RMX PHYSIOLOGICAL DATA RECORDER

K080375 · S.L.P. , Ltd. · MNR · Sep 2, 2008 · Anesthesiology

Device Facts

Record IDK080375
Device NameRMX PHYSIOLOGICAL DATA RECORDER
ApplicantS.L.P. , Ltd.
Product CodeMNR · Anesthesiology
Decision DateSep 2, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.2375
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The RMX is a physiological data recorder intended to collect and record data from multiple physiological channels and additional auxiliary inputs from undesignated channels. It is indicated for use by or on order of a physician. The RMX is intended for use in a supervised (hospital) or unsupervised (home) environment.

Device Story

RMX is an ambulatory, unattended physiological data recorder; collects signals including pulse rate, ECG, EEG, airflow, respiratory effort, snoring, body position, and pulse oximetry. Device worn on chest via inductive effort sensor band. Operates by digitizing and storing physiological data onto an SD memory card. Used in hospital or home settings; operated by or on order of a physician. Data transferred to PC at physician's clinic; converted to EDF format; analyzed by standard scoring software to generate reports. Provides clinicians with recorded data to assist in clinical decision-making; benefits patients by enabling sleep physiological data collection outside of intensive care units.

Clinical Evidence

Bench testing only. Validation performed by injecting known simulated signals into device inputs to verify accurate output replication in EDF format. Testing confirmed device ability to record multiple physiological signals at sufficient quality for expert analysis and verified integrity of patient/setup data.

Technological Characteristics

Ambulatory physiological data recorder. Components: main unit with digitizer, processor, SD memory card storage. Sensors: internal, wired, or wireless (e.g., chest inductive effort band, pulse oximeter). Connectivity: SD card transfer to PC. Software: setup parameters, firmware for A/D conversion, post-study conversion to EDF format. Materials: clothing-standard woven elastic laces for sensor band.

Indications for Use

Indicated for use by or on order of a physician to collect and record multiple physiological channels (pulse rate, ECG, EEG, airflow, respiratory effort, snoring sounds, body position/movement, leg movement, pulse oximetry) in hospital or home environments.

