ROCHE ELECSYS ANTI-CCP IMMUNOASSAY, ROCHE ELECSYS PRECICONTROL ANTI-CCP

K081338 · Roche Diagnostics · NHX · Sep 11, 2008 · Immunology

Device Facts

Record IDK081338
Device NameROCHE ELECSYS ANTI-CCP IMMUNOASSAY, ROCHE ELECSYS PRECICONTROL ANTI-CCP
ApplicantRoche Diagnostics
Product CodeNHX · Immunology
Decision DateSep 11, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5775
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K081338 · Sep 11, 2008ROCHE ELECSYS ANTI-CCP IMMUNOASSAY, ROCHE ELECSYS PRECICONTROL ANTI-CCPRoche DiagnosticsRetrospective clinical serum samplesRetrospective clinical samples were used to perform a method comparison between the Elecsys Anti-CCP assay and a predicate ELISA device to establish substantial equivalence.Method comparison; Retrospective clinical samples; Substantial equivalence

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective method comparisonClinical serum samples; Sample Size: 1606Eurodiagnostica IMMUNOSCAN RA Anti-CCP Test Kit (K052133)Positive Percent Agreement (94.3%), Negative Percent Agreement (98.4%), Total Percent Agreement (97.3%)

Indications for Use

Elecsys Anti-CCP immunoassay: Immunoassay for the in vitro semi-quantitative determination of human IgG autoantibodies to cyclic citrullinated peptides in human serum and plasma. The results of the assay are intended to be used as an aid in the diagnosis of rheumatoid arthritis in combination with other clinical and laboratory findings. The electrochemiluminescence immunoassay "ECLIA" is intended for use on Elecsys and cobas e immunoassay analyzers. Elecsys PreciControl Anti-CCP is used for quality control of the Elecsys Anti-CCP immunoassay on the Elecsys and cobas e immunoassay analyzers.

Device Story

The Elecsys Anti-CCP Immunoassay is an in vitro diagnostic test for detecting IgG autoantibodies to cyclic citrullinated peptides (CCP) in human serum and plasma. It utilizes a two-step IgG-capture principle with streptavidin-coated microparticles and electrochemiluminescence (ECLIA) detection. The device is intended for use on Elecsys and cobas e immunoassay analyzers in clinical laboratory settings. The system generates results via a calibration curve derived from a master curve provided by the reagent barcode and a 2-point calibration. Healthcare providers use the semi-quantitative results as an aid in diagnosing rheumatoid arthritis. The Elecsys PreciControl Anti-CCP is a lyophilized human serum-based control used to monitor assay performance. The device benefits patients by providing a standardized, automated method for identifying anti-CCP antibodies, which are clinical markers for rheumatoid arthritis.

Clinical Evidence

Clinical performance evaluated in a cohort of 792 confirmed RA patients, 420 asymptomatic healthy individuals, and 907 patients with other rheumatic/non-rheumatic disorders. Sensitivity was 67.7% and specificity was 97.0% at a cut-off of 17 U/mL. ROC curve AUC was 0.85. Method comparison against the predicate device (N=1606) showed 94.3% positive percent agreement and 98.4% negative percent agreement.

Technological Characteristics

Two-step IgG-capture immunoassay; electrochemiluminescence detection; streptavidin-coated microparticles. Measuring range: 8-1000 U/mL. Analyzers: Elecsys 2010, MODULAR ANALYTICS E170, cobas e 411, cobas e 601. Standardized against a commercially available anti-CCP assay. Complies with CLSI EP17-A (LoD/LoQ) and CLSI EP5-A2 (precision) guidelines.

Indications for Use

Indicated for the semi-quantitative determination of human IgG autoantibodies to cyclic citrullinated peptides in human serum and plasma to aid in the diagnosis of rheumatoid arthritis in conjunction with other clinical and laboratory findings. For use by trained laboratory professionals on Elecsys and cobas e immunoassay analyzers.

