The CG-6108 ACT-3L Continuous ECG Monitor and Arthythmia Detector is intended for use by palients who experience transient symptoms that may suggest cardiac arrhythmia. The device continuously monitors patient ECG, automatically generates an alarm triggered by an arrhythmia detection algorithm, or generates an alarm manually triggered by the patient, and transmits the recorded data transtele-phonically to a monitoring center. The monitoring center provides the ECG data to the medical practitioner for evaluation.
Device Story
System comprises chest-worn 3-lead ECG sensor and handheld device (PDA/Smartphone/PC) running proprietary medical application. Sensor acquires ECG signals via electrodes; transmits via Bluetooth to handheld unit. Handheld unit utilizes dual memory loop configuration for pre/post-event recording. Arrhythmia detection algorithm monitors ECG; triggers automatic alarm upon event detection; patient can also manually trigger alarm. Handheld device transmits recorded ECG data to remote Monitoring Center via cellular link or landline. Monitoring Center staff analyze data; provide results to medical practitioner for clinical decision-making. Enables remote cardiac monitoring for patients with transient symptoms; facilitates timely professional intervention.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by verification, validation, and testing (VV&T) documentation, including adherence to ANSI/AAMI EC57:1998 for cardiac rhythm and ST segment measurement algorithms and ANSI/AAMI EC38:1998 for ambulatory ECG.
Technological Characteristics
Chest-worn 3-lead ECG sensor with ASIC and Bluetooth transceiver. Handheld device (PDA/Smartphone/PC) with proprietary application. Dual memory loop configuration for signal processing. Connectivity via Bluetooth (sensor-to-handheld) and cellular/landline (handheld-to-center). Complies with IEC 60601-1, IEC 60601-2-27, and EN 60601-1-2 (EMC).
Indications for Use
Indicated for patients experiencing transient symptoms suggestive of cardiac arrhythmia. Used for continuous ECG monitoring, arrhythmia detection, and transmission of data to a monitoring center for professional evaluation.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Card Guard's CG-6108 Continuous ECG Monitor And Arrhythmia Detector (K071995)
Card Guard's CG-6550 Personal ECG Transmitter (K990478)
Submission Summary (Full Text)
{0}------------------------------------------------
P1/2
# MAY 2 9 2008
**CARD. GUARD.**
## CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector 510(k) Summary of Safety and Effectiveness
- IEC 60601-2-27 2005 Medical electrical equipment Part 2: Particular requirements for the safety of ● electrocardiographic monitoring equipment
- EN 475: Medical devices Electrically-generated alarm signals ; April 1995 .
- EN 980: Graphical symbols for use in the labeling of medical devices; August 2003 .
- EN 1041: Terminology, Symbols and Information provided with Medical Devices; Information supplied by the e manufacturer with medical devices; April 1998
- EN ISO 9001: Quality management systems Requirements; December 2000 ●
- EN ISO 13485: Quality systems Medical devices; August 2000 .
- EN 1SO 14971: Medical devices -- application of risk management to medical devices; March 2001 .
- EN ISO 10993 Biological evaluation of medical devices Part 1; Evaluation and testing; Dec. 1997 .
- EN 60601-1: Medical electrical equipment; Part 1: General requirements for safety; Sept. 2002 ●
- EN 60601-1-2: Medical electrical equipment; Part 1: 2. Collateral Std: EMC; requirements and tests; 2007 ●
- EN 60601-1-4+A1 2000: Medical electrical equipment; Part 1: 4. Collateral Std: Programmable electric � medical systems;
### Indications for Use, Intended Use
The CG-6108 ACT-3L Continuous ECG Monitor and Arthythmia Detector is intended for use by palients who experience transient symptoms that may suggest cardiac arrhythmia. The device continuously monitors patient ECG, automatically generates an alarm triggered by an arrhythmia detection algorithm, or generates an alarm manually triggered by the patient, and transmits the recorded data transtele-phonically to a monitoring center. The monitoring center provides the ECG data to the medical practitioner for evaluation.
#### Principles of operation
The chest-worn ECG sensor transmits signals via Bluetooth to the handheld device equipped with the Medical Application, which incorporates an algorithm specially developed for detection of AF events. A detected artifact triggers transmission of the Signal to the CG Monitoring Center for analysis.
## Substantially Equivalent Devices
The clearance for the CG-6108 ACT-3L is sought on the grounds of its claimed substantial equivalence (SE) to the following predicate devices:
- Card Guard's CG-6108 Continuous ECG Monitor And Arrhythmia Detector K071995 for being the 1. parent model
- Card Guard's CG-6550 Personal ECG Transmitter K990478 for the 3 leads/3 channel design. 2.
