← Product Code [NDY](/productcode/NDY) · K080811

# TINA-QUANT CYSTATIN C, CALIBRATOR F.A.S. CYSTATIN C AND CYSTATIN C ANDCYSTATIN C CONTROL SET (K080811)

_Roche Diagnostics Corp. · NDY · Jun 20, 2008 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K080811

## Device Facts

- **Applicant:** Roche Diagnostics Corp.
- **Product Code:** [NDY](/productcode/NDY.md)
- **Decision Date:** Jun 20, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1225
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K080811 · Jun 20, 2008 | TINA-QUANT CYSTATIN C, CALIBRATOR F.A.S. CYSTATIN C AND CYSTATIN C ANDCYSTATIN C CONTROL SET | Roche Diagnostics Corp. | Retrospective collection of banked frozen human plasma samples | Retrospective banked samples were used to perform a method comparison between the candidate device and the predicate device to demonstrate substantial equivalence. | Method comparison; Retrospective banked samples; Substantial equivalence |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective method comparison | Native human plasma samples from hospitalized and healthy blood donors; Sample Size: 94 | DAKO Cystatin C Immunoparticles (predicate device) | Correlation between candidate and predicate device (Passing/Bablock regression) |

## Indications for Use

Reagent: Immunoturbidimetric assay for the quantitative in vitro determination of cystatin C in human serum and lithium-heparin plasma on Roche automated clinical chemistry analyzers. Cystatin C measurements are used as an aid in the diagnosis and treatment of renal diseases. Calibrator: Cfas (Calibrator for automated systems) Cystatin C is for use in the calibration of quantitative Roche methods on Roche clinical chemistry analyzers as specified in the value sheets. Control: Cystatin C Control Set is for use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the value sheets.

## Device Story

Tina-quant Cystatin C is a ready-to-use dual reagent assay; utilizes particle-enhanced immunoturbidimetric principle. Input: human serum or lithium-heparin plasma samples. Operation: latex particles coated with rabbit anti-cystatin C antibodies agglutinate with cystatin C in sample; precipitate measured turbidimetrically on Roche Hitachi 917 analyzer. Output: quantitative cystatin C concentration calculated via interpolation on standard curve. Used in clinical chemistry laboratories by trained personnel. Results aid clinicians in diagnosing and monitoring renal disease. System includes C.f.a.s. Cystatin C calibrator and bi-level control set for quality monitoring.

## Clinical Evidence

Bench testing only. Precision studies (N=63 per material) showed total CVs 2.36-3.76%. Linearity confirmed 0.4-8.0 mg/L. Method comparison (N=94) against predicate yielded r²=0.9991. Interference testing showed no significant impact from common compounds or high-dose hook effect up to 20 mg/L. Reference range study conducted on 500 healthy donors.

## Technological Characteristics

Particle-enhanced immunoturbidimetric assay. Reagents: latex particles coated with anti-cystatin C antibodies. Calibrator/Control: recombinant human cystatin C produced in E. coli in delipidated human serum matrix. Storage: 2-8°C. Analyzers: Hitachi 917, MODULAR P, cobas c 501. Connectivity: Automated clinical chemistry analyzer integration.

## Regulatory Identification

A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

## Predicate Devices

- DakoCytomation Cystatin C Immunoparticles, Cystatin C Calibrator Kit and Cystatin C Control Set ([K041627](/device/K041627.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k080811

B. Purpose for Submission:
New device

C. Measurand:
Cystatin C

D. Type of Test:
Quantitative immunoturbidimetric assay

E. Applicant:
Roche Diagnostics Corp.

F. Proprietary and Established Names:
Tina-quant Cystatin C
Calibrator f.a.s. Cystatin C
Cystatin C Control Set

G. Regulatory Information:

|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  NDY | Class II | 21 CFR 862.1225 Test, Cystatin C | Clinical Chemistry  |
|  JIT | Class II | 21 CFR 862.1150 Calibrator | Clinical Chemistry  |
|  JJX | Class I | 21 CFR 862. 1660 Quality Control Material | Clinical Chemistry  |

H. Intended Use:

{1}

1. Intended use(s):

See indications for use statement below.

2. Indication(s) for use:

Immunoturbidimetric assay for the quantitative in vitro determination of cystatin C in human serum and lithium-heparin plasma on Roche automated clinical chemistry analyzers. Cystatin C measurements are used as an aid in the diagnosis and treatment of renal diseases.

