The Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3MQ1-2D is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm. The device detects the appearance of atrial fibrillation during measurement and gives a warning signal with the reading once the atrial fibrillation is detected.
Device Story
Automatic digital blood pressure monitor; uses inflatable upper arm cuff; electronic capacitive pressure sensor converts cuff pressure changes to electrical signals; oscillometric method determines systolic/diastolic pressure and pulse rate. Device includes atrial fibrillation (AF) detection module; displays AF symbol on LCD if detected. Used by adults; intended for clinical or home settings. Healthcare providers or patients use output to monitor blood pressure and identify potential AF. Benefits include non-invasive blood pressure measurement and automated screening for atrial fibrillation during routine checks.
Clinical Evidence
Clinical validation performed via two trials (TRIPPS 1.0 and 1.1) with 205 and 157 subjects respectively. Compared device AF readings against ECG rhythm. Combined sensitivity for individual readings was 96.1% (94-98% CI) and specificity 85.2% (83-88% CI). Combined sensitivity for three-reading sequential analysis was 97.7% (91-100% CI) and specificity 87.7% (83-91% CI). Algorithm also validated against MIT-BIH AFIB database (92% sensitivity, 96% specificity).
Technological Characteristics
Oscillometric measurement; electronic capacitive pressure sensor; automatic inflation/deflation via factory-set bleed valve. Powered by battery/external source. Software-based AF detection algorithm. Tested per ANSI/AAMI SP10:2002 and IEC 60601-1 safety standards.
Indications for Use
Indicated for adult individuals to measure systolic/diastolic blood pressure and pulse rate via non-invasive oscillometric upper arm cuff; includes detection and warning of atrial fibrillation.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP3BT0-1 (K013485)
{0}------------------------------------------------
## Exhibit#1
# 510(k) SUMMARY
# MAY 1 3 2008
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: k080337
### 1. Submitter's Identification:
Microlife Intellectual Property GmbH, Switzerland
Espenstrasse 139 9443 Widnau / Switzerland
Date Summary Prepared: Jan. 11, 2008
### 2. Name of the Device:
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3MQ1-2D
## Information for the 510(k) Cleared Device (Predicate Device): 3.
a. Microlife Upper Arm Automatic Blood Pressure Monitor, Model BP3BT0-1, K#013485, Microlife Corporation,
b. AfibAlbert™ , K#052767, Lechnologies Research, Inc.
## 4. Device Description:
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3MQ1-2D is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive MAM (Microlife Average Mode) technique in which an inflatable cuff is wrapped around the upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic capacitive pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method".
The device detects the appearance of atrial fibrillation during measurement and the atrial fibrillation symbol "~ " is displayed on the LCD screen if any atrial fibrillation signal has been detected.
page 1 OF 4
{1}------------------------------------------------
# 5. Intended Use:
The Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3MQ1-2D is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm.
The device detects the appearance of atrial fibrillation during measurement and gives a warning signal with the reading once the atrial fibrillation is detected.
### 6. Comparison to the 510(k) Cleared Device (Predicate Device):
The modified device model BP3MQ1-2D and the predicate device model BP3BT0-1 use the well-known oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate. An upper arm cuff is inflated automatically, deflate rate is controlled by a factory set bleed valve and the deflation pressures are transferred via tubing to a sensor.
The solely differences between the two models are the additional features such as atrial fibrillation detection function. However, the difference does not affect the accuracy and normal use of this device.
Atrial fibrillation detection function of BP3MQ1-2D is similar with what is used in predicate device AfibAlbert™, with 510(k) cleared number K#052767. The input data from the blood pressure signals (data) from BP3MQ1-2D will produce a substantially equivalent outcome as the predicate AfibAlert™ getting the input data from an ECG signal (data). They are processed by an algorithm and both output atrial fibrillation diagnositic information.
## 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Testing information demonstrating safety and effectiveness of the Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3MQ1-2D in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance requirements.
The following testing was conducted:
- Reliability Test Storage test a.
- Reliability Test Operating test ﻘ
- C. Reliability Test - Vibration test
- Reliability Test Drop test d.
- Reliability Test Life test e.
- EMC Test ಿ
- g. IEC 60601-1 Safety Test
None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was
page 2 of 4
{2}------------------------------------------------
our conclusion that Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3MQ1-2D tested met all relevant requirements of the aforementioned tests.
### 8. Discussion of Clinical Tests Performed:
For clinical validation the atrial fibrillation function, AfibAlert™ algorithm was tested using published available annotated MIT-BIH AFIB database, which showed a 92% sensitivity value and a 96% specificity value.