Regulatory Classification

Identification

A rheumatoid factor immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the rheumatoid factor (antibodies to immunoglobulins) in serum, other body fluids, and tissues. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k081338 B. Purpose for Submission: New Device C. Measurand: Human IgG autoantibodies to cyclic citrullinated peptides D. Type of Test: Semi-quantitative immunoassay E. Applicant: Roche Diagnostics F. Proprietary and Established Names: Elecsys Anti-CCP Immunoassay Elecsys PreciControl Anti-CCP G. Regulatory Information: 1. Regulation section: 21 CFR 866.5775 Rheumatoid factor immunological test system 21 CFR 862.1660 Quality control material (assayed and unassayed) 2. Classification: Class II (Device) Class I (Quality control) 3. Product codes: NHX, Antibodies, anti-cyclic citrullinated peptide JJX, Single (specified) analyte controls (assayed and unassayed) 4. Panel: Immunology H. Intended Use: 1. Intended use(s): See Indication for use 2. Indication(s) for use: Elecsys Anti-CCP Immunoassay Immunoassay for the in vitro semi-quantitative determination of human IgG autoantibodies to cyclic citrullinated peptides in human serum and plasma. The results of the assay are intended to be used as an aid in the diagnosis of rheumatoid arthritis in combination with other clinical and laboratory findings. The electrochemiluminescence immunoassay “ECLIA” is intended for use on Elecsys and cobas e immunoassay analyzers. Elecsys PreciControl Anti-CCP Elecsys PreciControl Anti-CCP is used for the quality control of the Elecsys Anti-CCP immunoassay on the Elecsys and cobas e immunoassay analyzers. 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: Elecsys 2010, MODULAR ANALYTICS E170, cobas e 411 and cobas e 601. {1} 2 I. Device Description: The Elecsys Anti-CCP reagent kit consists of a Reagent Pack (R1 and R2) and lyophilized calibrators 1 and 2. The Elecsys PreciControl Anti-CCP is a lyophilized product consisting of human serum with added Anti-CCP antibody (human) in two concentration ranges. During manufacture, the antibody is spiked into the matrix at the desired concentration levels. J. Substantial Equivalence Information: 1. Predicate device name(s): Eurodiagnostica IMMUNOSCAN RA Anti-CCP Test Kit 2. Predicate K number(s): k052133 3. Comparison with predicate: | Similiarities | | | | --- | --- | --- | | Item | Device | Predicate | | | Elecsys Anti-CCP Immunoassay | Eurodiagnostica IMMUNOSCAN RA Anti-CCP | | Intended Use / Indication for Use | Immunoassay for the in vitro semi-quantitative determination of human IgG autoantibodies to cyclic citrullinated peptides in human serum and plasma. The results of the assay are intended to be used as an aid in the diagnosis of rheumatoid arthritis in combination with other clinical and laboratory findings | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | | Elecsys Anti-CCP Immunoassay | Eurodiagnostica IMMUNOSCAN RA Anti-CCP | | Detection Protocol | Electrochemiluminescence | ELISA | | Sample Type | Human serum and plasma | Human serum | | Platform | Elecsys 2010, MODULAR ANALYTICS E170, cobas e 411, cobas e 601 | Spectrophotometer | | Measuring Range | 8-1000 U/mL | 1.6-1600 U/mL | | Expected Values | Positive: ≥ 17 U/mL | Negative: < 25 U/mL Positive: ≥ 25 U/mL | {2} 3 | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Precision | Elecsys 2010 and cobas e 411: Anti-CCP: 16.9 - 356 U/mL Total %CV: 2.5% - 4.5% Within-run %CV: 0.6% - 2.3% | Intra-assay: Anti-CCP: 33.6 – 1007.4 U/mL %CV: 4.3% - 12.8% Inter-assay: Anti-CCP: 33.3 – 1105.9 U/mL %CV: 6.0% - 17.7% | | Calibrator | Two levels | Five levels | | Control | Two levels | Reference, positive, and negative controls supplied with kit | K. Standard/Guidance Document Referenced (if applicable): CLSI EP17-A: “Protocol for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline” CLSI EP5-A2: “Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition” L. Test Principle: The Elecsys Anti-CCP immunoassay is a two step IgG-capture test principle immunoassay with streptavidin-coated microparticles and electrochemiluminescence detection. Results are determined using a calibration curve that is generated specifically on each instrument by a 2 point calibration and a master curve provided with the reagent bar code. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Precision of the Elecsys® Anti-CCP Test System was evaluated on Elecsys® 2010/cobas e® 411 Immunoassay Analyzer according to CLSI EP5-A2 guideline. The protocol consisted of testing 2 replicates of each control (PC = PreciControl) or sample (HS = human serum) per run, 2 runs per day for 21 days. Within run precision and total precision was calculated according to EP5-A2. | Sample | Anti-CCP (U/mL) | Within-run%CV | Total%CV | | --- | --- | --- | --- | | 1 | 16.9 | 0.6 | 3.1 | | 2 | 24.6 | 1.0 | 3.0 | | 3 | 137 | 1.4 | 2.5 | | 4 | 356 | 2.3 | 4.5 | A 10 day inter-assay precision study with a sample at 664 U/mL was also performed to support high end precision. Results are acceptable with %CV of 3.4%. {3} b. Linearity/assay reportable range: The linearity of the Elecsys® Anti-CCP was evaluated on the Elecsys® 2010/cobas e® 411 Immunoassay Analyzer by diluting three high through medium analyte level serum sample pools with varying amounts of low-level serum. Pure samples (0% and 100%) were run n=6 with mean measured value reported. The regression