Calibrator:

Cfas (Calibrator for automated systems) Cystatin C is for use in the calibration of quantitative Roche methods on Roche clinical chemistry analyzers as specified in the value sheets.

Control:

Cystatin C Control Set is for use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the value sheets.

3. Special conditions for use statement(s):

For prescription use only

4. Special instrument requirements:

Roche Hitachi 917

I. Device Description:

The Roche Tina-quant Cystatin C kit is a ready to use dual reagent assay. Reagent 1 is a polymer solution in MOPS- buffered saline, preservatives and stabilizers. Reagent 2 is a glycine buffer containing latex particles coated with anti-cystatin C antibody (rabbit), preservatives and stabilizers. Test principle is a particle enhanced immunoturbidimetric assay.

The C.f.a.s. Cystatin C calibrator is a liquid ready to use calibrator based on pooled delipidated human serum enriched with recombinant human cystatin C produced in E. coli. Each donor was tested and found negative for HIV 1 and 2, HCV and HBsAg by FDA-approved methods or by methods cleared in compliance with the European Directives 98/79/EC Annex II, List A.

The Cystatin C Control Set is a ready to use bi-level control that consists of four 1 mL vials of a low and high control. The controls are liquid pools of delipidated human serum enriched with recombinant human cystatin C produced in E. coli and

2

{2}

preserved with preservatives. Each donor was tested and found negative for HIV 1 and 2, HCV and HBsAg by FDA-approved methods or by methods cleared in compliance with the European Directives 98/79/EC Annex II, List A.

J. Substantial Equivalence Information:

|   | k080811- proposed device
Roche Tina-quant
Cystatin C, C.f.a.s.
Cystatin C
calibrators, Cystatin C Control Set | k041627 – predicate device Dako
Cytomation Cystatin C Immunoparticles, Cystatin C Control Set, Cystatin C Calibrator  |
| --- | --- | --- |
|  Similarities  |   |   |
|  Indications for use | Immunoturbidimetric assay for the quantitative in vitro determination of cystatin C in human serum and plasma on Roche automated clinical chemistry analyzers. Cystatin C measurements are used as an aid in the diagnosis and treatment of renal diseases. | Same  |
|  Analyte | Cystatin C | Same  |
|  Principle | Particle enhanced immunoturbidimetric assay. | Same  |
|  Differences  |   |   |
|  Measuring Range | 0.4 to 8.0 mg/L | 0.4 to 7.5 mg/L  |
|  Analyzers | Hitachi 917 | Hitachi 911, Hitachi 917, Modular P, Cobas Mira Plus and IMMAGE  |
|  Specimen | Serum and Lithium-heparinized plasma | Serum, heparinized plasma, EDTA plasma  |

{3}

4

# K. Standard/Guidance Document Referenced (if applicable):

|  STANDARDS  |   |   |   |
| --- | --- | --- | --- |
|  Title and Reference Number  |   |   |   |
|  None were referenced.  |   |   |   |
|  GUIDANCE  |   |   |   |
|  Document Title | Office | Division | Web Page  |
|  Guidance for Industry and FDA Staff; Replacement Reagent and Instrument Family Policy | OIVD |  | http://www.fda.gov/cdrh/oivd/guidance/950.html  |
|  Guidance for Industry and FDA Staff - Assayed and Unassayed Quality Control Material | OIVD |  | http://www.fda.gov/cdrh/oivd/guidance/2231.html  |
|  Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators; Final | OIVD |  | http://www.fda.gov/cdrh/ode/calibrator.html  |
|  CLSI EP17-A |  |  |   |

# L. Test Principle:

The Roche Tina-quant Cystatin C test principle is a particle enhanced immunoturbidimetric assay. Human cystatin C agglutinates with latex particles coated with anti-cystatin C antibodies. The precipitate is determined turbidimetrically. The signal generated is correlated with the concentration of cystatin C in the sample. By interpolation on a standard curve, the concentration of cystatin C in the sample is calculated.

