K080558 · Case Medical, Inc. · FRG · Dec 5, 2008 · General Hospital
Device Facts
Record ID
K080558
Device Name
STERITITE CONTAINER SYSTEM & MEDITRAY PRODUCTS
Applicant
Case Medical, Inc.
Product Code
FRG · General Hospital
Decision Date
Dec 5, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The SteriTite universal container system with MediTray products is a reusable sterilization container system intended to be used to enclose other medical devices (blades and lumens), which are to be sterilized, transported and stored by a health care provider. The SteriTite container system is intended for use in STERRAD 200. STERRAD NX and Ozone 125L sterilizers, as well as the sealed containers are intended for pre-vacuum steam flash sterilization (270°F for 4 minutes). The SteriTite perforated bottom container with disposable polypropylene filters (Polypro) must be used for STERRAD Sterilization. The paper filter is recommended for prevacuum steam flash sterilization. For ozone sterilization, either disposable naper or polypropylene filters can be used.
Device Story
Rigid, reusable sterilization container system; anodized aluminum base, passivated stainless steel hardware, silicone gaskets. Includes perforated or solid bases, lids, filter retention plates, and internal instrument trays/baskets (MediTray). Used in operating rooms/sterile processing; operated by healthcare staff. Devices placed inside; container sealed with disposable filters (polypropylene or paper) and tamper-evident locks. Protects instruments during sterilization, transport, and storage. Validated for specific lumen sizes and weights in STERRAD, Ozone, and steam flash cycles. Ensures sterility maintenance via bacterial barrier filtration.
Clinical Evidence
Bench testing only. Validation performed using the "overkill method" of sterility assurance. Microbial challenge testing utilized biological indicators and inoculated products in worst-case locations (lumens, opposing corners, under occlusion). Validated for maximum loads (22 lbs for STERRAD, 35 lbs for steam/ozone) and specific lumen dimensions (diameters 1-3mm, lengths 400-850mm) across multiple sterilization cycles in independent laboratories per AAMI standards.
Technological Characteristics
Rigid container system; anodized aluminum body; passivated stainless steel hardware; silicone gaskets. Disposable filter-based bacterial barrier (polypropylene or paper). Vent holes offset to prevent strike-through. Modular design with interchangeable lids/bases. Compatible with STERRAD 200, STERRAD NX, Ozone 125L, and pre-vacuum steam flash sterilization. Complies with ANSI/AAMI ST 77:2006 and ST 79:2006.
Indications for Use
Indicated for use by healthcare providers to enclose, sterilize, transport, and store medical devices, including full instrument sets and mixed loads (blades and lumens). Compatible with STERRAD 200, STERRAD NX, Ozone 125L, and pre-vacuum steam flash sterilization. Specific lumen diameter and length constraints apply based on the sterilization cycle.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Aesculap Sterilcontainer
One Tray
Submission Summary (Full Text)
{0}------------------------------------------------
K040558
DEC 05 2008
#### 510(k) Summary of Safety and Effectiveness 5.
SteriTite Universal Container System with MediTray Products for STERRAD 200, STERRAD NX, Ozone 125L Sterilization, and Prevacuum Steam Flash Sterilization
Date Prepared: October, 14 2008
Company Name: Case Medical, Inc. 65 Railroad Avenue Ridgefield, NJ 07657
Contact: Tania Lupu Phone: 201-313-1999 ext.229 Fax: 201-313-9090
| Trade Name: | SteriTite® universal container system |
|------------------------------------|-------------------------------------------------|
| Common Name: | Sterilization container with disposable filter, |
| Establishment registration number: | 2248608 |
| Classification name: | Sterilization Wrap |
| Class of Device: | Class II device, |
| Product Code: | 80FRG |
| Review Panel: | General Hospital |
#### Indications for Use:
The SteriTite universal container system with MediTray products is a reusable sterilization container system intended to be used to enclose other medical devices (blades and lumens), which are to be sterilized, transported and stored by a health care provider.
The SteriTite container system is intended for use in STERRAD 200. STERRAD NX and Ozone 125L sterilizers, as well as the sealed containers are intended for pre-vacuum steam flash sterilization (270°F for 4 minutes).
The SteriTite perforated bottom container with disposable polypropylene filters (Polypro) must be used for STERRAD Sterilization. The paper filter is recommended for prevacuum steam flash sterilization. For ozone sterilization, either disposable naper or polypropylene filters can be used.
