← Product Code [HIS](/productcode/HIS) · K081265

# THAI NIPPON MALE LATEX CONDOM (K081265)

_Thai Nippon Rubber Industry Co., Ltd. · HIS · Oct 16, 2009 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K081265

## Device Facts

- **Applicant:** Thai Nippon Rubber Industry Co., Ltd.
- **Product Code:** [HIS](/productcode/HIS.md)
- **Decision Date:** Oct 16, 2009
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The Thai Nippon Male Latex Condom is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted diseases)

## Device Story

Natural rubber latex sheath; covers erect penis; flared closed end with reservoir tip; nominal flat width 56mm, length 200mm, thickness 0.07mm. Lubricated with silicone (250 cps); cornstarch dressing. Used by men for contraception and prophylaxis against pregnancy and STDs. OTC use.

## Clinical Evidence

No clinical data. Device conforms to ASTM D3492:2008 and ISO 4074:2002 standards.

## Technological Characteristics

Natural rubber latex; silicone lubricant (250 cps); cornstarch dressing. Dimensions: 56mm nominal flat width, 200mm length, 0.07mm thickness. Conforms to ASTM D3492:2008 and ISO 4074:2002.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- Church & Dwight Male Latex Condom ([K001212](/device/K001212.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Abbreviated 510 {k} Notification for a Male Latex Condom

#### I. 510(k) SUMMARY

Submitted By: Thai Nippon Rubber Industry Co., Ltd. Laem Chabang Industrial Estate, 49-49/1 Export Processing Zone 1 Thungsukla, Sriracha, Chonburi Thailand

> 66 38 490258 Tel: Fax: 66 38 490414 Email: tnrfact(@tnrcondom.com

Contact Person: Eli J. Carter, Consultant PO Box 12139 Durham. NC 27709 Tel: 919 544 4098; Fax: 919 544 5849 Email: carterej@aol.com

Date Prepared: April 28, 2008

Proprietary Name: None

Common Name: Male Latex Condom: Flared, 56mm (nominal width)

Classification Name: Male Latex Condom

Predicate Devices: Church & Dwight Male Latex Condom - K001212

### Description of Device:

This condom is made of a natural rubber latex sheath, which completely covers the erect penis with a closely fitted membrane. This condom is colorless, smooth, flared at the closed end with a reservoir tip; nominal lav flat width 56mm, nominal length 200 mm, and nominal thickness 0.07mm. It is lubricated with silicone (viscosity 250 cps), and cornstarch is used as a dressing material. This condom conforms to current established national and international voluntary standards including ASTM D3492:2008 and ISO 4074:2002.

OCT 1 6 2009

{1}------------------------------------------------

## Intended Use of the Device:

This latex condom has the same intended use as the predicate condom. The condom is used for contraception and for prophylactic purposes to help prevent pregnancy and the transmission of sexually transmitted diseases, including HIV.

## Technological Characteristics:

This condom has the same technological characteristics as the predicate condoms identified above. It is made from natural rubber latex and the design is in conformance with current versions of ASTM D3492 and ISO 4074 Male Latex Condom Standards. The Thai Nippon condom is designed with a wider width of 56mm for the man who needs a larger size condom. This condom is intended for men who feel that current regular size condoms are too small.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of an abstract symbol, possibly representing a person or a flame.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Thai Nippon Rubber Industry c/o Mr. Eli Carter Consultant Eli Carter and Associates PO Box 12139 DURHAM NC 27709

# OCT 1 6 2009

Re: K081265

Trade/Device Name: Thai Nippon Malc Latex Condom 56mm (nominal width), flared, Colorless, silicone lubricated

Regulation Number: 21 CFR §884.5300 Regulation Name: Condom Regulatory Class: II Product Code: HIS Dated: September 28, 2009 Received: October 5, 2009

Dear Mr. Carter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 -

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Singerely yours

Janine M. Morris Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Information for a Male Latex Condom 510(k) Submission

#### II. INDICATIONS FOR USE STATEMENT

K081265 510(k) Number:

Device Name:

Thai Nippon Male Latex Condom 56mm (nominal width), flared, colorless, silicone lubricated

Indications for Use: The Thai Nippon Male Latex Condom is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted diseases)

### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEED Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Over-the-Counter Use or

X

formh

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number,

---

**Source:** [https://fda.innolitics.com/device/K081265](https://fda.innolitics.com/device/K081265)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
