← Product Code [JJX](/productcode/JJX) · K081478

# RNA MEDICAL GLUCOSE AND B-KETONE CALIBRATION VERIFICATION CONTROLS (K081478)

_Bionostics, Inc. · JJX · Oct 6, 2008 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K081478

## Device Facts

- **Applicant:** Bionostics, Inc.
- **Product Code:** [JJX](/productcode/JJX.md)
- **Decision Date:** Oct 6, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

RNA Medical® Brand Glucose and ß-Ketone Calibration Verification Controls are assayed materials for confirming the calibration and linearity of glucose and ß-ketone at the upper and lower limits of the reportable range and at three (3) points within the range. This product is for use with the Precision Xceed Pro Blood Glucose and ß-ketone Monitoring System, which uses Precision Xceed Pro Blood Glucose Test Strips and Precision Xceed Pro Blood B-Ketone Test Strips. It is not for use with the Precision PCx System or the i-STAT® 1 Analyzer, which use Precision PCx Test Strips, Precision PCx Plus Test Strips, or i-STAT 1 glucose cartridges. RNA Medical Glucose and ß-Ketone Calibration Verification Controls is intended for use by healthcare professionals. For In Vitro Diagnostic Use

## Device Story

RNA Medical Glucose and β-Ketone Calibration Verification Controls are 5-level, viscosity-adjusted, aqueous liquid controls; used by healthcare professionals to verify calibration and linearity of Precision Xceed Pro Blood Glucose and β-Ketone Monitoring Systems. Product packaged in plastic bottles with dropper tips for manual application to test strips. Green dye added for visual confirmation during dispensing. Controls contain D-Glucose and β-hydroxybutyrate in a buffered aqueous matrix. Healthcare professionals apply control solution to test strips; system measures analyte levels; results compared against expected values provided in package insert to ensure system accuracy across reportable range.

## Clinical Evidence

No clinical data. Bench testing only. Stability established via real-time studies; shelf life 24 months at 2-30°C. Open-vial stability demonstrated <10% change in glucose and acceptable ketone value variance over 90 days.

## Technological Characteristics

Aqueous solution containing glucose (0.03%-0.40%), β-hydroxybutyrate (0.01%-0.12%), viscosity enhancer, dye, buffer, and preservative. Five-level concentration set. Non-hazardous, non-biological. Standalone liquid control material.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Predicate Devices

- RNA GL4 Glucose Calibration Verification Control (k021624)
- Precision Control Solutions (k983504)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE

A. 510(k) Number:
k081478

B. Purpose for Submission:
New Device

C. Measurand:
Quality control materials for blood glucose monitoring systems - glucose and β-ketone

D. Type of Test:
Not applicable

E. Applicant:
Bionostics Quality Solutions

F. Proprietary and Established Names:
RNA Medical Glucose and β-Ketone Calibration Verification Controls

G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1660 Quality control material (assayed and unassayed)
2. Classification:
Class I (Reserved)
3. Product code:
JJX, single (specified) analyte controls (assayed and unassayed)
4. Panel:
Clinical Chemistry (75)

H. Intended Use:
1. Intended use(s):
RNA Medical Glucose and β-Ketone Calibration Verification Controls are assayed controls intended for use to confirm the calibration and linearity of glucose and β-ketone at the upper and lower limits of the reportable range and at three (3) points within the range. This product is for use with the Precision Xceed Pro™ Blood Glucose and β-Ketone Monitoring System, which uses Precision Xceed Pro™ Blood Glucose Test Strips and Precision Xceed Pro™ β-Ketone Test Strips. It is not for use with the Precision PCx System or the i-STAT 1 Analyzer, which use Precision PCx Test Strips, Precision PCx Plus Test Strips, or i-STAT 1 glucose cartridges.
The RNA Medical Glucose and β-Ketone Calibration Verification Controls is intended for use by healthcare professionals.
2. Indication(s) for use:
See intended use section above
3. Special conditions for use statement(s):
For In Vitro Diagnostic Use
For prescription use only

{1}

4. Special instrument requirements:
Abbott Diabetes Care Precision Xceed Pro™ Blood Glucose and β-Ketone Monitoring System

I. Device Description:
The RNA Medical Glucose and β-Ketone Calibration Verification Control is a five-level, viscosity-adjusted, aqueous liquid glucose and β-ketone control linearity set optimized for use with the Abbott Diabetes Care Precision Xceed Pro™ Blood Glucose and β-Ketone Monitoring System using Precision Xceed Pro Blood Glucose Test Strips and Precision Xceed Pro Blood β-Ketone Test Strips. The product is packaged in plastic bottles with dropper tips for application of the solution to test strips. The control has a green color to help users see the solution while dispensing onto a test strip.

J. Substantial Equivalence Information:

1. Predicate device name(s):
RNA GL4 Glucose Calibration Verification Control
Precision Control Solutions

2. Predicate 510(k) number(s):
k021624; k983504

3. Comparison with predicate:

|  Characteristic | Predicate Devices |   | Modified Device  |
| --- | --- | --- | --- |
|  Name: | RNA GL4 Glucose Calibration Verification Control | Precision Control Solutions | RNA Medical Glucose and β-Ketone Calibration Verification Controls  |
|  Number of levels: | 5
7 vials per kit, only 5 levels for use on specific devices | 3 | 5  |
|  Analytes: | Glucose | Glucose, Ketone | Glucose, Ketone  |
|  Container: | plastic bottle | plastic bottle | plastic bottle  |
|  Fill volume: | 4 mL | 3 mL | 4 mL  |
|  Color: | Red | Clear | Green  |
|  Matrix: | Buffered aqueous solution of D-Glucose, viscosity modifier, dye, preservative and other non-reactive ingredients. | Buffered aqueous solution of D-Glucose, Beta-hydroxybutyrate and other non-reactive ingredients. | Buffered aqueous solution of D-Glucose, β-hydroxybutyrate, viscosity modifier, dye, preservative and other non-reactive ingredients.  |

K. Standard/Guidance Document Referenced (if applicable):
ISO 15197 In-vitro diagnostic test systems – requirements for self-testing in managing diabetes
ISO 14971 Medical Devices, Application of risk management to medical devices
ISO 13485 Medical Devices-Quality Management Systems – Requirements for regulatory purposes

L. Test Principle:
Not Applicable

{2}

M. Performance Characteristics (if/when applicable):

1. Analytical performance:
a. Precision/Reproducibility:
Not applicable
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
No traceability was provided.

Product stability has been established based on real time studies on RNA Medical Glucose and β-Ketone Calibration Verification Controls. Samples are analyzed at predetermined intervals. The protocols for establishing shelf-life and open-vial stability were reviewed and are adequate. The real-time studies support the claimed shelf life, 24 months at room temperature (2° - 30°C) for unopened bottles. The open bottle testing demonstrated a less than 10% change in glucose (all levels) over the 90 day evaluation period. The open bottle testing demonstrated the change in ketone values over the 90 day evaluation met the stability specifications.

The expected values in the labeling are based on multiple determinations performed on randomly selected samples from each lot using several test strip lot numbers and monitors. The assigned values for the reference lots are determined by measurement of all levels of control on each blood glucose meter/strip configuration by Abbott Diabetes Care. Value assignment for each lot of control solution will be provided in the package insert delivered with the control.

d. Detection limit:
Not applicable

e. Analytical specificity:
Not applicable

f. Assay cut-off:
Not applicable

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):

{3}

Not applicable

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
Not applicable

**N. Proposed Labeling:**
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

**O. Conclusion:**
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K081478](https://fda.innolitics.com/device/K081478)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
