← Product Code [FOZ](/productcode/FOZ) · K081113

# COOK AND COOK SPECTRUM CENTRAL VENOUS CATHETERS (K081113)

_Cook, Inc. · FOZ · Jul 30, 2008 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K081113

## Device Facts

- **Applicant:** Cook, Inc.
- **Product Code:** [FOZ](/productcode/FOZ.md)
- **Decision Date:** Jul 30, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5200
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Cook Central Venous Catheter with or without Heparin is used for: - Continuous or intermittent drug infusions . - Central venous blood pressure monitoring (CVP) . - Acute hyperalimentation . - Blood sampling . - Delivery of whole blood or blood products . - Power injection of contrast media* . The device is a short-term use catheter. * The flow rate of the Cook Central Venous Catheters may not exceed 10 ml/sec. The Cook Spectrum and Spectrum Glide Central Venous Catheter is used for: - Continuous or intermittent drug infusions . - Central venous blood pressure monitoring (CVP) ● - Acute hyperalimentation ● - Blood sampling ● - Delivery of whole blood or blood products . - Power injection of contrast media* . The activity of the antimicrobial agents, minocycline and rifampin, is localized at the internal and external catheter surface and helps to provide protection against catheterrelated bloodstream infections (CRBSI). It is not intended for treatment of existing infection. The device is a short-term use catheter. * The flow rate of the Cook Spectrum and Spectrum Glide Central Venous Catheters may not exceed 10 ml/sec.

## Device Story

Central venous catheters with multiple non-communicating lumens; used for short-term vascular access. Spectrum/Spectrum Glide models impregnated with minocycline and rifampin to reduce risk of catheter-related bloodstream infections (CRBSI). Physician-operated in clinical settings; provides access for drug infusion, blood pressure monitoring, hyperalimentation, blood sampling, and power injection of contrast media. Output is physical access to central venous system; allows clinicians to manage patient fluid/medication needs. Benefits include reduced infection risk (Spectrum models) and versatile clinical utility for critical care.

## Clinical Evidence

No clinical data provided. Substantial equivalence based on physical identity to predicate devices and labeling expansion.

## Technological Characteristics

Multi-lumen catheter body; available in 7F to 10F sizes. Spectrum models impregnated with minocycline and rifampin. Designed for short-term use. Max power injection flow rate 10 mL/sec; max safety cut-off pressure 325 psi. No software or electronic components.

## Regulatory Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

## Predicate Devices

- Spectrum Central Venous Catheters with or without Hydrophilic Coating ([K033843](/device/K033843.md))
- Five Lumen Central Venous Catheter ([K032274](/device/K032274.md))
- Spectrum Five Lumen Central Venous Catheter ([K060174](/device/K060174.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### 5. 510(k) Summary

JUL 3 0 2008

## Cook Incorporated Central Venous Catheters 510(k) Summary 21 CFR 807.92

### 1. Submitter Information:

| Applicant:             | Cook Incorporated                                                                                  |
|------------------------|----------------------------------------------------------------------------------------------------|
| Address:               | 750 Daniels Way<br>Bloomington, IN 47404                                                           |
| Phone Number:          | (800) 468-1379                                                                                     |
| Fax Number:            | (812) 332-0281                                                                                     |
| Contact:               | Susanne Galin, Regulatory Affairs Specialist                                                       |
| Contact Address:       | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404                                      |
| Contact Phone Number:  | (812) 339-2235<br>(812) 332-0281                                                                   |
| 2. Device Information: |                                                                                                    |
| Trade Name:            | Cook Incorporated Central Venous Catheters and<br>Spectrum/Spectrum Glide Central Venous Catheters |
| Common Name:           | Central Venous Catheters                                                                           |
| Classification:        | Class II                                                                                           |
| Product Code:          | FOZ (21 CFR Part 880.5200)                                                                         |

#### 3. Predicate Device:

Cook Incorporated's Central Venous Catheters with their expanded indications for use statement are substantially equivalent to the Spectrum Central Venous Catheters with or without Hydrophilic Coating cleared under FDA 510(k) number K033843, the Five Lumen Central Venous Catheter cleared under 510(k) number K032274, and the Spectrum Five Lumen Central Venous Catheter cleared under 510(k) number K060174. They are similar to the Cook Incorporated Central Venous Catheters marketed prior to 1976.

{1}------------------------------------------------

# 4. Device Description:

Cook Incorporated's Central Venous Catheters incorporate separate, non-communicating vascular access lumens within a single catheter body. These catheters may be impregnated with minocycline and rifampin to help provide protection against catheterrelated blood stream infections (CRBSI).

