K080959 · Dornier Medtech America, Inc. · GEX · Jun 27, 2008 · General, Plastic Surgery
Device Facts
Record ID
K080959
Device Name
DORNIER MEDILAS D 30 LASER, MEDILAS D 1064
Applicant
Dornier Medtech America, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 27, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Dornier Medilas D 1064 Laser is indicated for use in medicine and surgery, in the following medical specialties: Urology, Plastic Surgery, General Surgery, Dermatology, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, ENT, Radiology. The Dornier Medilas D 1064 Laser is intended for use in cutting, vaporization, ablation and coagulation of soft tissue in conjunction with endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, gastroscopes, cystoscopes, and colonoscopies), in incision/excision, vaporization, ablation and coaqulation of soft tissue in contact and non-contact open surgery (with or without a handpiece), and in the treatment and/or removal of vascular lesions (tumors).
Device Story
Continuous-wave diode laser emitting 1064 nm radiation; used for soft tissue surgery (cutting, vaporization, ablation, coagulation) and vascular lesion removal. Operated by physicians in clinical settings (OR, endoscopy suites). System features graphic display for monitoring operating parameters, application modes, time functions, and system status. Laser energy delivered via fibers in contact or non-contact modes, with or without handpieces. Provides precise tissue interaction for various surgical specialties; benefits include controlled tissue destruction and hemostasis.
Clinical Evidence
Bench testing only. Device complies with Class IV performance standards (21 C.F.R. §1040.10 and §1040.11) and voluntary standards including IEC 60601-1, IEC 60601-1-6, IEC 60601-2-22, and IEC 60825-1. No clinical data provided.
Technological Characteristics
Continuous-wave diode laser; 1064 nm wavelength. Class IV laser. Graphic display interface for parameter control. Operates in Standard, Fibertom, LITT, and LPS modes. Complies with IEC 60601-1, IEC 60601-1-6, IEC 60601-2-22, and IEC 60825-1. Standalone system.
Indications for Use
Indicated for soft tissue cutting, vaporization, ablation, and coagulation in urology, plastic surgery, general surgery, dermatology, gynecology, pulmonary surgery, neurosurgery, gastroenterology, ENT, and radiology. Used in endoscopic procedures or open surgery (contact/non-contact). Also indicated for treatment/removal of vascular lesions (tumors). Prescription use only.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Dornier Medilas D family lasers (Medilas D LiteBeam, Medilas D LiteBeam+, Medilas D UroBeam, Medilas D MultiBeam, Medilas D FlexiPulse (K070536))
{0}------------------------------------------------
JUN 2-7-2008-
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92.
| Submitter Name and<br>Address: | Dornier MedTech America, Inc.<br>1155 Roberts Blvd.<br>Kennesaw, GA 30144 |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Theron Gober,<br>Quality & Regulatory Manager |
| Phone Number: | 770-514-6204 |
| Fax Number: | 770-514-6288 |
| Establishment<br>Registration Number: | 1037955 |
| Date Prepared: | February 23, 2007 |
| Device Trade Name(s): | Medilas D 30 Laser, model name Medilas D 1064 |
| Device Common Name: | Diode Laser System |
| Classification Name: | GEX - Laser Instrument, Surgical Powered |
| Predicate Device(s): | The Dornier Medilas D 1064 Laser is substantially equivalent<br>in terms of its technological performance to Dornier Medilas D family lasers, including Medilas D<br>LiteBeam, Medilas D LiteBeam+, Medilas D UroBeam,<br>Medilas D MultiBeam, Medilas D FlexiPulse<br>(K070536) Diomed Delta 15 and Diomed Delta 30 (K051996) The Dornier Medilas D 1064 Laser is substantially equivalent<br>in terms of its laser wavelength performance to Dornier Medilas Fibertome model 5100 (K964760) Diomed Delta 25 (K063828) Laserscope Lyra Surgical Laser System (K020021) |
| General Device<br>Description: | The Dornier Medilas D 1064 Laser is a continuous-wave<br>diode laser emitting laser radiation in the invisible range of<br>1064 nm. Each is calibrated during the manufacturing<br>process and during service calls. The end-user does not<br>calibrate fibers for this system. The Medilas D 1064 Laser<br>incorporate a graphic display panel, which shows laser<br>operating parameters, application modes, time functions,<br>system status and messages for the user. The Medilas D<br>1064 Laser features several operating modes, including |
{1}------------------------------------------------
Standard, Fibertom, LITT and LPS. The laser can be used in contact or non-contact open surgery with or without handpieces.
