K080845 · Cardinalhealth · JCX · Apr 30, 2008 · General, Plastic Surgery
Device Facts
Record ID
K080845
Device Name
ORWELL FLUID COLLECTION AND DISPOSAL SYSTEM
Applicant
Cardinalhealth
Product Code
JCX · General, Plastic Surgery
Decision Date
Apr 30, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The ORwell™ Fluid Collection and Disposal System is intended to be used in the Operating Rooms, Surgical Centers, and Doctors' Offices to collect and dispose of fluid waste.
Device Story
System collects surgical fluid waste and small debris from surgical sites; comprises mobile unit, disposable collection disc, and stationary disposal station. Mobile unit provides vacuum source; user regulates suction via interface knob. Disposable disc features four ports and flexible liner for fluid containment. Post-procedure, mobile unit docks with disposal station in utility area; station automatically empties disc contents into hospital wastewater system and collapses disc for disposal. Used in ORs, surgical centers, and doctors' offices by clinical staff. System protects personnel from biohazard exposure and cross-contamination by utilizing single-use disposable components. Benefits include reduced risk of splash/biohazard exposure and improved cleanliness compared to reusable canister systems.
Clinical Evidence
Bench testing only. Validated vacuum operating range (0-19" Hg), fluid suction flow rates (~5 LPM), HEPA filter efficiency (99.97% at 0.3 microns), and overflow/shutoff mechanisms. Hazard analysis and software validation performed. No clinical data required.
Technological Characteristics
Powered suction pump (21 CFR 878.4780). Components: Mobile unit (AC powered, UL 60601-1 compliant), disposable collection disc (manifold and liner), and stationary disposal station. Vacuum source: On-board pump. Regulation: Knob-controlled bleeder valve. Filtration: HEPA (99.97% efficiency @ 0.3 microns). Connectivity: Mechanical docking interface between mobile unit and disposal station.
Indications for Use
Indicated for use in operating rooms, surgical centers, and doctors' offices to collect and dispose of surgical fluid waste. No specific patient population or contraindications provided.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the Cardinal Health logo. The logo consists of a stylized graphic element above the company name. The graphic element is composed of several curved lines that intersect, creating a sense of movement and connection.
Medical Products and Services 1500 Waukegan Road McGraw Park, IL 60085 847.473.1500 tel 847.785.2461 fax
www.cardinal.com
## SMDA REQUIREMENTS - K080845
## 510 (K) SUMMARY OF SAFETY AND EFFECTIVENESS ORwell™ Fluid Collection and Disposal System
| Manufacturer: | Cardinal Health Corporation<br>1500 Waukegan Road MP/WM<br>Mc Gaw Park, IL 60085 |
|-----------------------------|-----------------------------------------------------------------------------------------------------|
| Regulatory Affairs Contact: | Lavenia Ford<br>1500 Waukegan Road MP/WM<br>Mc Gaw Park, IL 60085 |
| Telephone Number: | (847) 785-3323 |
| Date summary Prepared: | March 5, 2008 |
| Trade Name: | ORwell™ Fluid Collection and Disposal System |
| Classification: | Class II per 21 CFR § 878.4780 |
| Classification Name: | Powered Suction Pump |
| Predicate Device: | K012991 - Stryker Neptune Waste Management System and<br>K040239 - Neomedix Corp. Supraflow Console |
Description:
The Disposal Station is a stationary unit located in a utility area. The ORwell™ Fluid Collection and Disposal System consists of three (3) components:
- 1. Mobile Unit
- 2. Disposal Station
- Disposable Collection Disc (hereinafter referred to as "Disposable Disc," "Collection ന Disc," or "Disc")
The Mobile Unit is a portable unit that provides its own source of vacuum and houses the Disposable Collection Disc which is used to collect fluid waste and small debris from the surgical site. The amount of suction can be regulated by the end user via the simple user interface on the Mobile Unit.
The Disposable Collection Disc is a single use disposable that collects fluid waste from the surgical procedure. It has four ports to allow for the collection of fluid from various instruments during the procedure. Additionally, the Disc features an attached flexible liner in which collected fluid is contained.
The Disposal Station is a stationary unit located in a utility area plumbed with water inlet and outlet lines. After the surgical procedure, the Mobile Unit is moved to the utility area where it interfaces with and locks into the Disposal Station. After the Mobile Unit is locked into place, the
APR 3 0 2008
{1}------------------------------------------------
Disposal Station automatically empties the Collection Disc, sending the surgical fluid waste into the hospital's waste water system. The evacuation process also collapses the Disc to allow for easy disposal.
