← Product Code [EBF](/productcode/EBF) · K080480

# NANOCOMPOSITE RESTORATIVE KIT (K080480)

_Dmg USA, Inc. · EBF · May 30, 2008 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K080480

## Device Facts

- **Applicant:** Dmg USA, Inc.
- **Product Code:** [EBF](/productcode/EBF.md)
- **Decision Date:** May 30, 2008
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 872.3690
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Nanocomposite Restorative Kit consists of 2 types of nanocomposite restorative materials with a compatible bonding agent. The specific indications of each kit component are as follows: 1. Nanocomposite is indicated for: Restorations of all cavity classes Fabrication of direct inlays, onlays and indirect veneers Core build-up after tooth preparation with the root canal post Machinable Composite for Restorations Indirect Inlays/Onlays 2. Nanocomposite Flowable is indicated for: Small fillings of cavity classes III, IV and V Minimally invasive fillings for deciduous teeth (all cavity classes) Splinting of teeth 3. MDP One Bottle Bond is indicated for: Bonding resin-based materials (especially light-cure composite / compomer materials) to tooth structure (dentin and enamel)

## Device Story

Nanocomposite Restorative Kit comprises two nanocomposite materials and a bonding agent; used by dental professionals in clinical settings for tooth restoration and bonding. Nanocomposite used for cavity restorations, inlays, onlays, veneers, and core build-ups. Nanocomposite Flowable used for small fillings, deciduous teeth, and splinting. MDP One Bottle Bond facilitates adhesion of light-cure composites/compomers to tooth structure. Materials applied directly to prepared tooth surfaces; light-cured to harden. Benefits include structural repair of teeth and restoration of dental function.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Kit includes nanocomposite restorative materials and a bonding agent. Resin-based composition. Light-cure polymerization. Class II device per 21 CFR 872.3690.

## Regulatory Identification

Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.

## Predicate Devices

- SternOmega Composite LC ([K982692](/device/K982692.md))
- Clearfil tri-S bond ([K042913](/device/K042913.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summarv

|                      |                                                                                                                                  | MAY 30 2008 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------|-------------|
| Trade Name:          | Nanocomposite Restorative Kit                                                                                                    |             |
| Sponsor:             | DMG USA, Inc.<br>23 Frank Mossberg Drive<br>Attleboro, MA 02703<br>Registration # not yet assigned<br>Owner/Operator No. 9005969 |             |
| Device Generic Name: | Nanocomposite Restorative Kit                                                                                                    |             |
| Classification:      | According to Section 513 of the Federal Food, Drug, and<br>Cosmetic Act, the device classification is Class II.                  |             |

### Product Description:

The Nanocomposite Restorative Kit consists of 3 materials:

#### Nanocomposite is indicated for

- Restorations of all cavity classes .
- . Fabrication of direct inlays, onlays and indirect veneers
- Core build-up after tooth preparation with the root canal post .
- . Machinable Composite for Restorations
- . Indirect Inlays/Onlays

### Nanocomposite Flowable is indicated for

- . Small fillings of cavity classes III, IV and V
- Minimally invasive fillings for deciduous teeth (all cavity classes) .
- Splinting of teeth .

### MDP One Bottle Bond is indicated for:

- Bonding resin-based materials (especially light-cure composite / compomer . materials) to tooth structure (dentin and enamel)
### Predicate Devices:

The components of the proposed Nanocomposite Restorative Kit materials are substantially equivalent to several currently marketed dental restorative materials including the following:

### Nanocomposite:

| Product Name            | Predicates                     |
|-------------------------|--------------------------------|
| SternOmega Composite LC | K982692 (Sterngold / Implamed) |

{1}------------------------------------------------

#### Nanocomposite Flowable:

| Product Name            | Predicates                     |
|-------------------------|--------------------------------|
| SternOmega Composite LC | K982692 (Sterngold / Implamed) |

### MDP One Bottle Bond:

| Product Name        | Predicates                     |
|---------------------|--------------------------------|
| Clearfil tri-S bond | K042913 (Kuraray Medical inc.) |

### Safety and Performance:

This submission is an Abbreviated 510(k) as described in FDA's guidance document entitled "The New 510(k) Paradigm - Alternate Approaches to Demonstrating Substantial Equivalence in Premarket Notifications." In support of this 510(k), DMG-USA has provided information to demonstrate conformity with FDA's guidance document entitled Guidance for Industry and FDA Staff: Dental Cements - Premarket Notification (August 1998) and Guidance for Industry and FDA Staff: Dental Composite Resin Devices -Premarket Notification [510(k)] Submissions (October 2005).

#### Conclusion:

Based on the indications for use, technological characteristics, and comparison to the predicate devices, the Nanocomposite Restorative Kit has been shown to be safe and effective for its intended use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## MAY 30 2008

DMG USA, Incorporated C/O Ms. Pamela Papineau Regulatory Affairs Consultant Delphi Medical Device Consulting 5 Whitcomb Avenue Ayer, Massachusetts 01432

Re: K080480

Trade/Device Names: Nanocomposite Restorative Kit Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: II Product Codes: EBF and KLE Dated: May 5, 2008 Received: May 12, 2008

Dear Ms. Papineau:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Ms. Papineau

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Chin Liu, Ph.D.

Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

5 May 2008 Nanocomposite Restorative Kit

Page

0:0011

510(k) Number (if known): K080480

Device Name: Nanocomposite Restorative Kit

#### Product Indications for Use:

The Nanocomposite Restorative Kit consists of 2 types of nanocomposite restorative materials with a compatible bonding agent. The specific indications of each kit component are as follows:

- 1. Nanocomposite is indicated for:
- . Restorations of all cavity classes
- Fabrication of direct inlays, onlays and indirect veneers .
- Core build-up after tooth preparation with the root canal post .
- Machinable Composite for Restorations .
- Indirect Inlays/Onlays .
- 2. Nanocomposite Flowable is indicated for:
- Small fillings of cavity classes III, IV and V D
- Minimally invasive fillings for deciduous teeth (all cavity classes) ●
- Splinting of teeth .
- 3. MDP One Bottle Bond is indicated for:
- Bonding resin-based materials (especially light-cure composite / compomer . materials) to tooth structure (dentin and enamel)

| Prescription Use           | X |
|----------------------------|---|
| (Per 21 CFR 801 Subpart D) |   |

OR

| Over-the-Counter Use       |  |
|----------------------------|--|
| (Per 21 CFR 801 Subpart C) |  |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
![Signature](signature.jpg)
(Division Sign-Off)
Division of Anesthesiology, General Hospital- Infection Control, Dental Devices
510(k) Number: K080486

---

**Source:** [https://fda.innolitics.com/device/K080480](https://fda.innolitics.com/device/K080480)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
