratory Syncytial Virus subtype A and B, Parainfluenza 2, Parainfluenza 3, Parainfluenza 3 and Adenovirus, after examination using RVP be confirmed by cell culture. Negative results do not preclude respiratory virus infection and should not be used as the sole basis for diagnosis, treatment or other management decisions. Positive results do not rule out bacterial infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing (e.g. bacterial culture, immunofluorescence, radiography) and clinical presentation must be taken into consideration in order to obtain the final diagnosis of respiratory viral infection. Due to seasonal prevalence characteristics for Influenza A/H1 were established primarily with retrospective specimens. The RVP assay cannot adequately detect Adenovirus species C, or serotypes 7a and 41. The RVP primers for detection of rhinovirus cross-react with enterovirus. A rhinovirus reactive result should be confirmed by an alternate method (e.g. cell culture). Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary. If infections with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulenza viruses and sent to a state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. Prescription Use X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Sally atty Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) k081483
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.