M5 DIAGNOSTIC ULTRASOUND SYSTEM

K080640 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · IYN · Mar 21, 2008 · Radiology

Device Facts

Record IDK080640
Device NameM5 DIAGNOSTIC ULTRASOUND SYSTEM
ApplicantShenzhen Mindray Bio-Medical Electronics Co., Ltd.
Product CodeIYN · Radiology
Decision DateMar 21, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Intended Use

The device is intended for use by a qualified physician for ultrasound evaluation of fetal, abdominal, pediatric, neonatal cephalic, cardiac, small parts, transvaginal, transrectal, peripheral-vascular, muscular-skeletal(conventional and superficial).

Device Story

M5 is a portable, software-controlled diagnostic ultrasound system. It acquires and displays ultrasound data using linear and convex array probes (2.5–14 MHz). Modes include B-Mode, M-Mode, PW-Mode, Color-Mode, Power-Mode, Dirpower-Mode, and combined modes (e.g., B/M). Operated by qualified physicians in clinical settings. The system processes acoustic signals to generate real-time images for anatomical evaluation. Features include Tissue Harmonic Imaging, Smart3D, and iScape. Output is displayed on the system monitor for clinical assessment, aiding in diagnosis and patient management. The device is mobile, facilitating point-of-care use.

Clinical Evidence

Bench testing only. Acoustic output measured and calculated per NEMA UD 2 and NEMA UD 3 standards. Device conforms to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, and ISO 10993-1 safety standards.

Technological Characteristics

Portable ultrasound system; linear and convex array transducers (2.5–14 MHz). Imaging modes: B, M, PW, Color, Power, Dirpower. Features: Tissue Harmonic Imaging, Smart3D, iScape. Safety standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1. Acoustic output per NEMA UD 2/UD 3.