equations of the three pools are as follows: Sample 1: y=1.0029x-4.1733; r²=0.9992 Sample 2: y=0.9913x+0.5146; r²=0.9992 Sample 3: y=1.0261x+5.4971; r²=0.9944 The measuring range is 8-1000 U/mL (defined by the limit of detection and the maximum of the master curve). Performance information at the low end of the assay range is summarized in the Detection Limit section below. Roche has performed linearity testing to cover the low end of the measuring range. The linear regression is y=0.9338x+1.2276; r²=0.9904. c. Traceability, Stability, Expected values (controls, calibrators, or methods): There is no recognized material for anti-CCP. The Elecsys® Anti-CCP test is standardized against a commercially available anti-CCP assay. d. Detection limit: The limit of blank and limit of detection were determined in accordance with the CLSI EP17-A requirements. The limit of blank (≤ 7 U/mL) is the 95th percentile value from n ≥ 60 measurements of analyte free samples over several independent series. The limit of blank corresponds to the concentration below which analyte-free samples are found with a probability of 95%. The distribution of values for five zero-level human serum samples have been determined on two Elecsys® 2010 / cobas e® 411 Immunoassay Analyzer over 3 days, 2 runs per day. The limit of detection (≤8 U/mL) is determined based on the limit of blank and the standard deviation of low concentration samples. The limit of detection corresponds to the lowest analyte concentration which can be detected (value above the limit of blank with a probability of 95%). The distribution of values for five low-level human serum samples have been determined on two Elecsys® 2010 / cobas e® 411 Immunoassay Analyzer over 3 days, 2 runs per day. The limit of quantitation (8 U/mL) is the lowest analyte concentration that can be reproducibly measured with a between-run coefficient of variation of ≤ 20%. It has been determined using low concentration anti-CCP samples. Nine human serum sample pools with concentrations ranging from 6.57 to 27.2 U/mL were tested once per day, for 10 days. e. Analytical specificity: Effect on quantitation of analyte in the presence of endogenous interfering substances using the Elecsys® Anti-CCP Immunoassay was determined on 4 {4} Elecsys® 2010/cobas e® 411 Immunoassay Analyzer using natural (low analyte concentration) and spiked serum sample pools. The assay is unaffected by: - Bilirubin: &lt; 25 mg/dL - Hemoglobin: &lt; 0.5 g/dL - Intralipid: &lt; 1500 mg/dL - Biotin: &lt; 30 ng/mL - Rheumatoid Factors: &lt; 150 IU/mL Hook effect: No hook effect was detected up to concentration of 7000 U/mL. f. Assay cut-off: In cohorts of 792 confirmed rheumatoid arthritis patients, 420 asymptomatic healthy individuals and 907 patients with other rheumatic and non-rheumatic disorders, an optimal cut-off of 17 U/mL was determined. At this cut-off the sensitivity was calculated at 67.7% and the specificity at 97.0%. The calculated receiver operating characteristic (ROC) curve had an area under the curve (AUC) of 0.85. 2. Comparison studies: a. Method comparison with predicate device: A method comparison study was performed to compare the Elecsys Anti-CCP values obtained with both the Elecsys® 2010/cobas e® 411 and MODULAR ANALYTICS E170/cobas e® 601 Immunoassay Analyzer to the values obtained with the predicate device Eurodiagnostica IMMUNOSCAN RA Anti-CCP. Cohort: Non-RA cohort: Negative cohort consisting of 319 apparently healthy individuals and 673 patients with non-rheumatoid arthritis diseases; in total 992 patients from two sites. RA Cohort: Positive cohort consisting of 614 patients with clinical diagnosis of rheumatoid arthritis based on the ARA (American Rheumatology Association) 1987 Revised Criteria for the Classification of Rheumatoid Arthritis. 1606 samples tested using a cut-off of ≥ 17 U/mL for the Elecsys Anti-CCP assay. | N=1606 | Immunoscan Anti-CCP | | | | | --- | --- | --- | --- | --- | | | | + | - | Total | | Elecsys Anti-CCP | + | 428 | 18 | 446 | | | - | 26 | 1134 | 1160 | | | Total | 454 | 1152 | 1606 | {5} Positive Percent Agreement = 94.3% (95% CI = 91.7-96.2) Negative Percent Agreement = 98.4% (95% CI = 97.5-99.1) Total Percent Agreement = 97.3% (95% CI = 96.3-98.0) b. Matrix comparison: The effect on quantitation of analyte in the presence of anticoagulants with Elecsys Anti-CCP Immunoassay was determined on Elecsys 2010/cobas e 411 Immunoassay Analyzer by comparing values obtained from samples drawn into Serum, Li- and K3-EDTA-Heparin plasma primary tubes. Potential effects are assessed by Passing/Bablok regression analysis. The following was found: Serum/K3-EDTA: y=1.10x-0.977 (anti-CCP = 8.37-473 U/mL, n=31) Serum/Li-Heparin: y=1.03x-0.143 (anti-CCP = 8.37-769 U/mL, n=33) 3. Clinical studies: a. Clinical sensitivity and specificity: The clinical sensitivity and specificity were determined by testing 792 confirmed rheumatoid arthritis patients, 420 asymptomatic healthy individuals and 907 patients with other rheumatic and non-rheumatic disorders. Results showed sensitivity and specificity of 67.7% and 97.0% respectively. b. Other clinical supportive data (when a. is not applicable): Not applicable 4. Clinical cut-off: Results ≥ 17 U/mL are considered positive; see assay cut off 5. Expected values/Reference range: The expected value in the normal population is negative. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...