Intended Use: The Dornier Medilas D 1064 Laser is indicated for use in medicine and surgery, in the following medical specialties: Urology, Plastic Surgery, General Surgery, Dermatology, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, ENT, Radiology.
> The Dornier Medilas D 1064 Laser is intended for use in cutting, vaporization, ablation and coagulation of soft tissue in conjunction with endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, gastroscopes, cystoscopes, and colonoscopies), in incision/excision, vaporization, ablation and coaqulation of soft tissue in contact and non-contact open surgery (with or without a handpiece), and in the treatment and/or removal of vascular lesions (tumors).
Technological From a clinical perspective and comparing design specifications, the Dornier Medilas D 1064 Laser and the Characteristics: predicate devices are substantially equivalent. Based on the technological characteristics and overall performance of the devices, Dornier MedTech America, Inc. believes that no significant differences exist between the proposed diode laser and the predicate devices.
> Dornier MedTech America, Inc. believes the minor differences of the Medilas D 1064 Laser and its predicates lasers should not raise any concerns regarding the overall safety or effectiveness.
Performance Data: While no performance standards have been established for Diode lasers under Section 514 of the Federal Food, Drug, and Cosmetic Act, the Dornier Medilas D 1064 Laser is in compliance with class IV performance standards for light emitting products promulgated under the Radiation Control for Health and Safety Act of 1968. See 21 C.F.R. §1040.10 and §1040.11. The laser also complies with the applicable requirements of the following voluntary standards: IEC 60601-1, IEC 60601-1-6, IEC 60601-2-22, IEC 60825-1, and European Medical Device Directive (CE).
Conclusion: Based on the technological characteristics and overall performance of the devices, Dornier MedTech America, Inc. believes that the Medilas D 1064 Laser and the predicate
{2}------------------------------------------------
devices selected are substantially equivalent and that the differences between the devices are minor which do not raise new issues of safety or effectiveness.
Dornier MedTech America, Inc. Medilas D 1064 Laser 510(k) Submission February 23, 2007
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three overlapping, curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 7 2008
Dornier Medtech America, Inc. % Mr. Theron Gober Quality and Regulatory Manager 1155 Roberts Boulevard, Suite 100 Kennesaw, Georgia 30144
Re: K080959
Trade/Device Name: Medilas D 1064 Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical insgrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: June 9, 2008 Received: June 10, 2008
Dear Mr. Gober:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labcling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Theron Gober
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark M. Mullison
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
| 510(k) Number (if known): | K080954 |
|---------------------------|---------|
|---------------------------|---------|
Device Name: Medilas D 1064 Laser
## Indications for Use:
The Dornier Medilas D 1064 Laser model (trade name Medilas D30) is indicated for use in medicine and surgery, in the following medical specialties:
- Urology �
- Plastic Surgery �
- . General Surgery
- ◆ Dermatology
- Gynecology �
- � Pulmonary Surgery
- Neurosurgery ●
- . Gastroenterology
- . ENT
- Radiology .
The Dornier Medilas D 1064 Laser model (trade name Medilas D30) is intended for use in cutting, vaporization, ablation and coagulation of soft tissue in conjunction with endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, gastroscopes, cystoscopes, and colonoscopies), in incision/excision, vaporization, ablation and coagulation of soft tissue in contact and non-contact open surgery (with or without a handpiece), and in the treatment and/or removal of vascular lesions (tumors).
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | <div> <img alt="Checked" src="checkbox"/> </div> | AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |
|-------------------------------------------------|--------------------------------------------------|--------|------------------------------------------------|--|
|-------------------------------------------------|--------------------------------------------------|--------|------------------------------------------------|--|
Division of General, Restorative,
and Neurological Devices
| 510(k) Number | 2650982 |
|---------------|---------|
|---------------|---------|
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.