Indications for Use

Indicated for ultrasound evaluation of fetal, abdominal, pediatric, neonatal cephalic, cardiac, small parts, transvaginal, transrectal, peripheral-vascular, and musculoskeletal (conventional/superficial) anatomy in patients requiring diagnostic ultrasound imaging. Prescription use only.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## MAR 2 1 2008 ... ... ... ... ... ... ... ... .............................................................................................................................................. #### 510(K) SUMMARY Exhibit B This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92(c). The assigned 510(k) number is: ## 1. Submitter: Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China Tel: +86 755 2658 2888 Fax: +86 755 2658 2680 #### Contact Person: Li Dongling Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building. Keii 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China Date Prepared: January 31, 2008 ### 2. Device Name: M5 Diagnostic Ultrasound System ### Classification Regulatory Class: II Review Category: Tier II 21 CFR 892.1550 Ultrasonic Pulsed Doppler Imaging System' (90-IYN) 21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (90-IYO) 21 CFR 892.1570 Diagnostic Ultrasound Transducer (90-ITX) ### 3. Marketed Device: M5 Diagnostic Ultrasound System is substantially equivalent to the following devices: Mindray DC-6 (K#072164), GE Logiq 9 (K#061129) and GE Logiq E diagnostic ultrasound system (K#072797), Mindray DP-6600 (K#060949). ### 4. Device Description: {1}------------------------------------------------ The M5 is a portable general purpose, mobile, software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound data in B-Mode, M-Mode, PW-Mode, Color-Mode, Power-Mode, Dirpower-Mode or the combined mode (i.e. B/M-Mode).This system is a Track 3 device that employs an array of probes that include linear array and convex linear array with a frequency range of approximately 2.5 MHz to 14 MHz. ### 5. Intended Use: The device is intended for use by a qualified physician for ultrasound evaluation of fetal, abdominal, pediatric, neonatal cephalic, cardiac, small parts, transvaginal, transrectal, peripheral-vascular, muscular-skeletal(conventional and superficial). #### 6. Safety Considerations: The M5 Diagnostic Ultrasound System has been tested as Track 3 Device per the FDA Guidance document "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers" issued September 1997. The acoustic output is measured and calculated per NEMA UD 2 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment: 2004 and NEMA UD 3 Output Display Standard. The device conforms to applicable medical device safety standards, such as IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37 and ISO 10993-1. #### Conclusion: The conclusions drawn from testing of the M5 Diagnostic Ultrasound System demonstrate that the device is as safe and effective as the legally marketed predicate devices. > 0044 ﺴﺘﻌﻤﺎﺭ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤ {2}------------------------------------------------ Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the eagle in a circular fashion. **APR - 8 2008** . . Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. c/o Robert Mosenkis Citech 5200 Butler Pike Plymouth Meeting, PA 19462-1298 Re: k080640 Trade/Device Name: M5 Diagnostic Ultrasound System Regulation Number: 21 CFR 1550 Regulation Name: Ultrasonic pulsed Doppler imaging system Regulatory Class: Class II Product Code: IYN IYO ITX Dated: March 4, 2008 Received: March 6, 2008 Dear Mr. Mosenkis: This letter corrects our substantially equivalent letter of March 21, 2008. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. This determination of substantial equivalence applies to the following transducers intended for use with the M5 Diagnostic Ultrasound System, as described in your premarket notification: | Transducer Model Number | |-------------------------| | 3C5s | | 6C2s | | 6CV1s | | 7L4s | | 7L6s | {3}------------------------------------------------ | 10L4s | |-------| | 6LE7s | | 6LB7s | If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html If you have any questions regarding the content of this letter, please contact Robert Phillips, Ph.D., at (240) 276-3666. Sincerely yours, hryu Thhly Ancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosures {4}------------------------------------------------ ### Diagnostic Ultrasound Indications for Use Form System Model: Transducer × MS 510(k) Number(s) | | Mode of Operation | | | | | | | | | | |----------------------------------------------------------------------------------------------------|-------------------|---|---|---------|-----|------------------|----------------------|------------------------------|-----------------------|-----------------| | Clinical Application | A | B | M | PW<br>D | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other (specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | N | N | N | | N | N | | N | Note 1, 2, 3 | | Abdominal | | N | N | N | | N | N | | N | Note 1, 2, 3 | | Intraoperative (specify)* | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | N | N | N | | N | N | | N | Note 1, 2, 3 | | Small organ(specify)** | | N | N | N | | N | N | | N | Note 2, 3 | | Neonatal Cephalic | | N | N | N | | N | N | | N | Note 2, 3 | | Adult Cephalic | | N | N | N | | N | N | | N | Note 2, 3 | | Cardiac | | N | N | N | | N | N | | N | Note 2, 3 | | Transesophageal | | | | | | | | | | | | Transrectal | | N | N | N | | N | N | | N | Note 2 | | Transvaginal | | N | N | N | | N | N | | N | Note 2, 3 | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | N | N | N | | N | N | | N | Note 2, 3 | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal Conventional | | N | N | N | | N | N | | N | Note 2, 3 | | Musculo-skeletal Superficial | | N | N | N | | N | N | | N | Note 2, 3 | | Other (specify)*** | | N | N | N | | N | N | | N | Note 2, 3 | | N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | | | | | Additional comments: Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+ B, Power + PW +B. | | | | | | | | | | | | * Intraoperative includes abdominal, thoracic, and vascular etc. | | | | | | | | | | | ** Small organ-breast, thyroid, testes, etc. *** Other use includes Urology/Prostate. Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. Note 2: Smart3D Note 3: iScape (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation(ODE) Prescription USE (Per 21 CFR 801.109) Aoyez 722 (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number {5}------------------------------------------------ ## Diagnostic Ultrasound Indications for Use Form System Transducer × Model: 3C5s 510(k) Number(s) | | Mode of Operation | | | | | | | | | | |---------------------------------------------------------------------------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|-----------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other (specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | N | N | N | | N | N | | N | Note 1, 2, 3 | | Abdominal | | N | N | N | | N | N | | N | Note 1, 2, 3 | | Intraoperative (specify)* | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | N | N | N | | N | N | | N | Note 1, 2, 3 | | Small organ(specify)** | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | Other (Urology) | | | | | | | | | | | | N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | | | | | Additional comments: Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+B, Power + PW +B. | | | | | | | | | | | | * Intraoperative includes abdominal, thoracic, and vascular etc. | | | | | | | | | | | | ** Small organ-breast, thyroid, testes, etc. | | | | | | | | | | | Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. Note 2: Smart3D Note 3: iScape (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation(ODE) **Prescription USE (Per 21 CFR 801.109)** (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number K080640 0049 {6}------------------------------------------------ # Diagnostic Ultrasound Indications for Use Form