← Product Code [IYE](/productcode/IYE) · K080726

# CALYPSO 4D LOCALIZATION SYSTEM (K080726)

_Calypso Medical Technologies, Inc. · IYE · May 14, 2008 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K080726

## Device Facts

- **Applicant:** Calypso Medical Technologies, Inc.
- **Product Code:** [IYE](/productcode/IYE.md)
- **Decision Date:** May 14, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Calypso® 4D Localization System is intended for use as an adjunct in treatment planning and radiation therapy, to align and monitor the patient's position relative to the isocenter of a linear accelerator. The Calypso® System provides accurate, precise and continuous localization of a treatment isocenter by using two or more Beacon transponders. Beacon® transponders are indicated for use to radiographically and electromagnetically mark soft tissue for future therapeutic procedures. Permanent Beacon® transponders are indicated for permanent implantation in the prostate and the peri-prostatic tissue (i.e., prostatic bed).

## Device Story

System utilizes non-ionizing electromagnetic and optical technology for patient alignment and target position monitoring during radiation therapy; inputs consist of electromagnetic signals from passive implanted Beacon transponders; system transforms signals into 3D spatial coordinates of treatment target; output used by operator to align patient to linear accelerator isocenter and track target position during treatment; used in clinical radiation therapy settings; provides objective, continuous localization; benefits include improved accuracy and precision in target positioning, potentially enhancing radiation delivery to target while sparing surrounding tissue.

## Clinical Evidence

Bench testing only; no clinical data provided.

## Technological Characteristics

Utilizes non-ionizing electromagnetic and optical technology; passive Beacon transponders implanted in soft tissue; 3D localization capability; system integrates with linear accelerator workflows; Class II device.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- Calypso 4D Localization System ([K060906](/device/K060906.md))
- RadioMed Soft Tissue Marker ([K031206](/device/K031206.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Ko 80726

#### 510(K) SUMMARY OF SAFETY & EFFECTIVENESS 5.0

MAY 1 4 2008

This 510(k) Summary of Safety and Effectiveness is provided in accordance with 21 CFR 807.92.

March 13, 2008 Date of preparation:

Submitter information:

Calypso® Medical Technologies, Inc. 2101 Fourth Avenue, Suite 500 Seattle, WA 98121

Phone: 206-254-0600 Fax: 206-254-0606

- Sue Ridge Contact: Regulatory Affairs/Quality Systems Manager
- Calypso® 4D Localization System Device trade name:
- Patient localization system Common name:

Accelerator, Linear, Medical Classification name:

- CFR 892.5050 Classification: Class II Product code - IYE
- Calvoso® 4D Localization System Predicate devices: (K060906)

RadioMed™ Soft Tissue Marker (K031206)

### Indications for Use:

The Calypso® 4D Localization System is intended for use as an adjunct in treatment planning and radiation therapy, to align and monitor the patient's position relative to the isocenter of a linear accelerator. The Calypso® System provides accurate, precise and continuous localization of a treatment isocenter by using two or more Beacon transponders.

Beacon® transponders are indicated for use to radiographically and electromagnetically mark soft tissue for future therapeutic procedures.

Permanent Beacon® transponders are indicated for permanent implantation in the prostate and the peri-prostatic tissue.

{1}------------------------------------------------

#### Device Description:

The Calypso 4D Localization System (Calypso System) utilizes non-ionizing electromagnetic and optical technology to provide accurate, objective, and continuous localization of a treatment target for patient alignment and target position monitoring during radiation therapy. Use of the Calypso System for target localization is based on the system's detection of non-ionizing electromagnetic signals from passive implanted markers, called Beacon transponders. The Beacon transponders are implanted in or near the treatment target. When used with the Calypso System, the Beacon transponder signals enable objective measurement of the location of the treatment target in 3 dimensions. The system operator uses this information to align the patient's treatment target to the isocenter of the linear accelerator prior to treatment. This information can also be used to monitor (track) the position of the treatment target during radiation therapy treatment.

#### Summary of Performance Testing:

Performance testing demonstrated substantial equivalence.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## MAY 1 4 2008

Ms. Sue Ridge Regulatory Affairs/Quality Systems Manager Calypso Medical Technologies, Inc. 2101 Fourth Avenue, Suite 500 SEATTLE WA 98121

Re: K080726

Trade/Device Name: Calypso 4D Localization System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: March 13, 2008 Received: March 19, 2008

#### Dear Ms. Ridge:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Ant or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and liusting (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1050,

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to legally marketed predicate device results in a classification for your device of your device to results your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance a one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, release contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Shall Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html U

Sincerely vours.

Nancy C. Brigdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known):

K080726

Device Name:

Calypso® 4D Localization System

Indications For Use:

The Calypso® 4D Localization System is intended for use as an adjunct in treatment planning and radiation therapy, to align and monitor the patient's position relative to the isocenter of a linear accelerator. The Calypso® System provides accurate, precise and continuous localization of a treatment isocenter by using two or more Beacon transponders.

Beacon® transponders are indicated for use to radiographically and electromagnetically marks soft tissue for future therapeutic procedures.

Permanent Beacon® transponders are indicated for permanent implantation in the prostate and the peri-prostatic tissue (i.e., prostatic bed).

Prescription Use

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Armi N. Whang

Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

Page 1 of *_*_

---

**Source:** [https://fda.innolitics.com/device/K080726](https://fda.innolitics.com/device/K080726)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
