← Product Code [BZD](/productcode/BZD) · K081321

# MODIFICATION TO MIRAGE MICRO (K081321)

_Resmed, Ltd. · BZD · Oct 9, 2008 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K081321

## Device Facts

- **Applicant:** Resmed, Ltd.
- **Product Code:** [BZD](/productcode/BZD.md)
- **Decision Date:** Oct 9, 2008
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 868.5905
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The Mirage Micro channels airflow non-invasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system. The Mirage Micro is: - to be used by adult patients (>66lb / >30kg) for whom positive airway pressure has been prescribed. - intended for single patient re-use in the home environment and multi-patient re-use in the hospital/institutional environment.

## Device Story

Mirage Micro is a vented nasal mask for non-invasive delivery of positive airway pressure (CPAP or bilevel) to patient's nose. Device consists of dual-wall silicone interface, molded plastic components, and adjustable fabric headgear. Mask connects to air delivery hose via standard conical connectors (ISO 5356-1:2004); includes vent holes to flush CO2 and luer ports for oxygen/pressure sensing. Used in home (single-patient) or hospital (multi-patient) settings; operated by patient or clinician. Provides seal to maintain prescribed pressure; helps treat respiratory conditions requiring positive airway pressure. Healthcare providers use mask to facilitate therapy delivery; patient benefits from non-invasive respiratory support.

## Clinical Evidence

Bench testing only. Vented nasal masks for CPAP/Bilevel therapy are established technology. Performance data demonstrates CO2 flushing, pressure-flow characteristics, and flow impedance are substantially equivalent to the predicate device.

## Technological Characteristics

Dual-wall silicone interface; molded plastic components; fabric headgear. Standard conical connectors (ISO 5356-1:2004). Integrated vent holes for CO2 flushing. Luer ports for oxygen/pressure sensing. Materials compliant with ISO 10993-1. Non-sterile, reusable.

## Regulatory Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

## Predicate Devices

- Mirage Micro Mask ([K072940](/device/K072940.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K081321

