K081328 · Pathway Medical Technologies, Inc. · MCW · Jul 17, 2008 · Cardiovascular
Device Facts
Record ID
K081328
Device Name
PATHWAY PV ATHERECTOMY SYSTEM
Applicant
Pathway Medical Technologies, Inc.
Product Code
MCW · Cardiovascular
Decision Date
Jul 17, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4875
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pathway PVTM Atherectomy System is intended for use in atherectomy of the peripheral vasculature. It is not intended for use in coronary, carotid, iliac or renal vasculature.
Device Story
The Pathway PV Atherectomy System is a peripheral atherectomy device used by clinicians to remove plaque from peripheral vasculature. The system comprises a sterile, single-use PV Catheter with an electrically driven, differentially cutting, expandable-blade tip and a reusable PV Console. The console, mounted on an IV stand outside the sterile field, contains peristaltic pumps for aspiration and infusion, a power supply, and a system controller. During the procedure, the catheter is tracked over a 0.014" guidewire to the treatment site. The user controls blade deployment and device operation via a handheld Control Pod. The system provides continuous saline infusion to the site and simultaneous vacuum aspiration of excised tissue, thrombus, and fluid into a collection bag. This dual-action approach facilitates lesion treatment and debris removal, potentially improving blood flow in peripheral vessels.
Clinical Evidence
Performance testing included in vitro performance, biocompatibility, packaging, sterilization, electrical/EMC, animal, and clinical studies. All testing was conducted in accordance with in-house protocols and applicable recognized standards.
Technological Characteristics
System includes a sterile, single-use PV Catheter with a fluted, expandable-blade cutting tip and a reusable PV Console. Features include dual-lumen design for continuous saline infusion and vacuum aspiration. Compatible with 8F sheaths and 0.014" guidewires. Electrically driven via a handheld Control Pod with keypad and LED status indicators. Console utilizes peristaltic pumps. Sterilization method for the catheter is not specified.
Indications for Use
Indicated for atherectomy of the peripheral vasculature. Contraindicated for use in coronary, carotid, iliac, or renal vasculature.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
Predicate Devices
FoxHollow Technologies Inc. ReFORM Peripheral Catheter System (SilverHawk® Plaque Excision System) (K024243, K043553, K053460, K061063, K061188)
InterVentional Technologies, Inc. Transluminal Endarterectomy Catheter (TEC) (K890515, K922970)
Submission Summary (Full Text)
{0}------------------------------------------------
K081328 page 1 of 3
Image /page/0/Picture/1 description: The image contains the logo for Pathway Medical Technologies. The logo consists of a circular graphic on the left and the company name on the right. The circular graphic is divided into a black upper portion and a white lower portion, with a curved line separating the two. The company name is written in a simple, sans-serif font, with "PATHWAY" on the top line and "MEDICAL TECHNOLOGIES" on the bottom line.
JUL 17 2008
# SECTION 5: 510(k) SUMMARY
1
In accordance with the requirements of 21 CFR 807.92(c) Pathway Medical Technologies, Inc. (hereafter "Pathway Medical") has prepared this 510(k) Summary to provide information supporting the substantial equivalence of the Pathway PVTM Atherectomy System.
:
.
## General Information:
| Date of Summary Preparation: | May 9, 2008 |
|-----------------------------------|--------------------------------------------------------------------------------------------|
| Name and Address of Manufacturer: | Pathway Medical Technologies, Inc.<br>10801 120th Ave NE<br>Kirkland, Washington 98033 |
| Contact Person: | Brian Cleary<br>Director of Regulatory Affairs<br>Phone: 425-636-4079<br>Fax: 425-636-4001 |
| Trade Name: | Pathway PVTM Atherectomy System |
| Common Name: | Peripheral Atherectomy Catheter |
| Device Classification: | Catheter, Peripheral, Atherectomy |
| Classification Panel: | Cardiovascular |
| CFR Reference: | 870.4875 |
| Product Code: | MCW |
| Device Class: | Class II |
# 00018
{1}------------------------------------------------
FOR 1328
Predicate Devices: The Pathway PV Atherectomy System is substantially equivalent to the following legally marketed predicate device systems indicated for similar or identical usc:
- . FoxHollow Technologies Inc. ReFORM Peripheral Catheter System (marketed as the SilverHawk® Plaque Excision System) - K024243, K043553, K053460, K061063, K061188
- Boston Scientific Corporation Rotablator® Rotational Angioplasty System -. K901206, K954604, K970296, K993648
- . InterVentional Technologies, Inc. Transluminal Endarterectomy Catheter (TEC) -K890515, K922970
Indications for Use: The Pathway PV Atherectomy System is intended for use in atherectomy of the peripheral vasculature. It is not intended for use in coronary, carotid, iliac or renal vasculature.