#### Description of the Device:
The SteriTite universal container system consists of a family of rigid reusable containers and inserts that provide an effective sterilization packaging method for operating room instruments. The SteriTite® container has both solid and perforated base containers. The container is made out of anodized aluminum with passivated stainless steel hardware and
{1}------------------------------------------------
silicone gaskets. A stainless steel latching mechanism with handles on both ends secures the lid to the base and provide a method to incorporate tamper proof disposable locks. Each filter retention plate secures a disposable filter for bacterial barrier filtration. Various instrument trays as well as stacked baskets and inserts provide instrument protection and secure devices for sterilization within the container. The lids and bases as well as retention plates of the same model and size are compatible and interchangeable throughout its useful life. Vent holes are offset to prevent strike through. All components may be easily disassembled for cleaning.
#### Performance Data
Case Medical, Inc. follows the "overkill method" of sterility assurance to show an elimination of a biological challenge. Microbial challenge testing included placement of biological indicators and inoculated products in the most difficult-to-sterilize areas of the container in opposing corners, under the lid, inside of lumens, within insert boxes, and under occlusion within the slot of an instrument bracket securing a surgical instrument. The lumens tested are as follows:
| Sterilization Method | Max Diameter Lumen | Max Length Lumen | Material |
|--------------------------------------------------|--------------------|------------------|------------------------------------------------|
| STERRAD 200 Sterilization | 3 mm | 400 mm | Stainless Steel Blades, Lumens and Mixed Loads |
| STERRAD NX Standard cycle | 2 mm | 400 mm | Stainless Steel Blades, Lumens and Mixed Loads |
| STERRAD NX Advance cycle | 1 mm | 500 mm | Stainless Steel Blades, Lumens and Mixed Loads |
| | 1 mm | 850 mm | Porous Lumens (Flexible endoscope) |
| Pre-Vacuum Flash with paper filter Sterilization | 2 mm | 400 mm | Stainless Steel Blades, Lumens and Mixed Loads |
| | 3 mm | 400 mm | Porous Lumens |
| Ozone 125L Sterilization | 3 mm | 470 mm | Stainless Steel Blades, Lumens and Mixed Loads |
All containers were tested with inner baskets or trays representative of the MediTrav line of products.
Furthermore, sample containers were validated in worst-case scenarios. In STERRAD NX and STERRAD 200 Sterilization the maximum load was 22lbs. (STU 10 kg) including the container. In Pre-Vacuum Steam Sterilization and Ozone 125L Sterilization, the maximum load was 35lbs. (15.8 kg.) total weight. The test samples represented a worst-case scenario in the SteriTite line of product based on number of holcs (vents) per contamer volume. Internal stacking of trays and utilization of components was also validated under fractional and half
{2}------------------------------------------------
cycle conditions. Validation testing utilizing replicate cycles was conducted in independent laboratories in accordance with available AAMI standards.
#### Substantial Equivalence:
Case Medical's SteriTite container system is substantially equivalent to the company's previously device as well as Aesculap's Sterilcontainer in STERRAD 100s and Ozone 125L, as well as Aesculap's Sterilcontainer and One Tray for prevacuum steam sterilization. The SteriTite container system is the same container previously cleared. However, the STERRAD compatible Aesculap Sterilcontainer is made out of non-anodized aluminum and is thus a different Aesculap Sterilcontainer than that used for steam sterilization. All containers are of equivalent sizes, have gasketed lids that latch, and offer tamper evident features.
#### References:
ANSVAAMI ST 77:2006 - Containment Devices for Reusable Medical Devices Sterilization ANSVAAMI ST 79:2006 - Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with its wings spread, symbolizing protection and service. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", indicating the department's name and national affiliation.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Tania Lupu Quality Assurance - Quality Control Director Case Medical, Incorporated 65 Railroad Avenue Ridgefield, New Jersey 07657-0402
DEC 0 5 2008
Re: K080558
Trade/Device Name: SteriTite Universal Container System and MediTray Products Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization wrap Regulatory Class: II Product Code: FRG Dated: November 13, 2008 Received: November 28, 2008
Dear Ms. Lupu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Ms. Lupu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
**Sincerely yours,**
line, Chiu S. Lin. Ph. D
Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: SteriTite Universal Container System and MediTray products
Indications for Use:
The SteriTite universal container system with MediTray products is a reusable sterilization container system used to enclose other medical devices, which are to be sterilized, transported and stored by a health care provider. The SteriTite container system is intended for use in STERRAD 200, STERRAD NX and Ozone 125L sterilizers, as well as the sealed containers are intended for prevacuum steam flash sterilization (270°F for 4 minutes). The container may be used for sterilization of medical devices including full instrument sets and mixed loads.