Intended Use:
The ORwell™ Fluid Collection and Disposal System is intended to be used in the Operating Rooms, Surgical Centers and Doctors' Offices to collect and dispose of fluid waste.
#### Substantial Equivalence
The ORwell™ Fluid Collection and Disposal System is substantially equivalent to the Neptune Waste Management System in that:
- Intended use is a subset of that of the Neptune Waste Management System. ー Removal of the smoke evacuation feature limits use within the currently cleared indications of the predicate device. This change does not raise new risks to the patient or clinician, nor is there any significant effect on the device's safety or effectiveness with respect to surgical fluid collection.
- Performance attributes and safety components are substantially equivalent to the l predicate
- Vacuum range is the same as predicate 트
- I Both devices regulate vacuum via a knob-controlled bleeder valve
- . Both devices assure HEPA quality exhaust
- 트 Both devices contain an overflow prevention mechanism
- The ORwell™ System differentiates from the predicate in that its collection vessel is | a single use disposable component. This requires the user to dispose of all components that have come in contact with surgical fluid waste between procedures, limiting likelihood of cross-contamination. This key feature change increases the safety of the device and has no adverse effect on substantial equivalence.
{2}------------------------------------------------
.
The following performance attributes were found to be acceptable for the intended use.
Note tables below:
| Design<br>Feature | Proposed Device<br>Cardinal<br>ORwellTM<br>Surgical Fluid<br>Waste Collection<br>and Disposal<br>System | Predicate Device<br>Stryker Neptune<br>Waste<br>Management<br>System (K012991) | Predicate<br>Device<br>Neomedix<br>Corporation,<br>SupraFlow<br>Console<br>(K040239) | Comments |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Cardinal<br>ORwellTM System<br>is intended to be<br>used in the<br>Operating Room,<br>Surgical Centers,<br>and Doctor's<br>Offices to collect<br>and dispose of<br>surgical fluid<br>waste. | The Neptune Waste<br>Management<br>System is intended<br>to be used in the<br>Operating Room,<br>Surgical Centers<br>and Doctor's Offices<br>to collect and<br>dispose of surgical<br>fluid waste as well<br>as collect smoke<br>generated from<br>electrocautery or<br>laser devices. | The NeoMedix<br>SupraFlow<br>Console is an<br>aspiration pump<br>intended for use<br>in general<br>surgery<br>applications<br>where aspiration<br>of irrigation and<br>waste fluid from<br>a surgical site is<br>desired.<br>The device is not<br>indicated for<br>liposuction use. | The new device<br>and the<br>predicate<br>K040239 have<br>the same claim<br>of removing fluid. |
| Closed<br>System | Protects hospital<br>personnel from (1)<br>the splashes and<br>inadvertent<br>biohazard<br>exposure that are<br>possible with<br>suction canister<br>devices and (2)<br>the cross-<br>contamination and<br>cleanliness issues<br>that are possible<br>with fluid<br>collection and<br>disposal systems<br>that utilize a<br>reusable<br>collection<br>chamber | Protects hospital<br>personnel from<br>splashes and<br>inadvertent<br>biohazard exposure<br>that is possible with<br>suction canister<br>devices | | The addition of<br>protecting<br>against cross-<br>contamination<br>and cleanliness<br>issues does not<br>raise new<br>questions<br>regarding safety. |
| Fluid<br>Collection<br>Disposable | Manifold and<br>Collection Liner | Manifold and<br>Canister | | The addition of a<br>collection liner<br>for containment<br>of the waste fluid<br>does not raise<br>new questions<br>regarding safety. |
| Product Code | BTA, Powered<br>Suction Pump | FYD, Apparatus,<br>Exhaust, Surgical | BTA, Powered<br>Suction Pump | SAME |
| Design<br>Feature | Proposed Device<br>Cardinal<br>ORwell™<br>Surgical Fluid<br>Waste Collection<br>and Disposal<br>System | Predicate Device<br>Stryker Neptune<br>Waste<br>Management<br>System (K012991) | Predicate<br>Device<br>Neomedix<br>Corporation,<br>SupraFlow<br>Console<br>(K040239) | Comments |
| FDA<br>Regulation,<br>21 CFR | 878.4780 | 878.5070 | 878.4780 | SAME |
| Dimensions -<br>Mobile Unit | 23" width x 28"<br>depth x 47" height | 18" width x 25"<br>depth x 51" height | | Dimensional<br>differences do<br>not affect safety<br>or effectiveness |