# M. Performance Characteristics (if/when applicable):

1. Analytical performance:

All performance data was determined with the Hitachi 917 analyzer. Additional information can be obtained from k041627.

a. Precision/Reproducibility:

Precision for the Tina-quant Cystatin C was conducted on the Hitachi 917 analyzer with a single lot and single instrument. Within run imprecision of the Tina-quant Cystatin C test system was determined with two control materials and two serum samples run in duplicates of 21. Between-day imprecision was determined with two controls and two serum samples run in triplicate (median value used). Total imprecision was determined with two controls and two serum samples run once a day in triplicates for 21 days. Results for the studies are shown in the table below.

{4}

|  Within Run Precision  |   |   |   |   |
| --- | --- | --- | --- | --- |
|  Material used | control 1 | control 2 | sample 1 | sample 2  |
|  |   |   |   |   |
|  N | 21 | 21 | 21 | 21  |
|  MEAN (mg/L) | 4.48 | 0.95 | 0.75 | 5.14  |
|  SD (mg/L) | 0.04 | 0.01 | 0.01 | 0.03  |
|  CV (%) | 0.91 | 0.97 | 1.71 | 0.67  |
|  Between-Day Precision  |   |   |   |   |
|  Material used | control 1 | control 2 | sample 1 | sample 2  |
|  |   |   |   |   |
|  N | 21 | 21 | 21 | 21  |
|  MEAN (mg/L) | 4.37 | 0.4 | 0.73 | 4.98  |
|  SD (mg/L) | 0.09 | 0.03 | 0.02 | 0.10  |
|  CV (%) | 2.01 | 2.79 | 2.83 | 2.05  |
|  Total Precision  |   |   |   |   |
|  Material used | control 1 | control 2 | sample 1 | sample 2  |
|  |   |   |   |   |
|  N | 63 | 63 | 63 | 63  |
|  Total mean (mg/L) | 4.36 | 0.94 | 0.73 | 4.98  |
|  W-R SD(mg/L) | 0.1 | 0.0 | 0.0 | 0.0  |
|  W-R CV(%) | 1.4 | 1.9 | 2.0 | 0.9  |
|  Total SD(mgL) | 0.11 | 0.03 | 0.03 | 0.12  |
|  Total CV(%) | 2.5 | 3.13 | 3.76 | 2.36  |

# b. Linearity/assay reportable range:

Linearity was assessed via percent recovery of three serum samples (low, medium and high levels) of different ranges of cystatin C concentrations. Low and mid levels dilution series were prepared with human serum containing low concentrations of cystatin C and saline as a diluent. The high level dilution series were prepared using human serum spiked with Cystatin C and saline as a diluent. Samples were diluted in 9 steps with saline and were analyzed in with the Hitachi 917 analyzer. The sponsors acceptance criteria for samples ranging from 0.40 to  $1.00\mathrm{mg / L}$  is a recovery  $\leq 0.20\mathrm{mg / L}$  and for samples between 1.00 to 8.00 is a recovery  $\leq 10\%$ . The results are shown in the tables below and support the sponsors claimed linear range of 0.40 to 8.0 mg/L.

The sponsor conducted a high dose (hook-effect) study to determine if high does of cystatin C interfered with the assay. Human serum samples were

{5}

spiked to Cystatin C concentrations of 26.5 mg/L and 31.2 mg/L and diluted with saline. There was no interference observed for concentrations up to 20 mg/L for the Hitachi 917.