SteriTite Sealed Container system is recommended for surface and lumens:
-In STERRAD® 200 Sterilization, process only stainless steel lumened instruments of 3mm diameter or larger and a length up to 400 mm.
-In STERRAD NX standard cycle, process stainless steel lumened instruments of 2mm diameter or larger and up to 400 mm in length.
-In STERRAD NX advanced cycle, process stainless steel lumened instruments of 1mm diameter or larger and up to 500 mm in length and porous lumens (flexible endoscope) of 1mm diameter or larger and up to 850 mm in length.
-In Pre-vacuum Steam Flash Sterilization, process stainless steel lumened instruments of 2mm diameter or larger and a length of up to 400 mm as well as porous lumens of 3mm diameter or larger and a length up to 400 mm.
-In Ozone 125L Sterilization, process stainless steel lumened instruments of 3mm diameter or larger and a length up to 470 mm.
The attached tables identify which products with disposable filter may be sterilized in the respective sterilization cycles.
Over-The-Counter Use Prescription Use V AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
> Concurrence of O of Device Evaluation (ODE)
Shulie Murphy, MD
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of
510(k) Number: K080558
{6}------------------------------------------------
## SteriTite Universal Container System Compatibility Table in Prevacuum flash w/ paper filter
.
. . . . . . . . . . . . . . . . . . . . . . . .
1 .
:
:
:
.
.
and the comments of the count
:
:
.
- 上海
:
1
| Part<br>Number | Description | Pre-vacuum flash w/ paper<br>filter |
|----------------|----------------------------------------------|-------------------------------------|
| SC03M | 3" high Endo mini-size w/ solid base | X |
| SC03MG | 3" high Endo mini-size w/<br>perforated base | X |
| SC03Q | 3" high Endo mid-size w/ solid base | X |
| SC03QG | 3" high Endo mid-size w/ perforated<br>base | X |
| SC04F | 4" high Full-size w/ solid base | X |
| SC04FG | 4" high Full-size w/ perforated base | X |
| SC04H | 4" high Half -size w/ solid base | X |
| SC04HG | 4" high Half -size w/ perforated<br>base | X |
| SC04Q | 4" high Mid-size w/ solid base | X |
| SC04QG | 4" high Mid-size w/ perforated base | X |
| SC06F | 6" high Full-size w/ solid base | X |
| SC06FG | 6" high Full-size w/ perforated base | X |
| SC06H | 6" high Half -size w/ solid base | X |
| SC06HG | 6" high Half -size w/ perforated<br>base | X |
| SC06Q | 6" high Mid-size w/ solid base | X |
| SC06QG | 6" high Mid-size w/ perforated base | X |
| SC08F | 8" high Full-size w/ solid base | X |
| SC08FG | 8" high Full-size w/ perforated base | X |
| SC08H | 8" high Half -size w/ solid base | X |
| SC08HG | 8" high Half -size w/ perforated<br>base | X |
| SC08Q | 8" high Mid-size w/ solid base | X |
| SC08QG | 8" high Mid-size w/ perforated base | X |
Page 2 of _5__ ______
.
:
.
:
:
.
1000
:
:
:
. 1. 1
:
: .
