← Product Code [KXE](/productcode/KXE) · K081025

# ISOELASTICU (K081025)

_Remi Sciences, Inc. · KXE · Nov 21, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K081025

## Device Facts

- **Applicant:** Remi Sciences, Inc.
- **Product Code:** [KXE](/productcode/KXE.md)
- **Decision Date:** Nov 21, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3810
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The IsoelasticU™ is indicated for joint replacement arthroplasty of the ulnar head at the distal radioulnar joint (DRUJ) for the following indications - Rheumatoid arthritis with or without tendon ruptures - Degenerative arthritis or post traumatic arthritis o 1. Arthrofibrosis of the DRUJ 2. Reconstruction of the distal ulna post tumor resection 3. Failed Ulnar Head Resection 4. Distal ulna instability with xray or bone scan evidence of arthritic or inflammatory o changes 5. Revision following failed ulnar head arthroplasty o Ulna stems are intended for uncemented use

## Device Story

IsoelasticU is a modular ulnar head prosthesis for distal radioulnar joint (DRUJ) replacement. Device replaces distal ulnar head in patients with arthritis or instability. Implant consists of head and press-fit stem; intended for uncemented use. Surgeon performs arthroplasty procedure to restore joint function. Device provides mechanical replacement of ulnar head, mimicking function of predicate metallic and silicone elastomer implants. Benefits include pain relief and improved wrist joint stability.

## Clinical Evidence

No clinical data provided; substantial equivalence based on design, material, and intended use comparisons to predicate devices.

## Technological Characteristics

Modular prosthesis; materials include UHMWPE and Commercially Pure Titanium (CpTi); press-fit stem fixation; single-use; non-cemented; 19 head sizes and 9 stem sizes.

## Regulatory Identification

A wrist joint ulnar (hemi-wrist) polymer prosthesis is a mushroom-shaped device made of a medical grade silicone elastomer or ultra-high molecular weight polyethylene intended to be implanted into the intramedullary canal of the bone and held in place by a suture. Its purpose is to cover the resected end of the distal ulna to control bone overgrowth and to provide an articular surface for the radius and carpus.

## Predicate Devices

- SBI Uhead Implant
- Silastic Swanson Ulnar Head Implant

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the acronym "REMI" in large, bold letters above the word "SCIENCES" in a smaller font. To the right of "REMI" are the words "REGENERATIVE ENGINEERED MATERIAL INNOVATIONS" in a very small font. The text appears to be a logo or heading.

## 510 k SUMMARY K081025

Date of preparation: 11/11/08

NOV 21 2008

Applicant: REMI SCIENCES INC 1400 Civic Place, Suite 225 Southlake, Texas76092

Telephone: (817) 506-0835 Fax: (817) 608-9403 Contact: Sanjiv H.Naidu

| DEVICE NAME:           | Wrist joint ulnar (hemi-wrist) prosthesis              |
|------------------------|--------------------------------------------------------|
| DEVICE TRADE NAME:     | Isoclastic UTM                                         |
| DEVICE CLASSIFICATION: | Class II                                               |
| PREDICATE:             | SBI Uhead Implant, Silastic Swanson Ulnar Head Implant |

DEVICE DESCRIPTION:

The IsoelasticU ulnar head implant allows for replacement of the distal ulnar head much like the predicate metallic and silicone elastomer distal ulnar head implants.

### INDICATIONS FOR USE:

-

The IsoelasticU™ is indicated for joint replacement arthroplasty of the ulnar head at the distal radioulnar joint (DRUJ) for the following indications

- Rheumatoid arthritis with or without tendon ruptures -
	- Degenerative arthritis or post traumatic arthritis
		- o 1. Arthrofibrosis of the DRUJ
		- 2. Reconstruction of the distal ulna post tumor resection ୍
		- 3. Failed Ulnar Head Resection
		- 4. Distal ulna instability with xray or bone scan evidence of arthritic or inflammatory o changes
		- 5. Revision following failed ulnar head arthroplasty o

Ulna stems are intended for uncemented use

REMI Sciences, Inc. 1400 Civic Place, Suite 225 Southlake, Texas 76092

ph 817-608-9400 ext.1600 fax 817-608-9403 cell 214-641-5171

sanj@remisciences.com www.remisciences.com

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for REMI Sciences. The word "REMI" is in large, bold, black letters on the top left. To the right of "REMI" are the words "REGENERATIVE ENGINEERED MATERIALS INNOVATIONS" stacked on top of each other in smaller, gray letters. Below "REMI" are the words "SCIENCES" in gray letters.

Comparison to the Original Predicate Device:

IsoelasticU is substantially equivalent to the legally marketed SBI Small Bone Innovations U head implant.