System Transducer Model: 6C2s 510(k) Number(s) | | Mode of Operation | | | | | | | | | | |-------------------------------|-------------------|---|---|---------|-----|------------------|----------------------|------------------------------|-----------------------|-----------------| | Clinical Application | A | B | M | PW<br>D | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other (specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | N | N | N | | N | N | | N | Note 2, 3 | | Intraoperative (specify)* | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | N | N | N | | N | N | | N | Note 2, 3 | | Small organ(specify)** | | | | | | | | | | | | Neonatal Cephalic | | N | N | N | | N | N | | N | Note 2, 3 | | Adult Cephalic | | N | N | N | | N | N | | N | Note 2, 3 | | Cardiac | | N | N | N | | N | N | | N | Note 2, 3 | | Transesophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | Other (specify) | | | | | | | | | | | Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. . Note 2: Smart3D .. Note 3: iScape (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation(ODE) Prescription USE (Per 21 CFR 801.109) Arzola Mr. Why જ (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number 0050 {7}------------------------------------------------ ## Diagnostic Ultrasound Indications for Use Form System Transducer × Model: 6CV Is 510(k) Number(s) | Clinical Application | Mode of Operation | | | | | | | | | | |-------------------------------------------------------------------------------------------------------|-------------------|---|---|---------|-----|------------------|----------------------|------------------------------|-----------------------|-----------------| | | A | B | M | PW<br>D | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other (specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | N | N | N | | N | N | | N | Note 2, 3 | | Abdominal | | | | | | | | | | | | Intraoperative (specify)* | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small organ(specify)** | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | N | N | N | | N | N | | N | Note 2, 3 | | Transvaginal | | N | N | N | | N | N | | N | Note 2, 3 | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | Other (specify)*** | | N | N | N | | N | N | | N | Note 2, 3 | | N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | | | | | Additional comments: Combined modes: B+M, PW+B, Color + B, PW +Color+ B, PW +Color+ B, Power + PW +B. | | | | | | | | | | | | * Intraoperative includes abdominal, thoracic, and vascular etc. | | | | | | | | | | | | ** Small organ-breast, thyroid, testes, etc. | | | | | | | | | | | | *** Other use includes Urology. | | | | | | | | | | | Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. Note 2: Smart3D Note 3: iScape (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation(ODE) Prescription USE (Per 21 CFR 801.109) Joseph P. Wiley (Division Sign-Uff) Division of Reproductive, Abdominal and Radiological Galler : Sun Sum or 0051 {8}------------------------------------------------ # Diagnostic Ultrasound Indications for Use Form System Transducer × Model: 7L4s, 7L6s, 10L4s 510(k) Number(s) | | Mode of Operation | | | | | | | | | | |---------------------------------------------------------------------------------------------------|-------------------|---|---|---------|-----|------------------|----------------------|------------------------------|-----------------------|-----------------| | Clinical Application | A | B | M | PW<br>D | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other (specify) | | Ophthalrnic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative (specify)* | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small organ(specify)** | | N | N | N | N | | N | | N | Note 2, 3 | | Neonatal Cephalic | | N | N | N | N | | N | | N | Note 2, 3 | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | N | N | N | N | | N | | N | Note 2, 3 | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal Conventional | | N | N | N | N | | N | | N | Note 2, 3 | | Musculo-skeletal Superficial | | N | N | N | N | | N | | N | Note 2, 3 | | Other (specify) | | | | | | | | | | | | N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | | | | | Additional comments:Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+ B, Power + PW +B. | | | | | | | | | | | | * Intraoperative includes abdominal, thoracic, and vascular etc. | | | | | | | | | | | | ** Small organ-breast, thyroid, testes, etc. | | | | | | | | | | | | Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. | | | | | | | | | | | | Note 2: Smart3D | | | | | | | | | | | | Note 3: iScan | | | | | | | | | | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation(ODE) Prescription USE (Per 21 CFR 801.109) --- (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number 0052 ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ {9}------------------------------------------------ ## Diagnostic Ultrasound Indications for Use Form System n Transducer × Model: 6LE7s 510(k) Number(s) | Clinical Application | A | B | M | PW D | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|---|---|------|-----|---------------|-------------------|------------------------|--------------------|-----------------| | Ophthalmic | | | | | | | | | | | | Fetal | | N | N | N | | N | N | | N | Note 3 | | Abdominal | | | | | | | | | | | | Intraoperative (specify)* | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small organ(specify)** | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | N | N | N | | N | N | | N | Note 3 | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | Other (specify)*** | | N | N | N | | N | N | | N | Note 3 | | N=new indication; P=previously cleared by FDA; E=added under Appendix E<br>Additional comments:Combined modes: B+M, PW+B, Color + B, Power + B, PW +Color+ B, Power + PW +B.<br>*Intraoperative includes abdominal, thoracic, and vascular etc. | | | | | | | | | | | ** Small organ-breast, thyroid, testes, etc. ***Other use includes Urology. Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. Note 2: Smart3D Note 3: iScape (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation(ODE) Prescription USE (Per 21 CFR 801.109) Avanth M. Wz . (Division Sign-Off) (Division Sign-On) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number 0053 {10}------------------------------------------------ # Diagnostic Ultrasound Indications for Use Form × Transducer System 6LB7s Model: 510(k) Number(s) | | Mode of Operation | | | | | | | | | | |---------------------------------------------------------------------------------------------------|-------------------|---|---|---------|-----|------------------|----------------------|------------------------------|-----------------------|-----------------| | Clinical Application | A | B | M | PW<br>D | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other (specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative (specify)* | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small organ(specify)** | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | N | N | N | | N | N | | N | Note 2,3 | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | Other (specify) *** | | N | N | N | | N | N | | N | Note 2,3 | | N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | | | | | Additional comments: Combined modes: B+M, PW+B, Color + B, Power + B, PW+Color+ B, Power + PW +B. | | | | | | | | | | | | *Intraoperative includes abdominal, thoracic, and vascular etc. | | | | | | | | | | | | **Small organ-breast, thyroid, testes, etc. | | | | | | | | | | | | ***Other use includes Urology/Prostate. | | | | | | | | | | | | Note 1: Tissue Harmonic Imaging. The feature does not use contrast agents. | | | | | | | | | | | | Note 2: Smart3D | | | | | | | | | | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ### Concurrence of CDRH, Office of Device Evaluation(ODE) Prescription USE (Per 21 CFR 801.109) Arne M. Whtry (Division Sign-Off) (Division of Reproductive, Abdominal and Radiological Devices 510(k) Number J 0054
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...