G2 EXPRESS FILTER SYSTEM-FEMORAL (RF400F) AND JUGULAR/SUBCLAVIAN (RF400J) DELIVERY KITS
Applicant
C.R. Bard, Inc.
Product Code
DTK · Cardiovascular
Decision Date
Jul 30, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.3375
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The G2 EXPRESS™ Filter System – Femoral and Jugular/Subclavian Delivery Kits are indicated for use in the prevention of recurrent pulmonary embolism via permanent placement in the vena cava in the following situations: . Pulmonary thromboembolism when anticoagulants are contraindicated. . Failure of anticoagulant therapy for thromboembolic disease. Emergency treatment following massive pulmonary embolism where anticipated . benefits of conventional therapy are reduced. . Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated. . G2 EXPRESS™ Filter may be removed according to the instructions supplied under the section labeled: Optional Procedure for Filter Removal.
Device Story
Intravascular filter system for inferior vena cava (IVC) placement; prevents recurrent pulmonary embolism. Device consists of 12 shape-memory nitinol wires with snare tip; provides two-level emboli filtration (legs/arms). Delivered via femoral (7F catheter) or jugular/subclavian (10F sheath) approach using Seldinger technique. Physician-operated in clinical setting. Delivery kits include introducer, dilator, and preloaded filter. Snare tip added to facilitate retrieval. Radiopaque markers/tips assist placement. Output is physical filtration of emboli within IVC. Benefits patient by reducing risk of pulmonary embolism when anticoagulation is contraindicated or ineffective.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence supported by design verification and validation testing against predetermined acceptance criteria.
Technological Characteristics
Shape-memory nitinol wire filter; 12-wire configuration. Delivery systems: 7F (femoral) or 10F (jugular/subclavian) introducer/dilator sets. Radiopaque markers/tips. Hemostasis valve (jugular kit). Compatible with 0.038" guidewire. Supports contrast power injection (jugular kit, up to 800 psi). Sterilized device.
Indications for Use
Indicated for patients requiring prevention of recurrent pulmonary embolism via permanent vena cava placement. Applicable for patients with contraindications to anticoagulants, failure of anticoagulant therapy, or emergency treatment for massive pulmonary embolism. Filter may be removed per optional procedure instructions.
Regulatory Classification
Identification
A cardiovascular intravascular filter is an implant that is placed in the inferior vena cava for the purpose of preventing pulmonary thromboemboli (blood clots generated in the lower limbs and broken loose into the blood stream) from flowing into the right side of the heart and the pulmonary circulation.
Special Controls
*Classification.* Class II. The special controls for this device are:(1) “Use of International Standards Organization's ISO 10993 ‘Biological Evaluation of Medical Devices Part I: Evaluation and Testing,’ ” and
(2) FDA's:
(i) “510(k) Sterility Review Guidance and Revision of 2/12/90 (K90-1)” and
(ii) “Guidance for Cardiovascular Intravascular Filter 510(k) Submissions.”
Predicate Devices
Recovery® G2® Filter System - Femoral Delivery Kit (K073090)
Recovery® G2® Filter System - Jugular/Subclavian Delivery Kit (K073090)
Submission Summary (Full Text)
{1}------------------------------------------------
# G2 EXPRESS™ Filter System – Femoral and Jugular/Subclavian Delivery Kits 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a summary of the information upon which substantial equivalence determination is based as follows:
- 1. Submitter Information:
| Applicant: | Bard Peripheral Vascular, Inc<br>1625 West 3rd Street<br>PO Box 1740<br>Tempe, Arizona 85280 | JUL 30 2008 |
|------------|----------------------------------------------------------------------------------------------|-------------|
| Phone: | 480-303-2524 | |
| Fax: | 480-449-2546 | |
| Contact: | Genevieve Balutowski | |
Associate Project Manager, Regulatory Affairs
# 2. Subject Device:
| Device Trade Name: | G2 EXPRESS™ Filter System - Femoral Delivery Kit<br>and G2 EXPRESS™ Filter System -<br>Jugular/Subclavian Delivery Kit |
|-----------------------|------------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Filter, Intravascular, Cardiovascular |
| Classification: | Class II |
| Classification Panel: | Cardiovascular |
# 3. Predicate Devices:
Recovery® G2® Filter System - Femoral Delivery Kit (K073090)
Recovery® G2® Filter System - Jugular/Subclavian Delivery Kit (K073090)
Image /page/0/Picture/11 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, with a white fill. The word is horizontally oriented and appears to be a logo or brand name. The font style gives the word a strong and modern appearance.
{2}------------------------------------------------
#### 4. Summary of Change:
The modification from the predicate device, the Recovery® G2® Filter System - Femoral and Jugular/Subclavian Delivery System, to the subject device, the G2 EXPRESS™ Filter System Femoral and Jugular/Subclavian Delivery System, was adding a snare to the tip of the filter. Additionally, minor dimensional modifications were made to the delivery systems to accommodate the snare tip.
## 5. Device Description:
The G2 EXPRESS™ Filter consists of 12 shape-memory nitinol wires emanating from a central nitinol sleeve with a snare tip. These 12 wires form two levels of emboli filtration: the legs provide the lower level of filtration and the arms provide the upper level of filtration. The G2 EXPRESS™ Filter is intended to be used in the inferior vena cava with diameters less than or equal to 28 mm.
