K080391 · Medical Depot · CAW · Mar 18, 2008 · Anesthesiology
Device Facts
Record ID
K080391
Device Name
DRIVE SOLSTICE OXYGEN CONCENTRATOR
Applicant
Medical Depot
Product Code
CAW · Anesthesiology
Decision Date
Mar 18, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The intended function and use of the Drive Solistice Oxygen Concentrator (models 18050 and 18055) is to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve. It is not intended to sustain or support life.
Device Story
Drive Solstice Oxygen Concentrator (models 18050, 18055) is a portable PSA (Pressure Swing Adsorption) oxygen generator. Device draws room air through intake filters into a piston-type compressor; compressed air is directed via pneumatic valve into one of two molecular sieve beds. Nitrogen is adsorbed under pressure; oxygen concentrates at bed top. Enriched oxygen (90% ± 3%) is regulated and delivered to patient at 1-5 LPM. Second bed simultaneously exhausts nitrogen as pressure decreases. Dual solenoid, three-position, five-way valve manages cycle. Used in clinical or home settings; operated by patient or caregiver. Output monitored via optional oxygen purity sensor. Safety features include pressure relief, thermal protection, high/low pressure alarms, and power failure alerts. Benefits patients by providing supplemental oxygen for respiratory disorders.
Clinical Evidence
Bench testing only. Device performance verified against ASTM 1464 standards. Comparative data provided for concentration levels (90% ± 3%), flow rates (1-5 LPM), outlet pressure (8.5 psig), and sound levels (45-48 dBA).
Indicated for patients with respiratory disorders requiring supplemental oxygen. Not intended for life support.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
# 510(K) SUMMARY (as required by 807.92(c))
MAR 1 8 2008
| Submitter of 510(k): | Medical Depot<br>99 Seaview Blvd<br>Port Washington, NY 11050<br>USA |
|-------------------------|----------------------------------------------------------------------|
| | Phone: 877-224-0946<br>Fax: 516-998-4601 |
| Contact Person: | Randy Rosen |
| Date of Summary: | January 23, 2008 |
| Trade/Proprietary Name: | Drive Solstice Oxygen Concentrator |
| Classification Name: | Generator, Oxygen, Portable |
| Product Code: | CAW |
## Intended Use:
The intended function and use of the Drive Solistice Oxygen Concentrator (models 18050 and 18055) is to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve. It is not intended to sustain or support life.
## Device Description:
The Drive Solstice Oxygen Concentrator is a PSA (Pressure Swing Adsorption) system Oxygen Concentrator, the output of oxygen is 1 to 5 liter per minute. Room air enters the piston type compressor via a series of filters for removing dust particles. The output compressed air is directed by a pneumatic valve into one of the two sieve beds which is full of the adsorption material -molecular sieve. Nitrogen is adsorbed by the molecular sieve as the pressure increases; oxygen flows through the molecular sieve and concentrates at the sieve bed top. The enriched oxygen is divided into two streams; one stream enters a storage tank. The pressurized oxygen is regulated down to the suitable pressure, an adjustable flow meter and out to the patient. At the same time the second bed is in exhausted status, the molecular sieve desorbs nitrogen as the pressure decreases; another oxygen stream from first bed enters the top of the second bed, promotes purging the nitrogen and is exhausted into the atmosphere. Two sieve beds exchange the role of oxygen concentration and continue to produce 90% oxygen to the patient.
{1}------------------------------------------------
#### Predicate Device:
Invacare Platinum 5 Oxygen Concentrator - K020386 - Invacare Corporation A & J-POCA01 Oxygen Concentrator - K 071608 - Zhongshan A&J Medical Equipment Co., Ltd
### Substantial Equivalence:
Medical Depot claims the proposed device to be substantially equivalent to the device previously cleared by FDA in K071608. Medical Depot claims this equivalence because the proposed device has an equivalent intended use, manufacturing materials, operating principles, and physical, operational specification as compared to the predicate device. The Medical Depot Concentrator is identical to this predicate device except for the labels
The similarities and differences between the proposed and predicate devices have been identified and explained in the Comparison Matrix which has been included in Section 9 of this submission. Additionally, this matrix is included as an attachment to the 510(k) summary. These differences have no effect on safety and effectiveness.