{7}------------------------------------------------
| Part | Description | Ozone 125L w/ paper or<br>polypropylene disposable filter |
|--------|-------------------------------------------|-----------------------------------------------------------|
| Number | | |
| SC03M | 3" high Endo mini-size w/ solid base | X |
| SC03MG | 3" high Endo mini-size w/ perforated base | X |
| SC03Q | 3" high Endo mid-size w/ solid base | X |
| SC03QG | 3" high Endo mid-size w/ perforated base | X |
| SC04F | 4" high Full-size w/ solid base | X |
| SC04FG | 4" high Full-size w/ perforated base | X |
| SC04H | 4" high Half -size w/ solid base | X |
| SC04HG | 4" high Half -size w/ perforated base | X |
| SC04Q | 4" high Mid-size w/ solid base | X |
| SC04QG | 4" high Mid-size w/ perforated base | X |
| SC06F | 6" high Full-size w/ solid base | X |
| SC06FG | 6" high Full-size w/ perforated base | X |
| SC06H | 6" high Half -size w/ solid base | X |
| SC06HG | 6" high Half -size w/ perforated base | X |
| SC06Q | 6" high Mid-size w/ solid base | X |
| SC06QG | 6" high Mid-size w/ perforated base | X |
| SC08F | 8" high Full-size w/ solid base | X |
| SC08FG | 8" high Full-size w/ perforated base | X |
| SC08H | 8" high Half -size w/ solid base | X |
| SC08HG | 8" high Half -size w/ perforated base | X |
| SC08Q | 8" high Mid-size w/ solid base | X |
| SC08QG | 8" high Mid-size w/ perforated base | X |
.
:
.
.
#### SteriTite Universal Container System Compatibility Table Ozone 125L w/ paper & polypropylene disposable filter
1 - 1 - 1 - 1 - 1
and the comments of the
11
and the contraction of the comments of the count
: :
.
.
. . . . .
.
**Page 3 of 5**
. . . . .
网 - 亚洲足球
. .
:
.
:
:
:
:
:
.
....
. . . . . .
{8}------------------------------------------------
| SteriTite Universal Container System Compatibility in STERRAD 200 w/ | | |
|----------------------------------------------------------------------|---------------------------------|--|
| | polypropylene disposable filter | |
| Part | Description | STERRAD 200 w/<br>polypropylene<br>disposable filter |
|--------|-------------------------------------------|------------------------------------------------------|
| Number | | |
| SC03MG | 3" high Endo mini-size w/ perforated base | X |
| SC03QG | 3" high Endo mid-size w/ perforated base | X |
| SC04FG | 4" high Full-size w/ perforated base | X |
| SC04HG | 4" high Half-size w/ perforated base | X |
| SC04QG | 4" high Mid-size w/ perforated base | X |
| SC06FG | 6" high Full-size w/ perforated base | X |
| SC06HG | 6" high Half-size w/ perforated base | X |
| SC06QG | 6" high Mid-size w/ perforated base | X |
| SC08FG | 8" high Full-size w/ perforated base | X |
| SC08HG | 8" high Half-size w/ perforated base | X |
| SC08QG | 8" high Mid-size w/ perforated base | X |
Note: Part number with suffix "G" signifies perforated bottom container, which must be used with non-woven Polypro filter for STERRAD Sterilization.
#### SteriTite Universal Container System Compatibility in STERRAD NX w/ polypropylene disposable filter 1. 1
| Part<br>Number | Description | STERRAD NX w/<br>polypropylene<br>disposable filter |
|----------------|-------------------------------------------|-----------------------------------------------------|
| SC03MG | 3" high Endo mini-size w/ perforated base | X |
| SC03QG | 3" high Endo mid-size w/ perforated base | X |
| SC04FG | 4" high Full-size w/ perforated base | X |
| SC04HG | 4" high Half-size w/ perforated base | X |
| SC04QG | 4" high Mid-size w/ perforated base | X |
Note: For STERRAD NX Sterilization only the 3" high and 4" high containers can be used due to STERRAD NX chamber size
·
{9}------------------------------------------------
| MEDITRAY<br>PRODUCT | STERRAD 200 | STERRAD NX | Ozone 125L | Pre-vacuu<br>flash |
|---------------------|-------------|------------|------------|--------------------|
| Baskets | X | X | X | X |
| Trays | X | X | X | X |
| Insert Boxes | X | X | X | X |
| Metal Brackets | X | X | X | X |
| Metal Partitions | X | X | X | X |
| Posts | X | X | X | X |
| Silicone Brackets | X | X | X | X |
| Racks | X | X | X | X |
| Stringers | X | X | X | X |
### MediTray Products Compatibility Table
:
:
:
.
.
.
. . .
:
.
:
.
.
100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
: : :
:
11. 11. 11.
:
:
.
:
1. 1. 1. 1. 1.
:
1 - 1 - 1 - 1 - 1 - 1 -
. .
and the comments of the comments of
and the same of the same of the same
:
:
. T
Page 5 of _5__ Page 5 of _5___
ge 5 of 5
、「
:
:
:
.
:
:
:
1
. .
1. 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.