← Product Code [LZA](/productcode/LZA) · K080536

# MICRO-TOUCH NITRAFREE (K080536)

_Ansell Healthcare Products, LLC · LZA · Jun 19, 2008 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K080536

## Device Facts

- **Applicant:** Ansell Healthcare Products, LLC
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Jun 19, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

This is a medical glove to be worn on the hands of health care and similar personnel to prevent contamination between health care personnel and the patient's body, fluids, waste or environment and for use in handling chemotherapy drugs.

## Device Story

Micro-Touch NitraFree Nitrile Powder-Free Pink Examination Gloves are non-sterile, disposable nitrile gloves. Used by healthcare personnel in clinical settings to provide a protective barrier against contamination from patient body fluids, waste, or the environment; specifically indicated for handling chemotherapy drugs. The device functions as a physical barrier; it is worn on the hands. Benefits include reduced risk of cross-contamination and protection during chemotherapy drug handling. No complex processing or electronic output involved.

## Clinical Evidence

No clinical data. Bench testing only, including physical property testing per ASTM D6319-00a(2005), freedom from holes per ASTM D5151-06, and biocompatibility studies (ISO skin irritation, ISO maximization sensitization, and cytotoxicity).

## Technological Characteristics

Nitrile examination glove; powder-free (≤ 2 mg/glove). Meets ASTM D6319-00a(2005) for dimensions and physical properties. Freedom from holes meets ASTM D5151-06. Biocompatibility: passes ISO skin irritation, ISO maximization sensitization, and cytotoxicity (non-toxic at 24 hours for dilutions 1:2 through 1:32).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Ansel

K080536

JUN 1 9 2008

- [1] 510(k) SUMMARY
- [2] Ansell Healthcare Products LLC 1635 Industrial Road Dothan, AL 36303

Contact: Lon D. McIlvain, Vice President Regulatory and Quality Affairs Global Telephone: (334) 615-2562 Fax: (334) 615-2568

February 26, 2008

- [3] Trade Name: Micro-Touch® NitraFree™ Nitrile Powder-Free Pink Examination Gloves (Chemotherapy Use)
Common Name: Examination Gloves

Classification Name: Glove, Patient Examination, Nitrile

- [4] Micro-Touch® NitraFree™ Nitrile Powder-Free Pink Examination Gloves (Chemotherapy Use) meet all of the requirements of ASTM D 6319-00a(2005).
- Micro-Touch® NitraFree™ Nitrile Powder-Free Pink Examination Gloves [ર] (Chemotherapy Use) meet all of the current specifications of ASTM D6319-00a(2005), Standard Specification for Nitrile Examination Gloves for Medical Application.
- [6] Micro-Touch® NitraFree™ Nitrile Powder-Free Pink Examination Gloves (Chemotherapy Use) are non-sterile disposable devices to be worn on the hands of health care and similar personnel to prevent contamination between health care personnel and the patient's body, fluids, waste or environment and for use in handling chemotherapy drugs.
- Micro-Touch® NitraFree™ Nitrile Powder-Free Pink Examination Gloves [7] (Chemotherapy Use) are summarized with the following technological characteristics compared to ASTM or equivalent standards.

| Characteristics                                                                                                                                                | Standard                                                                      |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Dimensions                                                                                                                                                     | Meets ASTM D 6319-00a(2005)                                                   |
| Physical Properties                                                                                                                                            | Meets ASTM D 6319-00a(2005)                                                   |
| Freedom from Holes                                                                                                                                             | Meets ASTM D 6319-00a(2005)<br>Meets ASTM D 5151-06                           |
| Powder-Free                                                                                                                                                    | Powder content ≤ 2 mg per glove                                               |
| Biocompatibility:<br>ISO Skin Irritation Study<br>ISO Maximization Sensitization Study - Extract<br>Cytotoxicity Study Using the End-Point<br>Titration Method | Passes<br>Passes<br>Non-toxic at 24 hrs.<br>for dilutions 1:2 through<br>1:32 |

{1}------------------------------------------------

- [8] The performance test data of the non-clinical tests are the same as mentioned immediately above.
- [0] Clinical data is not needed for medical gloves or for most devices cleared by the 510(k) process.
- It is concluded that Micro-Touch® NitraFree™ Nitrile Powder-Free Pink Examination
Gloves (Chemotherapy Use) are as safe, as effective, and perform as well as the glove [10] performance standards referenced in Section 7 above and therefore meet:

ASTM listed standards. FDA hole requirements, and labeling claims for the product.

- This summary will include any other information reasonably deemed necessary by the [1] FDA.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird with three wing-like strokes, symbolizing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# JUN 1 9 2008

Mr. Lon D. McIlvain Vice president Regulatory and Quality Affairs Global Ansell Healthcare Products, LLC 1635 Industrial Road Dothan, Alabama 36303

Re: K080536

Trade/Device Name: Micro-Touch® NitraFree™ Nitrile Powder-Free Pink Examination Gloves (Chemotherapy Use) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA, LZC Dated: June 3, 2008 Received: June 4, 2008

Dear Mr. McIlvain:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. McIlvain

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Suette Y. Michie MD.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## INDICATIONS FOR USE

#### 510(K) Number (if known):

Device Name:

Micro-Touch® NitraFree™ Nitrile Powder-Free Pink Examination Gloves (Chemotherapy Use)

### Indications For Use:

This is a medical glove to be worn on the hands of health care and similar personnel to prevent contamination between health care personnel and the patient's body, fluids, waste or environment and for use in handling chemotherapy drugs.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation (ODE)

Shule A Mughts

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K080536

---

**Source:** [https://fda.innolitics.com/device/K080536](https://fda.innolitics.com/device/K080536)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