Device Description: The Pathway PV Atherectomy System consists of two primary components: {1} a PV Catheter and Control Pod and (2) a PV Console, which are packaged separately. Each of these system components is described generally as follows:
- PV Catheter and Control Pod: A sterile, single-use unit consisting of an . electrically driven PV Catheter and Control Pod. The PV Catheter utilizes a differentially cutting tip and includes both aspiration and infusion capabilities. The Control Pod provides a user interface with keypad controls and LED indicators for device operational status. The unit, its electrical connectors, tubing. and aspirant collection bag are packaged in a double pouched tray.
- Console: A reusable compact PV Console, with two (2) peristaltic pumps for . aspiration and infusion, power supply, system controller, keypad interface, and LED indicators for device operational status. The PV Console mounts on a standard I.V. stand and remains outside the sterile field during the procedure.
The Pathway PV Atherectomy System is compatible with 8F sheaths and exchange length (300cm or longer) 0.014" diameter guidewires.
Technological Characteristics: The Pathway PV Atherectomy System's PV Catheter is placed for treatment at the selected peripheral treatment site over a 0.014" exchange length guidewire. All device functions are then controlled by the user via the external handheld Control Pod. The Pathway PV Atherectomy System includes a fluted cutting tip designed with expandable blades. Expandable blade deployment and retraction is controlled by the user.
Separate lumens within the PV Catheter allow for continuous aspiration and infusion during device use. Excised tissue, thrombus, and fluid are aspirated from the treatment site through ports in the PV Catheter tip to a collection bag located on the PV Console.
{2}------------------------------------------------
The distal portion of the PV Catheter also possesses infusion ports that provide continuous infusion of sterile saline during the atherectomy procedure.
Substantial Equivalence: As compared to the predicate devices, the Pathway PV Atherectomy System has the identical indication for use as the FoxHollow SilverHawk Plaque Excision System and the identical device classification and device code as all of the selected predicate devices. The Pathway PV Athercctomy System is comprised of the same general patient contact materials and has undergone appropriate biocompatibility testing for its intended use. As with all of the listed predicate device systems, the PV Catheter component is provided sterile and is intended for single-use.
With respect to general device design and technological characteristics, the Pathway PV Atherectomy System is substantially equivalent to all of the selected predicate devices. since it similarly uses a cutting tip controlled by an external handheld control unit for the treatment of peripheral vascular lesions. The multilumen catheter design of the Pathway PV Atherectomy System allows for the infusion of sterile saline to the athercetomy treatment site (equivalent to the Boston Scientific Rotablator predicate device) and the removal of tissue and fluid from the patient via vacuum aspiration (equivalent to the IVT TEC predicate device).
There are no significant technological differences among the subject device and the predicate devices. Therefore, the Pathway PV Atherectomy System is substantially equivalent to the predicate devices.
Device Testing: Performance testing of the Pathway PV Atherectomy System included in vitro performance, biocompatibility, packaging, sterilization, electrical/EMC, animal, and clinical study. All testing was completed per Pathway Medical's in-house test methods, protocols, and requirements or in accordance with applicable recognized standards.
Conclusion: The information contained in this 510(k) summary establishes the substantial equivalence between the Pathway PV Atherectomy System and the selected predicate devices.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling an emblem. The emblem is a stylized representation of an eagle with three waves emanating from its body. The text reads, "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1 7 2008
Pathway Medical Technologies, Inc. C/O Brian Cleary Director of Regulatory Affairs 10801 120th Avenue NE Kirkland, WA 98033
Re: K081328
Trade/Device Name: Pathway PV Atherectomy System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal artery stripper Regulatory Class: Class II Product Code: MCW Dated: May 9, 2008 Received: May 12, 2008
Dear Mr. Cleary:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Foderal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
## Page 2 - Mr. Brian Cleary
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
R. Kirchner
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### SECTION 4: INDICATIONS FOR USE
510(k) Number (if known): K081328
Device Name: Pathway PVTM Atherectomy System
Indications for Use: The Pathway PVTM Atherectomy System is intended for use in atherectomy of the peripheral vasculature. It is not intended for use in coronary, carotid, iliac or renal vasculature.
Prescription Use ਮ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Evidence
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_ko81328
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.