← Product Code [DZE](/productcode/DZE) · K080559

# DIO SM IMPLANT SYSTEM (K080559)

_Dsi, Inc. · DZE · Oct 9, 2008 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K080559

## Device Facts

- **Applicant:** Dsi, Inc.
- **Product Code:** [DZE](/productcode/DZE.md)
- **Decision Date:** Oct 9, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The DIO SM Implant System is an endosseous dental implant is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading. Immediate loading is restricted to the anterior mandible, based on four interforminal placed implants, and not indicated for single unsplinted implants. Patients must be subject for dental treatment with endosseous implants.

## Device Story

DIO SM Implant System consists of endosseous dental implants, superstructures, and surgical/prosthetic instruments. Used by dental professionals in surgical settings to restore chewing function in partially or fully edentulous patients. Implants are surgically placed in jaw arches; support single or multiple-unit prosthetic appliances. Supports two-stage surgical process with transmucosal healing or single-stage process with immediate loading (restricted to anterior mandible with four interforaminal implants). Provides root-form foundation for prosthetic attachment. Benefits include restoration of oral function and aesthetics.

## Clinical Evidence

Bench testing only. Laboratory testing conducted to determine device functionality and conformance to design input requirements.

## Technological Characteristics

Material: CP Ti Gr4. Design: Internal type, Morse tapered, screw threads. Surface treatment: RBM (Resorbable Blast Media). Sterilization: Gamma. Diameters: 3.8/4.5/5.3 mm. Lengths: 7/8.5/10 mm.

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Predicate Devices

- SM Implant System ([K061797](/device/K061797.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Special 510(k): Device Modification

DIO SM Implant System

# OCT 0 9 2008

Attachment 4.

K080559

# 510(k) Summary

DIO SM Implant System

1. Submitter

DIO Department, DSI, Inc. 117 Kyo-Dong, Yangsan-City Kyungnam-Do, 626-210, Korea Phone: 82-55-383-7900 Fax : 82-55-363-3404

- 2. US Agent / Contact Person
CONTACT REPORT

3. Date Prepared

4. Device Name

5. Classification Name

6. Device Classification

7. Predicate Devices

8. Performance

9. Purpose

Hyungick Kim 3540 Wilshire Blvd. #1104 Los Angeles, CA 90010, USA Phone : 213-365-2875, Fax : 213-365-1595

January 08, 2008

### DIO SM IMPLANT SYSTEM

Endosseous Dental Implant System

Class II Dental Devices panel Regulation Number: 21 CFR 872.3640

### SM Implant System(510(k) No: K061797)

Laboratory testing was conducted to determine device functionality and conformance to design input requirements.

The purpose of this 510(k) is to modify the prior 510(k) submission for the SM Implant System

{1}------------------------------------------------

Charles of the Research Children of the Children Children of the Children

#### 10. Device Description

The DIO SM Implant System is comprised of dental implants, Superstructures, Instruments for prosthetics and Surgical Instruments.

The DIO SM Implant System is specially designed for use in dental implant surgery. A successfully osseointegrated implant will achieve a firm implant when surgically implanted under controlled conditions, per well known clinical studies. There are intended for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations.

#### 11. Packing / Labeling / Product Information

DIO SM Implant System follows the guidance of the 21 CFR 872.3640..

#### 12. Intended Use

The DIO SM Implant System is an endosseous dental implant is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading. Immediate loading is restricted to the anterior mandible, based on four interforminal placed implants, and not indicated for single unsplinted implants. Patients must be subject for dental treatment with endosseous implants.

### 13. Substantial Equivalence Comparison

|                          | Subject Device                   | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|--------------------------|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>Name     | DIO Department, DSI, Inc.        | DIO Department, DSI, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Device Name              | DIO SM Implant System            | SM Implant System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| 510(k) Number            | Not available yet                | K061797                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                          | Subject Device                   | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Intended Use             | Same with predicate device       | The DIO SM Implant System is an<br>endosseous dental implant is indicated for<br>surgical placement in the upper and lower<br>jaw arches, to provide a root form means<br>for single or multiple units' prosthetic<br>appliance attachment to restore a patient's<br>chewing function. Implants can be placed<br>with a conventional two stage surgical<br>process with an option for transmucosal<br>healing or they can be placed in a single<br>stage surgical process for immediate<br>loading. Immediate loading is restricted to<br>the anterior mandible, based on four<br>interforminal placed implants, and not<br>indicated for single unsplinted implants.<br>Patients must be subject for dental<br>treatment with endosseous implants. |
| Material                 | CP Ti Gr4                        | CP Ti Gr4                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Design                   | Internal Type and Morse Tapered  | Internal Type and Morse Tapered                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Screw Threads            | YES                              | YES                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Implant<br>Diameters(mm) | 3.8/4.5/5.3                      | 4.1/4.5/5.0/5.3                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Implant<br>Lengths(mm)   | 7/8.5/10                         | 7.0-12.5                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Surface<br>Treatment     | RBM (Resorbable Blast Media)     | RBM (Resorbable Blast Media)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Sterilization<br>Method  | GAMMA                            | GAMMA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Attachments              | Various abutments and components | Various abutments and components                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Code             | DZE                              | DZE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

## TECHNOLOGICAL CHARACTERISTIC COMPARISON

{2}------------------------------------------------

Special 510(k): Device Modification __________________________________________________________________________________________________________________________________________

. .

____DIO SM Implant System

. . .

Printer Production Provincial Province Program Program Program Program Program Program Program Program Program Program Program Program Program Program Program Program Program

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR & 807.93

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

OCT 0 9 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DIO Department, DSI, Incorporated C/O Mr. Hyungick Kim Manager DIO, USA 3540 Wilshire Boulevard, Suite 1104 Los Angeles, California 90010

Re: K080559

Trade/Device Name: DIO SM Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: September 19, 2008 Received: September 30, 2008

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Kim

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html,

Sincerely vours.

\$\mathbb{V}\$Ammelts-Lussards for \$\eta\$

Chiu S. Lin, Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Attachment 2.

# Indication for Use

510(K) Number (if known): K080559

Device Name: DIO SM Implant System

Indications For Use:

The DIO SM Implant System is an endosseous dental implant is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading. Immediate loading is restricted to the anterior mandible, based on four interforminal placed implants, and not indicated for single unsplinted implants. Patients must be subject for dental treatment with endosseous implants.

ts.

Suertunya

(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K080559

Prescription Use AND/OR Over - The-Counter Use (Part 21 CFR 801 Subpart D) (Per 21 CFR 801.109)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ·

Concurrence of CDRH, Office of Device Evaluation (ODE)

---

**Source:** [https://fda.innolitics.com/device/K080559](https://fda.innolitics.com/device/K080559)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
