← Product Code [MEG](/productcode/MEG) · K080877

# 1ML UNITRACT INSULIN SYRINGE 27G*1/2 , 1ML UNITRACT SYRINGE 29G*1/2 (K080877)

_Unilife Medical Solutions Limited · MEG · Oct 21, 2008 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K080877

## Device Facts

- **Applicant:** Unilife Medical Solutions Limited
- **Product Code:** [MEG](/productcode/MEG.md)
- **Decision Date:** Oct 21, 2008
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The syringe as supplied is a device with a small barrel, plunger and fixed needle, calibrated in units of insulin (U-100), to be used to administer (infuse) insulin subcutaneously. It incorporates features that include reuse prevention and sharps injury prevention.

## Device Story

1mL Unitract® Insulin Syringe is a sterile, single-use, fixed-needle syringe for subcutaneous insulin (U-100) delivery. Device features automatic needle retraction and auto-disable mechanisms to prevent sharps injuries and reuse. Operation involves standard filling, mixing, and injection steps; upon completion of the injection stroke, the plunger engages a spring-based mechanism to retract the needle into the barrel. Once retracted, the plunger locks with the outer collar to prevent further movement. Used in clinical or home settings by patients or healthcare providers. Benefits include reduced risk of accidental needlestick injuries and prevention of syringe reuse.

## Clinical Evidence

Bench testing only. Mechanical performance verified against ISO 8537:1991 and AS 1077:1992. Testing included peak activation forces for retraction, spring reaction forces, and auto-disable override forces. Chemical testing included extractables, residues, and pH. Biocompatibility testing (cytotoxicity, pyrogens) confirmed USP Class VI compliance.

## Technological Characteristics

Materials: USP Class VI polymers and stainless steel. Mechanism: Spring-based needle retraction and integrated plunger-collar auto-disable lock. Dimensions: 1mL capacity, 27G x 1/2" or 29G x 1/2" fixed needles. Sterilization: Ethylene oxide per DIN EN 550:1994. Connectivity: None (mechanical device).

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- 1cc Insulin SafePro® Safety Syringe ([K050134](/device/K050134.md))
- BD Integra 1mL Syringe ([K023752](/device/K023752.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "K0800877" at the top. Below the text is the logo for "UNILIFE medical solutions". The logo is black and white and features a stylized "U" and a curved arrow.

# OCT 21 2008 SECTION 5

## Abbreviated 510(k) Summary

This 510(k) summary has been prepared in compliance with 21 CFR 807.92.

| Applicant:           | Unilife Medical Solutions Limited<br>35 Clarence Street<br>Sydney NSW 2000<br>Australia<br>Tel: +61 2 8346 6500<br>Fax: +61 2 8346 6511                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:             | Michelle Gow<br>Email: michelle.gow@unilife.com.au                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Date Prepared:       | March, 2008                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Trade Name:          | 1mL Unitract® Insulin Syringe                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Common Name:         | Insulin Syringe                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Classification:      | Syringe, Piston; 21 CFR 880.5860; FMF; Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Predicate Devices:   | 1cc Insulin SafePro® Safety Syringe (K050134)<br>BD Integra 1mL Syringe (K023752)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Device Description:  | The 1mL Unitract® Insulin Syringe is a sterile, fixed needle, single use, insulin (U-100) syringe, consisting of a barrel and plunger assembly. The syringe is supplied in two needle sizes, being 27G x ½" and 29G x ½".<br>The devices function in a manner similar to standard syringes for filling and mixing insulin and subcutaneous injection, but also incorporates a sharps injury prevention (automatic needle retraction) and reuse prevention (auto disable) feature.<br>The sharps injury prevention feature allows for automatic and full retraction of the needle into the syringe barrel immediately after the full dose has been expelled; where the rate of retraction may be controlled by the user. The reuse prevention feature also engages immediately after retraction, which prevents further movement of the plunger in either direction. |
| Indications for Use: | The syringe as supplied is a device with a small barrel, plunger and fixed needle, calibrated in units of insulin (U-100), to be used to administer (infuse) insulin subcutaneously. It incorporates features that include reuse prevention and sharps injury prevention.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains the handwritten text "KO8087" at the top. Below that is the logo for "UNILIFE" in a stylized font. Underneath the logo, in smaller letters, is the text "medical solut on:", which appears to be part of the company's tagline or description.