Regulatory Classification

Identification

A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K080375 Image /page/0/Picture/1 description: The image shows a logo with a stylized cat inside a circle on the left, followed by the text "S.L.P." in large, bold letters. To the right of "P", the word "LTD." appears in smaller, subscript-like font. The cat design is abstract, with curved lines forming the shape of the cat's head and body within the circular frame. SEP - 2 -2008 RMX 510(k) Submission # SECTION 5 - 510 (K) SUMMARY #### 5.1 510(K) SUMMARY A 510(k) Summary is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990. This 510(k) Summary meets the requirements identified in 21 C.R.F. 807.92. ### RMX Physiological Data Recorder | Date of Preparation | Aug 26, 2008 | |----------------------------|----------------------------------------------------------------------------------------------| | Applicant's Name | S.L.P Ltd. | | | 62 Anilevitch Street<br>Tel-Aviv 67060. Israel<br>Tel: 972-3-537-1281<br>Fax: 972-3-537-1282 | | Establishment Reg. number | 9614500 | | Contact Person | N. Hadas, CTO (see footer for contact details) | | Trade Name | RMX | | Common Name | Physiological Data Recorder | | Classification Name | Ventilatory Effort Recorder | | Classification Reg. number | 21 CFR 868.2375 | | Regulatory Status | Class II | | Product Code | MNR | | Classification panel | Anesthesiology | {2}------------------------------------------------ Image /page/1/Picture/1 description: The image shows a logo with a stylized cat inside a circle on the left, followed by the text "S.L.P. LTD." on the right. The cat is white and appears to be curled up inside the black circle. The text is in a bold, sans-serif font, with "LTD." in a smaller font size as a subscript. #### 5.2 DEVICE DESCRIPTION The RMX is an ambulatory unattended recording device intended for the recording of sleep physiological data. The RMX functions exclusively as a data-recording device and is not intended to monitor Apnea or other vital signs in an intensive care unit or any other real time application. It is used by or on the order of a physician. The RMX data recording device offers a platform for collecting multiple channels of physiological signals such as pulse rate, ECG, EEG, airflow, thoracic and abdominal respiratory effort, sounds intensity (i.e. noise due to snoring), body position and movements, leg movement, pulse oximetry or similar physiological signals. The data collection box is normally attached over the clothes of the patient's chest using the chest inductive effort sensor band made of clothing-standard woven elastic laces. The device may use its internal sensors, standard wired physiological sensors or several wireless sensors (not covered in this application) to match the specific needs of the patient. Data collected is transferred on an SD memory card from the recorder to a PC at the physician's clinic, where it is converted into industry standard EDF format, after which it is analyzed by standard scoring software to generate a detailed final report with recommendations. The RMX data recording device includes 3 software elements; 1) the SETUP part that runs on a PC and uploads the recorder setup parameters to the SD (secure digital) card as determined by the physician, 2) The Firmware operating the main and auxiliary CPU hardware which control the A/D conversion and storage of the data, and 3) the Post-Study Software which runs on a PC, reads the raw data file from the SD card as a single file and converts it into the industry standard EDF format suitable for input to analytical systems (see section 16 for software description). #### 5.3 INTENDED USE The RMX is a physiological data recorder intended to collect and record data from multiple physiological channels and additional auxiliary inputs from undesignated channels. It is indicated for use by or on order of a physician. The RMX is intended for use in a supervised (hospital) or unsupervised (home) environment. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows a logo with a cat curled up in a circle on the left and the text "S.L.P. LTD." on the right. The cat is white and the circle is black. The text is black and in a sans-serif font. The letters "LTD." are smaller than the other letters. #### SUBSTANTIAL EQUIVALENCE 5.4 S.L.P Ltd. believes that the RMX is substantially equivalent to the following predicate devices without raising any new safety and/or effectiveness issues: - 1. Portable sleep data recorder, Pro-tech services Inc., WA, USA, cleared under K033402. - 2. DeVilbiss Sleep Recorder,DeVilbiss, PA USA, cleared under K012437. - 3. SleepSense Sleep Sensors, S.L.P. LTD., IL, K042253 - 4. SPO, Pulse Oximeter, SPO LTD., IL, K040178 Detailed description of the predicate devices is provided in Section 12 of this application. #### 5.5 TECHNOLOGICAL PROPERTIES The RMX is a basic physiological signal recorder. Similar to the predicate devices, the RMX uses a set of standard internal and/or external physiological sensors to collect and record multiple physiological parameters such as respiration, movement or sound. All signals are channeled to the main unit recorder which contains a digitizer and a processor to store them on a solid-state memory card. The predicate devices are identical to the RMX in that they contain similar functional modules, which perform the same functions. The sensors integrated in the RMX device, or connected externally to the system are based on sensor technology developed by SLP over the last 15 years of making standard sleep lab sensors, and are covered by 510K K042253. The SPO ltd. oximeter used in the system is covered by 510K K040178. Additional sensors or devices such as the pressure cannula are covered by their respective FDA 510K approvals. Therefore, in design and functionality aspects, the RMX does not raise any new questions of safety or effectiveness. Any minor differences in the design and operational method (intended to help the user perform the recording successfully) between the RMX and its predicate devices do not raise any safety or effectiveness issues. {4}------------------------------------------------ RMX 510(k) Submission Image /page/3/Picture/1 description: The image shows the logo for S.L.P. LTD. The logo consists of a stylized cat face inside of a circle on the left, followed by the text "S.L.P." in large, bold letters. The letters "LTD." are written in smaller letters as a subscript to the letter P. The logo is black and white. #### ડ.6 FUNCTIONAL TEST The technical specifications of the RMX were selected to match accepted non-formal standards in the industry, including sampling rates, sampling depths (# of bits per sample), sensor types and attachment location, as well as methods and overall operation protocol. Due to the sole function as a data recorder, validation encompassed a single setup in which the complete RMX device (software and hardware) was tested for its ability to generate, from known simulated signals injected to its inputs, an accurate output replica in a file in EDF format. The test also confirms that the original patient details and the loaded set-up data are correct. The results of this test confirm that the RMX can accurately record multiple physiological signals, at a sufficient quality for later analysis by trained experts. {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The logo is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP - 2 2008 S.L.P. Limited C/O Mr. J. G. van Vugt Responsible Third Party Official KEMA Quality B.V. 4377 County Line Road Chalfont, Pennsylvania 18914 Re: K080375 Trade/Device Name: RMX Physiological Data Recorder Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: August 15, 2008 Received: August 18, 2008 Dear Mr. van Vugt: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title, 21 Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ Page 2 - Mr. van Vugt Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal regaricies. You must comply with all the Act's requirements, including, but not limited to registers. and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing proctice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as of ancibed in your Section 510(k) premarket notification. The FDA finding of substantial aquivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrio's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. TK Samuel S-Sled, mr for " Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {7}------------------------------------------------ # SECTION 4- INDICATIONS FOR USE STATEMENT # (For 510k Substantially Equivalent Determination) ## 510(k) Number (if known): K080375 Device Name: RMX Physiological Data Recorder ## Indications for Use: The RMX is a physiological data recorder intended to collect and record data from multiple physiological channels and additional auxiliary inputs from undesignated channels. It is indicated for use by or on order of a physician. The RMX is intended for use in a supervised (hospital) or unsupervised (home) environment. Prescription Use __X (Part 21 CFR 801 Subpart D) AND/ OR Over the Counter Use (Part 21 CFR 801 Subpart C) (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K080375
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...