The BP3MQ1-2D was tested in two clinical trials which compared the atrial fibrillation (AF) readings from the BP3MQ1-2D to the rhythm determined by an electrocardiogram(ECG). The studies, Trial of Regular versus Irregular Pulse to Prevent Stroke (TRIPPS) 1.0 and 1.1, enrolled 205 and 157 subjects, respectively. All subjects in TRIPPS 1.0 and all, but one subject, in TRIPPS 1.1 had three atrial fibrillation readings performed with the BP3MQ1-2D. The results comparing the BP3MQ1-2D readings to the ECG readings read in a blinded fashion are shown in the table 1. The BP3MQ1-2D detected atrial fibrillation with a sensitivity ranging from 93% to 98% and a specificity ranging from 82-88%.
| BP3MQ1-2D reading | ECG reading | | Sensitivity | Specificity |
|-------------------|-------------|--------|-------------|-------------|
| | AF | Non-AF | | |
| TRIPPS 1.0 | | | 98.1% | 88.2% |
| AF | 156 | 54 | (96-99%) | (85-91%) |
| Non-AF | 3 | 402 | | |
| TRIPPS 1.1 | | | 92.9% | 81.7% |
| AF | 92 | 68 | (88-98%) | (78-86%) |
| Non-AF | 7 | 303 | | |
| Combined | | | 96.1% | 85.2% |
| AF | 248 | 122 | (94-98%) | (83-88%) |
| Non-AF | 10 | 705 | | |
Table 1. Results for individual readings by the device compared to ECG. The last two columns show the sensitivity and specificity and the 95% confidence intervals in parentheses.
In a second analysis of these results, the three sequential readings done in each subject were used to determine a single final reading based on which reading was most frequent. Thus if two or all three of the readings for a given patient showed atrial fibrillation then the combined final reading was atrial fibrillation. If none or only one of the three readings showed atrial fibrillation then the combined final reading
PAGE 3 OF 4
{3}------------------------------------------------
was not atrial fibrillation. The results of this analysis is shown in the table 2. Using the combined three sequential readings, the BP3MQ1-2D detected atrial fibrillation with a sensitivity ranging from 94% to 100% and a specificity ranging from 85-90%.
| BP3MQ1-2D reading | ECG reading | | Sensitivity | Specificity |
|-------------------|-------------|--------|-------------|-------------|
| | AF | Non-AF | | |
| TRIPPS 1.0 | | | 100.0% | 89.5% |
| AF | 53 | 16 | (92-100%) | (83-94%) |
| Non-AF | 0 | 136 | | |
| TRIPPS 1.1 | | | 93.9% | 85.4% |
| AF | 31 | 18 | (78-99%) | (78-91%) |
| Non-AF | 2 | 106 | | |
| Combined | | | 97.7% | 87.7% |
| AF | 84 | 34 | (91-100%) | (83-91%) |
| Non-AF | 2 | 242 | | |
Table 2. Results for the three sequential readings combined compared to the ECG. The last two columns show the sensitivity and specificity and the 95% confidence intervals in parentheses.
Based upon the aforementioned information, the requirement to develop an atrial fibrillation detection module for the BP3MQ1-2D to detect atrial fibrillation that is added to the existing firmware included in the BP3BT0-1 has been met.
### Software information: a.
Software validation was conducted in accordance with a moderate level of concern designation in accordance with the FDA November 2005 document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
## 10. Conclusions:
It has been demonstrated that there is no difference between the Microlife Upper Arm Automatic Digital Blood Pressure Monitor Model BP3MQ1-2D and the predicate model BP3BT0-1 in terms of safety and effectiveness based on electrical, mechanical and environmental test results, the FDA DCRND November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", and the ANSI/AAMI Voluntary Standard, SP10: 2002 test results.
{4}------------------------------------------------
Public Health Service
Image /page/4/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three stripes on its wing, symbolizing the department's commitment to health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 3 2008
Microlife Intellectual Property GmbH c/o Ms. Susan Goldstein-Falk Official Correspondent mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, NY 11021
Re: K080337
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3MQ1-2D Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II (two) Product Code: DXN Dated: April 17, 08 Received: April 18, 2008
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2 - Ms. Susan Goldstein-Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## EXHIBIT B
# Indications for Use
k080337 510(k) Number (if known):__
Device Name: Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3MQ1-2D
## Indications For Use:
The Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP3MQ1-2D is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm.
The device detects the appearance of atrial fibrillation during measurement and gives a warning signal with the reading once the atrial fibrillation is detected.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Islon Sign-Off
D. sion of Cardiovascular Devices
510(k) Number K680537
Page 1_of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.