MAPERSOL

K080432 · Preservation Solutions, Inc. · KDN · Aug 8, 2008 · Gastroenterology, Urology

Device Facts

Record IDK080432
Device NameMAPERSOL
ApplicantPreservation Solutions, Inc.
Product CodeKDN · Gastroenterology, Urology
Decision DateAug 8, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5880
Device ClassClass 2

Indications for Use

MaPerSol® organ preservation solution is intended for in-vitro flushing and continuous hypothermic machine perfusion of explanted kidneys.

Device Story

MaPerSol® is a sterile, non-toxic organ preservation solution used for explanted kidneys. It is formulated with soluble colloids, buffers, sodium and potassium salts, redox stabilizers, and ATP-regenerating compounds. The solution is supplied in 1-liter flexible bags. It is used in a clinical or laboratory setting to maintain organ cleanliness and viability during hypothermic machine perfusion prior to transplantation. The solution is administered by healthcare professionals during the organ preservation process. It benefits the patient by preserving the explanted kidney's viability for transplantation.

Clinical Evidence

Bench testing only. Testing included chemical, microbiological, and toxicological assessments. Shelf-life stability was validated over 12 months using Fourier transformed infrared spectrometry. Osmolarity/osmolality and pH measurements demonstrated consistency with predicate devices. Biocompatibility and particulate matter limits for large volume parenterals were verified.

Technological Characteristics

Formulation includes soluble colloids, buffers, sodium/potassium salts, redox stabilizers, and ATP-regenerating compounds. Packaged in 1-liter flexible solution bags with ports for aseptic connection. Sterilized via sterile filtration. No electronic or software components.

Indications for Use

Indicated for in-vitro flushing and continuous hypothermic machine perfusion of explanted kidneys.

Regulatory Classification

Identification

An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) SUMMARY ### SUBMITTER INFORMATION AUG - 8 2008 - A. Company Name: Preservation Solutions Inc. - B. Company Address: 980 Proctor Drive Elkhorn, Wisconsin 53121 - C. Company Phone: 262 723 6715 - D. Company Facsimile: 262 723 4013 - E. Contact Person: William Wagner Quality Assurance Director ## DEVICE IDENTIFICATION - A. Device Trade Name: MaPerSol® - B. Device Common Name: Organ Preservation Solution - C. Classification Name: Isolated kidney perfusion and transport system and accessories - D. Class II (21 CRF 876.5880) - E. Device Code: KDN ## IDENTIFICATION OF PREDICATE DEVICES MaPerSol® is a organ preservation solution for explanted kidneys, which is substantially equivalent to a both Belzer-MPS™, UW Machine Perfusion Solution (K972066) and KPS-1 Kidney Perfusion Solution (K022391). ## DEVICE DESCRIPTION Preservation Solutions, Inc. manufacturers MaPerSol® according to a "recipe" pioneered at the University of Wisconsin by Dr. Folkert O. Belzer and Dr. James Southard. The kidney flushing and perfusion storage solution may still be referred to as Belzer-MPS™, UW Machine Perfusion Solution, and is currently sold under that trade name (Trans-Med Corporation, Elk River, MN), as well as under the trade name KPS-1 (Organ Recovery Systems, Des Plaines, IL). The formulation includes soluble colloids, buffers, sodium and potassium salts, redox stabilizers, and compounds to aid tissue viability by enabling regeneration of adenosine triphosphate (ATP). {1}------------------------------------------------ MaPerSol® is a colorless, sterile, non-toxic solution for in-vitro flushing and continuous perfusion of explanted kidnevs. The solution is packaged in 1-liter bags. The Food and Drug Administration requlates such solutions as medical devices as referenced below. ## INDICATIONS FOR USE MaPerSol® organ preservation solution is intended for in-vitro flushing and continuous hypothermic machine perfusion of explanted kidneys. ## COMPARISON TO PREDICATE DEVICES MaPerSol® is identical in formulation to two (2) predicate products, KPS-1 Kidney Perfusion Solution (K022391) and Belzer-MPS™ UW Machine Perfusion Solution (K972066). The solutions maintain organ cleanliness and viability prior to transplantation. ## BIOCOMPATIBILITY, STERILIZATION, PACKAGING, AND BENCH TESTING MaPerSol®, mixed and filled according to written procedures under routine conditions at the manufacturer's (Preservation Solutions, Inc.) site, has been verified as biocompatible and meets the particulate matter limits for large volume parenterals (LVP's). Compounded and fully dissolved MaPerSol® is filter steritically filled into flexible solutions bags fitted with ports suitable for aseptic connection to standard infusion or administration sets. Sterilization processes have been validated following national and international standards and quidance documents. A shelf life of 1 year was confirmed using chemical, microbiological,and toxicological tests, as well as via Fourier transformed infrared spectrometry. MaPerSol® solution, fully packaged and labeled as if for sale and stored for approximately 12 months, compares favorably with fresh MaPerSol®. No deleterious effects due to aging occur in either the solution or its flexible pouch container. Routine chemical tests for osmolarity/osmolality and pH show consistent equivalence among the proposed device MaPerSol® and the two (2) predicates. ## CONCLUSION The above statements establish substantial equivalence between MaPerSol® and the cited predicates. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized design resembling three abstract human figures or a symbolic representation of health and human services. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG - 8 2008 Preservation Solutions, Inc. c/o Mr. Neil Burris Neil Burris & Associates 4250 Grove Street DENVER CO 80211 Re: K080432 > Trade/Device Name: MaPerSol® Regulation Number: 21 CFR §876.5880 Regulation Name: Isolated kidney perfusion and transport system and accessories Regulatory Class: II Product Code: KDN Dated: July 16, 2008 Received: July 17, 2008 Dear Mr. Burris: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Nancy Brogdon Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure Page 2 {4}------------------------------------------------ #### 4. Statement of Indication for Use Device Name: MaPerSol® Indication for Use MaPerSol® organ preservation solution is intended for in-vitro flushing and continuous hypothermic machine perfusion of explanted kidneys. Prescription Use XXXX (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) signature (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devic 510(k) Number MaPerSol Traditional 510(k) Preservation Solutions Inc. February 2008, page 17 of 241 Corrected July 2008 KD80432 PSI MaPerSol Solution
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...