Regulatory Class:II

| CATEGORY            | REMI ISOELASTIC U                                                                                                                                                                                                                                                                                                                                                                                         | SBI UHEAD                                                                                                                                                                                                                                                                                                                                                                                                              |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name                | IsoelasticU                                                                                                                                                                                                                                                                                                                                                                                               | Uhead                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Use                 | Single Use                                                                                                                                                                                                                                                                                                                                                                                                | Single Use                                                                                                                                                                                                                                                                                                                                                                                                             |
| Fixation            | Press fit stem                                                                                                                                                                                                                                                                                                                                                                                            | Press fit Stem                                                                                                                                                                                                                                                                                                                                                                                                         |
| Material            | UHMWPE/CpTitanium                                                                                                                                                                                                                                                                                                                                                                                         | Co-Cr/CpTi                                                                                                                                                                                                                                                                                                                                                                                                             |
| Sizes               | 19Head/9stems                                                                                                                                                                                                                                                                                                                                                                                             | 8head/4stem                                                                                                                                                                                                                                                                                                                                                                                                            |
| Indications for use | - Rheumatoid arthritis with or without tendon ruptures<br>- Degenerative arthritis or post traumatic arthritis<br>1. Arthrofibrosis of the DRUJ<br>2. Reconstruction of the distal ulna post tumor resection<br>3. Failed Ulnar Head Resection<br>4. Distal ulna instability with xray or bone scan evidence of arthritic or inflammatory changes<br>5. Revision following failed ulnar head arthroplasty | Is intended for replacement of the DRUJ following ulnar head resection arthroplasty: replacement of the distal ulnar head for rheumatoid, degenerative, or post-traumatic disabilities presenting the following: pain and weakness of the wrist joint not improved by conservative treatment instability of the ulnar head with xray evidence of dorsal subluxation and erosive changes and failed ular head resection |

Summary: Similarities of REMI Isoelastic U and SBI U head lies in the fact that they are both modular implants intended for use with press fit stems. Both the devices are intended for single use for implantation longer than 30 days. Both devices are made of standard industry materials with similar sizes and same indications.

> REMI Sciences, Inc. 1400 Civic Place, Suite 225 Southlake, Texas 76092

. . . .

ph 817-608-9400 ext.1600 fax 817-608-9403 cell 214-641-5171

sanj@remisciences.com www.remisciences.com

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure with three stylized wing or feather shapes.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

REMI Sciences, Inc. % Sanjiv H. Naidu, M.D., Ph.D. Chief Medical Officer 1400 Civic Place. Suite 225 Southlake, Texas 76092

NOV 2 1 2008

Re: K081025

Trade/Device Name: Isoelastic UTM Regulation Number: 21 CFR 888.3810 Regulation Name: Wrist joint ulnar (hemi-wrist) polymer prosthesis Regulatory Class: Class II Product Code: KXF Dated: November 1, 2008 Received: November 6, 2008

Dear Dr. Naidu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

· Page 2 - Sanjiv H. Naidu, M.D., Ph.D.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark H. Millman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for REMI Sciences. The logo is split into two lines, with "REMI" on the first line and "SCIENCES" on the second line. To the right of "REMI" are the words "REGENERATIVE ENGINEERED MATERIAL INNOVATIONS" stacked on top of each other in a smaller font.

# Indications for Use

## 510 (K) Number: K081025

#### Device Name: IsoelasticU™

Indications for Use:

The IsoelasticU™ is indicated for joint replacement arthroplasty of the ulnar head at the distal radioulnar joint (DRUJ) for the following indications

- -Rheumatoid arthritis with or without tendon ruptures
- -Degenerative arthritis or post traumatic arthritis
	- 1. Arthrofibrosis of the DRUJ o
	- 2. Reconstruction of the distal ulna post tumor resection o
	- 3. Failed Ulnar Head Resection 0
	- 4. Distal ulna instability with xray or bone scan evidence of arthritic or inflammatory O changes
	- 5. Revision following failed ulnar head arthroplasty O
- Ulna stems are intended for uncemented use -

Prescription Use × AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (Part 21 CFR 801 Subpart C)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

**510(k) Number** lo/r/o2x

Page 1 of 1

REMI Sciences, Inc.
1400 Civic Place, Suite 225
Southlake, Texas 76092

1976 33 33 35 35 45 45 45 45 35 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45 4

ph 817-608-9400 ext.1600 fax 817-608-9403 cell 214-641-5171

sanj@remisciences.com www.remisciences.com

---

**Source:** [https://fda.innolitics.com/device/K081025](https://fda.innolitics.com/device/K081025)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