The G2 EXPRESS™ Filter System - Femoral Delivery Kit consists of a 7 French inner diameter (I.D.) introducer catheter and dilator set and a storage tube preloaded with the G2 EXPRESS™ Filter and pusher system. The introducer catheter and dilator are used to gain access to the inferior vena cava via a femoral approach using the Seldinger technique. The dilator accepts a 0.038" guidewire and the introducer catheter has two radiopaque marker bands on the distal end to assist in filter delivery. The pusher system is designed to pass through the introducer catheter.
The G2 EXPRESS™ Filter System – Jugular/Subclavian Delivery Kit consists of a 10 French I.D. introducer sheath and dilator set and a delivery device preloaded with the G2 EXPRESS™ Filter. The introducer sheath and dilator are used to gain access to the inferior vena cava via a jugular approach using the Seldinger technique. The dilator accepts a 0.038" guidewire and enables a contrast medium power injection up to 800 psi maximum pressure. The introducer sheath contains a radiopaque tip and hemostasis valve with a side port for injecting contrast medium via a syringe. The delivery device fits within the introducer sheath and consists of a side port for saline infusion and a delivery mechanism to deploy the G2 EXPRESS™ Filter.
Image /page/1/Picture/8 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, with a white fill. The word is slightly distorted, with the "A" appearing to be tilted to the left.
{3}------------------------------------------------
#### 6. Indication for Use of Device:
The G2 EXPRESS™ Filter System – Femoral and Jugular/Subclavian Delivery Kits are indicated for use in the prevention of recurrent pulmonary embolism via permanent placement in the vena cava in the following situations:
- . Pulmonary thromboembolism when anticoagulants are contraindicated.
- . Failure of anticoagulant therapy for thromboembolic disease.
- Emergency treatment following massive pulmonary embolism where anticipated . benefits of conventional therapy are reduced.
- . Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated.
- . G2 EXPRESS™ Filter may be removed according to the instructions supplied under the section labeled: Optional Procedure for Filter Removal.
## 7. Technological Comparison to Predicate Device:
The technological characteristics of the subject device, the G2 EXPRESS™ Filter System - Femoral and Jugular/Subclavian Delivery Kits, are substantially equivalent to those of the predicate devices, the Recovery® G2® Filter System - Femoral and Jugular/Subclavian Delivery Kits (K073090), in terms of intended use, indications for use, application, user population, basic design, fundamental scientific technology, performance, and sterilization method.
## 8. Conclusions:
The G2 EXPRESS™ Filter System – Femoral and Jugular/Subclavian Delivery Kits met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The G2 EXPRESS™ Filter System - Femoral and Jugular/Subclavian Delivery Kits is substantially equivalent to the legally marketed predicate device, the Recovery® G2® Filter System - Femoral and Jugular/Subclavian Delivery Kits.
TRADE SECRET/CONFIDENTIAL INFORMATION Notify CR Bard Before Releasing this Document.
BARD
{4}------------------------------------------------
Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 30 2008
Bard Peripheral Vascular, Inc. c/o Genevieve Balutowski, RAC Associate Project Manager, Regulatory Affairs 1625 West 3rd Street P.O. Box 1740 Tempe, AZ 85280
Re: K080668
Trade/Device Name: G2 Express Filter System - Femoral Delivery Kit and G2 Express Filter System -Jugular/Subclavian Delivery Kit Regulation Number: 21 CFR 870.3375 Regulation Name: Cardiovascular intravascular filter Regulatory Class: Class II Product Code: DTK Dated: July 2, 2008 Received: July 3, 2008
Dear Ms. Balutowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or
{5}------------------------------------------------
Page 2 - Ms. Genevieve Balutowski
any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: G2 EXPRESS™ Filter System – Femoral and Jugular/Subclavian Delivery Kits
## Indications for Use:
The G2 EXPRESS™ Filter System – Femoral and Jugular/Subclavian Delivery Kits are indicated for use in the prevention of recurrent pulmonary embolism via permanent placement in the vena cava in the following situations:
- Pulmonary thromboembolism when anticoagulants are contraindicated. .
- Failure of anticoagulant therapy for thromboembolic disease. .
- Emergency treatment following massive pulmonary embolism where anticipated . benefits of conventional therapy are reduced.
- Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed . or is contraindicated.
- G2 EXPRESS™ Filter may be removed according to the instructions supplied . under the section labeled: Optional Procedure for Filter Removal.
| Prescription Use | X (Part21 CFR 801 Subpart D) |
|------------------|--------------------------------------------|
| | AND/OR |
| | Over-The-Counter Use (21CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
| 510(k) Number | K090668 |
|---------------|---------|
|---------------|---------|
Page 1 of 1
TRADE SECRET/CONFIDENTIAL INFORMATION Notify CR Bard Before Releasing this Document.
Image /page/5/Picture/14 description: The image shows the word "BARD" in a bold, outlined font. The letters are large and blocky, with a thick black outline. The word is presented in a straightforward, sans-serif typeface, making it easily readable.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.