← Product Code [JSS](/productcode/JSS) · K081309

# E.COLI/P. AERUGINOSA PNA FISH (K081309)

_Advandx, Inc. · JSS · Dec 19, 2008 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K081309

## Device Facts

- **Applicant:** Advandx, Inc.
- **Product Code:** [JSS](/productcode/JSS.md)
- **Decision Date:** Dec 19, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.2660
- **Device Class:** Class 1
- **Review Panel:** Microbiology

## Indications for Use

Escherichia coli/Pseudomonas aeruginosa PNA FISH is a multicolor, qualitative nucleic acid hybridization assay intended for identification of Escherichia coli and Pseudomonas aeruginosa on smears from positive blood cultures containing Gram-negative rods. The E.coli/P. aeruginosa PNA FISH assay is intended for use in conjunction with positive blood subcultures as an aid in the identification of E.coli and/or P. aeruginosa.

## Device Story

E. coli/P. aeruginosa PNA FISH is a qualitative nucleic acid hybridization assay; utilizes Peptide Nucleic Acid (PNA) probes for identification of E. coli and P. aeruginosa directly from positive blood culture smears containing Gram-negative rods. Performed in clinical laboratories by trained personnel. Input: smear from positive blood culture. Process: hybridization of fluorescently labeled PNA probes to target bacterial rRNA; visualization via fluorescence microscopy. Output: qualitative identification of E. coli (green fluorescence) and P. aeruginosa (red fluorescence). Results used as an aid to conventional subculture identification; enables faster pathogen identification to support clinical decision-making regarding antimicrobial therapy; potential benefit of reduced time-to-appropriate-treatment for patients with bloodstream infections.

## Clinical Evidence

No clinical data provided in the summary document. Substantial equivalence based on bench testing and performance validation of the PNA FISH hybridization assay.

## Technological Characteristics

Multicolor qualitative nucleic acid hybridization assay. Utilizes PNA probes targeting bacterial rRNA. Form factor: kit containing reagents for fluorescence in situ hybridization. Detection method: fluorescence microscopy. Standalone diagnostic test.

## Regulatory Identification

A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.

## Predicate Devices

- S. aureus PNA FISH ([K052636](/device/K052636.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE

A. 510(k) Number:
K081309

B. Purpose for Submission:
To determine substantial equivalence of the device for the identification of *E. coli* and/or *P. aeruginosa* on smears from positive blood cultures containing Gram-negative rods.

C. Measurand:
*E. coli* and *P. aeruginosa* specific ribosomal RNA sequences

D. Type of Test:
Fluorescence In Situ Hybridization (FISH) using protein nucleic acid (PNA) probes

E. Applicant:
AdvanDx, Inc

F. Proprietary and Established Names:
*E. coli/P. aeruginosa PNA FISH™*

G. Regulatory Information:
1. Regulation section:
866.2660
2. Classification:
Class I
3. Product code:
JSS
4. Panel:

{1}

83 Microbiology

## H. Intended Use:

1. Intended use(s):

Escherichia coli/Pseudomonas aeruginosa PNA FISH is a multicolor, qualitative nucleic acid hybridization assay intended for identification of Escherichia coli and Pseudomonas aeruginosa on smears from positive blood cultures containing Gram-negative rods. The E.coli/P. aeruginosa PNA FISH assay is intended for use in conjunction with positive blood subcultures as an aid in the identification of E.coli and/or P. aeruginosa.

2. Indication(s) for use:

Escherichia coli/Pseudomonas aeruginosa PNA FISH is a multicolor, qualitative nucleic acid hybridization assay intended for identification of Escherichia coli and Pseudomonas aeruginosa on smears from positive blood cultures containing Gram-negative rods. The E.coli/P. aeruginosa PNA FISH assay is intended for use in conjunction with positive blood subcultures as an aid in the identification of E.coli and/or P. aeruginosa.

3. Special conditions for use statement(s):

Prescription use only

4. Special instrument requirements:

Dual Band Filter (Cat. No. AC003)

Microscope Slides (Cat. No. AC001)

## I. Device Description:

A mixture of fluorescein-labeled, E. coli specific PNA probe and a Texas Red labeled, P. aeruginosa specific PNA probe is added to a smear prepared from a positive blood culture. Hybridization is performed at 55°C for 90 min. The hybridization is followed by a rinse step to remove the cover slip and a post-hybridization wash at 55°C for 30 min. with a stringent wash solution to remove unbound PNA probe. The smear is finally mounted with Mounting Medium for examination with fluorescence microscopy (Dual Band Filter). E. coli cells show green fluorescence whereas P. aeruginosa cells show red fluorescence.

