K080460 · Opko Health, Inc. · OBO · Nov 14, 2008 · Ophthalmic
Device Facts
Record ID
K080460
Device Name
SPECTRAL OCT/SLO
Applicant
Opko Health, Inc.
Product Code
OBO · Ophthalmic
Decision Date
Nov 14, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1570
Device Class
Class 2
Indications for Use
The Spectral OCT/SLO is a non-contact, high-resolution non-invasive tomographic and confocal imaging device. It is indicated for in vivo viewing, axial cross-sectional, and three-dimensional imaging and measurement of posterior ocular structures including: retina, macula, retinanserve fibre layer and optic disk. It is used as a diagnostic device to aid in the detection and management of ocular diseases affecting the posterior segment of the eye. In addition, corners and conjunctiva can be imaged with the system by changing the focal position.
Device Story
Spectral OCT/SLO is a computerized, non-contact ophthalmic imaging system. It uses low-coherence interferometry and superluminescent diode light to acquire simultaneous high-resolution cross-sectional OCT and confocal images of ocular structures. The system includes an optical assembly, light source, power supply, and computer for data analysis and image processing. It is used by clinicians to visualize and measure retinal, macular, and optic disc structures to assist in diagnosing and managing posterior segment ocular diseases. By adjusting focal position, it also images the cornea, sclera, and conjunctiva. Output is displayed to the clinician for diagnostic assessment, aiding in clinical decision-making regarding ocular pathology.
Clinical Evidence
Prospective comparative study of 36 subjects, including normal individuals and patients with retinal or glaucoma pathologies. Evaluated precision and agreement of retinal, RNFL thickness, and optic nerve disc ratio measurements between Spectral OCT/SLO and Stratus OCT. Results showed the device is equivalent or better in precision tests, with regression analysis confirming substantial equivalence to the predicate.
Technological Characteristics
Non-contact tomographic and confocal imaging system. Utilizes low-coherence interferometry with superluminescent diode light source. Hardware includes optical system, power supply, and computer for image processing. Key difference from predicate: uses spectrometer with optical grating and linear CCD camera for signal detection. Software performs data analysis and image processing.
Indications for Use
Indicated for in vivo viewing, axial cross-sectional, and 3D imaging and measurement of posterior ocular structures (retina, macula, retinal nerve fiber layer, optic disk) and anterior structures (cornea, sclera, conjunctiva) to aid in detection and management of ocular diseases.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
Predicate Devices
OCT Ophthalmoscope, Ophthalmic Technologies, Inc (K042885)
{0}------------------------------------------------
K080460
Image /page/0/Picture/1 description: The image shows the logo for OPKO Instrumentation/OTI, which is a division of OPKO Health, Inc. The logo is black and white and features the text "OPKO" in bold letters, followed by a slash and "OTI" in a different font. Above the text is the number 14.
510(k) Summary October 1, 2008
NOV 1 4 2008
1. Submitter Information
Name: OPKO Instrumentation/OTI 4400 Biscayne Boulevard Miami, FL 33137
| Telephone Number: | 305-575-6004 |
|-------------------|--------------|
| Fax Number: | 305-575-6016 |
Contact Person: Jane Hsiao e-mail: jhsiao@opko.com
Date Submitted: October 1, 2008
- 2. Name of Device
Trade Name: Spectral OCT/SLO Common Name: Ophthalmoscope using optical coherence tomography Classification Name: Ophthalmoscope, a-c powered, 800HE
- 3. Equivalent legally marketed devices.
K042885 OCT Ophthalmoscope, Ophthalmic Technologies, Inc K033123 Stratus OCT, Carl Zeiss Meditec System
- 4. Description
The OPKO/OTI Spectral OCT/SLO is a computerized instrument that employs non-invasive, low-coherence interferometry to acquire simultaneous highresolution cross-sectional OCT (Optical Coherence Tomography) and confocal images of ocular structure, including retina, retinal nerve fiber layer, macula and optic disc of the eye.
{1}------------------------------------------------
- 5. Indication for Use:
The Spectral OCT/SLO is a non-contact, high-resolution non-invasive tomographic and confocal imaging device. It is indicated for in vivo viewing. axial cross-sectional, and three-dimensional imaging and measurement of posterior ocular structures including: retina, macula, retina nerve fibre lancer and optic disk. It is used as a diagnostic device to aid in the detection and management of ocular diseases affecting the posterior segment of the eye. In addition, cornea, sclera and conjunctiva can be imaged with the system by changing the focal position.
- 6. Technological Characteristics:
The OPKO/OTI Spectral OCT/SLO and the predicate device, Zeiss Stratus OCT both utilize Optical Coherence Tomography, which relies upon interferometty of superluminescent diode light, reflected from the fundus of the eye, to produce cross-sectional image of the retina. Both systems have similar hardware system consist of an optical system, the light source, a power supply and a computer with software for data analysis and image processing. The two systems differ in the that OPKO/OTI Spectral OCT/SLO uses spectrometer with optical grading and Iinear CCD camera while Stratus OCT uses photo-detector to detect the regurn OCT light signal from the eve.
- 7. Performance Data
- a. Non-clinical tests
The Spectral OCT/SLO has had accuracy test, optical emissions analyses, image comparison, electrical safety test, electromagnetic compatibility test and software validation tests.
- b. Clinical tests:
Thirty six subjects, including normal and patients with retinal or glaucoma pathologies, participated in a controlled prospective comparative study to evaluate the performance (precision and agreement) of Spectral OCT/SLO and the predicate device Stratus OCT in the measurement of retinal, RNFL thickness and optical nerve disc ratio.
- c. Conclusions
The Spectral OCT/SLO is equivalent or better in the precision tests when compares to the predicate device. The regression analysis between the measurements are substantially equivalent between the test and the legally marketed predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines representing its body and wings. The eagle faces left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OPKO Health, Inc. c/o Jane Hsiao, Ph.D. OPKO Health, Inc. 4400 Biscayne Blvd. Miami, Florida 33137
NOV 1 4 2008
Re: K080460
Trade/Device Name: Spectral OCT/SLO Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: II Product Code: OBO Dated: November 11, 2008 Received: November 12, 2008
Dear Dr. Hsiao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-frec number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Maling B. Egeratas, und
Malvina B. Eydelman, M.L Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number: K080460 Device Name: Spectral OCT/SLO
Indications For Use:
The Spectral OCT/SLO is a non-contact, high-resolution non-invasive tomographic and confocal imaging device. It is indicated for in vivo viewing, axial cross-sectional, and three-dimensional imaging and measurcment of posterior ocular structures including: retina, macula, retinanserve fibre layer and optic disk. It is used as a diagnostic device to aid in the detection and management of ocular diseases affecting the posterior segment of the eye. In addition, corners and conjunctiva can be imaged with the system by changing the focal position.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
umtte
(Division Sign-Off)
Division of Ophthalmic and Ear. Nose and Throat Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.