| Dimensions -<br>Disposal<br>Station | 34" width x 15"<br>depth x 50" height | 25" width x 22"<br>depth x 21.5" height<br>(Measured - not<br>specified in 510(k)) | | Dimensional<br>differences do<br>not affect safety<br>or effectiveness |
| Mobile Unit<br>Electrical<br>Requirements | UL 60601-1<br>Compliant<br>AC Powered, 115<br>V, 12 Amps | UL 2601 Compliant<br>AC Powered. 110 V,<br>16 Amps | | Voltage and<br>amperage<br>differences do<br>not raise new<br>questions of<br>safety and<br>effectiveness. |
| Disposal<br>Station<br>Electrical<br>Requirements | 115 V, 4 Amps | 120 V, 3 Amps | | Voltage and<br>amperage<br>differences do<br>not raise new<br>questions of<br>safety and<br>effectiveness. |
| Suction<br>Source | On-Board Pump | On-Board Pump | | SAME |
| Mobile Unit<br>Vacuum<br>Range | 0-19" Hg | 0-19" Hg | | SAME |
| Flow Rate | On 100% vacuum<br>Fluid Flow Rate<br>must be $ ≥ ≥ $ 5.0<br>LPM | Unknown | | Requirement<br>met (using<br>water) for high<br>volume |
| Vacuum<br>Regulator | Knob-Controlled<br>Bleeder Valve | Knob-Controlled<br>Bleeder Valve | | SAME |
| Fluid Suction<br>Filter | HEPA, 99.97%<br>efficiency @ 0.3<br>microns | HEPA, 99.97%<br>efficiency @ 0.3<br>microns | | SAME |
| Emergency<br>Back-Up Port | Yes | Yes | | SAME |
| Design<br>Feature | Proposed Device<br>Cardinal<br>ORwell™<br>Surgical Fluid<br>Waste Collection<br>and Disposal<br>System | Predicate Device<br>Stryker Neptune<br>Waste<br>Management<br>System (K012991) | Predicate<br>Device<br>Neomedix<br>Corporation,<br>SupraFlow<br>Console<br>(K040239) | Comments |
| Single Patient<br>Use | Single | Single or Multiple<br>Patients | - | The indication<br>for single use in<br>the proposed<br>device increases<br>clinician/patient<br>safety by not<br>allowing<br>clinicians to<br>leave a<br>contaminated<br>device in the OR<br>for prolonged<br>periods of time.<br>This change<br>does not raise<br>new questions<br>regarding<br>effectiveness. |
| Canister<br>Capacity | 20 L | 20 L | - | SAME |
| Suction Mode | Continuous | Continuous | - | SAME |
| OTC or<br>Prescription | Prescription (RX<br>only) | OTC | - | Devices are<br>intended to be<br>used in<br>operating rooms,<br>surgical centers,<br>and doctor's<br>offices. |
| Overflow<br>Prevention<br>Mechanism | Yes | Yes | - | SAME |
| Pre-Filter | Single Use<br>Disposable | Single Use<br>Disposable | - | SAME |
| Suction Inlet<br>Port<br>Dimensions | $1/4$ inch inner<br>diameter | 3/8 inch inner<br>diameter | - | Port size and<br>design based off<br>of currently<br>acceptable<br>Cardinal Health<br>suction canister<br>design |
| Design<br>Feature | Proposed Device<br>Cardinal<br>ORwell™<br>Surgical Fluid<br>Waste Collection<br>and Disposal<br>System | Predicate Device<br>Stryker Neptune<br>Waste<br>Management<br>System (K012991) | Predicate<br>Device<br>Neomedix<br>Corporation,<br>SupraFlow<br>Console<br>(K040239) | Comments |
| Principles of<br>Operation | • Fluid suctioned<br>through suction<br>wand into<br>Disposable<br>housed in<br>Mobile Unit<br>• Mobile Unit is<br>moved to<br>Disposal Station<br>and fluid<br>connections<br>engage<br>automatically<br>• Contents are<br>pumped out of<br>Mobile Unit<br>directly into<br>hospital waste<br>line<br>• Disposable is<br>comprised of a<br>suction manifold<br>and attached<br>collection liner<br>• Disposable<br>connects to<br>vacuum line in<br>mobile unit and<br>is housed in<br>cylindrical | • Fluid and smoke<br>are suctioned from<br>a wand into the<br>Rover<br>• Rover is moved to<br>docking station<br>and fluid<br>connections<br>engage<br>automatically<br>• Contents are<br>pumped into the<br>Docking Station<br>tank<br>• Fluids pumped<br>down a drain<br>• Fluid is suctioned<br>through a manifold<br>with an anti-reflux<br>feature<br>• Manifold connects<br>to top of a 20 L<br>canister on the<br>Rover which has a<br>spray head inside | | The Manifold in<br>the predicate<br>has been<br>augmented into<br>a Disposable in<br>the proposed<br>device. This<br>change<br>enhances safety<br>of the clinician<br>and patient since<br>all portions of the<br>device which<br>contact surgical<br>fluid waste will<br>be disposed into<br>red bag waste<br>and not left in<br>the OR. The<br>proposed<br>change does not<br>affect device<br>effectiveness. |
{3}------------------------------------------------
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.