|  Low-level Dilution  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  No | Dilution low | Dilution high | Measured Value | Theoretical Value | Recovery [%]  |
|  0 | 8.50 | 1.50 | 0.330 | 0.360 | 91.667  |
|  1 | 8.00 | 2.00 | 0.430 | 0.443 | 97.065  |
|  2 | 7.00 | 3.00 | 0.640 | 0.610 | 104.918  |
|  3 | 6.00 | 4.00 | 0.770 | 0.777 | 99.099  |
|  4 | 5.00 | 5.00 | 0.980 | 0.943 | 103.924  |
|  5 | 4.00 | 6.00 | 1.140 | 1.110 | 102.703  |
|  6 | 3.00 | 7.00 | 1.310 | 1.277 | 102.584  |
|  7 | 2.00 | 8.00 | 1.450 | 1.443 | 100.485  |
|  8 | 1.00 | 9.00 | 1.600 | 1.610 | 99.379  |
|  9 | 0.00 | 10.00 | 1.730 | 1.777 | 97.355  |
|  Mid-level Dilution  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  No | Dilution low | Dilution high | Measured Value | Theoretical Value | Recovery [%]  |
|  0 | 10.00 | 0.00 | 0.760 | 0.935 | 81.283  |
|  1 | 9.00 | 1.00 | 1.560 | 1.670 | 93.413  |
|  2 | 8.00 | 2.00 | 2.350 | 2.405 | 97.713  |
|  3 | 7.00 | 3.00 | 3.250 | 3.140 | 103.503  |
|  4 | 6.00 | 4.00 | 4.090 | 3.875 | 105.548  |
|  5 | 5.00 | 5.00 | 4.780 | 4.610 | 103.688  |
|  6 | 4.00 | 6.00 | 5.400 | 5.345 | 101.029  |
|  7 | 3.00 | 7.00 | 6.190 | 6.080 | 101.809  |
|  8 | 2.00 | 8.00 | 6.690 | 6.815 | 98.166  |
|  9 | 1.00 | 9.00 | 7.210 | 7.550 | 95.497  |
|  High-level Dilution  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  No | Dilution low | Dilution high | Measured Value | Theoretical Value | Recovery [%]  |
|  0 | 10.00 | 0.00 | 1.300 | 1.315 | 98.859  |
|  1 | 9.00 | 1.00 | 2.450 | 2.450 | 100.000  |
|  2 | 8.00 | 2.00 | 3.690 | 3.585 | 102.929  |
|  3 | 7.00 | 3.00 | 4.940 | 4.720 | 104.661  |
|  4 | 6.00 | 4.00 | 6.000 | 5.855 | 102.477  |
|  5 | 5.00 | 5.00 | 6.990 | 6.990 | 100.000  |
|  6 | 4.00 | 6.00 | 7.880 | 8.125 | 96.985  |
|  7 | 3.00 | 7.00 | 8.920 | 9.260 | 96.328  |
|  8 | 2.00 | 8.00 | 9.990 | 10.395 | 96.104  |
|  9 | 1.00 | 9.00 | 10.810 | 11.530 | 93.755  |

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

{6}

Traceability

There is no internationally recognized reference standard for Cystatin C. The ready to use secondary calibrator materials are traceable to an in-house gravimetric (dry-mass) reference preparation of delipidated human serum enriched with recombinant human Cystatin C. The Cystatin C Control Set is traceable to the Cfas Cystatin C Calibrator (secondary calibrator materials).

Value assignment

The Cystatin C Calibrator and Control Set value assignments were determined by turbidimetry on the Hitachi 917. The reference preparation of pure recombinant human Cystatin C was prepared and is used to value assign a master calibrator or primary calibrators. The Primary calibrator is used to assign the value of the secondary calibrator (Cfas Cystatin C). A daily calibration curve is prepared from 6 dilutions of the master calibrator and of the secondary calibrator. All dilutions are measure in duplicate. This is repeated over 3 days and the final target value of the secondary calibrator is the grand mean of all determinations for the value with the highest concentration.

The Cystatin C Control Set is value assigned from the Cystatin C Calibrator based on six calibrations. The grand mean value is used to value assign the Cystatin C Control tested and the uncertainty for the grand mean is noted. The control from another lot is measured in duplicates. The recommended confidence interval for control accuracy is the assigned value +/- 15%.

Stability

Real-time stability for the C.f.a.s. Cystatin C and the Cystatin C Control set were conducted for 25 months and the results support the sponsor's shelf-life claims of 24 months.

Closed reagent stability was assessed with three lots of the Cystatin C latex reagent at day 0, month 2, 6,12,13,18, 24 and 25. The real-time stability testing is still on-going and the results support the sponsor's shelf-life claims of 24 months.

On-board reagent stability was assessed and the results support the sponsor's open vial claims of 8 weeks when stored on-board or at 2-8°C.

d. Detection limit:

The Limit of Blank (LoB) and Limit of Detection (LoD) were determined in accordance with the CLSI EP17-A requirements.

The LoB study was conducted to determine the highest observed measurement values for samples free of analyte. The LOB was determined as the 95th percentile of measurements of bank samples. A blank sample was assayed on two Hitachi 917 analyzers in six runs over 3 days per instrument for a total of 30 replicates. The LoB was determined to be ≤0.3 mg/L.