| Substantial<br>Equivalence:       | The 1mL Unitract® Insulin Syringe has identical indications for             use and similar features to the cited predicate # K050134             which is to infuse insulin subcutaneously; whilst incorporating             a sharps injury prevention feature. In addition, the spring             based mechanism of the cited predicate # K023752; and how             this mechanism is integrated within the design of the 1mL             Unitract® Insulin Syringe do not raise new questions of safety             and effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics: | The 1mL Unitract® Insulin Syringe has similar needle             retraction mechanisms to the cited predicate # K050134. The             syringe is filled; and at the end of the injection stroke once the             full dose has been expelled, the plunger engages the needle             to allow retraction into the syringe barrel. The cited predicate             device relies on manual retraction and plunger fracture, where             the 1mL Unitract® Insulin Syringe uses a spring based             mechanism to retract the needle. The auto disable feature of             the 1mL Unitract® Insulin Syringe is integrated within the             plunger design, where after full retraction, the plunger             engages with the collar outer to prevent movement of the             plunger in either direction. This feature requires no secondary             action by the user to disable the syringe.             The characteristics of the 1mL Unitract® Insulin Syringe spring             based mechanism are similar to the technology of the cited             predicate # K023752. The spring mechanism of this predicate             device is located at the distal end of the syringe which             facilitates the retraction of the needle into the plunger. The             1mL Unitract® Insulin Syringe spring mechanism is located on             the proximal end of the syringe to allow retraction into the             barrel.             The design of the 1mL Unitract® Insulin Syringe integrates the             technological design characteristics of the cited predicates,             and having similar operating procedures for filling and mixing             insulin, aspiration and injection, has been demonstrated to             have no significant differences. |
|                                   | The syringe components are manufactured from USP Class VI             materials and stainless steel.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Non Clinical<br>Performance Data: | Mechanical performance tests were conducted to verify the             device meets design specifications based on the requirements             of the FDA Consensus Standard ISO 8537:1991 Sterile             single-use syringes, with or without needle, for insulin and the             voluntary standard AS 1077:1992, Single-use syringes             (sterile) for the injection of 100 units per millilitre insulin (U-             100).             Additional performance testing was conducted on the safety             features relating to peak activation forces of the retraction             mechanism, spring reaction forces and forces to override the             auto disable feature.             Chemical testing included Reducing (Oxidisable) Matter,             Extractable Trace Metals, Non-Volatile Residue, pH and             chemical residues (EO Sterilisation).             Biocompatibility testing of the finished devices included             Cytotoxicity and Pyrogens; with all materials in the fluid path             meeting USP Class VI requirements.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

.

.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Unilife medical solutions. The logo is in black and white and features the word "Unilife" in a stylized font. Below the word "Unilife" is the phrase "medical solutions" in a smaller font. To the right of the word "Unilife" is a graphic element that appears to be a stylized letter "C".

.... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Sterilisation processes are validated to European Standard DIN EN 550:1994, Sterilisation of medical devices - Validation and routine control of ethylene oxide sterilisation to a sterility assurance level (SAL) of 108.

Manufacturing process control requires functional, chemical and biological tests be performed on each production lot number.

==============================================================================================================================================================================

Conclusion: The 1mL Unitract® Insulin Syringe with 27G x ½" and 29G x ½" needles submitted in this 510(k) is determined as substantially equivalent to the cited predicate devices based on performance testing, intended use and design principles of operation and technology, used for the subcutaneous infusion of insulin. Any differences cited do not raise any new questions of safety and effectiveness.

[INTENTIONALLY LEFT BLANK BELOW LINE]

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

#### OCT 2 1 2008

Ms. Michelle Gow Quality and Regulatory Affairs Director Unilife Medical Solutions Limited Level 5, 35 Clarence Street Sydney, NSW AUSTRALIA 2000

Re: K080877

Trade/Device Name: 1mL Unitract® Insulin Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: September 4, 2008 Received: September 8, 2008

Dear Ms. Gow:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Gow

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Unilife medical solutions. The logo is in black and white and features the word "UNILIFE" in a stylized font, with the words "medical solutions" underneath. Above the logo are some handwritten characters.

## SECTION 4

### Indications for Use

510(k) Number (if known):

Unknown

Device Name:

1mL Unitract® Insulin Syringe

Indications for Use:

The syringe as supplied is a device with a small barrel, plunger and fixed needle, calibrated in units of insulin (U-100), intended to be used to administer (infuse) insulin subcutaneously. It incorporates features that include reuse prevention and sharps injury prevention

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

C

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K080877

---

**Source:** [https://fda.innolitics.com/device/K080877](https://fda.innolitics.com/device/K080877)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