{2}------------------------------------------------
| Ver. 2.6.06 | | | |
|-------------------|-----------------------------|---------------------------------|---------------------------------|
| mfg. | Invacare | Drive | Zhongshan A&J Medical |
| product name | Invacare Platinum XL 5- | Solstice | A&J-POCA01 Oxygen |
| | Liter 02 Concentrator with | | Concentrator |
| | Sens 02 | | |
| model No. | IRC5LXO2 | 18050, 18055 | POCA01 |
| concentration | 95.6% to 87% at all flow | 1-5 LPM: 90% ± 3%) | 1-5 LPM: 90% ± 3%) |
| levels | rates | | |
| delivery rate | 0.5 to 5 LPM | 1 to 5 LPM | 1 to 5 LPM |
| Outlet pressure | 5 psi +/- 0.5 psi | 8.5 psig | 8.5 psig |
| | | | |
| alarms | battery-free power loss; | pressure relief / thermal | pressure relief / thermal |
| | sieve performance; oxygen | protection on compressor | protection on compressor |
| | monitor; low-flow; | high / low pressure | high / low pressure |
| | compressor 35 psi pressure- | power failure | power failure |
| | releif valve | low oxygen purity(optional) | low oxygen purity(optional) |
| | | current overload shoutdown | current overload shoutdown |
| | | | |
| electrical rating | 115V 60Hz | 115V/60Hz | 115V/60Hz |
| | | | |
| power | 4.3 amps average @ 5L/min. | 300W average | 300W average |
| consumption | (400W) | | |
| battery | | | |
| filters | Cabinet, outlet HEPA, | Cabinet, intake,outlet HEPA | Cabinet, intake, outlet HEPA |
| | compressor inlet | filter | filter |
| dimensions | 14-3/8 x 18-3/8 x 26-3/8 | 12x14x20 | 12x14x20 |
| (in. L x W x H) | | | |
| | | | |
| weight (lbs) | 51 | 38 | 38 |
| specs/ | | | |
| standards | | | |
| approvals | | Class II equip | Class II equip |
| | | double insulated | double insulated |
| | | Type B Applied Part | Type B Applied Part |
| operating | pressure based system | Timed cycle / pressure swing | Timed cycle / pressure swing |
| system | | | |
| sound level | | | |
| | 50dBA avg | 45~48dBA | 45~48dBA |
| operating | | 50 to 95 deg. F, Humidity: | 50 to 95 deg. F, Humidity: |
| environment | | 30% to 75% | 30% to 75% |
| warranty | 5 yrs | | |
| valve | unique designed (? Popet | The dual solenoid, three- | The dual solenoid, three- |
| | style valve | position, five-way valve | position, five-way valve |
| | | increases shift efficiency, | increases shift efficiency. |
| | | valve life, and reliability and | valve life, and reliability and |
| | | comes with a lifetime | comes with a lifetime |
| | | warranty | warranty |
| compressor | Thomas Based Double | GSE-280A compressor | GSE-280A compressor |
| | wobble, may be chineese | | |
| | made today | | |
| other | self-diagnostic electronics | | |
| | compatible w/HomeFill II O2 | | |
| | filling system | | |
| | | | |
| Oxygen Sensor | Yes (Optional) | Yes (Optional) | Yes (Optional) |
| | | | |
| HEPA Filter | Yes | Yes | Yes |
| ASTM 1464 | Meets Standard | Meets Standard | Meets Standard |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 8 2008
Medical Depot C/O Mr. Daniel W. Lehtonen Responsible Third Party Official Intertek Testing Services NA, Incorporated 2307 East Aurora Road, Unit B7 Twinsburg, Ohio 44087
Re: K080391
Trade/Device Name: Drive Solistice Oxygen Concentrator Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: March 3, 2008 Received: March 4, 2008
Dear Mr. Lehtonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Lehtonen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Sautte y. Michie M.D.
Chin Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Drive Solistice Oxygen Concentrator
Indications for Use:
The intended function and use of the Drive Solistice Oxygen Concentrator (models 18050 and 18055) is to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve. It is not intended to sustain or support life.
Thul
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Kolozal 510(k) Number:
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Drive Solstice Oxygen Concentrator Supplemental Response 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.