## OCT 0 9 2008

## RESMED

Mirage Micro
 Special 510k 

| <b>Special 510(k) SUMMARY</b><br>[As required by 21 CFR 807.92(c)] |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|--------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared                                                      | May 07, 2008                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Official Contact                                                   | David Thomson<br>Regulatory Affairs Director                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Device Trade Name                                                  | Mirage Micro™                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Device Common Name/<br>Classification Name                         | Vented Nasal Mask;<br>Accessory to Noncontinuous Ventilator (IPPB)                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Classification                                                     | 21 CFR 868.5905, 73 BZD (Class II)                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Predicate Device                                                   | Mirage Micro Mask (K072940)                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Description                                                        | The Mirage Micro provides seal such that airflow from a<br>positive pressure source is directed to the patient's nose.<br>The mask is held in place with adjustable headgear that<br>straps the mask to the face.<br><br>Mirage Micro is safe when used under the conditions and<br>purposes intended as indicated in the labeling provided<br>with the product.<br><br>Mirage Micro is a prescription device supplied nonsterile.                                                           |
| Intended Use                                                       | The Mirage Micro channels airflow non-invasively to a<br>patient from a positive airway pressure device such as a<br>continuous positive airway pressure (CPAP) or bilevel<br>system.<br><br>The Mirage Micro is:<br>- to be used by adult patients (>66lb / >30kg) for whom<br>positive airway pressure has been prescribed.<br>- intended for single patient re-use in the home<br>environment and multi-patient re-use in the<br>hospital/institutional environment.                      |
| Technological<br>Characteristics<br>comparison                     | Comparison with predicate Mirage Micro<br>The modified device and the predicate mask, provide seal<br>via dual wall silicone interface. Both masks are offered in<br>various sizes to ensure adequate fit over the extended<br>patient population.<br><br>Both the masks incorporate vent holes to provide<br>continuous air leak to flush out the dead space within the<br>mask and minimize the amount of CO2 rebreathed by the<br>patient. The design of the mask components is such that |
|                                                                    | the incorporation of these vent-holes does not interfere<br>with the intended performance of the masks.                                                                                                                                                                                                                                                                                                                                                                                      |
|                                                                    | Both the masks connect to conventional air delivery hose<br>between the mask and the positive airway-pressure<br>source via standard conical connectors (ref: ISO 5356-<br>1:2004)                                                                                                                                                                                                                                                                                                           |
|                                                                    | Both the masks have provisions for connecting oxygen<br>and pressure sensing tubing via luer ports.                                                                                                                                                                                                                                                                                                                                                                                          |
|                                                                    | Both the masks are constructed using molded plastic<br>components and fabric headgear. All the components of<br>both masks, including the modification to the swivel color,<br>are fabricated using materials deemed safe. (ref: ISO<br>10993-1).                                                                                                                                                                                                                                            |
|                                                                    | Both the modified device and the predicate device can be<br>reused in the hospital / institution environment.                                                                                                                                                                                                                                                                                                                                                                                |
| Clinical Data                                                      | Use of vented nasal masks with CPAP or Bilevel therapy<br>is proven technology and is well accepted by the medical<br>community. Bench testing is sufficient to demonstrate<br>safety and efficacy of the Mirage Micro, as was the case<br>with the predicate devices.                                                                                                                                                                                                                       |
| Performance Data                                                   | Comparison with predicate Mirage Micro<br>The CO2 performance of the modified device and the<br>predicate device are substantially equivalent.                                                                                                                                                                                                                                                                                                                                               |
|                                                                    | Both the modified device and the predicate are designed<br>to operate on the same standard flow generator setting.<br>The pressure-flow characteristics and flow impedance of<br>both the modified device and the predicate device are<br>substantially equivalent.                                                                                                                                                                                                                          |
| Substantial Equivalence<br>Conclusion                              | Modified Mirage Micro is substantially equivalent to the<br>predicate device:<br>- it has the same intended use;<br>- it has similar technological characteristics to both<br>predicates;<br>- it does not raise new questions of safety and<br>effectiveness;<br>- it is at least as safe and effective as the predicate                                                                                                                                                                    |

May 07, 2008

- .

{1}------------------------------------------------

.

May 07, 2008

・・

20

·

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 0 9 2008

ResMed Limited C/O Mr. David D'Cruz Vice President, Clinical & Regulatory Affairs ResMed Corporation 14040 Danielson Street Poway, California 92064-6857

Re: K081321

Trade/Device Name: Mirage MicroTM Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: September 19, 2008 Received: October 2, 2008

Dear Mr. D'Cruz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. D'Cruz

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Th. Samuels-Send me for//

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

**RESMed**

Mirage Micro Special 510k

## Indication for Use

| 510(k) Number (if known): | K08- |
|---------------------------|------|
| Device Name:              | MIRA |
| Indication for Use        |      |

1321 age Micro™

The Mirage Micro channels airflow noninvasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system.

The Mirage Micro is:

- to be used by adult patients (>66lb / >30kg) for whom positive airway pressure has been prescribed. י
- intended for single patient re-use in the home environment and multi-patient reuse in the hospital/institutional environment.

| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X | AND/OR | Over-The-Counter Use<br>(Part 21 CFR 807 Subpart C) |  |
|-------------------------------------------------|---|--------|-----------------------------------------------------|--|
|-------------------------------------------------|---|--------|-----------------------------------------------------|--|

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH; Office of Device Evaluation (ODE)

|                | <p>Division Sign-Off)</p>                                                        | Page 1 of 1 |
|----------------|----------------------------------------------------------------------------------|-------------|
|                | Division of Anesthesioloov: General Hospital<br>infection Control, Denta Devices |             |
| 710(k) Number: | K081321                                                                          |             |

---

**Source:** [https://fda.innolitics.com/device/K081321](https://fda.innolitics.com/device/K081321)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