7

{7}

The LoD was conducted to determine the lower limit for samples with analyte. The LoD was determined as the lowest amount of analyte in a sample that can be detected with 95% probability. Five human serum samples with low analyte were run on two Hitachi analyzers in six runs over 3 days per instrument for a total of 25 replicates. The LoD was determined to be ≤0.4 mg/L.

e. Analytical specificity:

Pooled human serum samples spiked with varying levels of interferent were tested on the Hitachi 917 analyzer. The effects of interference by hemoglobin, conjugated and unconjugated bilirubin, lipemia (intralipid) and rheumatoid factor were evaluated. The samples were tested in triplicate and the median values were used to calculate recovery. The sponsor states that no significant interference criterion is a recovery of +/-10% of the initial value. No significant interference was found on the Hitachi 917 for hemoglobin up to 700 mg/dL, conjugated and unconjugated bilirubin up to 60 mg/dL, lipemia up to an L index of 700 (corresponds to turbidity, however the sponsor states in the labeling there is a poor correlation between L index and triglycerides) and rheumatoid factor up to 1200 IU/mL.

Eighteen other compounds tested: Acetylsteine, ampicillin Na, ascorbic acid, Ca-Dobesilate, Cefoxitin-Na, Liquemin Na-Heparin, Levodopa, L-a-Methyldopa, Metronidazole, Phenylbutazone, Doxycycline HCl, Acetylsalicylic acid, Rifampicine, Acetaminophen, Ibuprofen, theophylline and cyclosporine were reported to have no significant interference or cross-reactivity at low levels of cystatin C.

Hook effect was evaluated by testing high dose analyte concentrations up to 31 mg/L on the Hitachi 917 analyzer. There was no high dose hook effect up to 20 mg/L. The sponsor states in the labeling that a high dose hook effect may occur at cystatin C concentrations &gt;20 mg/L.

f. Assay cut-off:

Not applicable.

2. Comparison studies:

a. Method comparison with predicate device:

Ninety-four native human plasma (Li heparin) samples from a retrospective collection of banked frozen samples, including multiple donors from both hospitalized and healthy blood donors were measured with the DAKO application (predicate device) on the Hitachi 917 (X) and with the Roche Tina-quant Cystatin C (candidate device) on the Hitachi 917 (Y). The Roche

8

{8}

Tina-quant Cystatin C reagent is the same reagent as the DAKO Cystatin C reagent. The samples ranged from 0.61 to 6.05 mg/L and the results were calculated using the Passing/Bablock regression analysis. The correlation was Y=1.009x+0.019 with a r²= 0.9991.

b. Matrix comparison:

An anticoagulation study was conducted to validate the use of Li-heparin plasma with the Tina-quant Cystatin C assay. Forty six paired samples were analyzed in singlicate on the Hitachi 917 and the concentrations ranged from 0.53 to 6.66 mg/L. Each plasma sample was compared to the respective serum sample and the percent recovery was calculated. The sponsor’s acceptance criterion was a deviation of less than 15%. The recoveries ranged from 93.09 to 113.89% and support the use of Li-heparin samples for the Roche Tina-quant Cystatin C assay.

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable.

b. Clinical specificity:

Not applicable.

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable.

4. Clinical cut-off:

Not applicable.

5. Expected values/Reference range:

The sponsor conducted a reference range study to determine the reference range for the Tina-quant Cystatin C assay. Serum samples were measured in singlicate on the Hitachi 917 analyzer. Samples were drawn from healthy, non-hospitalized donors and the cystatin C results are shown in the table below.

9

{9}

|  Ref. Panel Cystatin C |  | sorted by age |   | sorted by gender  |   |
| --- | --- | --- | --- | --- | --- |
|   |  All | 20-50 years | >50 years | Male | Female  |
|  N | 500 | 310 | 190 | 249 | 251  |
|  Min mg/L | 0.408 | 0.466 | 0.408 | 0.472 | 0.408  |
|  Perc (2.5) mg/L | 0.562 | 0.557 | 0.577 | 0.581 | 0.537  |
|  Median mg/L | 0.729 | 0.703 | 0.777 | 0.746 | 0.716  |
|  Perc (97.5) mg/L | 0.987 | 0.895 | 1.087 | 1.058 | 0.953  |
|  max mg/L | 1.445 | 1.098 | 1.445 | 1.445 | 1.091  |

# N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

# O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K080811](https://fda.innolitics.com/device/K080811)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