i
.
.
:
.
:
{4}------------------------------------------------
:
:
:
{5}------------------------------------------------
.
:
1999 - 1999
{6}------------------------------------------------
#### Summary of Testing:
The ORwell™ Fluid Collection and Disposal System is substantially equivalent to the predicate device. The testing consisted of all testing identified in the FDA's Guidance Document for Powered Suction Pumps, September 30, 1998 and Premarket Submission for Software Contained in Medical Device. A hazard analysis of the system and its software was performed and testing was conducted to validate the systems overall operations. The ORwell™ Fluid Collection and Disposal System has no direct patient contact therefore, biocompatibility testing is not applicable. A detailed list of testing is provided with test protocols and reports.
The following testing was completed on the ORwell™ System:
- 1. "Almost Full" Sensor – verification that light illuminates and audible alarm sounds at the appropriate pre-determined level
- 2. Vacuum Level Operating Range verification that, with regulator, vacuum operating range is within spec
- 3. Fluid Suction Flow Rates verification that typical worst case and maximum fluid suction flow rates are within spec
- 4. HEPA Filter Efficiency and Life - verification that filter assembly is HEPA efficient and has a life of ≥ 500 hours
- 5. Fluid Trap Shutoff fluid trap assembly tested for fluid ingress effectiveness
- 6. Overflow Valve Fluid Ingress overflow valve tested for fluid ingress effectiveness
- Instructions for Use ORwell™ system tested for safe operation when being used 7 per instructions for use
- 8. Disposal Station Valve Drip - verification that no dripping results after the fluid disposal process
- ல் Disposal Port Valve Opening Pressure – verification that the valve will prevent user from being exposed to biohazards
#### Performance Data
The ORwell™ Fluid Collection and Disposal System was tested against the predicate for vacuum level operating range and fluid suction flow rate. Vacuum level operating range was found to be the same at 0" – 19" Hg and maximum fluid suction flow rate was found to be comparable with ORwell™ performing at approximately 5 liters/minute and the predicate at approximately 4.9 liters/minute.
#### Conclusions Drawn from Testing
Data from testing demonstrates that the performance of the ORwell™ Fluid Collection and Disposal System is similar and substantially equivalent to that of Stryker's currently commercially available Neptune Waste Management System.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The seal is simple and iconic, representing the department's role in promoting health and well-being in the United States.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Cardinal Health % Underwriters Laboratories, Inc. Mr. Ned Devinc 333 Pfingsten Road Northbrook, Illinois 60062-2096
APR 3 0 2008
Re: K080845
Trade/Device Name: Orwell™ Fluid Collection and Disposal System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: II Product Code: JCX Dated: April 16, 2008 Received: April 17, 2008
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{8}------------------------------------------------
### Page 2 - Mr. Ned Devine
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systeros at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark N. Millman
Mark N. Melkerson Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use
510(k) Number (if known): ¥0808 ¥5
Device Name: ORwell™ Fluid Collection and Disposal System Indications for Use:
The ORwell™ Fluid Collection and Disposal System is intended to be used in the Operating Rooms, Surgical Centers, and Doctors' Offices to collect and dispose of fluid waste.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ Prescription Use X (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Net RP Qgl for mxm
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
510(k) Number_ _ 080895