#### Conclusions
All Verification, Validation and Testing (VV&T) documents - although not included in this dossier, constitute a part of the device DMR and are available upon request
The Truthful and Accurate Statement complies with 21 CFR 807.87(j).
We trust that the submitted information will enable the reviewer to process the material promptly. The time factor is crucial for Card Guard's commercial interests.
This dossier contains one paper copy of the 510(k) submission together with an electronic copy that is an exact duplicate of the paper copy.
{1}------------------------------------------------
| Submitter | Card Guard Scientific Survival Ltd., |
|-----------------|---------------------------------------------------------------|
| Address | 2 Pekeris St. P.O.B. 527 Rehovot 76100, Israel |
| Contact: | Alex Gonorovsky, RA Manager |
| Phone: | 972-8-9484019 Fax: 972-8-9484044 |
| E-mail: | galex@cardguard.com |
| Device | |
| Trade Name: | CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector |
| Classification: | detector and alarm, arrhythmia |
| Product Code: | DSI DXH |
| Regulation No: | 870.1025 |
| Class: | II |
## Device Definition
The CG-6108 ACT-3L is designed for self-testing by patients at home and for analysis by medical professionals at a remote Monitoring Center. It comprises a chest-worn ECG sensor and a handheld device with a proprietary application, configured to process and transmit the ECG recordings.
The chest-worn unit has 3 electrodes on a harness and it houses a battery, an ASIC and a Bluetooth transceiver for the acquisition, recording, and transmission of the ECG signal.
The ECG signals are transmitted via Bluetooth to the handheld device. When an event is detected it is wirelessly transmitted to the CG Monitoring Center for professional analysis. The handheld device is equipped with shared memory used to received from the sensor and to allow preand post processing options through the use of this memory in a dual memory loop configuration, both running in parallel. One loop is auto-triggered, with programmable that starts recording based on specific rhythms detected or manually activated by the patient. The second, and longer, recording loop is controlled to provide the physician with more information, when requested by the CG Monitoring Center.
The handheld device automatically transmits the recorded ECG, via cellular link, to the Monitoring Center. When cellular service is unavailable the patient can transmit via landline telephone.
#### Medical Application
The Application is designed for wireless mobile platforms, e.g. PDA, SmartPhone and for static platforms, such as PC. It is used to receive from the CG-6108 the test results and other medical data, to process and save these test results, and synchronize data and test results with the CG Medical Center. The Application is a part of a personal medical system solution. It performs the following activities:
- 1. Receives medical test inputs from the external accessories
- Collects medical test data and other related information as defined for each test 2.
- Accesses historical test and related data stored on the device 3.
- Transmits medical test data and other information to Center for professional evaluation/backup র্ধ
- క్. Receives data from Center
- Enables configuring GPRS data connection (based on mobile phone GPRS/CDMA capabilities), ర్. changing user name and password.
## Referenced Standards
- Arrhythmia Detector and Alarm Guidance for Industry and FDA Staff Class II Special Controls Guidance
- ANSVAAMI-EC 57:1998, Testing and Reporting Performance Results of Cardiac Rhythm and ST Segment � Measurement Algorithms
- ANSI/AAMI EC38:1998 Ambulatory Electrocardiograph �
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 9 2008
Card Guard Scientific Survival Ltd. c/o Mr. Alex Gonorovsky Manager, Regulatory Affairs 2 Pekeris St. P.O. Box 527 Rehovot 76101 ISRAEL
Re: K081257
CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Dctector Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm Regulatory Class: Class II (two) Product Code: DSI Dated: March 26, 2008 Received: May 2, 2008
Dear Mr. Gonorovsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Alex Gonorovsky
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050, This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial cquivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
B. Sommermeyer for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): Ko81257
## Device Name: CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector
Indications for Use:
The CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector is intended for use by patients who experience transient symptoms that may suggest cardiac arrhythmia. The device continuously monitors patient ECG, automatically generates an alarm triggered by an arrhythmia detection algorithm, or generates an alarm manually triggered by the patient, and transmits the recorded data transtelephonically to a monitoring center. The monitoring center provides the ECG data to the medical practitioner for evaluation.
Prescription Use 181 (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C).
..............................................................................................................................................................................
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blummmmor
(ion sign-om)
Division of Cardiovascular Devices
510(k) Number Korld57
Page 1 of 1
(Posted November 13, 2003)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.