Several sizes are available to allow physicians to choose the catheter that matches anatomical needs of the patient, as well as the number of lumens required for treatment.

| Catheter Size<br>(French) | Number<br>of<br>Lumens | Length (cm)    | Maximum Power<br>Injection Flow<br>Rate (mL/sec) | Maximum Safety<br>Cut-off<br>Pressure Limit (psi) |
|---------------------------|------------------------|----------------|--------------------------------------------------|---------------------------------------------------|
| 7                         | 3                      | 15, 20, 25     | 10                                               | 325                                               |
| 8                         | 2                      | 15, 20, 25     | 10                                               | 325                                               |
| 9                         | 3                      | 15, 20, 25     | 10                                               | 325                                               |
| 10                        | 5                      | 15, 20, 25, 30 | 10                                               | 325                                               |

There have been no changes in the design, dimensions, or materials of the device.

## 5a. Intended Use for the Cook Central Venous Catheter with or without Heparin:

The Cook Central Venous Catheter with or without Heparin is used for:

- Continuous or intermittent drug infusions .
- Central venous blood pressure monitoring (CVP) .
- Acute hyperalimentation .
- Blood sampling .
- Delivery of whole blood or blood products .
- Power injection of contrast media* .

The device is a short-term use catheter.

* The flow rate of the Cook Central Venous Catheters may not exceed 10 ml/sec.

{2}------------------------------------------------

Cook Incorporated-July 29, 2008 Cook Central Venous Catheters K081113-Response to FDA Inquiry

# 5b. Intended Use for the Cook Spectrum and Spectrum Glide Central Venous Catheters:

The Cook Spectrum and Spectrum Glide Central Venous Catheter is used for:

- Continuous or intermittent drug infusions .
- Central venous blood pressure monitoring (CVP) ●
- Acute hyperalimentation ●
- Blood sampling ●
- Delivery of whole blood or blood products .
- Power injection of contrast media* .

The activity of the antimicrobial agents, minocycline and rifampin, is localized at the internal and external catheter surface and helps to provide protection against catheterrelated bloodstream infections (CRBSI). It is not intended for treatment of existing infection. The device is a short-term use catheter.

* The flow rate of the Cook Spectrum and Spectrum Glide Central Venous Catheters may not exceed 10 ml/sec.

# 6. Technological Characteristics:

The Cook Central Venous Catheters described in this submission are physically identical to the predicate Central Venous Catheters in terms of technological characteristics (design, dimensions, and materials).

# 7. Reason for Filing

This submission is for an expansion in the indications for use-a change in labeling only-for Cook Central Venous Catheters. There has been no change to the design, dimensions, or materials of the existing on-market devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# JUL 30 2008

Ms. Susanne Galin, RAC Regulatory Affairs Specialist Cook Incorporated 750 Daniels Way Bloomington, Indiana 47404

Re: K081113

Trade/Device Name: Cook Spectrum and Spectrum Glide Central Venous Catheters Cook Central Venous Catheters with or without Heparin Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: July 8, 2008 Received: July 9, 2008

Dear Ms. Galin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Galin

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CHR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

### 4. Indications for Use Statement

### Indications for Use

510(k) Number (if known):

Device Name: Cook Central Venous Catheters with or without Heparin

Indications for Use:

The Cook Central Venous Catheter is used for:

- Continuous or intermittent drug infusions .
- Central venous blood pressure monitoring (CVP) .
- Acute hyperalimentation .
- Blood sampling .
- Delivery of whole blood or blood products .
- Power injection of contrast media* .

The device is a short-term use catheter.

* The flow rate of the Cook Central Venous Catheters may not exceed 10 ml/sec.

AND/OR Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# (PLEASE DO NOT WRITE BELOW THIS LINE---- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

signature

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number. K081113

Company Confidential

{6}------------------------------------------------

Cook Incorporated-July 29, 2008 Cook Central Venous Catheters K081113-Response to FDA Inquiry

### 4. Indications for Use Statement

### Indications for Use

510(k) Number (if known): K081113

Device Name: Cook Spectrum and Spectrum Glide Central Venous Catheters

Indications for Use:

The Cook Spectrum and Spectrum Glide Central Venous Catheter is used for:

- Continuous or intermittent drug infusions ●
- Central venous blood pressure monitoring (CVP) .
- . Acute hyperalimentation
- Blood sampling .
- Delivery of whole blood or blood products .
- Power injection of contrast media* .

The activity of the antimicrobial agents, minocycline and rifampin, is localized at the internal and external catheter surface and helps to provide protection against catheterrelated bloodstream infections (CRBSI). It is not intended for treatment of existing infection. The device is a short-term use catheter.

* The flow rate of the Cook Central Venous Catheters may not exceed 10 ml/sec.

Prescription Use AND/OR × (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Clen
(Division Sign Off)

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K061113

Company Confidential

---

**Source:** [https://fda.innolitics.com/device/K081113](https://fda.innolitics.com/device/K081113)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