## J. Substantial Equivalence Information:

1. Predicate device name(s):

{2}

E. faecalis/OE PNA FISH

2. Predicate 510(k) number(s):

K063127

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Technology | Fluorescence In Situ Hybridization (FISH) using protein nucleic acid (PNA) probe | Same  |
|  Sample | Positive blood culture | Same  |
|  Interpretation of Results | Qualitative Fluorescence microscope | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Function | Identification of Escherichia coli and Pseudomonas aeruginosa | Identification of E. faecalis and other Enterococci  |
|  Control organisms | Pos control: E. coli and P. aeruginosaNeg control: Klebsiella spp | Pos control: E. faecalis and E. faeciumNeg control: Streptococcus spp  |
|  PNA Probes | E. coli and P. aeruginosa | E. faecalis and other Enterococci  |

## K. Standard/Guidance Document Referenced (if applicable):

Non applicable

## L. Test Principle:

A mixture of a fluorescein-labeled, E.coli specific PNA probe and a Texas Red-label, P. aeruginosa specific PNA probe is added to a smear prepared from a positive blood culture. Hybridization is performed at 55°C for 90 minutes. The hybridization is followed by a water rinse at 55°C to remove the cover slips followed by a wash at 55°C for 30 min with a stringent Wash Solution. Finally, the smear is mounted with Mounting Medium and examined by fluorescence microscopy.

## M. Performance Characteristics (if/when applicable):

1. Analytical performance:

{3}

a. Precision/Reproducibility:

A Reproducibility study for *E. coli/P. aeruginosa* PNA FISH assay was performed by using ten reference isolates of Gram negative rods, once per day with positive and negative controls, over a period of three days at three different sites, by one operator at each site. Results showed 100% precision and reproducibility between and within sites.

b. Linearity/assay reportable range:

Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Positive and negative control slides were performed at each testing site. All results were as expected.

d. Detection limit:

The detection limit was determined to be approximately 10⁵ CFU/mL by serial dilutions of *E. coli* or *P. aeruginosa* positive cultures. Serial dilutions of exponentially growing cultures were prepared. Then 0.01 mL was plated on growth media and 0.01 mL was used for preparation of smears. The smears were run through PNA FISH and scored positive or negative. The following day, colonies were counted on the plates and the average number of colonies per dilution was calculated. The data sets showed a minimum of 10⁵ CFU/mL to produce a positive result for the *E. coli/P. aeruginosa* PNA FISH™ assay.

e. Analytical specificity:

The analytical specificity of the *E. coli/P. aeruginosa* PNA FISH™ was determined by BLAST search and sequence alignments and experimentally by testing of well characterized laboratory and reference strains comprising of 14 *E. coli* and 17 *P. aeruginosa*, 76 additional Gram negative, 12 Gram positive organisms and 7 yeasts. All *E. coli* and *P. aeruginosa* were correctly identified. *Brevundimonas diminuta*, *Herbaspirillum huttiense*, *Pseudomonas nitroreducens* and *Pseudomonas fulva* cross reacted to create a red signal; *Shigella* spp, (serogroup A, B, C, or D), *Escherichia fergusonii* and *Escherichia albertii* cross reacted to create a green signal. All other strains were negative.

Interference

A study consisting of 15 Gram negative rods were tested on BACTEC Plus blood culture bottles for the interference of resin. No interferences were observed. Current peer-reviewed publications indicated that sodium polyanetholesulfonate (SPS) or charcoal does not cause interference with PNA FISH assays.

{4}

f. Assay cut-off:

Not applicable

2. Comparison studies:

a. Method comparison of device to conventional methods, as the reference method:

Performance results compare to routine identifications, based on conventional methods following subculture.

b. Matrix comparison:

Not applicable

3. Clinical studies:

a. Clinical Sensitivity:

Clinical Study was conducted at four sites in the United States and Europe, directly on blood culture bottles containing Gram negative rods (GNR). A total of 240 GNR-positive blood bottles, from two commercial continuously monitoring blood culture systems (BacT/Alert and BACTEC) were included in the study. Performance results compare to routine identifications, based on conventional methods following subculture, was summarized below.

|  Study | Sensitivity E. coli (Green) | Sensitivity P. aeruginosa (Red) | Specificity | Blood Culture System  |
| --- | --- | --- | --- | --- |
|  A | 100% | 100% | 100% | BacT/Alert  |
|   |  (14/14) | (11/11) | (25/25) |   |
|   |  95%CI (80.7-100) | 95%CI (76.2-100) | 95%CI (88.7-100) |   |
|  B | 100% | 100% | 96.6% | BacT/Alert  |
|   |  (35/35) | (9/9) | (28/29) |   |
|   |  95%CI (91.8-100) | 95%CI (71.7-100) | 95%CI (92.2- 99.9) |   |
|  C | 100% | 92.3% | 93.1% | BACTEC  |
|   |  (20/20) | (12/13) | (27/29) |   |
|   |  95%CI (86.1-100) | 95%CI (64.0-99.8) | 95%CI (77.7-99.2) |   |
|  D | 100% | 100% | 100% | BacT/Alert  |
|   |  (32/32) | (2/2) | (21/21) |   |
|   |  95%CI (91.1-100) | 95%CI (22.4-100) | 95%CI (86.7-100) |   |
|  Total | 100% | 97.1% | 97.1% |   |
|   |  (101/101) | (34/35) | (101/104) |   |
|   |  95% CI (97.1 - 100) | 95% CI (85.1 - 99.9) | 95% CI (91.8 - 99.4) |   |

b. Clinical specificity:

{5}

Refer to table above

c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
E. coli cells: green fluorescence
P. aeruginosa cells: red fluorescence

The expected positive rate from positive blood culture bottles is 37% for E.coli and 13% for P. aeruginosa.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The information submitted in this premarket notification is complete and supports a substantial equivalence decision

---

**Source:** [https://fda.innolitics.com/device/K081309](https://fda.innolitics.com/device/K081309)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
