K080451 · AngioDynamics, Inc. · GEI · Jul 3, 2008 · General, Plastic Surgery
Device Facts
Record ID
K080451
Device Name
UNIBLATE ELECTROSURGICAL DEVICE
Applicant
AngioDynamics, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 3, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AngioDynamics Uniblate Electrosurgical Device is intended for coagulation and ablation of tissue during percutaneous, laparoscopic, and intraoperative surgical procedures, such ar martial or complete ablation of non-resectable liver lesions, osteoid osteoma, and palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard therapy.
Device Story
Monopolar radiofrequency (RF) electrosurgical probe; used for tissue coagulation and ablation. Input: RF energy from RITA Medical 1500X generator; saline infusion from IntelliFlow peristaltic pump. Device features variably insulated electrode (1-2.5 cm active length) for scalable ablation zones. Operated by physicians in percutaneous, laparoscopic, or intraoperative settings. Provides RF energy and temperature feedback to generator/pump system. Benefits patients by enabling targeted destruction of non-resectable liver lesions, osteoid osteoma, and palliative treatment of metastatic bone pain.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and intended use comparisons to legally marketed predicate devices.
Technological Characteristics
Monopolar RF electrosurgical probe; variably insulated electrode (1-2.5 cm active length); single-use. Integrates with external RF generator and peristaltic pump for saline infusion. Provides temperature monitoring and local fluid delivery. Energy source: RF generator.
Indications for Use
Indicated for coagulation and ablation of tissue in patients with non-resectable liver lesions, osteoid osteoma, or metastatic bone lesions who have failed or are not candidates for standard therapy. Used in percutaneous, laparoscopic, or intraoperative procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
KO80451
JUL - 3 2008
# 510(k) Summary
Date: February 15, 2008
## Sponsor of the 510(k)
AngioDynamics, Inc. 603 Queensbury Ave Queensbury, NY 12801 Establishment Registration number 1319211 Contact: Brian Kunst, Vice President, Regulatory Affairs and Quality Assurance 518-798-1215, x1123
### Device Identification:
Proprietary Name: Common Name: Classification Name:
Classification Number: Classification Panel: Product Code: Regulatory Class:
Uniblate Electrosurgical Device Radiofrequency probe Electrosurgical cutting and coagulation device and accessories 21 CFR §878.4400 General and Plastic Surgery GEI II
## Legally marketed device to which equivalence is claimed:
Angiodynamics Uniblate Electrosurgical Device, 510(k) K070101 Angiodynamics (formerly RITA Medical Systems) Starburst Electrosurgical Device 510(k) K030967 Valleylab Cool-Tip Electrosurgical probe, 510(k) K052796
### Intended Use / Indications
The AngioDynamics Uniblate System is intended for coagulation and ablation of tissue during percutaneous, laparoscopic, and intraoperative surgical procedures, such ar martial or complete ablation of non-resectable liver lesions, osteoid osteoma, and palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard therapy.
Page 1 of 2
{1}------------------------------------------------
# K08045
### Device Description
The UniBlate Electrosurgical Device is a monopolar radiofrequency (RF) device that consists of a variably insulated electrode that provides an adjustable ablation zone of 1 cm-3 cm. The device has an attached electrical cable and infusion tubing set which connects the it directly to the RTA Medical 1500X RF generator and the IntelliFlow peristaltic pump. The electrical cable provides RF energy and temperature feedback to the generator and the pump infuses normal salive through the infusion tubing set. The UniBlate Electrosurgical Device is a single use device.
### Technological Differences:
The AngioDynamics UniBlate, AngioDynamics Starburst Electrosurgical device (K030967), and Valleylab Cool-Tip (K052796) are monopolar electrodes used to deliver RF energy during open, laparoscopic or percutaneous procedures to ablate and coagulate soft tissue. The UniBlate is designed to provide a scaleable coagulation zone. The UniBlate electrode has a single active electrode that can be exposed from 1 cm to 2.5 cm by retracting an insulating sheath. The Cool-Tip has a fixed length active electrode, but is available in different sizes of 1cm, 2cm, and 3cm. The UniBlate allows for local fluid delivery as well as temperature monitoring. The Starburst SDE is a mulit-tined probe that provides an adjustable ablation zone up to 2cm.
## Intended Use
The AngioDynamics Uniblate Electrosurgical Device is intended for coagulation and ablation of tissue during percutaneous, laparoscopic, and intraoperative surgious procedures, such as partial or complete ablation of non-resectable liver sugars, osteoid osteoma, and palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard therapy
The Angiodynamics Starburst electrosurgical device is indicated for coagulation and ablation of tissue during percutaneous, laparoscopic, and intraoperative surgional procedures, such as partial or complete ablation of non-resectable liver lessions and palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard therapy
The Valleylab Cool-tip RF System (generator and accessories) is intended for the use in percutaneous, laparoscopic, intraoperative coagulation and ablation of tissue, such as partial or complete ablation of non-resectable liver lesions and osteoid osteoma tumors within bone.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird-like figure, composed of several curved lines. The symbol is positioned in the center of the seal.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AngioDynamis, Inc. % Mr. Brian Kunst 603 Queensbury Avenue Qucensbury, New York 12804
JUL - 3 2008
Re: K080451
Trade/Device Name: Uniblate Electrosurgical Device Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: May 24, 2008 Received: May 29, 2008
Dear Mr. Kunst:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Brian Kunst
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial cquivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Prostmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events, (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html,
Sincerely vours.
Mark N Millham
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K080451
Device Name: Uniblate Electrosurgical Device
Indications for Use:
The AngioDynamics Uniblate Electrosurgical Device is intended for coagulation and ablation of tissue during percutaneous, laparoscopic, and intraoperative surgical procedures such as partial or complete ablation of non-resectable liver lesions, osteoid osteoma, and palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard therapy
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807
Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.